Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K212894

    Validate with FDA (Live)

    Device Name
    PhasTIPP
    Date Cleared
    2022-03-31

    (202 days)

    Product Code
    Regulation Number
    870.4885
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PhasTIPP System is indicated for use in ambulatory phlebectomy procedures for the resection and ablation of varicose veins. The Illuminator is also indicated for use without the Resector for visualization of varicose veins and infusion of tumescent solution during an ambulatory phlebectomy procedure.

    Device Description

    The PhasTIPP System consists of an Illuminator and a Resector. During the surgical procedure, the illuminator shaft is placed through an incision under the skin with its light-emitting end to provide intense light for a better visualization of the varicose veins. A port that runs along the illuminator shaft is used to infuse tumescence solution to cause the contraction of varicosities, which aids vein visualization by creating a subcutaneous fluid pocket through which the illuminator's light can disperse. A powered resector, positioned through a different incision can then be used to morcellate and aspirate the varicosities.

    The PhasTIPP Illuminator consists of two devices: a reusable Illuminator Handpiece and a disposable Illuminator.

    The Handpiece provides illumination controls and contains an LED to provide intense light. The Handpiece is powered with new batteries for each new surgical procedure (Duracell CR 123a cell). The handpiece is provided non-sterile and must be covered with the microbial barrier sheath attached to the Disposable Illuminator in the sterile field before the procedure begins.

    The Disposable Illuminator includes, in addition to the microbial barrier sheath, a distal stainless steel fiber optic light shaft. When connected to a peristaltic pump, the Disposable Illuminator can also infuse tumescence.

    The PhasTIPP Resector consists of two devices: a reusable Resector Handpiece and a Resector Disposable (available in two diameters, 4.5mm and 5.5mm). The Disposable Resector also connects to a peristaltic suction pump to remove the resected varicosities.

    The Resector Handpiece provides controls for the operation of the rotation blades on the Disposable Resector and is powered with a set of new batteries (TLM-1550 HPM cell), for each surgical procedure. The Resector Handpiece is provided non-sterile and must be covered with the microbial barrier sheath attached to the Disposable Resector in the sterile field.

    The Control Unit that was a large equipment in the predicate device is displaced in the subject device. Instead, the Resector Handpiece and the Illuminator Handpiece control the disposable resector and illuminator in the subject device.

    AI/ML Overview

    This document describes the PhasTIPP System, an external vein stripper. The FDA's 510(k) summary provides details about the device, its intended use, and the testing conducted to establish its substantial equivalence to a predicate device.

    1. A table of acceptance criteria and the reported device performance

    The provided document doesn't explicitly state "acceptance criteria" in a tabulated format alongside "reported device performance" in the way one might expect for a clinical study with primary endpoints. Instead, it lists various design validation, system validation, and performance tests that were completed to demonstrate the device meets product performance specifications and for substantial equivalence.

    For the purpose of this request, I will infer the "acceptance criteria" from the types of tests performed and the "reported device performance" as the overall statement that the device meets these specifications.

    Acceptance Criteria (Inferred from Tests)Reported Device Performance (Summary)
    Illuminator Design Validation:The Illuminator's design, including shaft dimensions, flow rate, light output, leak testing, sheath integrity and usability, and tensile/torque strengths of components, met specifications.
    Shaft diameter, length, IV spike to pump, pump tube, pump to hub lengthsMet specifications
    Flow Rate (ml/min)Met specifications
    Light Output (lumens)Met specifications
    Leak TestingPassed
    Sheath Length (in)Met specifications
    Usability Through SheathConfirmed
    Sheath Maintains Microbial BarrierConfirmed
    Tensile Strength: Hub to Shaft (lbf), Disposable (lbf), Pump tube to Connectors, IV Spike to Tubing (lbf)Met specifications
    Torque Strength: Disposable (in-lb)Met specifications
    Kink Test: Infusion TubingPassed
    Illuminator System Validation:The Illuminator's system functions, including basic operation and low battery performance, were verified.
    Basic FunctionVerified
    Low Battery FunctionalityVerified
    Fault Insertion and Power Supply Fuse testingVerified
    Resector System Validation:The Resector's system functions, including voltage regulation, motor RPM, home position, communications, diagnostics, embedded processor, and battery critical function, were verified.
    Voltage RegulationVerified
    Motor RPMVerified
    Home Position and Home Position DriftVerified
    Communications and Diagnostic LoggingVerified
    Embedded Processor Hardware Basic FunctionVerified
    Jam ClearVerified
    Battery Critical FunctionVerified
    Embedded processor/Software loadVerified
    Resector Software Integration Validation:The Resector's software integration for timer, low battery, hardware configuration, motor state machine, oscillation, anti-stall, high current fault, motor time save/retrieve, and RTC were validated.
    Timer functionalityValidated
    Low battery functionalityValidated
    Low level hardware configuration of the micro-controller hardwareValidated
    Motor state machine runningValidated
    Functionality of the oscillation feature and softwareValidated
    Fault states related to the "anti-stall routineValidated
    High motor current fault by simulating excessive motor currentValidated
    Motor time save and retrieve functionValidated
    Real Time Clock (RTC) against the system clockValidated
    Battery critical lockout featureValidated
    Resector Design Validation:The Resector's design, including shaft diameters and lengths, aspiration tube characteristics, sterile sheath integrity, and tensile/torque strengths of components, met specifications.
    Shaft Diameter (4.5mm & 5.5mm)Met specifications
    Shaft Length (4.5mm & 5.5mm)Met specifications
    Aspiration Tube LengthMet specifications
    Aspiration Tube tensile strength to ConnectorMet specifications
    Kink Testing & Leak TestingPassed
    Bond Tubing to Back LidMet specifications
    Aspiration Flow Rate Comparison (4.5 & 5.5)Met specifications
    Motor Functionality DataMet specifications
    Sterile Sheath: Adequate length, Microbial Barrier Testing, Maintains Contamination BarrierConfirmed
    Tensile Strength: Disposable Hub to Main Shaft (4.5mm & 5.5mm), Driven Gear to the resector inner shaft (4.5mm & 5.5mm), Disposable Assembly to HandpieceMet specifications
    Torque Strength: Disposable Hub to Shaft assembly, Driven Gear to the resector inner shaftMet specifications
    Bonded joint between proximal hub and the backlidMet specifications
    Distal Hub Pinned to proximal resectorMet specifications
    Microbial Sheath Barrier Design Validation:The sheath's bubble leak test confirmed its barrier integrity.
    Sheath Bubble Leak TestPassed
    PhasTIPP Illuminator Injection Line Design Validation:The injection line's pressurized leak testing met specifications.
    Pressurized Leak TestingPassed
    PhasTIPP Illuminator Injection Line Shelf Life Validation (2 years Accelerated aging):The injection line maintained performance after accelerated aging.
    Pressurized Leak TestingPassed
    Flow Rate MeasurementMet specifications
    Luer Fitting Tensile Strength testMet specifications
    Overall LengthMet specifications
    PhasTIPP Performance Testing:System performance was confirmed in a 3D in-vitro bench test.
    System performance in 3D in-vitro bench testConfirmed
    PhasTIPP Illuminator Heat Evaluation:The heat produced by the Illuminator was evaluated.
    Evaluation of heat produced by PhasTIPP IlluminatorEvaluated (implies within acceptable limits for a medical device)
    PhasTIPP Testing to IEC 60601 Medical Electrical Equipment:The device met relevant IEC 60601 standards for electromagnetic compatibility.
    Radiated EmissionsCompliant
    Electro-Static DischargeCompliant
    Radiated Field ImmunityCompliant
    Power Frequency Magnetic FieldCompliant
    Biocompatibility (ISO 10993-1:2018):The device was found to be biocompatible for its intended use.
    External Communicating Device, Contact Circulating Blood, limited use (<24h)Demonstrated biocompatible
    Sterilization (EtO gas):The sterilization process was validated according to industry standards.
    Per EN ISO 11135-1:2014 and AAMI TIR28:2016Validated

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide specific sample sizes for the individual engineering and performance tests listed. The data provenance is not specified beyond being "product testing" and "verification activities." These are typically conducted in-house by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This type of information is not applicable to the tests described in this 510(k) summary. The listed tests are primarily engineering and performance validations, not clinical studies requiring expert interpretation of medical images or patient outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods are typically relevant for clinical studies involving human interpretation or subjective assessments. This 510(k) submission focuses on device engineering and performance validation, so "adjudication method" is not applicable here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such MRMC study was mentioned. The device described is an external vein stripper, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the PhasTIPP System is a physical medical device (an external vein stripper) used in surgery, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the engineering and performance tests described, the "ground truth" or reference standard would be the established engineering specifications, industry standards (e.g., IEC 60601, ISO 10993-1, EN ISO 11135-1), and the performance of the predicate device. These are objective measures or recognized benchmarks. For example:

    • Mechanical properties (tensile strength, torque strength): Established material specifications and engineering design tolerances.
    • Flow rate, light output: Measured against design specifications.
    • Sterilization: Validated against sterility assurance levels defined by standards.
    • Biocompatibility: Assessed against cytotoxicity, sensitization, irritation tests as per ISO 10993.

    8. The sample size for the training set

    The concept of a "training set" is generally applicable to machine learning or AI models. This 510(k) summary does not concern an AI/ML device, so a training set as such is not relevant. The device undergoes design verification and validation testing.

    9. How the ground truth for the training set was established

    As there is no "training set" in the context of this device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041453

    Validate with FDA (Live)

    Date Cleared
    2004-07-22

    (51 days)

    Product Code
    Regulation Number
    870.4885
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Smith & Nephew Tumescent Catheter Inversion System is intended to provide a system for removal of incompetent greater saphenous veins.

    Device Description

    The Smith & Nephew Tumescent Catheter Inversion System provides a system for removal of incompetent greater saphenous veins. The system also provides the capability to flush the tunnel left behind from the removal of the vein with tumescent solution.

    AI/ML Overview

    The provided text is a 510(k) summary for the Smith & Nephew Tumescent Catheter Inversion System. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain any information regarding specific acceptance criteria, a study that proves the device meets those criteria, or any of the detailed performance metrics commonly associated with such studies (like sample sizes, ground truth establishment, or expert qualifications).

    The document explicitly states:

    H. Summary Performance Data

    "There are no known performance standards or special controls promulgated under section 514 of the Act for this device has been tested and found to be in compliance with ISO 10993-1 and at the time of commercialization will be in compliance with applicable sterilization standards. This device has been tested and found to be safe and effective for its intended use."

    This statement indicates that while the device was tested for safety and effectiveness and compliance with ISO 10993-1 (biocompatibility) and sterilization standards, the submission does not include a detailed study with acceptance criteria and reported performance data in the manner typically seen for complex diagnostic or AI-based devices.

    Therefore, I cannot populate the table or answer the specific questions about acceptance criteria and study details because that information is not present in the provided text. The submission focuses on demonstrating substantial equivalence to a pre-amendment device based on similar design, materials, function, and intended use, rather than presenting a detailed performance study against quantified acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032387

    Validate with FDA (Live)

    Device Name
    TRIVEX SYSTEM
    Date Cleared
    2003-10-29

    (86 days)

    Product Code
    Regulation Number
    870.4885
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for use in ambulatory phlebectomy procedures for resection and ablation of varicose veins.

    Device Description

    The Smith & Nephew TriVex™ System consists of a control unit mounted on a pedestal stand, dual peristaltic pumps, an illuminator, resector handpiece and footswitches.

    AI/ML Overview

    The Smith & Nephew TriVex™ System is indicated for use in ambulatory phlebectomy procedures for the resection and ablation of varicose veins.

    Here's an analysis of the provided text regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria / Performance MetricReported Device Performance
    Biocompatibility requirements (ISO 10993-1)Meets biocompatibility requirements.
    Electrical safety standards for BF type equipment (UL 60601-1 and IEC 60601-1)Meets electrical safety standards. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Substantial Equivalence to Predicate DevicesThe device is deemed "substantially equivalent in Intended Use and Fundamental Scientific Technology" and "substantially equivalent in design, materials, function and intended use" to the predicate devices: Dyonics® Varicose Vein Ablation System (K990723), Subcutaneous Illuminator (K991323), Smith & Nephew Xenon Light Source (K994084), and H.K. Surgical Klein Pump (K012044). No specific numerical performance metrics are provided for the TriVex system's efficacy in vein ablation or resection beyond this statement of equivalence.

    2. Sample Size and Data Provenance for Test Set:

    The provided document does not contain any information about sample sizes for a test set or the provenance of any clinical data (e.g., country of origin, retrospective or prospective). The submission primarily focuses on substantial equivalence based on technological characteristics and safety standards.

    3. Number of Experts and Qualifications for Ground Truth of Test Set:

    This information is not provided in the document. As no clinical test set performance data is detailed, information about experts for ground truth establishment is absent.

    4. Adjudication Method for Test Set:

    This information is not provided in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    The document does not mention or describe any MRMC comparative effectiveness study comparing human readers with and without AI assistance. This device is not an AI-powered diagnostic system; it's a surgical tool.

    6. Standalone (Algorithm Only) Performance Study:

    The document does not describe any standalone algorithm performance, as the device is a physical surgical system, not an algorithm.

    7. Type of Ground Truth Used:

    Based on the provided text, the "ground truth" for the device's acceptance is based on its compliance with established safety and biocompatibility standards (ISO 10993-1, UL 60601-1, IEC 60601-1) and its substantial equivalence to existing legally marketed predicate devices in terms of intended use, design, materials, and function. There is no mention of clinical outcome ground truth in a study.

    8. Sample Size for Training Set:

    This information is not applicable and not provided. The TriVex™ System is a medical device, not an AI model requiring a training set in the conventional sense. Its "training" would be through engineering design, material selection, and manufacturing processes, guided by established standards and predicate device characteristics, not data-driven machine learning.

    9. How Ground Truth for Training Set Was Established:

    This information is not applicable and not provided. The "ground truth" for the development of such a device stems from engineering principles, regulatory compliance requirements, and the established safety and efficacy of the predicate devices it aims to be equivalent to.

    In summary:

    The acceptance of the Smith & Nephew TriVex™ System by the FDA, as indicated by the 510(k) clearance (K032387), is based on its demonstration of substantial equivalence to existing predicate devices and its compliance with relevant biocompatibility and electrical safety standards. The submission focuses on these aspects rather than detailed clinical performance studies with specific patient or image-based test sets, which are typically required for diagnostic algorithms or novel therapeutic devices requiring extensive clinical validation.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1