Search Results
Found 223 results
510(k) Data Aggregation
(47 days)
The Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Circuit Tubing Set is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures for ≤ 6 hours.
The Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing is a tubing pack intended for use during cardiopulmonary bypass. The tubing is DOP free polyvinylchloride (PVC) with an inner diameter of 9.5mm coated with Smart Reactors' proprietary Camouflage coating provided in a length of 180cm.
This device is intended for use in the adult patient population.
N/A
Ask a specific question about this device
(146 days)
These cannulae are indicated for CPB procedures requiring delivery of cardioplegia in a retrograde direction through the coronary sinus for up to 6 hours in duration.
The Retrograde Coronary Sinus Perfusion Cannulae models in scope of this 510(k) consist of a wire-wound silicone cannula body with a beveled tip (6-Fr models have a non-wire wound cannula body). Two side holes are present near the tip. The back of the cannula body terminates in a locking female luer. A pressure monitoring line is an integral part of the cannula body, beginning at the tip and terminating in a locking female luer fitting or 3-way stopcock with a locking female luer fitting. An inflatable balloon is located at the distal beveled tip. The inflation assembly is located at the back of the cannula body and contains a female slip luer and a one-way valve assembly. Cannulae have a guidewire or solid stylet with a male luer handle or a TruTouch handle.
N/A
Ask a specific question about this device
(70 days)
The Dual Stage Venous Return Cannula is indicated for single tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery for up to six hours.
Dual Stage Venous Return Cannulae are single-use, non-toxic, non-pyrogenic fluid path devices and supplied sterile and individually packaged.
The devices are composed of the cannula: an open lumen PVC polymer tube incorporating wire reinforcement in distal section. In some models is present a malleable obturator inserted into the cannula to allow the placement of the cannula along the vein. Once the cannula is positioned onto the vessel the surgeon removes the obturator from the cannula tubing and connects it to the venous line.
The distal end of the cannula has a lighthouse-shaped tip, allowing the venous blood to flow from the vessel into the cannula, in addition the distal sections of the cannula are perforated with multiple holes at multiple stages to better allow the fluid flows from outside into the cannula body. The clear proximal section is not reinforced to allow clamping; the proximal end typically does not have any pre-mounted connector and can accommodate a barbed connector for standard cardiopulmonary bypass tubing (½" diameter). In some models a pre-mounted 1/2" barbed connector is present.
The following cannula models are available:
| Device trade name | Model | French size (Outside diameter) | Characteristics |
|---|---|---|---|
| Dual Stage Venous Cannulae | RDS-61040 | Proximal: 40 Fr Distal: 32 Fr. | Without connector |
| Dual Stage Venous Cannulae | RDS-61034 | Proximal: 46 Fr Distal: 34 Fr. | Without connector |
| Dual Stage Venous Cannulae | RDS-61046 | Proximal: 46 Fr Distal: 36 Fr. | Without connector |
| Dual Stage Venous Cannulae | RDS-61050 | Proximal: 50 Fr Distal: 36 Fr. | Without connector |
| Dual Stage Venous Cannulae | RDS-61140 | Proximal: 40 Fr Distal: 32 Fr. | With connector |
| Dual Stage Venous Cannulae | RDS-61134 | Proximal: 46 Fr Distal: 34 Fr. | With connector |
| Dual Stage Venous Cannulae | RDS-61146 | Proximal: 46 Fr Distal: 36 Fr. | With connector |
| Dual Stage Venous Cannulae | RDS-61150 | Proximal: 50 Fr Distal: 36 Fr. | With connector |
The Dual Stage Venous Return Cannulae are the modified version of the disposables currently marketed under K943934.
Both modified and unmodified cannulae models are recommended for use as Venous cannula during cardiopulmonary bypass up to six hours.
N/A
Ask a specific question about this device
(39 days)
The device is indicated for temporary use during the creation of anastomosis for medical conditions requiring beating heart coronary artery bypass procedures.
The disposable intracoronary shunts consist of a flexible tube. Both ends of the tube feature a teardrop shaped tip made from radiopaque material with 14 mm between the ends. Tip diameters range from 1.00 mm to 3.00 mm. A tether with a tag made of radiopaque material is permanently attached to the shunts. These devices are sterile, nonpyrogenic, disposable and intended for single use only. They are sterilized using ethylene oxide and are categorized as external communicating devices, with limited contact (<=24 hours) circulating blood contact.
N/A
Ask a specific question about this device
(30 days)
The AngioVac Cannula is indicated for use as a venous drainage cannula and for removal of fresh, soft thrombi or emboli during extracorporeal bypass for up to 6 hours.
The AngioVac Cannula is a venovenous cannula with a nitinol basket reinforced, self-expandable funnel shaped distal tip collapsed using an over-sheath that can be advanced through a 26 Fr sheath and over a guidewire into the venous system percutaneously or via a surgical cut-down. During use, the cannula is connected to an extracorporeal circuit, an AngioVac Circuit, a commercially available centrifugal pump and bubble trap. A commercially available reinfusion cannula is placed for venous return (typically within internal jugular or one of the common femoral veins) and connected to the extracorporeal circuit. The funnel tip is actuated by advancing the AngioVac Cannula out of the sheath deploying the self-expanding nitinol reinforced funnel shaped tip at the desired tip angle. Once optimal flow rate is achieved, the AngioVac Cannula is advanced under image guidance towards the undesirable intravascular (i.e. thrombus or emboli) until it is engaged, suctioned into the cannula and removed from the vasculature. The blood is then circulated though the filter and returned to the patient via the venous return cannula. A benefit of the AngioVac Cannula is that it allows for removal of thrombus and embolic material, while minimizing blood loss via recirculation of blood through a standard extracorporeal (venovenous) bypass circuit. Target vessels for the thrombus/embolus extraction include but are not limited to, the iliofemoral vein, Inferior Vena Cava (IVC), Superior Vena Cava (SVC) and Right Atrium (RA). The device is provided in ~20° (AngioVac C20) and ~180° (AngioVac C180) angled configurations.
N/A
Ask a specific question about this device
(160 days)
The Remote Access Perfusion (RAP) femoral venous cannula is intended for use as a venous drainage cannula during cardiopulmonary bypass up to six hours
RAP Femoral venous cannulae are single-use, non-toxic, non-pyrogenic fluid path devices and supplied sterile and individually packaged.
The device is composed of a cannula and a malleable obturator inserted into the cannula to allow the placement of the cannula along the femoral vein till the vena cava. The cannula and the obturator are packaged within the same pouch and sold together.
The cannula is an open lumen PVC polymer tube incorporating wire reinforcement in distal sections. The distal sections of the cannula are perforated with multiple holes at multiple stages to allow fluid flow. The clear proximal section is not reinforced to allow clamping; the proximal end terminates in a 3/8" to ½" barbed connector for standard cardiopulmonary bypass tubing.
The RAP femoral arterial cannulae are the modified version of the disposables currently marketed as the RAP Femoral Venous cannula (K052081).
This document, an FDA 510(k) Clearance Letter, is for a medical device (a cannula) and not for a software or AI-based medical device. Therefore, it does not contain the information requested in the prompt regarding acceptance criteria and studies that prove a device meets acceptance criteria for an AI/software device.
The questions in your prompt are highly specific to AI/Machine Learning device validation, such as:
- Acceptance criteria for AI performance: (e.g., sensitivity, specificity, AUC)
- Sample size, provenance, expert ground truth, adjudication: These are standard for evaluating AI model performance.
- MRMC study and effect size: How AI assists human readers.
- Standalone performance: Algorithm without human input.
- Type of ground truth: Pathology, outcomes, expert consensus.
- Training set details: Sample size and ground truth establishment.
None of this information is relevant or present in the provided 510(k) clearance letter for the RAP Femoral Arterial Cannulae, which is a physical device subject to mechanical, material, and biocompatibility testing.
The document confirms the device meets acceptance criteria through:
- Non-Clinical Performance Data: Extensive verification and validation testing of the physical device components and function (e.g., visual inspection, connector testing, flow rate, kink test, pull strength, blood trauma characterization, biocompatibility tests).
- No Clinical Performance Data: The submission explicitly states "No clinical testing was conducted in support of the RAP Femoral venous cannulae, as the indications for use and technical characteristics are equivalent to those of the predicate devices, which have been on the market for several years with proven safety and efficacy of use."
Therefore, I cannot extract the information asked in your prompt from this particular document.
Ask a specific question about this device
(119 days)
The EasyFlow Aortic Cannula is intended for use in the perfusion of the aorta during cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass for up to 6 hours.
The EasyFlow Duo Cannula with guidewire is intended for use with cardiopulmonary bypass as an arterial return femoral cannula for up to 6 hours.
Easyflow & Easyflow Duo are single-use, non-toxic, non-pyrogenic fluid path devices and supplied sterile and individually packaged.
The Easyflow & Easyflow Duo are composed of the cannula and a malleable obturator inserted into the cannula to allow the placement of the cannula along the Artery.
The cannula is an open lumen PVC polymer tube incorporating wire reinforcement in distal sections. The distal sections of the cannula are perforated with multiple holes at multiple stages to allow fluid flow. The clear proximal section is not reinforced to allow clamping; the proximal end terminates in a 3/8" barbed connector for standard cardiopulmonary bypass tubing
In the Easyflow Duo models the device has a 0.035" hole punched in the very distal tip of the cannula and the obturator has an approx. 0.035" internal diameter central lumen to accommodate a 0.035" guide wire used to drive the cannula along the vessel till its final placement. Once the cannula is positioned onto the vessel the surgeon removes the obturator from the cannula tubing and connects it to the arterial line.
The Easyflow is packaged as sterile stand-alone device while the Easyflow Duo is packaged into a tray containing also a Guidewire, a Needle and Vascular Dilators in order to have a kit for vascular access packaged together with the cannula.
The Easyflow & Easyflow Duo are the modified version of the disposables currently marketed as the Easyflow cannula (K060101) and the Easyflow Duo cannula (K091542). Both modified and unmodified cannulae models are recommended for use as an arterial return cannula during cardiopulmonary bypass up to six hours
The provided FDA 510(k) Clearance Letter for the Easyflow Cannulae does not contain any information related to a study proving the device meets acceptance criteria for an AI/ML (Artificial Intelligence/Machine Learning) powered medical device.
This document describes a medical device (cannulae for cardiopulmonary bypass) which is a physical, single-use, non-toxic, non-pyrogenic fluid path device. The clearance is based on "substantial equivalence" to predicate devices, meaning it has similar indications for use and technological characteristics, with the primary change being the removal of phthalates (DEHP and DnHP) from the PVC materials.
Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria in the context of an AI/ML device, as this information is not present in the provided text.
The information you asked for, such as sample size, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth types, and training set details, are all highly specific to the validation of AI/ML algorithms, which are not relevant to the Easyflow Cannulae as described in this FDA clearance.
In summary, based on the provided document, there is no AI/ML component to the Easyflow Cannulae, and thus no such study or acceptance criteria as you've outlined would be applicable or present.
Ask a specific question about this device
(58 days)
This cannula is intended for use in venous drainage via the right atrium and inferior vena cava simultaneously during cardiopulmonary bypass surgery up to six hours or less.
The MC2™ Two-Stage Venous Cannula and MC2X™ Three-Stage Venous Cannula models feature wire wound polyvinyl chloride (PVC) bodies with side ports in the distal tip, a ported atrial basket drainage site located along the length of the cannula body, and a 3/8-inch (0.95 cm) to 1/2-inch (1.27 cm) connection site. The overall length of each cannula body is approximately 15¼ inch (38.7 cm). Insertion depth marks are provided to aid in positioning of the cannula. Each cannula is nonpyrogenic, is intended for single use, and has been sterilized using ethylene oxide.
This document is a 510(k) clearance letter for a medical device: the MC2™ Two-Stage Venous Cannula and MC2X™ Three-Stage Venous Cannula. It does not contain any information about an AI/ML-driven medical device, nor does it discuss acceptance criteria, test sets, ground truth establishment, or human reader studies related to AI performance.
The clearance is for a physical device used in cardiopulmonary bypass surgery, and the summary of performance data refers to pre-clinical bench testing related to material formulation changes, not algorithmic performance.
Therefore, I cannot provide the requested information based on the provided text. The prompt asks for details about AI/ML device performance, which are entirely absent from this 510(k) clearance.
Ask a specific question about this device
(28 days)
The Quantum Perfusion Dual Lumen Cannula is intended for use as a single cannula for both venous drainage and reinfusion of blood via the internal jugular vein during extracorporeal life support procedures for periods of maximum 6 hours.
Quantum Perfusion Dual Lumen Cannula product family are single use devices intended to be used in medical procedures providing cardiac and/or respiratory support up to 6 hours. Devices are characterized by two coaxial catheters and are designed to be coupled with an extracorporeal circuit for artificial oxygenation of the blood and carbon dioxide removal.
Devices exploit a single site insertion which reduces the invasiveness of extracorporeal procedures. It can be inserted via a percutaneous approach through the right internal jugular vein. Blood is drained through the lumen of the outer catheter and returned through the lumen of the inner catheter.
Devices feature multiple inflow openings in the outer catheter limiting the circulation of unoxygenated blood.
Devices are non-toxic, non-pyrogenic, sterilized by ethylene oxide and packaged in a double pouch. All the devices' surfaces in contact with blood are coated with a phosphorylcholine-based biocompatible material.
The provided document is a 510(k) summary for a medical device called the "Quantum Perfusion Dual Lumen Cannula 31F-V1." It does not describe an AI/ML powered device, therefore the requested information regarding acceptance criteria, study details, and human-AI interaction is not applicable. The document focuses on regulatory approval for an updated version of an existing medical cannula.
Here's a breakdown of why this information isn't available in the document:
- Device Type: The device is a physical medical cannula used in cardiopulmonary bypass, not an AI/ML software or algorithm.
- Approval Type: The document is a 510(k) premarket notification, which is for demonstrating substantial equivalence to a legally marketed predicate device, not for a novel AI/ML application that would require extensive performance studies as requested.
- Performance Data: The "Performance Data" section explicitly states "NON-CLINICAL TESTING" and lists updates to labeling and performance tests according to recognized consensus standards for the physical device. It also states "No animal studies have been performed" and "No clinical data have been included."
Therefore, I cannot provide the requested information for acceptance criteria and study proving device meets said criteria as it pertains to AI/ML devices.
Ask a specific question about this device
(29 days)
The Aortic root cannulae are intended for use in adult and pediatric patients to cannulate the aortic root for cardioplegic solution delivery into the coronary arteries and for venting of the heart during cardiopulmonary bypass surgery for periods of up to six hours.
The R501-R502 Aortic Root Cannula (hereinafter referred to as R501-R502 cannulae) are aortic root cannulae intended to be used to cannulate the aortic root, to vent and to deliver cardioplegic solution during heat bypass operations.
The device is available in two styles: with a vent line (R502) and without a vent line (R501). Both models consist of a flexible Polyvinylchloride (PVC) tubing body, a PVC flanged tip and luer connector. The introducer is made of stainless steel with a PVC obturator. The vent line model have an additional feature constituted by a vent line provided as integral part of the device. This attribute makes it possible to vent and also introduce cardioplegic solutions using the same device. These cannulae are available in a range of sizes as per table below:
| Model | Tip size (internal diameter) | Product designation |
|---|---|---|
| R501 (aortic root without vent line) | 1.5 mm | R501-15 |
| 2.0 mm | R501-20 | |
| 2.6 mm | R501-26 | |
| R502 (aortic root with vent line) | 1.5 mm | R502-15 |
| 2.0 mm | R502-20 | |
| 2.6 mm | R502-26 |
The R501-R502 cannulae are a modified version of the currently marketed R501-R502 Aortic Root cannulae (K200612).
The provided text is a 510(k) premarket notification summary for a medical device, the R501-R502 aortic root cannulae. This document focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving that the device meets a specific set of acceptance criteria through a clinical study or a multi-reader, multi-case (MRMC) study.
Therefore, many of the requested details, such as acceptance criteria, expert ground truth, adjudication methods, MRMC studies, and standalone performance, are not applicable to this type of regulatory submission. The primary method of demonstrating safety and effectiveness in this document is comparison to a predicate device via non-clinical performance data and a statement of substantial equivalence.
Here's an explanation based on the provided text:
1. A table of acceptance criteria and the reported device performance:
This document does not contain a table of explicit acceptance criteria with corresponding device performance results in the way one would expect from a clinical trial or a formal validation study for a novel device. Instead, the performance is demonstrated through a comparison to a predicate device and adherence to applicable voluntary standards.
The non-clinical performance data section states: "The device passed all the testing in accordance with national and international standards." This implies that the acceptance criteria for these tests were meeting the requirements of the standards, but the specific numerical criteria and results are not detailed.
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not applicable. The submission relies on non-clinical performance data (testing) and a comparison to a predicate device. There is no "test set" in the context of clinical data or AI model evaluation.
- Data Provenance: Not applicable. As there are no clinical studies or AI evaluations described, there is no "data provenance" in terms of subject demographics or study design. The non-clinical testing was conducted by Sorin Group Italia s.r.l.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. No ground truth was established by experts for a test set in the context of this submission, as it focuses on substantial equivalence through non-clinical testing and comparison to a predicate.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable. No test set requiring expert adjudication is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC comparative effectiveness study was done. This is explicitly stated in the document under "Clinical Performance Data": "No clinical testing was conducted in support of the R501-R502 Cannulae, as the indications for use and technical characteristics are equivalent to those of the predicate device, which has been on the market for several years with proven safety and efficacy of use."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This device is a physical medical device (cannula), not an AI algorithm. Therefore, no standalone algorithm performance was assessed.
7. The type of ground truth used:
Not applicable. As no clinical studies or AI performance evaluations are described, the concept of "ground truth" (e.g., expert consensus, pathology, outcomes data) does not apply here. The submission relies on the established safety and efficacy of the predicate device and the results of non-clinical engineering and materials testing.
8. The sample size for the training set:
Not applicable. This is a physical medical device. There is no "training set" in the context of machine learning or AI models.
9. How the ground truth for the training set was established:
Not applicable. There is no training set for this device.
Ask a specific question about this device
Page 1 of 23