Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K142994

    Validate with FDA (Live)

    Date Cleared
    2015-04-21

    (187 days)

    Product Code
    Regulation Number
    862.1045
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS-iSYS Aldosterone assay (IS-3300) is a device intended for use in clinical laboratories for the quantitative determination of Aldosterone in human EDTA plasma on the IDS-iSYS Multi-Discipline Automated System. Aldosterone measurements are used in the diagnosis and treatment of primary aldosteronism (a disorder caused by excessive secretion of Aldosterone by the adrenal gland), hypertension caused by primary aldosteronism, selective hyperaldosteronism, edematous states and other conditions of electrolyte balance.

    The IDS-iSYS Aldosterone Control Set (IS-3330) is intended for use as assaved quality control samples to monitor the accuracy of the IDS-iSYS Aldosterone assay on the IDS-iSYS Multi-Discipline Automated System.

    The IDS-iSYS Aldosterone Calibration Verifiers (IS-3335) are intended for medical purposes for use in the quantitative verification of calibration of the IDS-iSYS Aldosterone assay on the IDS-iSYS Multi-Discipline Automated System.

    Device Description

    The IDS-iSYS Aldosterone assay is based on chemiluminescence technology. A biotinylated monoclonal anti-Aldosterone antibody is incubated with the sample, after an incubation step an Aldosterone acridinium conjugate is added and after a further incubation step streptavidin coated magnetic particles are added. Following a third incubation step the particles are "captured" using a magnet. After a washing step and addition of trigger reagents, the light emitted by the acridinium label is inversely proportional to the concentration of Aldosterone in the original sample.

    AI/ML Overview

    Here's an analysis of the provided text to extract the requested information about acceptance criteria and the supporting study:

    The provided document is a 510(k) Summary for the IDS-iSYS Aldosterone assay, control set, and calibration verifiers. It focuses on demonstrating substantial equivalence to a predicate device, as required for FDA clearance. The document details performance characteristics but does not explicitly state acceptance criteria in a pass/fail format typical of formal acceptance criteria documents. Instead, it reports performance values found during validation studies.

    Here's the breakdown of the information you requested, based on what's available in the document:


    1. Table of Acceptance Criteria and Reported Device Performance

    As mentioned, explicit "acceptance criteria" are not listed in a single table. However, the performance characteristics are reported, and implicitly, these values are considered acceptable for demonstrating substantial equivalence. The table below compiles the reported performance data from the document.

    Note: The document does not provide a column for "Acceptance Criteria" as a separate, pre-defined target. The "Reported Device Performance" is the outcome of the studies aiming to demonstrate acceptable performance.

    Performance CharacteristicImplicit or Explicit Acceptance Threshold (Not explicitly stated as "acceptance criteria" but implied targets from regulatory guidance or industry standards)Reported Device Performance (IDS-iSYS Aldosterone Assay)Source of Data
    Precision (CV%)Implied: Generally low CV% indicating good reproducibility (e.g., <15-20% for clinical assays, lower for critical ranges)Within-Run: 2.2% - 8.4% (at 98.7 ng/dL to 7.5 ng/dL)Table (p.8)
    Total: 5.2% - 12.8% (at 98.7 ng/dL to 7.5 ng/dL)Table (p.8)
    Linearity (R²)Implied: High R² value (e.g., >0.98 or >0.99) indicating a strong linear relationship.R² = 1.00Text (p.9)
    Regression Equation (y=mx+b)Implied: Slope (m) close to 1, intercept (b) close to 0.y = 1.00x - 1.24 (slope = 1.00, intercept = -1.24)Text (p.9)
    Limit of Blank (LoB)Implied: Low value to correctly identify absence of analyte.2.0 ng/dLTable (p.11)
    Limit of Detection (LoD)Implied: Low value to correctly detect presence of analyte.3.2 ng/dLTable (p.11)
    Limit of Quantitation (LoQ)Implied: Low value with acceptable precision (e.g., typically ≤20% CV).3.9 ng/dL (at 20% CV)Table (p.11)
    Interference (Concentration Bias)Explicitly stated criterion in study: ≤10% concentration bias to the unspiked sample.All tested interferents met this criterion.Text (p.11)
    Cross-ReactivityImplied: Low percentage for non-target analytes, 100% for target analyte.Aldosterone: 100%. Other analytes: 0.0003% to 3.1% (3α, 5β-Tetrahydroaldosterone: 3.1%)Table (p.13)
    Method Comparison (Slope vs. Predicate)Implied for substantial equivalence: Slope close to 1, intercept close to 0, high R².Linear Regression: Slope = 1.053, Intercept = 0.09 ng/dL, R² = 0.980Text (p.14)
    Passing-Bablok: Slope = 1.070, Intercept = -0.29Text (p.14)

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision/Reproducibility:

      • Sample Size: Six EDTA plasma samples (9 "samples" were run over 20 days, generating 80 replicates per sample concentration).
      • Data Provenance: Not explicitly stated, but clinical laboratory setting is implied. The phrase "two sites using three analyzers" could suggest multi-center testing, potentially from different locations/countries, but this is not specified. The studies were performed for the manufacturer for regulatory submission.
    • Linearity/Assay Reportable Range:

      • Sample Size: One high plasma sample diluted with a low sample to create 11 dilution samples. Run in quadruplicate.
      • Data Provenance: Not specified, likely internal laboratory data for the manufacturer.
    • Detection Limit (LoB, LoD, LoQ):

      • Sample Size:
        • LoB (Lot 1): One zero-aldosterone plasma sample, 10 replicates for 5 days (total 50 replicates).
        • LoD/LoQ (Lot 1): 7 samples, 2 replicates per sample, once per day for 8 days.
        • LoB (Lot 2): One zero-aldosterone plasma sample, 6 replicates for 5 days (total 30 replicates).
        • LoD (Lot 2): 8 samples, 2 replicates per sample, once per day for 5 days.
        • LoQ (Lot 2): 7 samples, 2 replicates per sample, once per day for 5 days.
      • Data Provenance: Not specified, likely internal laboratory data for the manufacturer. Two different sites and two different analyzers were used.
    • Analytical Specificity (Interference and Cross-Reactivity):

      • Sample Size:
        • Interference: Two base plasma samples ("Low" and "High" Aldosterone concentrations), spiked with various potential interferents. For each interferent, 26 replicates for blank and spiked samples were compared.
        • Cross-reactivity: Stock solutions of various compounds diluted serially to create 7-point standard curves for each substance.
      • Data Provenance: Not specified, likely internal laboratory data for the manufacturer.
    • Method Comparison:

      • Sample Size: 161 samples (including 12 altered samples).
      • Data Provenance: Not specified, implied to be clinical or patient samples. Whether these were retrospective or prospective, or their country of origin, is not mentioned.
    • Reference Range Study (Expected Values):

      • Sample Size: 228 Caucasian adult samples.
      • Data Provenance: Collected in the US, prospective (samples collected under specific conditions: 7-10 am after overnight fasting, upright/supine positions).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information is generally NOT applicable to in vitro diagnostic (IVD) devices like the Aldosterone assay described. For IVD assays, "ground truth" is typically established by:

    • Reference methods/predicate devices (e.g., Diasorin Liaison Aldosterone assay for method comparison).
    • Gravimetric preparation of standards and calibrators for traceability.
    • Defined sample characteristics (e.g., "zero aldosterone plasma," "high plasma sample").

    There were no human experts assessing images or making diagnoses that would require adjudication.


    4. Adjudication Method for the Test Set

    Not applicable, as this is an IVD assay, not a device requiring human interpretation of results in a diagnostic context that would call for adjudication of different interpretations.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an IVD assay for quantitative determination of a biomarker; there are no "human readers" interpreting results in the way an MRMC study would be designed for an imaging AI device.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies conducted (Precision, Linearity, LoD/LoQ, Analytical Specificity, Method Comparison) represent the standalone performance of the IDS-iSYS Aldosterone assay system. The device quantifies aldosterone in plasma samples directly, without human interpretation of results being a variable in its core analytical performance.


    7. The Type of Ground Truth Used

    The ground truth used for various studies includes:

    • Reference materials: Aldosterone (≥95% HPLC; A9477Sigma-Aldrich) dissolved in Dioxane, with concentration calculated by UV quantitation using molar extinction coefficient (for calibrators and traceability).
    • Predicate device results: The Diasorin Liaison Aldosterone assay (K130321) for method comparison.
    • Spiked samples: For linearity, interference, and cross-reactivity studies, known concentrations of analyte or interferents were added to base samples to create "true" concentrations.
    • Patient samples: For method comparison and reference range studies, patient samples were used. Their "truth" for method comparison was established by the predicate device. For reference ranges, the "truth" was the measured value distributed among a healthy population.

    8. The Sample Size for the Training Set

    The document describes performance studies (validation) but does not explicitly mention a "training set" in the context of machine learning. For an IVD assay, the development process involves reagent optimization, calibration curve fitting, etc., which conceptually use data, but this is not typically termed a "training set" like in AI/ML.

    However, the "Master curve" and "two-point calibration" system is mentioned. The "logistic parameters of the Master calibration curve" are generated using "data of 20 runs Internal Reference Calibrators" (p.9). This could be considered analogous to a training process for establishing the core measurement algorithm, but it's not a training set for an AI inference model.


    9. How the Ground Truth for the Training Set Was Established

    If we consider the generation of the "Master calibration curve" as a "training" process:

    • Ground Truth: The "Internal Reference Calibrators" referenced (p.9) would serve as the ground truth. These calibrators are traceable to gravimetrically prepared Aldosterone standards using high-purity Aldosterone and UV quantitation (p.9).
    • Establishment Method: The "new kit calibrator sets are run as 'unknowns' in duplicate in at least 20 assays on one analyser" (p.9). "The data of 20 runs Internal Reference Calibrators are used to generate the logistic parameters of the Master calibration curve by using Prism software package" (p.9).
    Ask a Question

    Ask a specific question about this device

    K Number
    K130321

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-04-09

    (60 days)

    Product Code
    Regulation Number
    862.1045
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIAISON® Aldosterone assay uses chemiluminescent immunoassay (CLIA) technology and is intended for the quantitative determination of Aldosterone in human serum. EDTA plasma and treated urine samples. Aldosterone measurements are intended for use in the diagnosis and treatment of primary aldosteronism (a disorder caused by excessive secretion of aldosterone by the adrenal gland), hypertension caused by primary aldosteronism, selective hypoaldosteronism, edematous states and other conditions of electrolyte imbalance. The test has to be performed on the LIAISON® Analyzer.

    The DiaSorin LIAISON® Aldosterone Control Set is intended for use as assayed quality control samples to monitor the accuracy of the DiaSorin LIAISON® Aldosterone assay on the LIAISON® Analyzer.

    The DiaSorin LIAISON® Aldosterone Calibration Verifiers are assayed quality control materials intended for the quantitative verification of calibration and reportable range of the LIAISON® Aldosterone assay when performed on the LIAISON® Analyzer.

    Device Description

    The LIAISON® Aldosterone assay is a competitive modified 2 step chemiluminescent assay that uses sheep monoclonal antibody for capture of the Aldosterone molecule. Results are determined by a 2 point calibration conversion of the master curve to a working curve. The light signal is measured by a photomultiplier as relative light units (RLU) and is inversely proportional to the concentration of aldosterone present in the calibrators, controls or samples.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the LIAISON® Aldosterone assay, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with specific pass/fail thresholds for the LIAISON® Aldosterone assay. However, it provides performance data that would implicitly be compared against desired clinical or analytical requirements. Below is a table summarizing the reported performance, which serves as evidence of the device meeting its intended analytical characteristics.

    Performance MetricReported Device Performance (LIAISON® Aldosterone)Clinical Context / Implied Acceptance
    Method Comparison (vs. Predicate RIA)High correlation and agreement with established method
    - Serum Samples (n=144)Slope: 0.98 (95% CI: 0.94 to 1.02)Slope near 1.0, indicating proportional agreement.
    Intercept: 1.10 ng/dL (95% CI: 0.43 - 1.49)Intercept near 0, indicating minimal constant bias.
    R: 0.988High correlation (close to 1).
    - Urine Samples (n=104)Slope: 0.98 (95% CI: 0.91 to 1.05)Slope near 1.0, indicating proportional agreement.
    Intercept: 34 ng/dL (95% CI: 11.43 to 56.7)Intercept near 0 (though 34 ng/dL for urine is noted).
    R: 0.948High correlation (close to 1).
    Limit of Blank (LoB)Serum: 0.97 ng/dLAbility to distinguish analyte absence from presence.
    Urine: 1.26 ng/dL
    Limit of Detection (LoD)Serum: 1.45 ng/dLLowest concentration detectable with reasonable certainty.
    Urine: 2.00 ng/dL
    Limit of Quantitation (LoQ)Serum: 3.0 ng/dLLowest concentration quantifiable with acceptable precision and accuracy.
    Urine: 2.80 ng/dL
    Precision (20-Day Reproducibility)Consistent and reproducible results across runs, days, sites, and reagent lots.
    - Serum (Aldo-S1 to Aldo-S6, n=480 each)Total %CV: 5.6% - 10.5%Low Coefficient of Variation (CV) indicating high precision.
    - Urine (Aldo-U1 to Aldo-U3, n=480 each)Total %CV: 8.6% - 9.8%Low Coefficient of Variation (CV) indicating high precision.
    Dilution LinearitySerum: Observed = 0.994(Expected) + 0.71, R = 1.000Linear response across dilution range. (Slope near 1, intercept near 0, R near 1)
    EDTA Plasma: Observed = 1.01(Expected) + 1.43, R = 0.998
    Urine: Observed = 0.996(Expected) + 0.69, R = 0.999
    Interfering SubstancesNo interference observed at specified concentrations for various substances.Robustness against common physiological and therapeutic interferences.
    Cross-Reactivity< 0.02% cross-reactivity for all tested substances.High specificity, minimizing false positives from similar compounds.

    2. Sample Sized Used for the Test Set and the Data Provenance

    • Method Comparison Test Set:

      • Serum: 155 samples (144 included in analysis as 11 were below measuring range).
      • Urine: 106 samples (104 included in analysis as 2 were above measuring range).
      • Data Provenance: Samples were "collected from apparently healthy individuals." A portion (approximately 15%) were spiked with aldosterone to span the measuring range. The study was conducted using the LIAISON® Aldosterone assay and the predicate RIA assay. Although not explicitly stated, the context of a 510(k) submission strongly implies these are likely prospective or retrospective clinical samples (for the healthy individuals) used in a controlled laboratory setting, potentially from the region where DiaSorin Inc. operates (Stillwater, MN, USA), though this is not explicitly stated.
    • LoB/LoD/LoQ Test Set: Not specified, but generally refers to a set of blank and low-concentration samples. The determination was performed according to CLSI EP17-A2.

    • Reference Range/Expected Values Test Set:

      • Serum and EDTA Plasma: 126 apparently healthy subjects.
      • Urine (24 hour): 91 individuals.
      • Data Provenance: Samples collected from apparently healthy subjects aged 21-65 years with normal blood pressure and normal fasting glucose levels. These are prospectively collected clinical samples following specific patient preparation protocols (fasting, upright/supine positions, 24-hour collection). No specific country of origin is mentioned.
    • Reproducibility/Precision Test Set:

      • Panel: 6 frozen serum samples and 3 frozen urine samples.
      • Controls: 2 levels of controls.
      • Data Provenance: A "coded panel" prepared by DiaSorin Inc. Tested at "DiaSorin Inc. and 2 external sites." This indicates laboratory-prepared samples tested in a multi-site, controlled study.
    • Dilution Linearity Test Set: Not specified, but samples of "each sample type" (serum, EDTA plasma, urine) were diluted. This implies laboratory-prepared or clinical samples serially diluted.

    • Interfering Substances & Cross-Reactivity Test Set: Not specified, but involved samples spiked with various substances at defined concentrations. These are laboratory-prepared samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of immunoassay (measuring a specific analyte concentration) does not typically involve human experts to establish "ground truth" in the same way an imaging or diagnostic AI does. The "ground truth" for this device's performance studies is established by:

    • Predicate Device: For method comparison, the Siemens Coat-A-Count® Aldosterone RIA assay (K831178) served as the reference standard. This is a previously cleared and established clinical assay, representing the accepted "truth" for aldosteronemeasurements at the time.
    • Known Concentrations: For studies like LoB/LoD/LoQ, Dilution Linearity, Interfering Substances, and Cross-Reactivity, the "ground truth" is based on known, prepared concentrations of the analyte and potential interfering/cross-reacting substances. These are analytically derived standards.
    • Statistical Analysis: CLSI guidelines (EP9-A2, EP17-A2, EP6-A) provide the statistical and methodological frameworks for establishing these analytical performance characteristics.

    Therefore, no information about human experts (e.g., radiologists) with specific qualifications or experience in establishing ground truth for this analytical device is provided or expected.

    4. Adjudication Method for the Test Set

    No adjudication method (e.g., 2+1, 3+1) is mentioned or applicable. As a quantitative immunoassay, the output is a numerical concentration. The comparison is between the new device's numerical result and the predicate device's numerical result (or known spike concentrations), not subjective interpretations requiring adjudicated consensus.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No MRMC comparative effectiveness study was performed or is applicable to this type of in vitro diagnostic device (immunoassay). This device directly measures an analyte concentration, and there is no "human reader" involved in interpreting its output in the context of an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies presented (Method Comparison, LoB/LoD/LoQ, Precision, Dilution Linearity, Interfering Substances, Cross-Reactivity) all represent standalone performance of the LIAISON® Aldosterone assay. The device provides a quantitative measurement directly, without human interpretation for the primary output. Its performance is evaluated purely on its analytical accuracy, precision, and specificity against established methods or known standards.

    7. The Type of Ground Truth Used

    The ground truth used in these studies consists of:

    • Predicate Device Measurements: For method comparison, the results from the Siemens Coat-A-Count® Aldosterone RIA assay served as the comparative ground truth.
    • Known/Expected Concentrations: For LoB/LoD/LoQ determinations, dilution linearity, interference, and cross-reactivity studies, the ground truth was based on analytically derived and known concentrations of aldosterone or other substances.
    • Biological Referencing: For the reference range studies, the "ground truth" was established by measuring aldosterone levels in clinically defined healthy populations under controlled conditions.

    8. The Sample Size for the Training Set

    No information about a "training set" is provided. This type of immunoassay (CLIA technology) is a chemical and biological measurement system, not a machine learning or AI algorithm that typically has a distinct training phase requiring a training dataset. The device's operational parameters (e.g., calibration, reagent concentrations, reaction times) are established through analytical development and optimization, rather than machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" in the context of a machine learning algorithm, this question is not applicable. The device's functionality is based on its underlying chemical and immunoassay principles, which are developed and validated through a series of analytical studies as described in the performance data section, rather than through a machine learning training process.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050784

    Validate with FDA (Live)

    Date Cleared
    2005-06-01

    (65 days)

    Product Code
    Regulation Number
    862.1045
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nichols Advantage Aldosterone Assay is intended for in vitro diagnostic laboratory use with the Nichols Advantage® Specialty System for quantitative measurement of aldosterone in human serum, EDTA plasma, and extracted urine. Aldosterone measurements are intended for use in the diagnosis and treatment of primary aldosteronism (a disorder caused by excessive secretion of aldosterone by the adrenal gland), hypertension caused by primary aldosteronism, selective hypoaldosteronism, edematous states, and other conditions of electrolyte imbalance.

    Device Description

    The Nichols Advantage® Aldosterone Assay is a competitive immunochemiluminometric in vitro diagnostic laboratory immunoassay (IVD device) that utilizes a biotinylated mouse monoclonal anti-aldosterone antibody as the capture reagent and an acridinium ester labeled aldosterone as a tracer reagent. This Aldosterone IVD device immunoassay is intended for use for the measurement of aldosterone in human serum, EDTA plasma, and extracted urine, as an extended diagnostic method utilized within the Nichols Advantage® Specialty System.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Nichols Advantage® Aldosterone Assay, focusing on acceptance criteria and study details:

    Acceptance Criteria and Device Performance

    The submission doesn't explicitly state quantitative acceptance criteria in a dedicated section. Instead, the performance characteristics are presented as evidence of the device's capabilities and are implicitly compared to the predicate device or general laboratory expectations. The method comparison to a predicate device serves as the primary "acceptance criterion" for demonstrating substantial equivalence.

    Implicit Acceptance Criteria and Reported Device Performance:

    Performance MetricAcceptance Criteria (Implicit / Contextual)Reported Device Performance (Nichols Advantage® Aldosterone Assay)Predicate Device (DSL-8600 ACTIVE® Aldosterone Coated-Tube Radioimmunoassay Kit)
    Method Comparison (Urine)
    Correlation (Pearson's r)High correlation (e.g., >0.90) with predicate device0.96 (95% CI: 0.94 to 0.97)N/A (this is the comparator)
    Slope (Passing Bablok)Should be close to 1 (indicating proportional agreement)1.23 (95% CI: 1.2 to 1.28)N/A
    Intercept (Passing Bablok)Should be close to 0 (indicating constant agreement)-1.19 (95% CI: -1.43 to -0.81)N/A
    Range of Values (Urine)Comparable to clinical needs and predicate device0.4 to 66.7 µg/24 Hr0.8 to 80.2 µg/24 Hr
    Analytical Performance
    Analytical SensitivitySufficient for clinical measurement (comparable or better than predicate)1.2 ng aldosterone/dL0.7 ng aldosterone/dL
    Within-run (%CV)Low variability (e.g., <10%)1.6 to 6.1%3.3 to 7.4%
    Total Precision (%CV)Low variability (e.g., <15-20%)10.2 to 15.3%5.1 to 6.4%
    Recovery (%)Close to 100% (e.g., 90-110%)93% to 110%92% to 138%
    Linearity (%)Close to 100% (e.g., 80-120%)85% to 116%89% to 111%
    SpecificityLimited or no cross-reactivity with structurally similar compoundsUndetectable for most tested compounds; 0.85% for EstradiolNot specified in detail, but assumed to be acceptable
    Urine Extraction EfficiencyClose to 100% recovery for spiked samples98% to 103%Not explicitly detailed for predicate
    Expected Values (Reference Range)Establishment of a normal reference range0.7 to 23.0 µg/24 hours (for 24-hr urine)Not explicitly detailed for predicate

    Study Information:

    The primary "study" described is a method comparison and performance characteristic evaluation against a legally marketed predicate device.

    1. Sample sizes used for the test set and data provenance:

      • Method Comparison (Urine): n = 118 urine samples.
      • Expected Values (Urine): n = 80 healthy adults (41 females, 39 males, age: 18 to 78 years).
      • Reproducibility for Urine: 4 urine samples, assayed in duplicate, once per day over 20 days.
      • Parallelism for Urine: 5 urine samples, serially diluted and assayed in duplicate.
      • Recovery for Urine: 3 sets of high and low urine samples mixed in various ratios, assayed in duplicate.
      • Urine Extraction Efficiency: 3 spiked urine samples, tested in n=8 replicates per level.
      • Data Provenance: Not explicitly stated, but given the manufacturer (Nichols Institute Diagnostics) and context, it is highly likely that the data was collected retrospectively or prospectively as part of an internal validation study for the device, presumably in the US or a region with comparable regulatory standards. It is retrospective in the sense that existing urine samples might have been used, but the testing itself would have been a prospective application of both assays.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This is an in vitro diagnostic immunoassay, not an AI/medical imaging device that requires radiologist or clinician interpretation for establishing ground truth for individual cases. The "ground truth" for the test set (the 118 urine samples) is the measurement obtained from the predicate device (DSL-8600 ACTIVE® Aldosterone Coated-Tube Radioimmunoassay Kit), which is itself a laboratory assay.
      • For the Expected Values study, "80 healthy, prescription expected forchoo fangedults (41 females and 39 males, age:18 to 78 years)" were used. The 'ground truth' here is their healthy status, presumably determined by routine medical screening, not by expert consensus on specific diagnostic cases.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. This is not a study involving human readers or expert consensus on case interpretation that would require adjudication. The method comparison uses quantitative measurements from two different laboratory assays.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is an IVD device, not an AI algorithm for image interpretation or diagnosis that would involve human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, this is an intrinsic performance study of the standalone device (the Nichols Advantage® Aldosterone Assay). Its performance is evaluated objectively through analytical characteristics (precision, recovery, linearity, specificity) and compared to a predicate device. There is no human-in-the-loop component in the measurement process itself; a laboratory technician operates the system, but the device provides the quantitative result.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the method comparison study, the "ground truth" for the 118 urine samples was the measurement obtained from the predicate device (DSL-8600 ACTIVE® Aldosterone Coated-Tube Radioimmunoassay Kit).
      • For other analytical performance studies (precision, recovery, linearity, specificity, extraction efficiency), the "ground truth" is established through known concentrations of analytes (e.g., spiked samples, reference materials) or by the absence/presence of cross-reactants.
      • For establishing the Expected Values, the "ground truth" was the clinical health status (healthy individuals) of the sampled population.
    7. The sample size for the training set:

      • Not applicable in the conventional sense of machine learning. This is a traditional immunoassay, not an AI/ML algorithm that requires a "training set" to learn from data. The assay's performance is based on its chemical and immunological design.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" in the context of an immunoassay. The device is built based on established biochemical principles.
    Ask a Question

    Ask a specific question about this device

    K Number
    K032188

    Validate with FDA (Live)

    Date Cleared
    2003-07-31

    (14 days)

    Product Code
    Regulation Number
    862.1045
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nichols Advantage® Aldosterone assay is intended for use with the Nichols Advantage® Specialty System to quantitatively measure aldosterone in human serum and EDTA plasma. Aldosterone measurements are used in the diagnosis and treatment of primary aldosteronism (a disorder caused by excessive secretion of aldosterone by the adrenal gland), hypertension caused by primary aldosteronism, selective hypoaldosteronism, edematous states, and other conditions of electrolyte imbalance.

    Device Description

    The Nichols Advantage® Aldosterone assay contains sufficient reagents for 100 tests. The assay is a competitive binding assay for aldosterone in human serum or plasma.

    AI/ML Overview

    The Nichols Advantage® Aldosterone assay is a competitive binding immunoassay intended for quantitative measurement of aldosterone in human serum and EDTA plasma using the Nichols Advantage Specialty System. Aldosterone measurements are used in the diagnosis and treatment of conditions such as primary aldosteronism, hypertension caused by primary aldosteronism, selective hypoaldosteronism, edematous states, and other electrolyte imbalances.

    The device's performance was compared to a predicate device, the DPC Coat-A-Count Aldosterone RIA (K831178), to establish substantial equivalence.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a quantitative sense for all features. However, it provides comparative performance characteristics between the Nichols Advantage Aldosterone and the predicate device. The implied acceptance is that the Nichols Aldosterone performs comparably to the predicate.

    FeaturePredicate Device (DPC Aldosterone)Nichols Advantage Aldosterone
    Comparison Study (Quantitative)
    Range (Method X - DPC)2.7 to 125 ng/dLNot directly stated (Y=1.04X+0.1)
    Range (Method Y - Nichols)Not directly stated2.7 to 120 ng/dL
    Regression (Passing Bablok)Y = 1.04X + 0.1N/A
    (95% CI Slope)(0.98 to 1.10)N/A
    (95% CI Intercept)(-1.0 to +1.1)N/A
    Regression (Deming)Y = 1.09X - 0.6N/A
    (95% CI Slope)(1.03 to 1.15)N/A
    (95% CI Intercept)(-3.2 to +2.1)N/A
    Pearson's Correlation (r)N/A0.96
    Performance Characteristics
    Within-Run Precision (%CV)2.3-5.4%2.9-14.0%
    Total Precision (%CV)3.8-15.7%4.9-18.6%
    Recovery86-111%88-110%
    Linearity100-119%91-116%
    Analytical Sensitivity1.1 ng/dL1.2 ng/dL

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: One hundred three (103) remnant serum samples.
    • Data Provenance: The clinical diagnosis for these samples was unknown. The document does not specify the country of origin, but given the manufacturer's location (San Clemente, CA) and the FDA submission, it's highly likely to be U.S.-based. The samples were "remnant," suggesting a retrospective collection.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    Not applicable. This device is an in vitro diagnostic immunoassay measuring an analyte concentration. The "ground truth" for the test set was the measurement result obtained by the predicate device (DPC Coat-A-Count Aldosterone RIA), not expert consensus on an image or clinical observation.

    4. Adjudication Method for the Test Set:

    Not applicable. The study involved a direct comparison of quantitative measurements from two immunoassay methods on the same samples. There was no need for expert adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not conducted as this is an in vitro diagnostic device for quantitative measurement, not an imaging device requiring human reader interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, the core of the submission is a standalone performance evaluation of the Nichols Advantage Aldosterone assay, comparing its quantitative results to a legally marketed predicate device without human-in-the-loop interaction for result interpretation beyond running the assay.

    7. The Type of Ground Truth Used:

    The ground truth for comparison was the quantitative measurement obtained from the predicate device (DPC Coat-A-Count Aldosterone RIA) on the same serum samples. This is a common approach for demonstrating substantial equivalence for new IVD devices by comparing them to an established, legally marketed assay.

    8. The Sample Size for the Training Set:

    Not applicable. This document describes the validation of an immunoassay kit, not a machine learning algorithm that requires a "training set" in the conventional sense. The development of the assay would have involved internal optimization and validation, but this specific submission focuses on the performance comparison to the predicate.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no mention of a "training set" in the context of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1