Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K994409

    Validate with FDA (Live)

    Date Cleared
    2000-07-10

    (195 days)

    Product Code
    Regulation Number
    862.3640
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Instant-View™ Morphine Urine Dip Strip Test is a qualitative immunoassay device intended to detect morphine only in human urine at a cutoff level of 300 ng/ml. It is for health care professional use only.

    Instant-View Morphine Urine Dip Strip Test is a qualitative one step lateral flow, competitive binding immunoassay device intended to be used to detect morphine only in human urine at a cutoff level of 300 ng/ml. It is intended for health care professional use only.

    Device Description

    This test is a one-step lateral flow chromatographic immunoassay.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Instant-View™ Morphine Urine Dip Strip Test, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryCriteria/MethodReported Device Performance
    Correlation with PredicateComparison to predicate device (QuikStrip One Step Opiate Test)Higher than 91.3% correlation with the legally marketed test device.
    Agreement across POL sites92.5% agreement across all three POL sites.
    Accuracy (Overall Agreement)Agreement with expected results96.2% agreement with expected results.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample size for the test set. It mentions "results from the Clinical Laboratory and three Physician's Offices." Without specific numbers, it is not possible to determine the precise sample size.

    Regarding data provenance:

    • Country of Origin: Not specified, but likely the USA given the FDA 510(k) submission.
    • Retrospective or Prospective: Not explicitly stated. The phrasing "results from the Clinical Laboratory and three Physician's Offices" suggests a prospective study, but this is not confirmed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the number of experts used to establish ground truth or their specific qualifications. It only states that the accuracy evaluation was "conducted by persons with diverse educational background and working experience." This implies that the "expected results" (ground truth) were established by a reference method rather than expert consensus on interpretation of the test device results.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method like 2+1 or 3+1. The accuracy was determined by comparing the device's results to "expected results," which likely implies a single reference method result was used as the ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This device is a rapid diagnostic test (lateral flow immunoassay) for detecting morphine, not an AI-assisted diagnostic tool that aids human interpretation of complex medical images or data.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device itself is a standalone, qualitative immunoassay. The performance described (correlation and accuracy) is its intrinsic performance for detecting morphine in urine. There is no "algorithm" in the traditional sense of AI, but rather a biochemical reaction. The "standalone" performance here refers to the test strip's ability to produce a result without further human interpretation beyond reading the C and T lines.

    7. The Type of Ground Truth Used

    The ground truth used for accuracy evaluation was referred to as "expected results." While not explicitly stated, for drug screening devices, this typically refers to:

    • Known concentrations: Spiking urine samples with known concentrations of morphine both above and below the cutoff.
    • Reference method: Comparison to a gold standard analytical method like Gas Chromatography/Mass Spectrophotometry (GC/MS), which is explicitly mentioned as the "preferred confirmatory method" for preliminary positive results.

    8. The Sample Size for the Training Set

    The document does not provide information about a "training set." This type of immunoassay device does not typically involve a machine learning training phase in the same way an AI algorithm would. Its development is based on biochemical principles and validation.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a training set in the context of machine learning, there is no information on how a ground truth for such a set was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K994404

    Validate with FDA (Live)

    Date Cleared
    2000-07-05

    (190 days)

    Product Code
    Regulation Number
    862.3640
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Instant-View™ Morphine Urine Cassette Test is a qualitative immunoassay device intended to detect morphine only in human urine at a cutoff level of 300 ng/ml. It is for health care professional use only.

    Device Description

    This test is a one-step lateral flow chromatographic immunoassay.

    AI/ML Overview

    Here's an analysis of the provided text, outlining the acceptance criteria and the study details:

    Acceptance Criteria and Study Details for Instant-View™ Morphine Urine Cassette Test

    1. Acceptance Criteria and Reported Device Performance

    The provided text implies acceptance criteria for two key areas:
    i) Correlation with a legally marketed predicate device, and
    ii) Accuracy compared to expected results.

    Acceptance Criteria CategorySpecific Metric (Implied)Acceptance Threshold (Implied)Reported Device Performance
    Correlation with Predicate DeviceAgreement/Correlation with predicate device> 91.6% (implied from "higher than 91.6 %")"higher than 91.6 %"
    Agreement Across Different Testing Sites/UsersAgreement of results across multiple POL sitesNot explicitly stated"The results from all three POL sites agreed 92.5%."
    Accuracy (Comparison to Expected Results)Accuracy compared to expected results (clinical and Physician's Offices)Not explicitly stated"agreed 96.5 % with the expected results."

    Note: The document states "The correlation of results from the Instant-View ™ Morphine Urine Cassette Test, and the legally marketed test device compared, is higher than 91.6 %." This phrasing can be interpreted as the acceptance criterion itself, and the device met it. Similarly, "The Accuracy evaluation results...agreed 96.5 % with the expected results" indicates the performance achieved.


    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact numerical sample size used for the test set. However, it mentions:

    • "The correlation of results from the Instant-View ™ Morphine Urine Cassette Test, and the legally marketed test device compared, is higher than 91.6 %." This suggests a comparison was made on a set of samples.
    • "The results from all three POL sites agreed 92.5%." This indicates testing was conducted at multiple Physician's Office Laboratories (POLs) on a set of samples.
    • "The Accuracy evaluation results from the Clinical Laboratory and three Physician's Offices conducted by persons with diverse educational background and working experience agreed 96.5 % with the expected results." This implies a set of samples was tested in a clinical laboratory and at the three POLs.

    Data Provenance: The data appears to be prospective in nature, as it describes a study conducted to evaluate the device. The country of origin is not explicitly stated within the provided text, but the submission is to the U.S. FDA, suggesting the study was likely conducted in the U.S. or followed U.S. regulatory guidelines.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number of experts used to establish the ground truth for the test set or their specific qualifications. It mentions "persons with diverse educational background and working experience" conducted the accuracy evaluation, but these are presumably the test operators, not necessarily the experts establishing the ground truth if a confirmatory method was used.


    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method for the test set. It mentions "The correlation of results" with a legally marketed device and "Accuracy evaluation results" with "expected results." This implies a gold standard or reference method was used to determine the "expected results" or for comparison with the predicate device, but the process of resolving discrepancies (adjudication) is not detailed. The intent for positive results is that "A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrophotometry (GC/MS) is the preferred confirmatory method." This suggests GC/MS would serve as the ultimate ground truth, but how initial discrepancies in the study were handled is not detailed.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This device is a qualitative immunoassay for detecting morphine in urine, intended for standalone use by healthcare professionals. The study focuses on its performance characteristics, correlation with a predicate, and accuracy against expected results. It does not involve human readers interpreting images, nor does it assess the improvement of human readers with AI assistance.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, a standalone performance study was done. The Instant-View™ Morphine Urine Cassette Test is a "qualitative immunoassay device" designed for direct use to produce a result. The performance data presented (correlation, agreement across sites, accuracy) reflects the performance of the device itself, providing a direct interpretation (appearance of lines) without human interpretation beyond reading the indicated result. It is described as a "one-step lateral flow chromatographic immunoassay."


    7. The Type of Ground Truth Used

    The type of ground truth used is implied to be:

    • Comparison with a predicate device: For the correlation study, the results from the predicate device (QuikStrip One Step Opiate Test) served as a reference.
    • Expected results: For the accuracy evaluation, results were compared against "expected results," which, given the context of drug testing, would typically refer to results obtained from a highly accurate, quantitative reference method. The document explicitly states that for confirmation, "Gas Chromatography/Mass Spectrophotometry (GC/MS) is the preferred confirmatory method," strongly suggesting that a similar method or a prepared reference standard was used to define these "expected results" for the study samples. This aligns with chemical analysis/quantitative reference method as the ground truth.

    8. The Sample Size for the Training Set

    The document does not mention a training set or its sample size. The Instant-View™ Morphine Urine Cassette Test is described as a "lateral flow chromatographic immunoassay," implying a chemical/biological test rather than a machine learning or AI-based algorithm that would require a distinct training set. The study detailed focuses on evaluating the performance of the manufactured device.


    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a training set in the context of this immunoassay device, the method for establishing its ground truth is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1