Search Filters

Search Results

Found 25 results

510(k) Data Aggregation

    K Number
    K232788

    Validate with FDA (Live)

    Device Name
    INSPIRA ART100
    Date Cleared
    2024-05-24

    (256 days)

    Product Code
    Regulation Number
    870.4380
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INSPIRA™ ART100 system is intended for use in an extracorporeal perfusion circuit to pump blood during short duration cardiopulmonary bypass procedures lasting six hours or less.

    Device Description

    The INSPIRA™ ART100 system is a device for pumping blood without direct contact in an extracorporeal blood circulation circuit. The device is prescribed (Rx) by specialists to be used in Cardiopulmonary Bypass (CPB) procedures. The INSPIRA™ ART100 system consists of the following components: Controller Unit, Pump Drive Unit, Manual Pump Drive Unit, Blood Flow and Bubble Detection Sensor, Temperature Sensor, Cable adapters for Pressure Sensors, Cable adapters for Temperature Sensor, Oxygenator and Pump Drive Holders. The system is compatible with a Non-Roller Centrifygal Pump (Capiox iCP Centrifyqal Pump). The Inspira Pump Drives are magnetically coupled exclusively with the Capiox iCP Centrifugal Pump, designed and manufactured by Terumo Cardiovascular Systems Corporation, and can be supplied by Inspira. The INSPIRA™ ART100 system is intended to be used in conjunction with other previously 510(k)-cleared devices and sterile accessories such as cannulas, catheters, oxygenators, luer connectors, pressure sensors, supplied by hospital users or other manufacturers.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the INSPIRA™ ART100 device. While it details the device's technological characteristics, intended use, and comparison to a predicate device, it does not include specific acceptance criteria for performance metrics (e.g., accuracy, sensitivity, specificity) or a detailed study proving the device meets these criteria in the context of an AI/human-in-the-loop system.

    The device, Inspira Art100, is a cardiopulmonary bypass pump speed control system. The testing performed is related to ensuring the device functions as intended and meets safety standards, rather than evaluating specific clinical performance metrics typical of AI/ML-driven diagnostic or prognostic devices.

    Therefore, many of the requested details about acceptance criteria and performance studies for an AI device cannot be extracted from this document.

    However, I can extract information regarding non-clinical tests performed to demonstrate safety and effectiveness for this type of medical device:

    1. Table of Acceptance Criteria and Reported Device Performance
    The document states: "All testing met predetermined acceptance criteria." However, the specific acceptance criteria and detailed reported performance values for each non-clinical test are not explicitly provided in a table format. The reported performance is implicitly that the device passed these tests.

    Acceptance Criteria CategoryReported Device Performance
    Software Verification and ValidationMet predetermined acceptance criteria.
    Functional Design Verification and ValidationMet predetermined acceptance criteria.
    Electrical Safety, EMI/EMCMet predetermined acceptance criteria and certified by accredited third-party laboratories to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8.
    Interoperability Evaluation (with accessories/compatible devices)Met predetermined acceptance criteria.
    Reliability TestingMet predetermined acceptance criteria.
    Simulated Use TestingMet predetermined acceptance criteria.
    Cleaning ValidationMet predetermined acceptance criteria.
    Packaging and Shipping TestingMet predetermined acceptance criteria and certified to ASTM D4332-22, ASTM D4169-22.
    Usability/Human FactorsCompleted human factors validation study in accordance with IEC 62366-1, ISO 14971, and FDA guidance. All residual use-error risks evaluated and minimized; overall residual risk acceptable.
    Risk ManagementIn accordance with ISO 14971:2019.
    Software Life Cycle ProcessesIn accordance with IEC 62304:2006/AMD 1:2015.
    Battery Safety/TransportCertified to IEC / UL 62133-2:2020 Ed.1 / CSA C22.2 and UN38.3.

    2. Sample size used for the test set and the data provenance: Not applicable. The tests performed are non-clinical, bench, and simulated use tests for a pump speed control device, not a diagnostic or prognostic AI system using patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically defined for AI models (e.g., disease presence) is not relevant for the type of device and testing described. Expert involvement would be in the design, testing, and validation of the device's engineering and safety aspects.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, this type of study was not done. The device is a cardiopulmonary bypass pump speed control system, not an AI assistance tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The document describes standalone device testing (bench tests) for its functional and safety performance, but not in the context of an AI algorithm vs. human performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the non-clinical tests described, "ground truth" refers to engineering specifications, performance standards (e.g., for pump speed, flow rate accuracy), and compliance with recognized industry standards (e.g., IEC, ISO, ASTM).

    8. The sample size for the training set: Not applicable. This device is not an AI/ML model that requires a training set of data.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230698

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-04-18

    (36 days)

    Product Code
    Regulation Number
    870.4380
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Anivia SG1000 Pump Console is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass (up to 6 hours).

    Device Description

    The Anivia SG1000 Pump Console is a device for pumping blood without direct contact in an extracorporeal blood circulation circuit. The device is prescribed (Rx) by specialists to be used in Cardiopulmonary Bypass (CPB) procedures. The Anivia SG1000 Pump Console consists of the following modules: Display & Control Module, Pump Driver Module, Flow Bubble Sensor Module, Backup Battery Module, and Cart Module. It is intended to be used in conjunction with other previously 510(k)-cleared devices and sterile accessories such as cannulas, catheters, oxygenators, luer connectors and single use centrifugal pump heads, supplied by hospital users or other manufacturers.

    AI/ML Overview

    Here's the information regarding the acceptance criteria and study proving the device meets them, based on the provided text:

    Device Name: Anivia SG1000 Pump Console (K230698)
    Regulation Name: Cardiopulmonary Bypass Pump Speed Control


    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document does not explicitly list "acceptance criteria" in a quantitative performance table format. Instead, it describes various non-clinical tests conducted and states that "All testing met predetermined acceptance criteria." The table below summarizes the types of tests performed and the general outcome.

    Acceptance Criteria CategoryReported Device Performance / Outcome
    Software PerformanceSoftware verification and validation testing met predetermined acceptance criteria.
    Functional DesignFunctional design verification and validation testing met predetermined acceptance criteria.
    Electrical SafetyElectrical safety testing met predetermined acceptance criteria and the IEC 60601-1:2005/AMD1:2012 standard.
    Electromagnetic Compatibility (EMI/EMC)Electromagnetic interference and compatibility (EMI/EMC) testing met predetermined acceptance criteria and the IEC 60601-1-2:2014 standard.
    InteroperabilityInteroperability evaluation with specified accessories met predetermined acceptance criteria.
    ReliabilityReliability testing met predetermined acceptance criteria.
    Simulated UseSimulated use testing met predetermined acceptance criteria.
    Cleaning ValidationCleaning validation met predetermined acceptance criteria and the ISO 17664-2:2021 standard.
    Battery PerformanceBackup Battery Module (LiFePO4) provides a minimum of 1 hour and up to 3 hours of backup power depending on speed and flow. Certified to IEC 62133-2:2017/AMD1:2021 and UN38.3.
    Packaging & ShippingPackaging and ship testing met predetermined acceptance criteria and the ASTM D4169:2022, DC13, Level I standard.
    Risk ManagementComplies with ISO 14971:2019.
    UsabilityUsability testing was conducted. (Outcome: met predetermined acceptance criteria, implied by "All testing met predetermined acceptance criteria.")
    AlarmsVisual and auditory alarms are present for abnormal conditions (Speed, Flow Rate, Back Flow, Bubble, Pressure, Temperature).
    Pump SpeedVaries by pump driver module, e.g., 0-4500 RPM (Medtronic BPX-80 compatible), 0-5000 RPM (Maquet Rotaflow® RF-32 compatible), 0-4000 RPM (Medtronic Affinity CP compatible), 0-3500 RPM (LivaNova Revolution® compatible).
    Blood Flow RateDependent on external circuit, up to 0 – 9.9 L/min.
    External Standards ComplianceCertified to various IEC, ISO, UL, UN, and ASTM standards (listed in detail in section 9.3).

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: The document does not specify a distinct "test set" sample size in terms of number of patient cases or images, as this is a device for regulating pump speed in cardiopulmonary bypass, not an AI/imaging diagnostic device. The testing involved various engineering and systems-level tests.
    • Data Provenance: Not applicable in the context of typical AI/imaging studies. The testing described is performance testing of the device itself (hardware, software, electrical, mechanical, reliability) against engineering standards and internal criteria. There is no mention of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The device is a cardiopulmonary bypass pump console, not a diagnostic tool requiring expert interpretation of medical data (like images or clinical reports) to establish ground truth for a test set. The "ground truth" here is adherence to engineering specifications and performance standards, validated through bench testing, not expert consensus on clinical cases.


    4. Adjudication method for the test set:

    Not applicable, as there is no "test set" in the context of clinical interpretation subject to adjudication. The performance was assessed against predetermined engineering and functional criteria.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a cardiopulmonary bypass pump console, which is an electromechanical medical device, not an AI-assisted diagnostic or interpretation tool for human readers. No MRMC study was performed.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This is not an AI algorithm. The device operates with human involvement (physicians, perfusionists, etc.) for setup, monitoring, and control during cardiopulmonary bypass procedures. The "standalone" performance refers to the device's functional integrity as a machine.


    7. The type of ground truth used:

    The "ground truth" for this device's performance is established by:

    • Engineering specifications and design requirements.
    • International consensus standards (e.g., IEC 60601 series, ISO 14971).
    • Internal predetermined acceptance criteria for functional, electrical, mechanical, software, reliability, and safety tests.

    8. The sample size for the training set:

    Not applicable. This device does not involve a "training set" in the context of machine learning or AI models. It's an electromechanical medical device.


    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221491

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-02-03

    (256 days)

    Product Code
    Regulation Number
    870.4380
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Anivia SG1000 Pump Console is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass (up to 6 hours).

    Device Description

    The Anivia SG1000 Pump Console is a device for pumping blood without direct contact in an extracorporeal blood circulation circuit. The device is prescribed (Rx) by specialists to be used in Cardiopulmonary Bypass (CPB) procedures. The Anivia SG1000 Pump Console consists of the following modules: Display & Control Module, Pump Driver Module, Flow Bubble Sensor Module, and Backup Battery Module. It is intended to be used in conjunction with other previously 510(k)-cleared devices and sterile accessories such as cannulas, catheters, oxygenators, luer connectors and single use centrifugal pump heads, supplied by hospital users or other manufacturers.

    AI/ML Overview

    The provided text describes the Anivia SG1000 Pump Console, a device intended to pump blood through an extracorporeal bypass circuit for cardiopulmonary bypass procedures (up to 6 hours). The 510(k) summary focuses on demonstrating substantial equivalence to a predicate device (Medtronic Bio-console® 560 Extracorporeal Blood Pumping Console, K131964).

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list a table of acceptance criteria with specific numerical targets and corresponding reported device performance values for the Anivia SG1000 Pump Console. Instead, it broadly states that "All testing met predetermined acceptance criteria." The performance characteristics are described in comparison to the predicate and reference devices in the "Technological Characteristics" table, implying that the acceptance criteria for these characteristics are met by being "Same as predicate" or by having differences that "do not raise different questions of safety and effectiveness."

    Here's a breakdown of the performance characteristics provided, and where the device's performance aligns with or is considered equivalent to the predicate:

    Acceptance Criteria (Implied from Predicate/Reference)Reported Device Performance (Anivia SG1000 Pump Console)
    Indications for Use: Pump blood through extracorporeal bypass circuit for up to 6 hours.Met: Intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass (up to 6 hours). (Same as predicate)
    Duration of Use: Up to 6 hours.Met: Up to 6 hours (limited by disposables, not limited by Console electro-mechanical modules). (Same as predicate)
    Intended Use: Pump Speed Control, CPB Machine Console.Met: Pump Speed Control, CPB Machine Console (not including sterile, blood-contacting accessories). (Same as predicate)
    Intended Users: Cardiopulmonary physicians, perfusionists, trained physician assistants.Met: Cardiopulmonary physicians, perfusionists, trained physician assistants. (Same as predicate)
    Intended Use Environment: Cardiopulmonary procedure rooms, intensive care units.Met: Cardiopulmonary procedure rooms, intensive care units. (Same as predicate)
    Intended Patients: As prescribed by cardiopulmonary specialists.Met: As prescribed by cardiopulmonary specialists. (Same as predicate)
    Compatible Disposable Centrifugal Pump Heads: Medtronic BPX-80, BP-50, RotaFlow® RF-32.Met: Medtronic BPX-80, BP-50 centrifugal pump heads (w/ Pump Driver Module SG1000-PDM-001); RotaFlow® RF-32 centrifugal pump heads (w/ Pump Driver Module SG1000-PDM-002). (Matches/expands on predicate and reference)
    Centrifugal Pump Driver: Non-contact, magnetically coupled.Met: Non-contact, magnetically coupled. (Same as predicate)
    Control Function: Speed Control.Met: Speed Control, Flow Control (Stability), Mechanical Knob and Touch Display. (Similar to predicate's speed control, adds flow control and touch display, considered not to raise new safety/effectiveness questions).
    Pump Speed (RPM): 0 – 4500 (predicate), 0 – 5000 (reference).Met: 0 – 4500 for CS-0210-ASY-00002 (Medtronic BPX-80 compatible); 0 – 5000 for CS-0210-ASY-00010 (Maquet Rotaflow RF-32 compatible). (Matches/exceeds predicate)
    Visual and Auditory Alarms on Abnormal Conditions: Yes, preset limits for Speed, Flow Rate, Back Flow, Bubble, Pressure, Temperature.Met: Yes, preset limits: Speed, Flow Rate, Back Flow, Bubble, Pressure, Temperature. (Same as predicate)
    Blood Flow Rate (L/min): Up to 0 – 9.9 L/min.Met: Dependent on external circuit, up to 0 – 9.9 L/min. (Same as predicate)
    Interface to Blood Flow Sensor: Yes (Qty 1).Met: Yes (Qty 1). (Same as predicate)
    Air Bubble Detector: Yes (Qty 1), integrated with Blood Flow Sensor.Met: Yes (Qty 1), integrated with Blood Flow Sensor. (Integrated with BFS, which is implicitly similar to predicate's separate detector for safety/effectiveness)
    Blood Flow and Bubble Detector Sensor Technology: Non-contact, various methods.Met: Non-contact, ultrasound Clamp-On around blood tube. (Different technology than predicate's electrical conductivity, but implicitly determined not to raise new safety/effectiveness questions, aligning with reference device technology).
    Pressure Sensors: Yes (Qty 2, external), previously 510(k)-cleared.Met: Yes (Qty 2, external), previously 510(k) cleared accessories manufactured by third parties. (Same as predicate)
    Temperature Sensors: Yes (Qty 2, external), previously 510(k)-cleared.Met: Yes (Qty 2, external), previously 510(k) cleared accessories manufactured by third parties. (Same as predicate)
    Power Input: Universal 90-264 VAC/50-60 Hz, 2.5A / 1.3A, up to 250 W.Met: Universal 90-264 VAC/50-60 Hz, 2.5A / 1.3A, up to 250 W. (Stated as "Same as predicate – single universal power supply, instead of different models, no effect on safety and effectiveness.")
    Backup Battery: Lead acid (predicate), NiCad (reference).Met: LiFePO4. (Improvement in technology stated to "not raise question on safety and effectiveness.")
    Backup Battery Capacity: 24 VDC (predicate), 24 VDC, 5 AH, 120 WH (reference).Met: 25.6 VDC, 12 AH, 307 WH, minimum 1 hour, up to 3 hours depending on speed and flow. (Stated as "Same as predicate" in terms of function, despite numerical differences, indicating functional equivalence for backup power duration).
    Backup Pump: Hand-crank.Met: Yes, backup electrical Pump Driver Module on standby, and hand-crank. (Improvement over predicate by adding an electrical backup, stated as "Same as predicate" for safety/effectiveness).
    Pump Motor Technology: Brushless DC motor.Met: Brushless DC motor. (Same as predicate)
    Display Screen Size: 26 cm (10.4") (predicate).Met: 31 cm (12.1"). (Difference stated to have "no impact on safety and effectiveness.")
    Dimensions/Weight: Varied.Met: Dimensional and weight differences in Display & Control Module (smaller/lighter than predicate) are stated to have "no impact on safety and effectiveness" due to newer technology.
    Non-Clinical Tests: Software verification, functional design, electrical safety, EMI/EMC, interoperability, reliability, simulated use, cleaning validation, packaging, shipping.Met: All listed tests were conducted and "All testing met predetermined acceptance criteria." Compliance with numerous consensus standards (e.g., IEC 60601 series, ISO 14971, IEC 62304, IEC 62366, UL 2054, etc.) was also achieved.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of clinical data for algorithmic performance, as this device does not involve AI or machine learning that would typically use such a test set. The non-clinical tests involved "system bench testing, mechanical testing, electrical safety and electromagnetic interference and compatibility testing, software testing, reliability, and usability testing." The sample size for these engineering and performance tests is not quantified (e.g., number of devices tested, number of test runs).

    Data provenance is not applicable since no clinical data or patient data was used to assess the device's performance in this 510(k) submission. All tests mentioned are non-clinical, benchtop, or simulated use.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device does not involve a "ground truth" derived from expert review or clinical annotations for an algorithm. The evaluation relies on engineering tests and compliance with recognized standards.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    Not applicable, as there is no "test set" in the context of human-reviewed data where adjudication would be necessary.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    Not applicable. This device is a cardiopulmonary bypass pump console, not an AI/ML-driven diagnostic or interpretative tool that would involve human readers or an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is electro-mechanical hardware with software controls, not a standalone algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    Not applicable. For this type of medical device (a pump console), "ground truth" is established through engineering specifications, validated test methods (e.g., measuring pump speed, flow rate, alarm accuracy), and compliance with international consensus standards. There is no biological or imaging "ground truth" involved.

    8. The Sample Size for the Training Set

    Not applicable. This device does not involve machine learning or AI, and therefore does not have a "training set." The software embedded in the device undergoes traditional software verification and validation.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192838

    Validate with FDA (Live)

    Date Cleared
    2020-04-13

    (193 days)

    Product Code
    Regulation Number
    870.4380
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Quantum Centrifugal Drive is a cardiopulmonary bypass speed control device indicated for use for exclusively with the Quantum PureFlow Centrifugal Blood Pump for speedcontrolled pumping through the extracorporeal circuit for typical durations of six hours or less.

    Device Description

    The Quantum Centrifugal Drive is a cardiopulmonary speed control device intended to be used for pumping arterial blood through the cardiopulmonary bypass circuit. The Quantum Centrifugal Drive is available in two models - High Pressure and Low Pressure.

    The Quantum Centrifugal Drive is designed to work with the Quantum Pump Manager software application, as part of the Quantum Pump Console (K173834). The Quantum Centrifugal Drive is magnetically coupled exclusively with Quantum Pureflow Centrifugal Blood Pumps, designed and manufactured by Qura s.r.l. The operation of the Quantum Centrifugal Drive is managed from the Quantum Workstation (K163657).

    AI/ML Overview

    The provided document, an FDA 510(k) summary for the Quantum Centrifugal Drive, focuses on establishing substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria for an AI/device's diagnostic or predictive capabilities.

    Therefore, the document does not contain the information requested regarding:

    1. A table of acceptance criteria and reported device performance related to diagnostic accuracy.
    2. Sample sizes and data provenance for a test set.
    3. Number and qualifications of experts for ground truth establishment.
    4. Adjudication methods for a test set.
    5. MRMC comparative effectiveness study results.
    6. Standalone algorithm performance.
    7. Type of ground truth used (e.g., pathology, outcomes data).
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    The document states:

    • "No animal testing was submitted to support the substantial equivalence of the Quantum Centrifugal Drive to the Sorin/LivaNova Stöckert Centrifugal Pump."
    • "The following non-clinical testing was performed to support the substantial equivalence of the Quantum Centrifugal Drive to the legally marketed predicate:
      • Electrical safety
      • Electromagnetic compatibility (EMC)
      • Electrosurgery interference
      • Hardware testing
      • Software verification and validation"
    • "No clinical data were submitted to support the substantial equivalence of the Quantum Centrifugal Drive to the Sorin/LivaNova Stöckert Centrifugal Pump."

    This indicates that the substantial equivalence was primarily based on non-clinical engineering and software validation tests (electrical safety, EMC, electrosurgery interference, hardware, and software V&V) rather than a study evaluating diagnostic performance parameters against ground truth established by experts or other clinical gold standards. The device is a "Cardiopulmonary bypass pump speed control," which is a mechanical device, not typically a diagnostic AI tool that would involve ground truth derived from expert consensus or pathological findings.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200306

    Validate with FDA (Live)

    Date Cleared
    2020-03-06

    (29 days)

    Product Code
    Regulation Number
    870.4380
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc.)

    The CentriMag Return (Arterial) Cannula is indicated for use as an arterial return cannula with an extracorporeal bypass circuit for extracorporeal circulatory support for periods up to six hours.

    The CentriMag Drainage (Venous) Cannula is indicated for use with an extracorporeal bypass circuit for extracorporeal circulatory support for periods up to six hours.

    Device Description

    The CentriMag™ Circulatory Support System, hereafter referred to as the CentriMag System, was designed to provide temporary mechanical circulatory support. The CentriMag system in the United States (US) is indicated as a component of an extracorporeal bypass circulatory support circuit for use during cardiopulmonary bypass (CPB) surgery (up to 6 hours). The CentriMag System provides circulatory assistance for patients in acute hemodynamic compromise, a population whose treatment options are limited.

    The CentriMag System is composed of:

    • CentriMag Primary Console
    • CentriMag Motor
    • CentriMag Blood Pump
    • CentriMag Flow Probe
    • Mag Monitor (optional)

    The Thoratec CentriMag Return (Arterial) Cannula Kit consists of the return (arterial) cannula and several accessories used in the surgical placement procedure. The cannula and all kit accessories are sterile, single-use, disposable devices. The cannula kit is designed for use with the Thoratec CentriMag® Extracorporeal Blood Pumping System.

    The Thoratec CentriMag® Drainage (Venous) Cannula Kit consists of the drainage (venous) cannula and several accessories used in the surgical placement procedure. The cannula and all kit its accessories are sterile, single use, disposable devices. The cannula kit is designed for use with the Thoratec CentriMag® Extracorporeal Blood Pumping System.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (CentriMag Acute Circulatory Support System, CentriMag Return (Arterial) Cannula Kit, and CentriMag Drainage (Venous) Cannula Kit).

    Based on the content of the document, the device did not undergo a study to prove its performance against acceptance criteria in the typical sense of a new or modified device requiring such evidence. This 510(k) submission is a "Special 510(k)" which implies that the changes being made are minor and do not affect the device's fundamental performance, design, or intended use. Specifically, the document states:

    • "The technological characteristics of the CentriMag Acute Circulatory Support System, CentriMag Return (Arterial) Cannula Kit and CentriMag Drainage (Venous) Cannula Kit have not changed from the predicate devices."
    • "Components of the CentriMag Acute Circulatory Support System, CentriMag Return (Arterial) Cannula Kit and CentriMag Drainage (Venous) Cannula Kit are not affected by the labeling changes proposed in this Special 510(k) premarket notification. The labeling changes make the contraindications consistent among all labeling, add and/or strengthen warnings and precautions, and generally align and update the content for consistency."
    • "The performance characteristics of the CentriMag Acute Circulatory Support System, CentriMag Return (Arterial) Cannula Kit and CentriMag Drainage (Venous) Cannula Kit have not changed. The labeling changes included in this 510(k) are for the Thoratec (now Abbott's) devices (predicates listed above) and do not require performance data to evaluate the change."
    • "Clinical testing was not necessary for the labeling changes stated herein since the changes do not impact the design and performance of the devices."

    Therefore, I cannot provide a table of acceptance criteria, reported device performance, or details of a study with sample sizes, expert ground truth, adjudication methods, or MRMC studies, as these types of studies were explicitly stated as not necessary for this specific 510(k) submission.

    The document argues for substantial equivalence primarily based on the lack of change in technological characteristics and performance from legally marketed predicate devices (K191557, K152161, and K152190).

    Ask a Question

    Ask a specific question about this device

    K Number
    K191557

    Validate with FDA (Live)

    Date Cleared
    2019-07-10

    (28 days)

    Product Code
    Regulation Number
    870.4380
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).

    Device Description

    The CentriMag™ Circulatory Support System, hereafter referred to as the CentriMag System, was designed to provide temporary mechanical circulatory support. To date, the CentriMag system in the United States (US) is indicated as a component of an extracorporeal bypass circulatory support circuit for use during cardiopulmonary bypass (CPB) surgery (up to 6 hours). The CentriMag System provides circulatory assistance for patients in acute hemodynamic compromise, a population whose treatment options are limited.

    The CentriMag System is composed of:

    • CentriMag Primary Console
    • CentriMag Motor
    • CentriMag Blood Pump
    • CentriMag Flow Probe
    • Mag Monitor (optional, not shown)

    The CentriMag Motor is a reusable, non-sterile component of the CentriMag Acute Circulatory Support System. The CentriMag Motor holds the blood pump and drives the impeller inside the blood pump. The motor turns the magnet (and impeller) within the blood pump (Full MagLev™ technology) at a speed that is set on the console by the user. When the impeller is rotated, a pressure gradient develops between the center and outside edge of the pump, causing blood to flow from the inflow to the outflow port of the pump. The amount of flow through the pump depends on the speed of the impeller, and the difference between the inlet and outlet pressures.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a design modification to the CentriMag™ Acute Circulatory Support System, specifically a redesigned motor cable with enhanced bend protection. This is a medical device, and the submission focuses on proving the substantial equivalence of the modified device to its predicate, rather than establishing clinical effectiveness for a new-to-market device or an AI algorithm.

    Therefore, the typical metrics and study designs used for AI/algorithm performance (e.g., sensitivity, specificity, MRMC studies, expert consensus for ground truth of an AI algorithm) do not apply in this context. The acceptance criteria and "study" are focused on mechanical and electrical performance to ensure the modified cable is at least as safe and effective as the original, and does not introduce new safety concerns.

    Here's a breakdown of the requested information based solely on the provided text, recognizing its limited scope for AI-specific metrics:


    Acceptance Criteria and Device Performance (for a Medical Device Component Modification)

    The acceptance criteria for this device modification are focused on maintaining or improving the mechanical and electrical integrity of the motor cable compared to the predicate device.

    Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance (as stated in the text)
    Motor Cable Endurance: Durability and functionality under simulated severe handling to meet expected service life. Integrity of outer surface after bending.Passed: Successfully passed the cable endurance (bend) test protocol. The redesigned robust motor cable with extended bend relief successfully passed (no failures), whereas the original cable design was less robust under high stress. Visual inspection showed the integrity of the outer surface of the motor cable or bend protection in the area of bending. This demonstrates meeting the expected service life under worst-case conditions.
    Motor Cable Kink Test: Tolerance to excessive kinking/bending with small radius for targeted expected service life. Absence of defects after kinking.Passed: Successfully passed the kink test protocol. Visual inspection showed no defects on the surface of the motor cable or bend protection in the area of bending. This proves it can tolerate excessive kinking or bending with small radius for the targeted expected service life.
    Electrical / Safety Compliance: Compliance with relevant electrical safety standards.Passed: Compliance to IEC 60601-1, Medical electrical equipment, Part 1: General requirements for basic safety and essential performance for the redesigned CentriMag Motor cable and new bend protection assembly has been confirmed, and it is therefore, substantially equivalent. The design modifications enhance cable flexibility and durability, and reduce the impact of rough handling and the occurrence of cable breaks and short circuiting, indicating improved electrical integrity and safety over the predicate.
    Maintain Substantial Equivalence: No new safety or effectiveness concerns compared to predicate.Met: "The redesigned CentriMag Motor cable assembly with longer bend protection has indication for use, technological characteristics, and performance an characteristics which are substantially equivalent to the predicate CentriMag Motor cable. Due to this, the redesigned CentriMag Motor cable structure with enhanced bend protection raises no new safety or effectiveness concerns." "The redesigned Motor cable offers enhanced cable flexibility and durability and improved protection from bending and kinking at the bend protection over the current commercial cable design."

    Study Details (Bench Testing for a Device Component Modification):

    1. Sample sizes used for the test set and data provenance:

      • Sample Size: The text states "Comparison bench testing was performed" and refers to "both designs" (redesigned and current/original cable designs). It does not specify the number of individual cables or test repetitions for each test (Endurance and Kink test).
      • Data Provenance: The testing was carried out by Thoratec (now part of Abbott) as part of their 510(k) submission, implied to be internal company testing. The country of origin of the data is not explicitly stated beyond the company's US address (Pleasanton, CA). This was not a human clinical study, but rather laboratory bench testing. The term "retrospective or prospective" doesn't directly apply here; it's a controlled bench test evaluating a manufactured component.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable (N/A) / None specified: For these types of mechanical and electrical bench tests, "ground truth" is established by direct measurement and observation of physical parameters, not by expert interpretation in the way it is for diagnostic imaging or clinical outcomes. Engineering specifications and industry standards (like IEC 60601-1) serve as the "truth" against which performance is measured.
      • Qualifications of personnel conducting the tests are not provided, but would typically be engineers or technicians trained in relevant mechanical and electrical testing protocols.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not Applicable (N/A): Adjudication methods like 2+1 or 3+1 refer to consensus-building among human readers for subjective data interpretation (e.g., reading medical images). This was a series of objective bench tests with pass/fail criteria based on physical performance and visual inspection. Any "adjudication" would be part of standard quality control and engineering verification processes, not a multi-reader review.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, Not Applicable (N/A): This was a 510(k) for a physical medical device component (motor cable) and not an AI algorithm. Therefore, an MRMC study comparing human reader performance with or without AI assistance was not conducted, nor would it be relevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No, Not Applicable (N/A): This is a physical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Engineering Specifications and Industry Standards: The "ground truth" for these tests consisted of predefined pass/fail criteria derived from engineering design specifications and compliance with recognized international standards (e.g., IEC 60601-1). Performance data (e.g., integrity of cable after specified bends, absence of short circuits) formed the direct evidence.
    7. The sample size for the training set:

      • Not Applicable (N/A): This device modification did not involve machine learning or AI, so there was no "training set."
    8. How the ground truth for the training set was established:

      • Not Applicable (N/A): As there was no training set, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131964

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-07-24

    (27 days)

    Product Code
    Regulation Number
    870.4380
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medtronic centrifugal blood pumping system is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass procedures (up to 6 hours).

    Device Description

    The Bio-Console® 560 Extracorporeal Blood Pumping Console is an electromechanical, software-controlled device intended to allow the user to control a Bio-Pump® or Affinity® CP centrifugal blood pump that is used to return the blood to the body of a patient on cardiopulmonary bypass. As part of the blood pumping circuit, the patient's blood is typically pumped through an oxygenator and a filter on its return path to the patient. The Bio-Console 560 is a reusable non-sterile device.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Bio-Console® 560 Extracorporeal Blood Pumping Console:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Maintain the same intended use as the predicate device.The Bio-Console 560 has the same intended use as the previously cleared Bio-Console 560.
    Maintain the same hardware design as the predicate device.The Bio-Console 560 has the same hardware design as the previously cleared Bio-Console 560.
    Maintain the same basic software as the predicate device.The Bio-Console 560 has the same basic software as the previously cleared Bio-Console 560.
    Incorporate software enhancements to correct anomalies.Software enhancements were incorporated to correct anomalies.
    Incorporate IEC 60601-1 (3rd Edition) compliant alarm and alert tones.IEC 60601-1 (3rd Edition) compliant alarm and alert tones were incorporated.
    Software verification and validation testing confirms the function of the device."Software verification and validation testing confirms the function of the Bio-Console 560..."
    Software-controlled functional characteristics are substantially equivalent to the predicate device."...and its software-controlled functional characteristics are substantially equivalent to the predicate device."
    All test data obtained satisfy documented product and performance specifications."All test data obtained satisfied the documented product and performance specification."

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the given document. The document mentions "software verification and validation testing" and "all test data obtained," but does not specify the sample size (e.g., number of test cases, number of runs) or the provenance (country of origin, retrospective/prospective nature) of the data used for this testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the given document. The document describes a technical evaluation of software enhancements, not a study involving human experts establishing ground truth for a test set in the context of diagnostic interpretation or similar medical AI applications.

    4. Adjudication Method for the Test Set

    This information is not provided in the given document. As there's no mention of human experts establishing ground truth with potential disagreements, an adjudication method would not be relevant in this context.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This submission is for software enhancements to an existing device and focuses on substantial equivalence based on technical verification and validation, not on evaluating human performance with or without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    Yes, a standalone performance evaluation was implicitly done. The "software verification and validation testing" assessed the "software-controlled functional characteristics" of the device in isolation, confirming its function and substantial equivalence. The enhancements relate to internal software corrections and alarm tones, which are tested as part of the device's inherent functionality.

    7. Type of Ground Truth Used

    The "ground truth" for this submission appears to be derived from documented product and performance specifications and the functional characteristics of the predicate device. The testing aimed to confirm that the updated software met these pre-defined specifications and maintained performance equivalent to the predicate.

    8. Sample Size for the Training Set

    This information is not provided in the given document. The device is described as an electromechanical, software-controlled device, but there is no indication that it utilizes machine learning or AI that would require a "training set" in the conventional sense of an AI model. The software enhancements are described as corrections and updates to existing functionality.

    9. How the Ground Truth for the Training Set was Established

    This information is not provided in the given document. As no "training set" is mentioned or implied for a machine learning model, the establishment of ground truth for such a set is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131179

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-07-18

    (84 days)

    Product Code
    Regulation Number
    870.4380
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 2nd Generation CentriMag Primary Console and Mag Monitor are indicated for use with the CentriMag Extracorporeal Blood Pumping System. The CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).

    Device Description

    The 2nd Generation CentriMag Primary Console is a microprocessor based device. The microprocessor generates the primary motor control signal, monitors system sensors, generates display outputs, interprets the inputs through the front keypad, provides alarm functions and handles the Mag Monitor interface. The Console is intended to be operated on single phase AC power; however, it also has a built-in rechargeable battery with a battery charger. The rechargeable battery is field replaceable. In addition, an optional external modular battery (UPS) can be used to power the CentriMag Primary Console. The human interface of the CentriMag Primary Console consists of a graphical screen to display data and system options and touch pads to change the system status. The 2nd Generation CentriMag Primary Console is intended to be used together with the Mag Monitor: however it can be also operated as a stand-alone unit.

    The Mag Monitor provides a redundant user interface containing a display and touch pads. The Mag Monitor is a non-sterile, reusable device that is designed to work only with the 2nd Generation CentriMag Primary Console. The Mag Monitor is a 12V DC-powered device and receives its power directly from 200 Generation CentriMag Primary Console via a power connector that mates with a connector on the back-panel of 2nd Generation CentriMag Primary Console. Based on the design of its power connector, the Mag Monitor cannot be plugged into a hospital AC power outlet. The Mag Monitor's core function is to provide multi-color alphanumerical and graphical displays of information it receives from the 200 Generation CentriMag Primary Console.

    The Mag Monitor is a processor based device, with a flat color screen (e.g. LCD). In addition, it is equipped with touch pad to allow the user to enter commands. In the hospital-setting configuration, the user is able to operate and monitor the performance of the CentriMag System through the Monitor. The Monitor can control the functions of the 2nd Generation CentriMag Primary Console, and therefore, of the CentriMag System. The Monitor only displays data and stores user commands without interacting directly with the primary motor control which is managed by the 2nd Generation CentriMag Primary Console. The Mag Monitor is connected to the 200 Generation CentriMag Primary Console through one cable, which includes data and power lines. Power is provided by the 200 Generation CentriMag Primary Console.

    The redesigned Mag Monitor is compatible with both the current 200 Generation Console and the redesigned 2nd Generation Console. However, when plugged into the current Console, the redesigned Mag Monitor will only operate when the Console is connected to AC power. It will not operate when the current Console is run on Battery power

    AI/ML Overview

    The provided text describes a 510(k) summary for the Thoratec 2nd Generation CentriMag Primary Console and Mag Monitor. This documentation focuses on demonstrating substantial equivalence to predicate devices for regulatory approval, rather than detailing a clinical study with specific acceptance criteria and performance metrics for an AI-powered diagnostic device.

    Therefore, the information required to answer your prompt, particularly regarding AI device performance, sample sizes for test/training sets, expert adjudication, MRMC studies, standalone performance, and ground truth establishment, is not present in the provided 510(k) summary.

    This document focuses on the device itself (a cardiopulmonary bypass pump console and monitor), its design, intended use, and a comparison of its technological characteristics and performance (functional testing, safety, environmental, etc.) against existing predicate devices. It explicitly states: "Clinical testing was not performed." (Section I. Clinical Performance).

    As such, I cannot create the requested table or provide details about AI device performance from this document. If you have a different document describing an AI-powered diagnostic device, please provide it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112225

    Validate with FDA (Live)

    Date Cleared
    2011-09-20

    (48 days)

    Product Code
    Regulation Number
    870.4380
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sorin Centrifugal Pump (CP5) is a cardiopulmonary bypass speed control device indicated for use exclusively with the Sorin Revolution®, for speedcontrolled pumping through the cardiopulmonary bypass circuit for typical durations of six hours or less.

    The CP5 has been qualified only for typical durations of six hours or less, appropriate to cardiopulmonary bypass procedures and has not been qualified through in vitro, in vivo, or clinical studies, for long term use as a bridge to transplant, pending recovery of the natural heart, or extracorporeal membrane oxygenation (ECMO) procedures.

    Device Description

    The Sorin CP5 is intended for use as a component part of or optional accessory for the Sorin heart lung machine consoles and with the Stöckert Air Purge (APC) the Borin heads they System. The Sorin CP5 consists of the drive unit, a mounting receptacle for the Dystein. The Cretaining key, splash protection, and a 3-joint mast holder with fast clamp connector. It integrates the functional capabilities of the SCP, the SCP Plus, and the Sorin Pulse Mode Control within a single device.

    The Sorin CP5 consists of hardware, firmware, and electronics that are used to drive the centrifugal pump for surgical procedures requiring cardiopulmonary bypass for typical durations of six (6) hours or less. The SCP drive unit uses magnetic coupling with the disposable centrifugal pump.

    The Sorin CP5 with the loaded centrifugal pump recirculates the contents of the extracorporeal circuit during cardiopulmonary bypass.

    The design, materials, and physical properties of the CP5 are essentially unchanged from that of the parent/predicate devices already reviewed and cleared by the FDA.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Sorin Centrifugal Pump (CP5). However, it explicitly states that no clinical testing and no comparative nonclinical testing served as the basis for substantial equivalence. The summary focuses on technological characteristics compared to predicate devices and internal validation.

    Therefore, the requested information regarding an acceptance criteria table, detailed study parameters, ground truth establishment, expert involvement, and MRMC studies cannot be extracted from the provided text, as such studies were not performed or are not reported.

    Here's what can be extracted based on the information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document states: "Testing of the Sorin Centrifugal Pump (CPS) demonstrates that the CP5 fulfills prospectively defined performance criteria and meets user needs." However, the specific quantitative or qualitative acceptance criteria and the corresponding reported device performance values are not detailed in this summary.

    2. Sample size used for the test set and the data provenance:

    Not applicable. The document states: "No comparative nonclinical testing served as the basis for substantial equivalence." and "No clinical testing served as the basis for substantial equivalence." The testing performed involved "integration testing of hardware and firmware, functional testing, testing of organized (IEC 60601-1 including deviations for the US) and associated tooling, success (IEC 60601-1-2), and validation testing (simulated use and in use)." This describes internal validation, not a test set based on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable, as no external test set with ground truth established by experts is described.

    4. Adjudication method for the test set:

    Not applicable, as no external test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a cardiopulmonary bypass centrifugal pump, not an AI-assisted diagnostic tool for human readers. No MRMC study was performed or is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device is a hardware component (centrifugal pump) with firmware, not a standalone algorithm. Performance testing was internal validation of the hardware and firmware.

    7. The type of ground truth used:

    Not applicable in the context of clinical or diagnostic performance. The "ground truth" for the internal validation would be the design specifications and expected functional behavior.

    8. The sample size for the training set:

    Not applicable, as this is not an AI/machine learning device with a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as this is not an AI/machine learning device with a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103468

    Validate with FDA (Live)

    Date Cleared
    2011-01-20

    (57 days)

    Product Code
    Regulation Number
    870.4380
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stockert Centrifugal Pump System is a cardiopulmonary bypass speed control device (21CFR 870.4380) that is indicated for use with the Cobe Revolution® Pump Head/Dideco Synergy™/ECC.OTM for for speed controlled pumping through a cardiopulmonary bypass circuit for typical durations of six hours or less. The SCP Plus is intended for use as a component part of or optional accessory to the S3 and S5 System (or any compatible system using the SS firmware versions of 3.0. or greater) and the Stöckert Air Purge Control (APC) System.

    The SCP Plus has been qualified only for typical durations of six hours or less, appropriate to cardiopulmonary bypass procedures and has not been qualified through in vitro, in vivo, or clinical studies for long-term use as a bridge to transplant, pending recovery of the natural heart, or extracorporeal membrane oxygenation (ECMO) procedures.

    Device Description

    The SCP Plus System is intended for use as a component part of or optional accessory to the S3 and S5 System (or any compatible system using the S5 firmware versions of 3.0. or greater) and with the Stöckert Air Purge (APC) System. The System consists of the drive unit, a mounting receptacle for the pump head, a retaining key, splash protection, and a 3-joint mast holder with fast clamp connector.

    The Stöckert Centrifugal Pump (SCP) Plus System consists of hardware, firmware, and electronics that are used to drive the centrifugal pump for surgical procedures requiring cardiopulmonary bypass for typical durations of six (6) hours or less. The SCP drive unit uses magnetic coupling with the centrifugal pump.

    AI/ML Overview

    This document describes the acceptance criteria and accompanying study for the Sorin Group Deutschland GmbH Stöckert Centrifugal Pump (SCP) Plus System.

    No information regarding a multi-reader multi-case (MRMC) comparative effectiveness study, standalone algorithm performance, or training set details is available in the provided text, as this is a traditional 510(k) for a medical device (a pump), not an AI/ML-driven device.


    1. Acceptance Criteria and Reported Device Performance

    The SCP Plus System was evaluated against prospectively defined performance criteria. The document states that the device's performance characteristics remain unchanged from its predicate device (K091008). Therefore, the acceptance criteria are implicitly tied to maintaining the established performance and safety standards of the predicate device.

    Acceptance Criteria CategoryReported Device Performance
    SafetyTested in accordance with IEC60601-1 (with National Deviations), demonstrating safety.
    Electromagnetic Compatibility (EMC)Tested in accordance with IEC60601-1-2, demonstrating EMC compliance.
    PerformanceFulfills prospectively defined functional acceptance tests and meets user needs. The system performs in an identical manner to the predicate device integrated with a different system (Sorin C5).
    Intended Use DurationsQualified for typical durations of six hours or less for cardiopulmonary bypass procedures.
    Technological CharacteristicsIdentical in hardware, firmware, electronics, and components (drive unit, control panel, flow probe, emergency drive unit, flexible drive shaft, connection cables) to the predicate device (K091008).

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify a "test set" in the context of diagnostic performance evaluation with a defined sample size for images or patient data. This is a submission for a physical medical device (a pump), not a diagnostic algorithm.

    The testing conducted was primarily non-clinical performance testing (safety, EMC, functional acceptance tests) of the physical device and its components. The document does not describe any studies involving human subjects or data derived from patients.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable as the document does not describe a "test set" requiring ground truth established by experts (e.g., radiologists, pathologists). The evaluation focused on engineering and performance criteria for a physical pump.


    4. Adjudication Method for the Test Set

    This section is not applicable as there was no test set requiring expert adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The provided text describes a 510(k) submission for a physical medical device (a centrifugal pump), not an AI/ML-driven diagnostic or assistive tool. Therefore, there is no mention of human readers improving with or without AI assistance.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone performance study (in the context of an algorithm) was not done. The device is a physical pump designed to operate within a cardiopulmonary bypass circuit, and its performance is assessed through engineering and functional tests, not as an "algorithm only."


    7. The Type of Ground Truth Used

    For the non-clinical performance testing, the "ground truth" was established by:

    • Compliance with international and national standards (IEC60601-1 for safety, IEC60601-1-2 for EMC).
    • Adherence to formal prospectively defined functional acceptance test specifications.
    • Meeting user needs and performance identical to the predicate device.

    There was no "ground truth" derived from expert consensus, pathology, or outcomes data in the context of diagnostic accuracy, as this is not a diagnostic device.


    8. The Sample Size for the Training Set

    This section is not applicable. The SCP Plus System is a hardware device with firmware, not a machine learning model, so there is no concept of a "training set" for an algorithm.


    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the same reasons as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3