Search Filters

Search Results

Found 28 results

510(k) Data Aggregation

    K Number
    K240987

    Validate with FDA (Live)

    Device Name
    Access sTfR
    Date Cleared
    2024-07-03

    (84 days)

    Product Code
    Regulation Number
    866.5880
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access sTfR assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of soluble transferrin receptor (sTR) levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of Iron Deficiency Anemia (IDA), and for the differential diagnosis of IDA and Anemia of Chronic Disease (ACD).

    This assay may also be used in conunction with an Access Ferritin measurement to provide a calculated sTR log ferritin index. This index is intended as an aid in the diagnosis of IDA, and for the differential diagnosis of IDA and ACD.

    Device Description

    Access sTfR: The sTfR assay reagent pack consists of two specific reagents: (R1a) paramagnetic particles coated with streptavidin:biotinylated soluble transferrin receptor monoclonal antibody, proteins (mouse, goat, bovine), bovine serum albumin (BSA), 0.1% sodium azide, and 0.17% ProClin 300; and (R1b) Monoclonal mouse anti-human soluble transferrin receptor alkaline phosphatase (bovine) conjugate, BSA, 0.1% sodium azide and 0.17% ProClin 300. Two assay packs containing 50 tests per pack are provided for a total of 100 assay determinations.

    The Access sTfR assay is a sequential two-step immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel along with paramagnetic particles coated with anti-sTfR antibody. During incubation, the sTfR antigen in the sample binds to the immobilized anti-sTfR antibody on the solid phase. Alkaline phosphatase conjugated anti-sTfR antibody is then added and reacts with a different antigenic site on the sTfR molecule.

    After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Access sTfR device, based on the provided FDA 510(k) summary:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device PerformanceMeets Criteria?
    Method ComparisonSlope: 0.99 – 1.01 (95% CI)Slope: 1.00 (95% CI: 0.99 – 1.01)Yes
    Intercept: -0.14 - 0.15 (95% CI)Intercept: -0.016 (95% CI: -0.14 - 0.15)Yes
    Correlation Coefficient (R): Not explicitly stated as acceptance, but comparison shows R = 1.00Correlation Coefficient (R): 1.00Yes (Excellent)
    Imprecision≤ 0.72 nmol/L SD at concentrations ≤ 9 nmol/LSample 1 (9.1 nmol/L): 0.24 nmol/L SDYes
    ≤ 8.0% CV at concentrations > 9 nmol/LSample 2 (17 nmol/L): 2.6% CV < 8%Yes
    Sample 3 (20 nmol/L): 2.4% CV < 8%Yes
    Sample 4 (33 nmol/L): 1.8% CV < 8%Yes
    Sample 5 (67 nmol/L): 2.0% CV < 8%Yes
    Sample 6 (118 nmol/L): 2.2% CV < 8%Yes
    Linearity"The assay demonstrated linearity across the measuring interval.""The assay demonstrated linearity across the measuring interval."Yes (Direct statement)
    Detection CapabilityNot explicitly stated as acceptance criteria, but listed as targets to be determined.LoB: 0.04 nmol/LAchieved targets
    LoD: 0.05 nmol/LAchieved targets
    LoQ: 0.05 nmol/L (≤ 20% within-lab CV)Achieved targets

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Method Comparison: N = 200 patient samples.
      • Imprecision: N = 6 samples (Sample 1-6) tested in duplicate, in 2 runs/day for a minimum of 20 days (total N for each sample is 80 measurements).
      • Linearity & Detection Capability: Sample sizes for these specific studies are not explicitly stated, but the studies were performed on the Dxl 9000 Access Immunoassay Analyzer.
      • Data Provenance: Not specified in the provided document (e.g., country of origin, retrospective or prospective).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This device is an in vitro diagnostic (IVD) immunoassay for the quantitative determination of soluble transferrin receptor (sTfR) levels. The performance studies (method comparison, imprecision, linearity, detection capability) do not involve human experts establishing a "ground truth" in the way an imaging AI might. Instead, the "ground truth" (or reference standard) for method comparison is the measurement from the predicate device (Access sTfR Assay on Access 2 Immunoassay System). For imprecision, linearity, and detection capability, the ground truth is established by the analytical measurement procedures themselves against defined statistical targets.
    3. Adjudication method for the test set:

      • Not applicable. As noted above, this is an IVD immunoassay, not a system requiring human adjudication of results in the traditional sense. The reference method (predicate device) serves as the comparator for the method comparison study.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is an IVD immunoassay, not an imaging AI or a device that assists human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance studies (method comparison, imprecision, linearity, detection capability) represent the standalone performance of the Access sTfR assay on the Dxl 9000 Access Immunoassay Analyzer. Its output is a quantitative sTfR value.
    6. The type of ground truth used:

      • Method Comparison: Measurements from the predicate device (Access sTfR Assay on Access 2 Immunoassay System). This serves as the reference for comparison against the new device.
      • Imprecision, Linearity, Detection Capability: Analytical measurements are compared against pre-defined statistical performance targets (e.g., SD, CV, LoB, LoD, LoQ) established according to CLSI guidelines.
    7. The sample size for the training set:

      • Not applicable. This is a conventional immunoassay, not a machine learning/AI device requiring a separate "training set" in that context. The device uses established biochemical reactions and a stored calibration curve.
    8. How the ground truth for the training set was established:

      • Not applicable (see above). The device establishes its "calibration" using internal calibrators and controls to create a stored calibration curve, as is typical for immunoassays. This is distinct from machine learning model training with labeled ground truth data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190495

    Validate with FDA (Live)

    Device Name
    Transferrin
    Manufacturer
    Date Cleared
    2020-03-16

    (382 days)

    Product Code
    Regulation Number
    866.5880
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Not Found

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text is a standard FDA clearance letter and does not contain information about the acceptance criteria, study details, or performance of the "Transferrin" device mentioned. The letter confirms that the device is substantially equivalent to a predicate device and can be marketed, but it does not include the technical or clinical study data.

    To answer your request, I would need a document that describes the actual performance study and its results.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182095

    Validate with FDA (Live)

    Date Cleared
    2018-11-05

    (94 days)

    Product Code
    Regulation Number
    866.5880
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Tina-quant Transferrin ver.2 (urine application) assay is an in vitro test for the quantitative determination of transferrin in human urine on Roche/Hitachi cobas c systems.

    A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in urine. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.

    Device Description

    The Tina-quant Transferrin ver.2 (urine application) assay is a two reagent assay for the in vitro quantitative determination of transferrin in human urine on automated clinical chemistry analyzers. It is an immunoturbidimetric assay in which human transferrin forms a precipitate with a specific antiserum which is determined turbidimetrically.

    Engineering drawings, schematics, and figures are not pertinent to describe the device, as the device is a reagent.

    AI/ML Overview

    The document provided is a 510(k) Premarket Notification for an in vitro diagnostic device, the "Tina-quant Transferrin ver.2 (urine application) assay." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving clinical effectiveness through extensive human studies often seen with novel medical devices. Therefore, the information regarding acceptance criteria and study design will be primarily focused on analytical performance validation rather than multi-reader multi-case clinical studies involving human interpretation of images, as this is a laboratory reagent.

    Here's an analysis of the provided text in the context of your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are largely implicit in the fact that "All data passed the predetermined acceptance criteria" for each analytical study. The performance is reported as the results of these studies.

    Performance CharacteristicAcceptance Criteria (Implicit: "All data passed the predetermined criteria")Reported Device Performance
    PrecisionCV% and SD values within predefined limits.Repeatability (within-run precision)
    - PreciControl ClinChem Multi 1Not explicitly stated, but results passed.Mean: 1.98 mg/dL, SD: 0.0140 mg/dL, CV: 0.7%
    - PreciControl ClinChem Multi 2Not explicitly stated, but results passed.Mean: 3.05 mg/dL, SD: 0.0242 mg/dL, CV: 0.8%
    - Human Urine 1Not explicitly stated, but results passed.Mean: 0.435 mg/dL, SD: 0.00979 mg/dL, CV: 2.3%
    - Human Urine 2Not explicitly stated, but results passed.Mean: 0.737 mg/dL, SD: 0.00920 mg/dL, CV: 1.2%
    - Human Urine 3Not explicitly stated, but results passed.Mean: 1.27 mg/dL, SD: 0.0107 mg/dL, CV: 0.8%
    - Human Urine 4Not explicitly stated, but results passed.Mean: 2.52 mg/dL, SD: 0.0184 mg/dL, CV: 0.7%
    - Human Urine 5Not explicitly stated, but results passed.Mean: 3.30 mg/dL, SD: 0.0252 mg/dL, CV: 0.8%
    Intermediate Precision (within-lab precision)Not explicitly stated, but results passed.
    - PreciControl ClinChem Multi 1Not explicitly stated, but results passed.Mean: 1.98 mg/dL, SD: 0.0158 mg/dL, CV: 0.8%
    - PreciControl ClinChem Multi 2Not explicitly stated, but results passed.Mean: 3.05 mg/dL, SD: 0.0267 mg/dL, CV: 0.9%
    - Human Urine 1Not explicitly stated, but results passed.Mean: 0.435 mg/dL, SD: 0.0111 mg/dL, CV: 2.5%
    - Human Urine 2Not explicitly stated, but results passed.Mean: 0.737 mg/dL, SD: 0.0112 mg/dL, CV: 1.5%
    - Human Urine 3Not explicitly stated, but results passed.Mean: 1.23 mg/dL, SD: 0.0130 mg/dL, CV: 1.1%
    - Human Urine 4Not explicitly stated, but results passed.Mean: 2.52 mg/dL, SD: 0.0215 mg/dL, CV: 0.9%
    - Human Urine 5Not explicitly stated, but results passed.Mean: 3.30 mg/dL, SD: 0.0289 mg/dL, CV: 0.9%
    Analytical Sensitivity
    - Limit of Blank (LoB)LoB Claim: 0.10 mg/dL (highest measurement for blank sample with stated prob.)Lot #1: 0.0150 mg/dL, Lot #2: 0.0130 mg/dL, Lot #3: 0.0150 mg/dL
    - Limit of Detection (LoD)LoD Claim: 0.15 mg/dL (lowest detectable analyte concentration with 95% prob.)Lot #1: 0.0382 mg/dL, Lot #2: 0.0331 mg/dL, Lot #3: 0.0353 mg/dL
    - Limit of Quantitation (LoQ)LoQ Claim: 0.22 mg/dL with 20% CV (lowest quantifiable concentration)Lot #1: 0.124 mg/dL, Lot #2: 0.137 mg/dL, Lot #3: 0.143 mg/dL
    Linearity/Assay Reportable RangeLinear relationship across the measuring range.
    - Measuring Range Claim0.22 to 3.5 mg/dL.Confirmed.
    - Pearson Correlation Coefficient (r)Close to 1.Lot 1: 0.9999, Lot 2: 0.9999, Lot 3: 0.9997
    Endogenous InterferencesNo interference at specified concentrations.
    - AlbuminNo interference ≤ 5000 mg/L.Passed (tested at 5 g/L).
    - CalciumNo interference ≤ 8 mmol/L.Passed (tested up to 9.92/9.80 mmol/L).
    - CitrateNo interference ≤ 10 mmol/L.Passed (tested up to 11 mmol/L).
    - CreatinineNo interference ≤ 44 mmol/L.Passed (tested up to 88 mmol/L).
    - GlucoseNo interference ≤ 111 mmol/L.Passed (tested up to 388 mmol/L).
    - HemoglobinNo significant interference up to 100 mg/dL.Passed (tested up to 146/149 mg/dL).
    - Immunoglobulin (IgG)No interference ≤ 500 mg/L.Passed (tested up to 1.1 g/L).
    - MagnesiumNo interference ≤ 75 mmol/L.Passed (tested up to 75 mmol/L).
    - OxalateNo interference ≤ 2.2 mmol/L.Passed (tested up to 3.75 mmol/L).
    - PhosphateNo interference ≤ 40 mmol/L.Passed (tested up to 130 mmol/L).
    - UreaNo interference ≤ 1000 mmol/L.Passed (tested up to 1500/1800 mmol/L).
    - Uric AcidNo interference ≤ 6 mmol/L.Passed (tested up to 6 mmol/L).
    - UrobilinogenNo interference ≤ 15 mg/dL.Passed (tested up to 15 mg/dL).
    Exogenous Interferences – DrugsNo interference at therapeutic concentrations (except noted).
    - AcetaminophenNot explicitly stated, but results passed up to 3000 mg/L.No interference up to 3000 mg/L.
    - Ascorbic acidNot explicitly stated, but results passed up to 4000 mg/L.No interference up to 4000 mg/L.
    - CefoxitinNot explicitly stated, but results passed up to 12000 mg/L.No interference up to 12000 mg/L.
    - Gentamicin sulfateNot explicitly stated, but results passed up to 400 mg/L.No interference up to 400 mg/L.
    - IbuprofenNot explicitly stated, but results passed up to 500 mg/L.No interference up to 500 mg/L.
    - LevodopaNot explicitly stated, but results passed up to 1000 mg/L.No interference up to 1000 mg/L.
    - MethyldopaNot explicitly stated, but results passed up to 2000 mg/L.No interference up to 2000 mg/L.
    - N-AcetylcysteineNot explicitly stated, but results passed up to 10 mg/L.No interference up to 10 mg/L.
    - OfloxacineNo interference.Interference observed (artificially high results). Claim adjusted.
    - PhenazopyridineNot explicitly stated, but results passed up to 50 mg/L.No interference up to 50 mg/L.
    - Salicyluric acidNot explicitly stated, but results passed up to 100 mg/L.No interference up to 100 mg/L.
    - TetracyclineNot explicitly stated, but results passed up to 300 mg/L.No interference up to 300 mg/L.
    Method Comparison to PredicatePredetermined acceptance criteria met. (e.g. slope near 1, intercept near 0, high r)y = 1.007x + 0.0052, r = 0.995 (Passing Bablok Regression)

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision and Analytical Sensitivity (LoB, LoD, LoQ):
      • Precision (Repeatability and Intermediate Precision): 5 human urine sample pools and 2 control samples. Tested for 21 days, 1 run/day, with 2 aliquots per sample in singlicate per part. This setup generates a large number of individual measurements over time to assess variability.
      • LoB: One analyte-free sample, measured 10-fold per run across 6 runs (over 4 days) on 3 reagent lots, resulting in 60 measurements per lot.
      • LoD: Five human urine samples with low-analyte concentration, measured 2-fold per run across 6 runs (over 4 days) on 3 reagent lots, resulting in 60 measurements per lot.
      • LoQ: A low-level sample set prepared by diluting 5 human urine samples, tested in 5 replicates per sample on 4 days, 1 run per day.
    • Linearity/Assay Reportable Range: Dilution series prepared using human urine sample pools (number of pools not specified, but likely at least one concentrated pool), with 13 concentrations measured on 3 lots in triplicate.
    • Endogenous Interferences: Two human urine pools (at two transferrin concentrations) were used for each interferent. Each pool was divided into two aliquots (spiked with interferent vs. solvent control). A dilution series of 11 steps was prepared and 3 aliquots per level were tested.
    • Exogenous Interferences – Drugs: Two human urine sample pools (spiked with approximately 0.433 and 2.48 mg/dL transferrin concentrations) were used for each drug. Each pool was divided into two aliquots (drug spiked vs. solvent control), measured in triplicate.
    • Method Comparison to Predicate: One hundred and seven (107) routine fresh, never-frozen human urine samples. Two samples were excluded (pH >8, value outside measuring range), so 107 samples were truly used in the comparison.
    • Data Provenance: The document explicitly states "human urine samples" or "human urine sample pools." For the method comparison, samples were "routine fresh, never-frozen human urine samples." There is no specific mention of the country of origin, but Roche Diagnostics Operations (RDO) is located in Indianapolis, Indiana, USA, and Roche Diagnostics GmbH, Mannheim, Germany is also mentioned as having the establishment registration. The studies are prospective analytical validation studies conducted with collected samples, not retrospective analysis of clinical patient data in the typical sense for imaging.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This type of in vitro diagnostic device (IVD) aims for quantitative measurement of a biomarker. Therefore, the "ground truth" is established through analytical reference methods, certified reference materials, or highly rigorous internal validation processes tied to metrological traceability, rather than human expert consensus on subjective findings (as would be the case for an imaging AI).

    • The ground truth for the test set is established by the analytical measurement on the reference method (for method comparison study), and by the known concentrations/dilutions of samples prepared for analytical studies (e.g., linearity, sensitivity, interference).
    • No "experts" in the sense of radiologists interpreting images were involved in establishing the ground truth for these analytical performance studies. The accuracy of measurements is verified against the reference standard or predetermined analytical values.

    4. Adjudication Method for the Test Set

    Not applicable for this type of analytical performance study. Adjudication methods (e.g., 2+1, 3+1) are common in clinical studies where multiple human readers independently assess data (like images) and then a consensus or tie-breaking mechanism is needed to establish ground truth or assess agreement. For an IVD, the "truth" is typically defined by the analytical method itself or a reference method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is a quantitative laboratory assay (a reagent for measuring transferrin in urine), not an AI imaging algorithm that assists human readers. No MRMC study was performed as it is irrelevant to the device's function.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the analytical performance studies (Precision, Analytical Sensitivity, Linearity, Interference, Method Comparison) represent the "standalone" performance of the assay itself (the reagent/instrument system) without human intervention in the result generation beyond operating the analyzer according to instructions. This is the primary form of performance evaluation for an IVD.

    7. The Type of Ground Truth Used

    • For Precision, Analytical Sensitivity, Linearity, and Interference studies: The ground truth is effectively derived from known concentrations in prepared samples (e.g., analyte-free samples, low-concentration samples, dilution series, spiked samples) or reference materials/controls with established values.
    • For Method Comparison: The ground truth is the measurement obtained from the predicate device, the "N Antisera to Human Transferrin (Siemens) on the BN ProSpec analyzer." This establishes substantial equivalence to an already legally marketed device.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that requires "training data" in the conventional sense. It's a chemical reagent for an established analytical method (immunoturbidimetry). The development process would have involved formulation and optimization, but not "training" on a data set in the way an AI model would be.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of IVD, which relies on chemical and immunological principles rather than machine learning from data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120236

    Validate with FDA (Live)

    Date Cleared
    2013-02-13

    (384 days)

    Product Code
    Regulation Number
    866.5880
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Binding Site Human Transferrin Kit is intended for the quantitative determination of human transferrin using the Binding Site SPAPLus turbidimetric analyzer in human serum. The measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection and iron deficiency anemia. This test should be used in conjunction with other laboratory and clinical findings

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for a diagnostic kit, specifically "Human Transferrin kit for use on the SPAPlus™". This type of document primarily confirms that a new device is substantially equivalent to a legally marketed predicate device, and it typically does not include detailed study reports with specific acceptance criteria, performance data, or details about training and test sets in the way that would be provided for an AI/ML medical device submission.

    The information requested in the prompt (acceptance criteria, sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, standalone performance, etc.) is characteristic of the rigorous clinical validation studies required for AI/ML devices, not typically for a traditional in vitro diagnostic kit like the one described here.

    Therefore, I cannot extract the requested information from the provided text because the document does not contain details about:

    • A table of acceptance criteria and reported device performance in the context of an AI/ML device. The "performance" mentioned for this kit would usually refer to analytical performance characteristics (e.g., accuracy, precision, linearity) which are not detailed here.
    • Sample sizes for test sets or data provenance.
    • Number of experts or their qualifications.
    • Adjudication methods.
    • MRMC comparative effectiveness studies.
    • Standalone performance (as it's not an AI/ML algorithm).
    • Type of ground truth used (beyond implying the "gold standard" for transferrin measurement).
    • Sample size for training sets.
    • How ground truth for training was established.

    The document is a regulatory approval letter, specifically stating the device's substantial equivalence and its intended use for the quantitative determination of human transferrin to aid in the diagnosis of malnutrition, acute inflammation, infection, and iron deficiency anemia using the Binding Site SPAPlus™ turbidimetric analyzer.

    Ask a Question

    Ask a specific question about this device

    K Number
    K080634

    Validate with FDA (Live)

    Date Cleared
    2008-08-11

    (158 days)

    Product Code
    Regulation Number
    866.5880
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access sTfR assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of soluble transferrin receptor (sTfR) levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of Iron Deficiency Anemia (IDA), and for the differential diagnosis of IDA and Anemia of Chronic Disease (ACD).

    This assay may also be used in conjunction with a ferritin measurement to provide a calculated sTfR/log ferritin index. This index is intended as an aid in the diagnosis of IDA, and for the differential diagnosis of IDA and ACD.

    The Access sTfR Calibrators are intended to calibrate the Access sTfR assay for the quantitative determination of soluble transferrin receptor levels in human serum and plasma (heparin) using the Access Immunoassay Systems.

    The Access sTfR QC is intended for monitoring system performance of the Access sTfR assay.

    Device Description

    The Access sTfR reagent, calibrators, controls, and the Access Immunoassay Analyzers (Access, Access 2, Synchron LXi 725, UniCel DxC 600i, UniCel Dxl 600, and UniCel Dxl 800) comprise the Access Immunoassay Systems for the quantitative determination of soluble transferrin receptor in human serum and plasma.

    Automated; Paramagnetic particles coated with mouse monoclonal antibody against sTfR. Uses the same mouse monoclonal antibodies against sTfR in the capture phase and signal phase as the predicate device.

    Utilizes dioxetane-based chemiluminescent substrate; measures light production from a chemiluminescent reaction.

    Calibrators are comprised of natural sTfR at 6 levels (0, 3, 10, 30, 80, and 150 nmol/L) in a buffered matrix.

    QCs are human sTfR provided as a liquid at 3 levels (~10, ~25, ~90 nmol/L) in a buffered matrix.

    AI/ML Overview

    Here's an analysis of the provided text to fulfill your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (Access sTfR)
    Imprecision (Total precision)≤ 8% CV for concentrations > 9 nmol/L2.6 to 5.4% CV for samples > 9 nmol/L
    Imprecision (Total precision)< 0.72 nmol/L SD for concentrations < 9 nmol/L0.08 to 0.38 nmol/L SD for samples < 9 nmol/L
    Imprecision (Within run precision)Not explicitly stated, but generally lower than total precision1.6 to 5.2% CV for samples > 9 nmol/L
    Imprecision (Within run precision)Not explicitly stated, but generally lower than total precision0.04-0.22 nmol/L SD for samples ≤ 9 nmol/L
    Analytical Sensitivity (Lowest detectable level)Not explicitly stated, but should be distinguishably non-zero< 0.05 nmol/L
    Dilution Recovery (Mean recovery)100 ± 15%Within 100 ± 15%
    Dilution Recovery (Individual mean recovery)At least 92% within 100 ± 20%At least 92% within 100 ± 20%
    Methods Comparison (Correlation with Predicate)High correlation (e.g., r > 0.95 or similar)r = 0.96 (y=0.8901X + 0.6853)
    Analytical Specificity (Interference avoidance)No significant interference from therapeutic drugs, similar compounds, or common contaminantsNo significant interference from therapeutic drugs, similar compounds, bilirubin, total protein, hemoglobin, triglycerides, or rheumatoid factor (up to 850 IU/mL)
    Reagent Stability (Opened)Not explicitly stated, but ensures practical shelf-life28 days
    Calibrator Stability (Opened)Not explicitly stated, but ensures practical shelf-life90 days
    Control Stability (Opened)Not explicitly stated, but ensures practical shelf-life90 days
    Calibration Curve StabilityNot explicitly stated, but ensures practical shelf-life28 days
    Clinical Sensitivity (sTfR for IDA/ACD+IDA)Optimized sensitivity with reasonable specificity86% sensitivity at 21 nmol/L cutoff
    Clinical Specificity (sTfR for IDA/ACD+IDA)Optimized specificity with reasonable sensitivity49.1% specificity at 21 nmol/L cutoff
    Clinical Sensitivity (sTfR/log ferritin index for IDA/ACD+IDA)Optimal sensitivity80.7% sensitivity at 14 cutoff
    Clinical Specificity (sTfR/log ferritin index for IDA/ACD+IDA)Optimal specificity82.5% specificity at 14 cutoff

    2. Sample Size Used for the Test Set and Data Provenance

    • Methods Comparison (Analytical Study):

      • Sample Size: 271 samples
      • Data Provenance: Not explicitly stated, but implies clinical samples used for comparing the new assay to the predicate. The "External Site" suggests multiple locations.
      • Retrospective/Prospective: Not specified.
    • Clinical Studies (Clinical Trial):

      • Sample Size: Not explicitly stated for the clinical trial itself, but implied to be sufficient to derive sensitivity and specificity values.
      • Data Provenance: "Prospective multicenter clinical trial." This indicates data was collected forward-in-time from multiple clinical sites. The country of origin is not specified.
      • Retrospective/Prospective: Prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. For the clinical studies, "diagnosis of Iron Deficiency Anemia (IDA), and for the differential diagnosis of IDA and Anemia of Chronic Disease (ACD)" would typically involve clinical experts (e.g., hematologists, clinical pathologists), but the number and qualifications are not mentioned.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. For clinical diagnoses used as ground truth, an adjudication process (e.g., consensus by multiple experts) is often used, but it's not detailed here.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There was no MRMC comparative effectiveness study described. This device is an in vitro diagnostic (IVD) assay for measuring a biomarker (sTfR), not an AI-powered image analysis or diagnostic aid that would assist human readers in interpreting images or other complex data. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply to this specific device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies described are for the device's standalone performance as an in vitro diagnostic assay. The "Access sTfR reagents, calibrators, controls, and the Access Immunoassay Analyzers" perform the quantitative determination of sTfR levels. There isn't a human-in-the-loop component in the actual measurement and result generation of the sTfR value itself. The clinical interpretation of that sTfR result (and sTfR/log ferritin index) is then performed by a clinician.

    7. The Type of Ground Truth Used

    • Analytical Studies (Methods Comparison): The ground truth for method comparison was values obtained from the predicate device (Quantikine® IVD® sTfR ELISA). This is a common approach for demonstrating substantial equivalence for quantitative assays.
    • Clinical Studies: The ground truth for the clinical study was the diagnosis of Iron Deficiency Anemia (IDA) and Anemia of Chronic Disease (ACD). This would typically be established through a combination of clinical assessment, laboratory tests (beyond just sTfR), and potentially bone marrow biopsy or response to iron therapy, which constitutes outcomes data or expert consensus-based clinical diagnosis. The document does not specify the exact methods for defining IDA and ACD beyond the general diagnoses.

    8. The Sample Size for the Training Set

    The document does not explicitly describe a "training set" in the context of an algorithm or machine learning model. This is an immunoassay (laboratory test), not a software/AI device that typically undergoes separate training and test phases.

    • The "analytical studies" and "clinical studies" described serve as validation studies for the device's performance, much like a traditional device's verification and validation.
    • There's no mention of a development phase where a model would be "trained" on a specific dataset.

    9. How the Ground Truth for the Training Set Was Established

    As there is no explicit "training set" mentioned in the context of typical machine learning, this question is not applicable. The development and optimization of the immunoassay reagents and protocols are based on scientific principles of immunology and chemistry, rather than an algorithmic training process on ground truth data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081299

    Validate with FDA (Live)

    Date Cleared
    2008-08-08

    (92 days)

    Product Code
    Regulation Number
    866.5880
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TRF method is an in vitro diagnostic test for the quantitative measurement of transferrin in human serum, heparinized plasma, EDTA plasma or urine on the Dimension Vista® System. Measurements of transferrin aid in the diagnosis of malnutrition, acute inflammation, infection, and iron deficiency anemia.

    Protein 1 Calibrator is an in vitro diagnostic product for the calibration of the Dimension Vista® System for: α₁-Acid Glycoprotein (A1AG), α₁-Antitrypsin (A1AT), β₂-Microglobulin (B2MIC), C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Haptoglobin (HAPT), Hemopexin (HPX), Homocysteine (HCYS), Immunoglobulin A (IGA), Immunoglobulin E (IGE), Immunoglobulin G (IGG) [serum/plasma] and (IGG-C) [cerebrospinal fluid], Immunoglobulin G Subclass 1 (IGG1), Immunoglobulin G Subclass 2 (IGG2), Immunoglobulin G Subclass 3 (IGG3), Immunoglobulin G Subclass 4 (IGG4), Immunoglobulin M (IGM), Prealbumin (PREALB), Retinol Binding Protein (RBP), soluble Transferrin Receptor (STFR), Transferrin (TRF) [serum/plasma] and (TRF-U) [urine].

    PROT3 CON is an assayed, low level intralaboratory quality control for assessment of precision and analytical bias on the Dimension Vista® System in the determination of α -Microglobulin (A1MIC), specialty Albumin (sALB*), Immunoglobulin G (IGG -C*), Microalbumin (MALB) and Transferrin (TRF-U**). * For Cerebrospinal fluid (CSF) ** For urine

    Device Description

    Proteins contained in human body fluids form immune complexes in an immunochemical reaction with specific antibodies. These complexes scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.

    Protein 1 Calibrator is a multi-analyte, liquid human serum based product containing q acid glycoprotein, a-antitrypsin, ß2-microglobulin, C3 complement, C4 complement, ceruloplasmin, haptoglobin, hemopexin, homocystine, immunoglobulin A, immunoglobulin E, immunoglobulin G, immunoglobulin G subclass 1, immunoglobulin G subclass 2, immunoglobulin G subclass 3, immunoglobulin G subclass 4, immunoglobulin M, prealbumin, retinol binding protein, soluble transferrin receptor and transferrin.

    Protrein 3 Control is a multi-analyte, lyophilized, polygeline and rabbit plasma albumin based product containing a - Microglobulin, albumin, immunoglobulin G and transferrin.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Siemens Healthcare Diagnostics Inc. 510(k) Notification for the Dimension Vista® System TRF Flex® reagent cartridge, Protein 1 Calibrator, and Protein 3 Control, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary primarily focuses on establishing substantial equivalence through a method comparison study. It does not explicitly state pre-defined acceptance criteria in terms of performance metrics like sensitivity, specificity, accuracy against a gold standard, or specific numerical targets for regression analysis. Instead, the "acceptance criteria" can be inferred as achieving a strong correlation and suitable regression statistics (slope, intercept, correlation coefficient) when compared to the legally marketed predicate device.

    Acceptance Criteria (Inferred)Reported Device Performance
    Strong correlation with predicate device (Dade Behring N Antisera to Human Transferrin on BN ProSpec® System) for Transferrin (TRF) measurements in urine.Regression Analysis: Equation provided: (Specific regression equation not fully legible in the provided text, but mentioned as having been performed).
    Suitable regression statistics (slope, intercept, correlation coefficient) demonstrating substantial equivalence.Demonstrated substantial equivalence based on correlation studies and provided information.
    The modified Dimension Vista® System TRF assay, Protein 1 Calibrator, and Protein 3 Control are substantially equivalent to the legally marketed devices.Conclusion states that the devices are substantially equivalent based on the correlation studies.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The study evaluated urine samples with concentrations ranging from 2.0 to 24.4 mg/L. The exact number of samples is not explicitly stated in the provided text.
    • Data Provenance: The text does not specify the country of origin of the data. The study appears to be prospective in nature, comparing the new device against an existing one in a method comparison.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this submission. The "ground truth" here is not established by expert consensus on clinical interpretation, but rather by the measurements obtained from the legally marketed predicate device. This is a measurement device comparison, not a diagnostic interpretation device.

    4. Adjudication Method for the Test Set

    This is not applicable as the study is a method comparison of quantitative measurements, not a diagnostic interpretation that typically involves adjudication of reader disagreements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This is not applicable. The device is an in vitro diagnostic (IVD) reagent and calibrator/control system for quantitative measurement, not an AI-assisted diagnostic imaging or interpretation tool that would involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    The device itself is a standalone measurement system (reagent, calibrator, control used on an automated analyzer). The comparison study implicitly tests the standalone performance of the Dimension Vista® System TRF assay against the predicate device. The performance data presented directly reflects the algorithmic (or analytical) output of the device.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is the measurements obtained from the legally marketed predicate device: the Dade Behring N Antisera to Human Transferrin assay on the BN ProSpec® System. This is a form of comparative analytical performance against an established method.

    8. The Sample Size for the Training Set

    This is not applicable. This submission describes an in vitro diagnostic test system (reagent, calibrator, control). These types of devices do not typically involve "training sets" in the same way machine learning algorithms do. The development of the assay would involve internal optimization and validation, but not a distinct "training set" for an algorithm in the sense of AI.

    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, the concept of a "training set" and associated ground truth is not applicable to this type of IVD device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063086

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-08-08

    (302 days)

    Product Code
    Regulation Number
    866.5880
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The transferrin test system is intended for the quantitative in-vitro diagnostic determination of transferrin in serum or plasma using T60 Clinical chemistry Analyzers. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, acute infection and iron deficiency anemia.

    SpeciCal protein calibrator is used as a stock calibrator for both quantification of specific proteins in serum and plasma by immunoturbidimetry and for antigen excess detection using methods defined by Thermo Electron Oy

    SpeciTrol is intended to be used as an assayed control serum to monitor precision of specific protein tests defined by Thermo Electron Ov

    Specitrol High is intended to be used as an assayed control serum to monitor precision of specific protein tests defined by Thermo Electron Oy

    Device Description

    Not Found

    AI/ML Overview

    The provided text does not contain detailed information about the acceptance criteria or a study that proves the device meets those criteria. The document is primarily a 510(k) clearance letter from the FDA for a device named "Transferrin, Specical Calibrator, Specitrol Control and Specitrol High Control."

    The letter confirms that the device is substantially equivalent to legally marketed predicate devices, allowing it to proceed to market. However, it does not include:

    • A table of acceptance criteria and reported device performance.
    • Information on sample sizes used for test sets, data provenance (country, retrospective/prospective).
    • Details about the number or qualifications of experts used to establish ground truth.
    • Adjudication methods.
    • Results from a multi-reader multi-case (MRMC) comparative effectiveness study, nor the effect size of AI assistance.
    • Results from a standalone algorithm-only performance study.
    • The type of ground truth used (e.g., pathology, outcomes data).
    • The sample size for the training set or how its ground truth was established.

    The document states the intended use of the devices, which is for the quantitative in-vitro diagnostic determination of transferrin in serum or plasma using T60 Clinical Chemistry Analyzers, aiding in the diagnosis of malnutrition, acute inflammation, acute infection, and iron deficiency anemia. It also specifies the use of SpeciCal as a stock calibrator and SpeciTrol/SpeciTrol High as assayed control serums to monitor precision.

    Therefore, I cannot fulfill the request for information on acceptance criteria, study details, or AI-related metrics based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063766

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-04-23

    (124 days)

    Product Code
    Regulation Number
    866.5880
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Good Biotech Corp. Transferrin test system is intended to be used for the quantitative determination of transferrin in human serum by turbidimetric immunoassay (TIA). Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.

    Good Biotech Corp. Transferrin Calibrator Set is intended to be used with Transferrin TIA for the quantitative determination of transferrin in serum samples.

    Good Biotech Corp. Transferrin Controls are intended to be used as the assayed quality control material for transferrin analysis.

    For In Vitro Diagnostic Use.

    Device Description

    Good Biotech Corp. Transferrin TIA is a ready to use reagent for the quantitative determination of transferrin in human serum by turbidimetric immunoassay (TIA). When transferrin of the serum sample encounters with duck anti-transferrin antibody, the aqqlutination based on the antigen-antibody reaction increases the turbidity of the sample. The value of the absorbance change at 505 nm is proportional to the transferrin concentration of the sample and is recorded by a general chemistry autoanalyzer. Then, the actual transferrin concentration of the serum sample is determined by interpolation of the calibration curve obtained by standard samples with known transferrin concentrations.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state pre-defined acceptance criteria in terms of specific numeric thresholds for slope, intercept, and correlation coefficient. Instead, it relies on a "high correlation coefficient" as an indicator of substantial equivalence. The predicate device's performance characteristics (slope, intercept, correlation coefficient) likely serve as an implicit benchmark for "high correlation."

    MetricAcceptance Criteria (Implicit from Predicate/General Expectation for Substantial Equivalence)Reported Device Performance
    SlopeClose to 1.0 (indicating similar proportionality)0.927
    Intercept (mg/dL)Close to 0.0 (indicating similar baseline/bias)19.784
    Correlation CoefficientHigh (e.g., >0.95 or >0.90, indicating strong linear relationship)0.992

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 82 serum samples
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. It only mentions "serum samples."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable to this kind of diagnostic device (immunoassay). The "truth" for the test set is established by the measurements obtained from the predicate device, which is an already legally marketed and validated assay. No human experts are used to interpret the results or establish "ground truth" in the way they would for imaging or clinical diagnosis.

    4. Adjudication Method for the Test Set

    Not applicable. The comparison is between two quantitative assays, not subjective interpretations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for devices that involve human interpretation, such as imaging AI, where the impact of AI assistance on human reader performance is evaluated. This submission is for a quantitative immunoassay.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    Yes, a standalone study was performed. The device's performance was compared directly against the predicate device without human intervention or interpretation as part of the primary measurement. The "algorithm" here refers to the immunoassay's ability to quantitatively determine transferrin levels.

    7. Type of Ground Truth Used

    The "ground truth" in this context is the quantitative determination of transferrin levels by the predicate device (Roche Tina-quant Transferrin ver.2). This is a common approach for establishing substantial equivalence for new diagnostic assays, where a well-established and legally marketed device serves as the reference standard.

    8. Sample Size for the Training Set

    The document does not specify a training set. This is a pre-market submission for an immunoassay, not a machine learning algorithm that typically requires a distinct training and test set in the same way. The development and optimization of the immunoassay reagents and methodology would have been an iterative process by the manufacturer, but the term "training set" as commonly applied to AI/ML is not relevant here. The studies described are for validation/testing.

    9. How the Ground Truth for the Training Set Was Established

    As no training set (in the AI/ML sense) is explicitly mentioned, this question is not directly applicable. The "ground truth" for the comparative performance study was established by the predicate device's measurements.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063663

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-03-21

    (103 days)

    Product Code
    Regulation Number
    866.5880
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dimension Vista™ STFR Flex® reagent cartridge: The STFR method is an in vitro diagnostic test for the quantitative determination of soluble transferrin receptor in human serum and heparinized plasma on the Dimension Vista System. Measurements of soluble transferrin receptor aid in the diagnosis of malnutrition, acute inflammation, infection, and iron deficiency anemia.

    Dimension Vista™ Protein 1 Calibrator: PROT1 CAL is an in vitro diagnostic product for the calibration of the C3 Complement (C3), C4 Complement (C4), Immunoglobulin A (IGA), Immunoglobulin G (IGG), Immunoglobulin M (IGM), Prealbumin (PREALB) and Soluble Transferrin Receptor (STFR), methods on the Dimension Vista System.

    Dimension Vista™ Protein 1 Control L, M and H: PROT1 CON L, M and H are assayed intralaboratory quality controls for assessment of precision and analytical bias in the determination of C3 complement (C3), C4 complement (C4), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM), prealbumin (PREALB) and soluble transferrin receptor (STFR) on the Dimension Vista System.

    Device Description

    Dimension Vista™ STFR Flex reagent cartridge: Polystyrene particles coated with monoclonal antibodies specific to human soluble transferrin receptor are aggregated when mixed with samples containing soluble transferrin receptor. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.

    Dimension Vista™ Protein 1 Calibrator: Protein 1 Calibrator is a multi-analyte, liguid, human serum based product containing C3 Complement, C4 Complement, immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB) and Soluble Transferrin Receptor (STFR).

    Dimension Vista™ Protein 1 Control L. M and H: Protein 1 Control L, M and H are multi-analyte, liquid, human serum based products containing C3 complement, C4 complement, immunoqlobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB) and soluble transferrin receptor (STFR).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Dimension Vista™ STFR Flex® reagent cartridge, Dimension Vista™ Protein 1 Calibrator, and Dimension Vista™ Protein 1 Control L, M, and H:


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes a method comparison study between the new device and a legally marketed predicate device. The acceptance criteria are implicitly defined by the reported regression analysis results, which demonstrate a strong correlation and "equivalent performance" to the predicate device.

    Acceptance Criteria (Implicit)Reported Device Performance (Dimension Vista™ STFR Flex® compared to N Latex sTFR on BN ProSpec® System)
    Strong correlationCorrelation Coefficient: 0.997
    Slope close to 1Slope: 1.053
    Intercept close to 0Intercept: -0.078 mg/L
    Equivalent performance"These studies demonstrate correlation and equivalent performance"

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 153 serum and plasma samples.
    • Data Provenance: The document does not explicitly state the country of origin or if the data was retrospective or prospective. It only mentions that "serum and plasma samples with concentrations ranging from 0.21 mg/L to 4.13 mg/L" were evaluated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the given summary. The ground truth in this context is implicitly established by the measurements obtained from the predicate device (Dade Behring N Latex sTFR assay on the BN ProSpec® System).

    4. Adjudication Method for the Test Set

    This information is not applicable and therefore not provided in the given summary. The study is a quantitative method comparison between two automated diagnostic devices, not a study requiring expert adjudication of results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a method comparison between two in vitro diagnostic (IVD) devices for quantitative measurement of an analyte, not a study involving human readers or interpretation of medical images.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance study was done. The study directly assesses the performance of the Dimension Vista™ STFR assay on the Dimension Vista System against a predicate device, without involving human interpretation or modifications to the results.

    7. The Type of Ground Truth Used

    The "ground truth" for this study was established through the measurements obtained from the legally marketed predicate device, the Dade Behring N Latex sTFR assay on the BN ProSpec® System. This is a common practice in IVD device submissions to demonstrate substantial equivalence.

    8. The Sample Size for the Training Set

    This information is not provided in the given summary. The document describes a comparison study for the finalized device, not the development or training of an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided in the given summary, as the submission focuses on the performance of the final device rather than its developmental stages or training data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063322

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-01-24

    (82 days)

    Product Code
    Regulation Number
    866.5880
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dimension Vista™ TRF Flex® reagent cartridge: The TRF method is an in vitro diagnostic test for the quantitative determination of transferrin in human serum, heparinized plasma or EDTA plasma on the Dimension Vista® System. Measurements of transferrin aid in the diagnosis of malnutrition, acute inflammation, infection, and iron deficiency anemia.

    Dimension Vista " Protein 1 Calibrator: PROT1 CAL is an in vitro diagnostic product for the calibration of the C3 complement (C3), C4 complement (C4), Immunoglobulin A (IGA), Immunoglobulin G (IGG), Immunoglobulin M (IGM), Prealbumin (PREALB) and Transferrin (TRF) methods on the Dimension Vista® System.

    Dimension Vista™ Protein 1 Control L, M and H: PROT1 CON L, M and H are assayed intralaboratory quality controls for assessment of precision and analytical bias in the determination of C3 Complement (C3), C4 Complement (C4), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM), prealbumin (PREALB) and transferrin (TRF) on the Dimension Vista "System.

    Device Description

    Dimension Vista™ TRF Flex® reagent cartridge: Proteins contained in human body fluids form immune complexes in an immunochemical reaction with specific antibodies. These complexes scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.

    Dimension Vista" Protein 1 Calibrator: Protein 1 Calibrator is a multi-analyte, liquid, human serum based product containing C3 complement, C4 complement, immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM), prealbumin (PREALB) and Transferrin.

    Dimension Vista™ Protein 1 Control L, M and H: Protein 1 Control L, M and H are multi-analyte, liquid, human serum based products containing C3 complement, C4 complement, immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM), prealbumin (PREALB) and transferrin.

    AI/ML Overview

    This document describes the Dimension Vista™ TRF Flex® reagent cartridge, Dimension Vista™ Protein 1 Calibrator, and Dimension Vista™ Protein 1 Control L, M, and H, which are in vitro diagnostic medical devices. The core of your request focuses on the Dimension Vista™ TRF Flex® reagent cartridge for the quantitative determination of transferrin.

    Here's an analysis of the provided information based on your request:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of pre-defined thresholds. Instead, it presents a method comparison study comparing the new device (Dimension Vista™ TRF assay) with a legally marketed predicate device (Dade Behring N Antisera to Human Transferrin assay on the BN ProSpec® System). The reported device performance from this study serves as the evidence of equivalence to the predicate.

    Performance MetricAcceptance Criteria (Implied by equivalence)Reported Device Performance (Dimension Vista™ TRF vs. N Antisera to Human Transferrin on BN ProSpec®)
    Correlation CoefficientHigh correlation (e.g., >0.95 or >0.975)0.992
    SlopeClose to 1 (e.g., 0.95 - 1.05)1.087
    InterceptClose to 00.060

    Note: The document states, "These studies demonstrate correlation and equivalent performance between the Dade Behring N Antisera to Human Transferrin assay and the Dimension Vista™ TRF assay." This implies that the observed correlation, slope, and intercept values were considered acceptable by the manufacturer for demonstrating substantial equivalence.


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 180 samples (n = 180)
    • Data Provenance: The document does not explicitly state the country of origin. It also does not explicitly state if the data was retrospective or prospective. However, given that it's a method comparison study for an in vitro diagnostic device, it is typically conducted on clinical samples, which are generally considered prospective or retrospective clinical samples collected for research purposes. No specific information on the demographics or collection methods is provided.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This type of in vitro diagnostic device (immunoassay for transferrin) does not typically involve human experts for establishing "ground truth" in the same way an imaging or pathology device would. The "ground truth" for method comparison studies like this is the measurement obtained from the predicate device (Dade Behring N Antisera to Human Transferrin assay on the BN ProSpec® System), which is itself an established laboratory method. Therefore, no human experts were involved in establishing the ground truth for this test set; rather, it relied on the analytical performance of a reference method.


    4. Adjudication Method for the Test Set

    Not applicable. As explained above, the "ground truth" for this method comparison study is the measurement obtained from the predicate device, not an interpretation requiring expert adjudication.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This is an in vitro diagnostic assay, not an imaging or interpretive device that would involve multiple human readers. Therefore, an MRMC comparative effectiveness study is not relevant here.


    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, in a sense. The study directly compares the analytical performance of the new automated Dimension Vista™ TRF assay (device only) against the predicate automated assay (device only). This is an evaluation of the algorithm/device's performance in generating quantitative results without human interpretive input affecting the result itself. The device is designed to provide a quantitative measurement.


    7. Type of Ground Truth Used

    The "ground truth" for the performance comparison was the measurements obtained from the legally marketed predicate device, the Dade Behring N Antisera to Human Transferrin assay on the BN ProSpec® System. This is a common approach for demonstrating substantial equivalence for new in vitro diagnostic assays.


    8. Sample Size for the Training Set

    The document does not provide information about a separate "training set." For in vitro diagnostic assays, especially those based on established immunochemical principles, the development often relies on well-understood chemical reactions and calibration procedures rather than a large, continuously-trained machine learning model in the typical sense. Data used for initial assay development and optimization would precede this validation study, but specifics are not mentioned.


    9. How the Ground Truth for the Training Set Was Established

    As with point 8, the document does not detail a "training set" or its ground truth establishment. Development of such assays generally involves using characterized reference materials and calibrators, often with values traceable to international standards or established reference methods, to ensure accuracy during the assay's development and optimization phases.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3