Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    K Number
    K232904

    Validate with FDA (Live)

    Device Name
    Access Ostase
    Manufacturer
    Date Cleared
    2024-04-15

    (210 days)

    Product Code
    Regulation Number
    862.1050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Ostase assay is a paramagnetic particle, chemiluminescent immunoassay for use with the Access Immunoassay Systems for the quantitative measurement of bone alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum and plasma. This device is intended to be used as an aid in the management of postmenopausal osteoporosis and Paget's disease.

    Device Description

    The Access Ostase assay is a one-step sandwich immunoenzymatic assay. The Access Ostase assay consists of the reagent pack, calibrators and QCs. Other items needed to run the assay include substrate and wash buffer. The Access Ostase assay reagent pack, Access Ostase assay calibrators, Access Ostase QCs, along with the UniCel Dxl Wash Buffer II are designed for use with the Dxl 9000 Access Immunoassay Analyzer in a clinical laboratory setting.

    AI/ML Overview

    The provided text describes the performance of the Beckman Coulter Access Ostase assay on the Dxl 9000 Access Immunoassay Analyzer, comparing it to the predicate device (Access Ostase on Access 2 Immunoassay System).

    Here's an analysis of the acceptance criteria and the studies performed, structured as requested:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" for each performance characteristic in a separate, clear table format. However, it does outline the assay's design goals for imprecision and then presents the results. For other parameters like linearity and detection limits, it states the claimed values or that linearity was demonstrated. I will infer the acceptance criteria for imprecision from the "designed to have" statement and present the reported performance.

    ParameterAcceptance CriteriaReported Device Performance (Access Ostase on Dxl 9000)
    Method ComparisonImplicit: Close statistical agreement (slope near 1, intercept near 0, high correlation) with the predicate device (Access Ostase on Access 2 Immunoassay System). While not explicitly stated as an "acceptance criteria," the results indicate substantial equivalence.N: 163 samples Concentration Range: 0.34 - 108 µg/L Slope: 0.95 (95% CI: 0.93 - 0.98) Intercept: 0.53 (95% CI: 0.32 - 0.75) Correlation Coefficient (R): 1.00
    Imprecision≤ 0.2 µg/L SD at concentrations ≤ 3 µg/L ≤ 7.0% CV at concentrations > 3 µg/L (within-laboratory imprecision)Sample 1 (Mean 1.8 µg/L): SD 0.1, %CV 4.7 Sample 2 (Mean 9.1 µg/L): SD 0.4, %CV 4.1 Sample 3 (Mean 25 µg/L): SD 1.1, %CV 4.4 Sample 4 (Mean 38 µg/L): SD 1.6, %CV 4.3 Sample 5 (Mean 98 µg/L): SD 4.1, %CV 4.2 All reported within-laboratory %CVs are ≤ 7.0% and the SD for Sample 1 (1.8 µg/L) is 0.1 µg/L, meeting the criteria.
    LinearityImplicit: Demonstrate linearity across the measuring interval.Demonstrated linearity across the measuring interval.
    Limit of Blank (LoB)Claimed LoB of 0.1 µg/L.Claimed LoB for Access Ostase assay is 0.1 µg/L on Dxl 9000 Access Immunoassay Analyzer.
    Limit of Detection (LoD)Claimed LoD of 0.1 µg/L.Claimed LoD for Access Ostase assay is 0.1 µg/L on Dxl 9000 Access Immunoassay Analyzer.
    Limit of Quantitation (LoQ)Claimed LoQ of 0.3 µg/L.Claimed LoQ for Access Ostase assay is 0.3 µg/L on Dxl 9000 Access Immunoassay Analyzer.

    2. Sample size used for the test set and the data provenance:

    • Method Comparison Test Set:

      • Sample Size: 163
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the study involved comparing results from two different immunoassay systems (Access 2 and Dxl 9000) using patient samples, implying real-world or simulated clinical samples.
    • Imprecision Test Set:

      • Sample Size: 80 for each of the 5 samples tested (total of 400 individual measurements of samples).
      • Data Provenance: Not explicitly stated. The study involved testing "multiple samples in duplicate in 2 runs per day for a minimum of 20 days," which suggests a controlled laboratory setting.
    • Linearity, LoB, LoD, LoQ: The sample sizes for these studies are not explicitly mentioned, nor is their specific provenance beyond being performed "on the Dxl 9000 Access Immunoassay Analyzer."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable to an in vitro diagnostic (IVD) assay that measures quantitative analytes like bone alkaline phosphatase. The "ground truth" for such devices is typically established through reference methods, certified reference materials, or statistical comparison to a legally marketed predicate device (as done here). There are no "experts" in the human perception or diagnostic interpretation sense used to establish ground truth for this type of test.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation (e.g., radiology reads) where discrepancies between experts need to be resolved to establish ground truth. This is a quantitative immunoassay.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an in vitro diagnostic immunoassay, not an AI-assisted diagnostic tool for human readers. No MRMC study was performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This device is an automated immunoassay system that provides quantitative results. The performance studies described (Method Comparison, Imprecision, Linearity, LoB/LoD/LoQ) inherently represent the standalone performance of the device (including its reagents and analyzer software) in generating these quantitative results. It operates without direct human-in-the-loop interpretation of the primary measurement signal for the reported value. A human reviews the final numerical result, but the device itself generates that result primarily through an automated process.

    7. The type of ground truth used:

    • Method Comparison: The ground truth was established by comparison to the predicate device (Access Ostase on Access 2 Immunoassay System). This assumes the predicate device's measurements represent a valid "truth" for substantial equivalence purposes. Patient samples were used, and their values from one system were compared to the other.
    • Imprecision: The ground truth for evaluating imprecision is the inherent variability of the measurement process itself, determined by repeated measurements of samples with known (or established) concentrations.
    • Linearity, LoB, LoD, LoQ: These are determined using samples of known concentrations (dilutions, blank samples, low-concentration samples) and statistical methods based on CLSI guidelines.

    8. The sample size for the training set:

    Not applicable. This device is a quantitative immunoassay system, not a machine learning or AI model that requires a "training set" in the conventional sense. The development of such assays involves extensive research, development, and optimization of reagents and protocols, but this is distinct from "training data" for an algorithm.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no "training set" in the context of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200475

    Validate with FDA (Live)

    Date Cleared
    2020-09-30

    (217 days)

    Product Code
    Regulation Number
    862.1050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS-iSYS Ostase® BAP assay is an in vitro diagnostic device intended for the quantitative determination of bone-specific alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum on the IDS system. Results are to be used in conjunction with other clinical and laboratory data to aid the clinician in the management of postmenopausal osteoporosis and Paget's disease.

    Device Description

    The IDS-iSYS Ostase® BAP assay consists of one reagent cartridge and one set of calibrators (CAL A & CAL B).

    The reagent cartridge contains multiple reagents:

    • MPM1 (Magnetic particles coated with streptavidin in a phosphate buffer with sodium azide as preservative);
    • Ab-BIOT Monoclonal anti-BAP labelled with biotin, in buffer containing horse serum with bovine and mouse proteins and sodium azide as a preservative (<0.1 %)
    • -SUBS (p-nitrophenyl phosphate in a stabilising buffer containing preservatives).

    Calibrators A and B are buffered bovine protein matrix containing human BAP with sodium azide as preservative (<0.1 %).

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a medical device called "IDS-iSYS Ostase® BAP." This document is primarily concerned with demonstrating the substantial equivalence of the new device to a predicate device, rather than focusing on the acceptance criteria and study proving performance for an AI/ML-based medical device.

    Therefore, the requested information regarding AI/ML-specific acceptance criteria and validation studies (like sample size for test set, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) is not present in this document because it describes an in vitro diagnostic (IVD) test for quantitative determination of bone-specific alkaline phosphatase, not an AI/ML-driven imaging or diagnostic algorithm.

    However, I can extract the relevant performance characteristics that are analogous to "acceptance criteria" for this type of IVD device and the study data proving it meets those.


    Device Description

    The IDS-iSYS Ostase® BAP assay is an in vitro diagnostic device intended for the quantitative determination of bone-specific alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum on the IDS system. Results are to be used in conjunction with other clinical and laboratory data to aid the clinician in the management of postmenopausal osteoporosis and Paget's disease.


    1. Table of Acceptance Criteria (Performance Characteristics) and Reported Device Performance

    For an in vitro diagnostic device, "acceptance criteria" are typically defined by various analytical performance characteristics that demonstrate the device's accuracy, precision, and reliability for its intended use. While explicit acceptance ranges are not always presented as a separate "criteria" table in a 510(k), the studies conducted implicitly aim to demonstrate performance within acceptable ranges for IVDs. The comparison to the predicate device and adherence to CLSI guidelines are key for demonstrating substantial equivalence.

    Here's a table summarizing the analytical performance characteristics and the reported device performance, which serve as the "proof" that the device meets the implied acceptance criteria for an IVD:

    Performance CharacteristicAcceptance Criteria (Implied / Comparator)Reported Device Performance (IDS-iSYS Ostase® BAP)
    PrecisionTypically aims for low %CV (Coefficient of Variation) within and between runs, indicating reproducibility. Compared to predicate: - Predicate Within Run: 2.6% to 6.5% (7.4 to 79.5 μg/L) - Predicate Between Run: 2% to 6.4% (8.4 to 81.1 µg/L)Repeatability (Within Run): From 1.7% to 2.8% in the concentration range 6.2 to 59.8 µg/L (N=80 data points for each sample, for one representative lot). From 1.7% to 2.8% for combined 3 lots across samples 1-10 (6.2-59.8 μg/L). Within Laboratory (Between Run/Total Precision): From 3.0% to 7.6% in the concentration range 6.2 to 59.8 µg/L (N=80 data points for each sample, for one representative lot). From 3.0% to 7.2% for combined 3 lots across samples 1-10 (6.2-59.8 μg/L). Overall, shows robust precision comparable or improved relative to the predicate.
    Linearity / Reportable RangeThe assay should demonstrate linearity across its claimed measuring range. Expected a high correlation coefficient (R²) close to 1.0. - Predicate Range: 0.7 – 90 µg/L.Linear Range: 0.9 to 78.5 µg/L. Measuring (Reportable) Range: 3 to 70 µg/L. Regression: Observed = 0.98 x (Expected) - 0.9 ng/mL. Regression coefficient R²: 1.00. The high R² demonstrates excellent linearity across the range.
    Detection Limits (LoB, LoD, LoQ)Low values for LoB (Limit of Blank), LoD (Limit of Detection), and LoQ (Limit of Quantitation) are desirable to demonstrate the ability to detect very low concentrations of the analyte. - Predicate: LoD 0.7 µg/L. Other values N/A.LoB (Limit of Blank): 0.3 µg/L LoD (Limit of Detection): 0.4 µg/L LoQ (Limit of Quantitation): 0.5 µg/L Demonstrates improved or comparable sensitivity to the predicate.
    Analytical Specificity (Interference)Bias due to common interfering substances should be non-significant, typically defined as <10% bias between test and control samples.Non-significant interference (<10% bias) observed for: Acetaminophen, Alendronate, Bilirubin (Conjugated & Unconjugated), Biotin, Calcium Chloride, Cholesterol, Estradiol, Etidronate, Haemoglobin, HAMA, Ibuprofen, Pamidronate, Progesterone, PTH 1-34, PTH 1-84, Raloxifene, Red Blood Cells, Rheumatoid Factor (RF), Risedronate, Salicylic Acid (Asprin), Salmon Calcitonin, Total Protein, Triglycerides, 25-hydroxyvitamin D. Tested at specified high concentrations (e.g., Bilirubin 40 mg/dL, Biotin 400 ng/mL, Haemoglobin 300 mg/dL etc.).
    Analytical Specificity (Cross-Reactivity)Low percent cross-reactivity with structurally similar substances is desired to ensure specificity to BAP.Low cross-reactivity observed for: - Liver ALP (745 µg/L spiked): 0.1% - Placental ALP (90 U/L spiked): 0.5% - Intestinal ALP (500 µg/L spiked): Undetectable Demonstrates high specificity to BAP.
    Method ComparisonStrong correlation and agreement with the legally marketed predicate device (Ostase® BAP EIA) is required for substantial equivalence. A slope close to 1.0 and an intercept close to 0, with a high correlation coefficient (r) close to 1.0, are expected. - Predicate comparison to Tandem-R Ostase: slope 1.02, intercept 0.28, r = 0.97.Comparison against Ostase® BAP EIA: - N: 150 samples - Slope: 0.99 (95% CI: 0.97 to 1.02) - Intercept: 0.17 µg/L (95% CI: -0.1 to 0.5) - Correlation Coefficient (r): 0.99 These results demonstrate excellent agreement and strong correlation, supporting substantial equivalence.
    StabilityThe device should maintain its performance over a defined shelf life.Shelf Life: 12 months, determined based on real-time studies.
    Expected Values / Reference RangeEstablished ranges for different populations. Consistency with established clinical understanding.Population-specific BAP Concentrations (µg/L) for IDS-iSYS Ostase® BAP: - Males (N=140): Mean 13.7, Median 13.0, Observed Range 7.9 to 23.5 - Pre-menopausal (N=140): Mean 11.5, Median 11.1, Observed Range 5.9 to 20.5 - Post-menopausal (N=139): Mean 15.7, Median 14.3, Observed Range 7.9 to 34.2 These values are consistent with the general understanding of BAP levels in these populations, though laboratories are advised to establish their own ranges.

    Study Details (Based on the Provided Document)

    As this is an IVD device, not an AI/ML system, many of the AI/ML-specific questions are not applicable. However, I will address what is implied or directly stated in the context of an IVD submission.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Precision/Reproducibility:
      • 10 native serum samples.
      • Each sample assayed 80 replicates (in duplicate, twice per day for 20 days) on one system (representative lot) and 240 replicates (combined from 3 lots, 3 systems).
      • Data provenance not explicitly stated (e.g., country of origin, retrospective/prospective), but implied to be laboratory-based analytical studies.
    • Linearity:
      • Not specified how many serum samples (a high and a low human serum sample were used to create 11 dilutions, but the N for testing these dilutions is not given).
    • Detection Limits (LoB, LoD, LoQ):
      • LoB: 60 replicates of blank per kit lot (total 180 over 3 lots).
      • LoD & LoQ: 10 low BAP concentration samples tested per kit lot (details on number of replicates not fully clear, mentions tested in duplicate for 5 days per lot).
    • Analytical Specificity (Interference/Cross-Reactivity):
      • For interference: Two serum samples at two different BAP concentrations were spiked with potential interferents. For each condition, 26 replicate assays were performed for spiked and control samples.
      • For cross-reactivity: Not explicitly stated, but substances were spiked into serum samples and measured.
    • Method Comparison:
      • N = 150 samples.
      • Samples were "selected to represent a wide range of BAP concentrations [3.0 to 67.6 ug/L]".
      • Data provenance not explicitly stated (e.g., country of origin, retrospective/prospective).
    • Expected Values/Reference Range:
      • N = 419 apparently healthy donors from the United States.
      • 140 males (35 to 75 years of age)
      • 140 pre-menopausal women (35 to 45 years of age)
      • 139 post-menopausal women (55 to 75 years of age).
      • This is a prospective collection of reference interval data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Not applicable in the AI/ML sense. For IVD tests measuring a biomarker like BAP, the "ground truth" is typically the quantitative value obtained from a reference method or the assigned values of traceable calibrator materials. No human expert "adjudication" or "ground truth labeling" of the BAP level in a sample is described as a part of establishing test set ground truth; rather, it’s about the measured BAP concentration.
    • The calibrators for the new device are traceable to the predicate device via internal standards that were value-assigned using the predicate assay procedure. This method of traceability establishes the "truth" for calibration.

    4. Adjudication Method for the Test Set:

    • Not applicable. As above, for an IVD, the measurement itself is the "truth" within the context of the assay's analytical performance. No human adjudication is involved in determining the concentration of BAP in a blood sample for these types of studies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No. MRMC studies are specific to image-based diagnostic aids where multiple human readers' performance is evaluated and compared with and without AI assistance. This document describes an in vitro diagnostic assay that quantifies a substance in human serum, not an imaging device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Yes, in essence. For an IVD, the "standalone" performance is the analytical performance of the assay itself – its precision, linearity, detection limits, specificity, and comparison to a predicate device. The performance characteristics described in Section 1 (Precision, Linearity, Detection Limits, Analytical Specificity, Method Comparison) represent the standalone performance of the IDS-iSYS Ostase® BAP assay. It is designed to provide a quantitative result without direct human "interpretation" of the assay's output in the way an AI would assist a radiologist.

    7. The Type of Ground Truth Used:

    • Analytical Ground Truth (for analytical performance): For precision, linearity, detection limits, and specificity, the "ground truth" refers to:
      • The known properties of control materials or spiked samples.
      • The measured values from samples used in reproducibility studies.
      • The established values/performance of the predicate device (for comparison studies).
    • Reference Intervals (for clinical context): For expected values, the "ground truth" is derived from a defined cohort of apparently healthy individuals, and ranges are statistically determined (e.g., 2.5th to 97.5th percentile).

    8. The Sample Size for the Training Set:

    • Not applicable (in the AI/ML sense). This device is an IVD assay, not an AI/ML model that requires training on a vast dataset. The "training" for an IVD involves assay development, reagent optimization, and establishing manufacturing controls, not statistical model training based on patient data.
    • The calibrators for the device are value-assigned using a method traceable to the predicate device. This process involves testing calibrators in multiple runs (minimum 20 assay runs on one system) to determine their assigned values. This could be considered analogous to "training" the assay system to accurately report concentrations based on known standards.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable (in the AI/ML sense). However, for an IVD, the "ground truth" for calibrator and control materials is established through:
      • Traceability: The IDS-iSYS Ostase® BAP kit calibrators are value assigned against in-house secondary standards (IRs). These IRs are themselves value assigned against the predicate device (Ostase® BAP EIA assay) using the predicate assay procedure. This establishes traceability to a previously cleared, legally marketed device.
      • Verification: The assigned kit calibrator values are then verified by running the assay on three different IDS systems and analyzing internal quality control (IQC) samples of known values across the range of the assay.
    Ask a Question

    Ask a specific question about this device

    K Number
    K043125

    Validate with FDA (Live)

    Date Cleared
    2005-05-04

    (173 days)

    Product Code
    Regulation Number
    862.1050
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Orion Diagnostica UniQ™ PINP RIA is a quantitative radioimmunoassay designed for the measurement of intact aminoterminal propeptide of type I procollagen, an indicator of osteoblastic activity, in human serum. The test is intended to be used an aid in the management of postmenopausal osteoporosis. For in vitro diagnostic use.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is related to an FDA 510(k) clearance for a diagnostic device, the Orion Diagnostica UniQ™ PINP RIA. However, it does not contain any information regarding acceptance criteria, device performance, specific study details (sample size, data provenance, expert qualifications, adjudication methods), MRMC studies, standalone performance, or training set details.

    The document is a clearance letter and an "Indications for Use" statement. It confirms that the device has been found substantially equivalent to a predicate device but does not present the data from the studies that led to this determination.

    Therefore, I cannot provide the requested information based on the given input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020625

    Validate with FDA (Live)

    Date Cleared
    2002-05-06

    (69 days)

    Product Code
    Regulation Number
    862.1050
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The test system will be used in laboratories to separate serum ALP isoenzymes. Elevated bone isoenzyme can be found in conditions of physiologic bone growth, metastatic carcinoma and as rickets. Increased liver isoenzyme values can be seen in hepatitis and cirrhosis of the liver. Intestinal isoenzymes can be seen in patients with intestinal disease in particular after a meal. Intestinal Isoenzymes occur in blood type O or B after a meal.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for the SPIFE 2000/3000 ALP device. It does not contain information about the acceptance criteria or a study that proves the device meets the acceptance criteria.

    The letter states that the FDA has determined the device to be substantially equivalent to legally marketed predicate devices. This determination is based on a comparison to existing devices, not on a new study providing specific performance metrics against pre-defined acceptance criteria.

    Therefore, I cannot extract the requested information from the provided text. The document is primarily an administrative letter confirming regulatory clearance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K011113

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2001-06-13

    (63 days)

    Product Code
    Regulation Number
    862.1050
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K994278

    Validate with FDA (Live)

    Date Cleared
    2000-03-28

    (99 days)

    Product Code
    Regulation Number
    862.1050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Beckman Coulter, Inc's Access Ostase Assay is a paramagnetic particle, chemiluminescent immunoassay for use with the Access Immunoassay System for the quantitative measurement of bone alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum and plasma. This device is intended to be used as an aid in the management of postmenopausal osteoporosis and Paget's disease.

    Device Description

    The Access Ostase Assay is a paramagnetic particle, chemiluminescent immunoassay for use with the Access Immunoassay System for the quantitative measurement of bone alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum and plasma.

    AI/ML Overview
    {
      "acceptance_criteria_and_device_performance": {
        "title": "Acceptance Criteria and Reported Device Performance for Access® Ostase® Assay",
        "table": {
          "headers": ["Performance Characteristic", "Acceptance Criteria (Implicit via predicate comparison)", "Reported Device Performance"],
          "rows": [
            ["Method Comparison (Slope vs. Predicate)", "Close to 1", "0.9756"],
            ["Method Comparison (Intercept vs. Predicate)", "Close to 0", "-0.5987"],
            ["Method Comparison (Correlation r vs. Predicate)", "Close to 1", "0.9895"],
            ["Mean % Difference (Osteoporosis vs. Predicate)", "Low Percentage Difference", "-5.96%"],
            ["Mean % Difference (Paget's vs. Predicate)", "Low Percentage Difference", "-7.01%"],
            ["Between-Run %CV (Imprecision)", "Low Percentage (e.g., <10%)", "3.3% to 5.9%"],
            ["Within-Run %CV (Imprecision)", "Low Percentage (e.g., <10%)", "1.5% to 2.6%"],
            ["Total %CV (Imprecision)", "Low Percentage (e.g., <10%)", "3.6% to 6.4%"],
            ["Recovery (Spike and Recovery)", "Close to 100%", "Average 92.4% (range 89.9% to 95.3%)"],
            ["Recovery (Specimen Dilution Average)", "Close to 100%", "Average 92.2% (range 78.1% to 106.1%)"],
            ["Interference (Absence of significant interference)", "No significant interference", "No significant interference from tested substances"]
          ]
        },
        "notes": "The acceptance criteria for the Access® Ostase® Assay are implicitly defined by its substantial equivalence claim to the Tandem®-R Ostase® Assay predicate device. The reported performance data demonstrate close agreement, high correlation, and low variability, indicating that the device performs comparably to the predicate and meets general expectations for diagnostic assays."
      },
      "study_details": {
        "sample_size_test_set": {
          "method_comparison": "172 samples (88 osteoporosis, 84 Paget's)",
          "imprecision": "Across various concentrations (exact number of individual samples not specified, but multiple runs and concentrations were evaluated).",
          "recovery_spike": "15 samples",
          "recovery_dilution": "10 samples",
          "interfering_substances": "Not explicitly stated for the test set, but substances were tested at specified concentrations.",
          "data_provenance": "Not explicitly stated for the test set. However, a crucial statement indicates that all clinical studies were generated using the Tandem-R Ostase Assay (the predicate device), implying that the clinical utility data, rather than the analytical performance data, came from studies on the predicate."
        },
        "number_of_experts_ground_truth_test_set": "Not applicable; this is a quantitative immunoassay comparing a new device to a predicate device, not a qualitative diagnostic interpretation requiring expert consensus on images or similar data.",
        "qualifications_of_experts": "Not applicable.",
        "adjudication_method": "Not applicable.",
        "multi_reader_multi_case_comparative_effectiveness_study": "No, this is an in vitro diagnostic (IVD) device for quantitative measurement of a biomarker, not an imaging or interpretive device requiring human-in-the-loop assessment.",
        "standalone_performance": "Yes, the study presents the standalone performance of the Access® Ostase® Assay by comparing its measurements directly to those of the predicate device (Tandem®-R Ostase® Assay) and evaluating its analytical characteristics (imprecision, recovery, interference).",
        "type_of_ground_truth_test_set": "The 'ground truth' for the analytical performance studies (method comparison, imprecision, etc.) is the quantitative measurement provided by the established predicate device (Tandem®-R Ostase® Assay) and established analytical methodologies. For clinical utility, the 'ground truth' for indications (managing osteoporosis and Paget's disease) was established by studies performed using the predicate device, which were then referenced for the new device through the substantial equivalence pathway.",
        "sample_size_training_set": "Not applicable for this type of IVD device. The device is not based on a machine learning algorithm that requires a dedicated training set in the typical sense. Performance is established through analytical validation against a predicate.",
        "ground_truth_for_training_set_established": "Not applicable."
      }
    }
    
    Ask a Question

    Ask a specific question about this device

    K Number
    K981373

    Validate with FDA (Live)

    Date Cleared
    1998-05-27

    (41 days)

    Product Code
    Regulation Number
    862.1050
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Measurement of BAP is intended for use as an aid in the:

    management of postmenopausal osteoporosis and Paget's disease.

    monitoring of postmenopausal women on hormonal or bisphosphonate therapy

    prediction of skeletal response to hormonal therapy in postmenopausal women.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text from the FDA 510(k) letter for the Alkphase-B® Assay Kit does not contain information about the acceptance criteria, study details, sample sizes, expert qualifications, or ground truth establishment. The document is an FDA letter granting substantial equivalence, not a detailed study report. Therefore, I cannot generate the requested table and paragraphs based on this input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K972666

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-09-11

    (57 days)

    Product Code
    Regulation Number
    862.1050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tandem-MP Ostase Immunoenzymetric Assay is an in vitro device indicated for the quantitative measurement of skeletal alkaline phosphatase (sALP), an indicator of osteoblastic activity, in human serum. This device is intended to be used as an aid in the management of postmenopausal osteoporosis and Paget's disease.

    Device Description

    Tandem-MP Ostase is an in vitro device for the quantitative measurement of skeletal alkaline phosphatase (sALP) in human serum. The assay is a solid-phase, immunoenzymetric assay. Serum samples containing sALP are reacted in a microwell with the biotinylated capture antibody. Following binding of the biotinylated antibody/antigen complex to the streptavidin coated wells, the microwells are washed and incubated with an enzyme substrate. The captured sALP enzyme turns over the substrate and the amount of sALP bound to the microwell is determined colorimetrically by measuring the absorbance of the quenched reaction at 405 nm in a microplate reader. The calculation of the sALP concentration in the sample is based on concurrent testing of the Ostase Calibrators and Zero/Diluent.

    AI/ML Overview

    The provided text describes the Tandem-MP Ostase Assay, an in vitro device for measuring skeletal alkaline phosphatase (sALP). However, it is a 510(k) premarket notification and focuses on demonstrating substantial equivalence to a predicate device (Tandem-R Ostase Immunoradiometric Assay, K961573) rather than a study providing acceptance criteria and performance data in the typical sense of a new device validation.

    Therefore, many of your requested points cannot be directly extracted from this document, as it's not a performance study report for a novel AI device. This document is a regulatory filing for a laboratory assay.

    Here's a breakdown of what can and cannot be answered based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    This information is not provided in the given text. A 510(k) submission primarily focuses on demonstrating substantial equivalence to a predicate device, not necessarily on presenting new, specific acceptance criteria and performance metrics for the new device in isolation. The document states that the Tandem-MP Ostase Assay "is substantially equivalent" to the predicate, implying its performance should be comparable.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided. The text mentions "human serum" as the specimen matrix, but no details about sample size, origin, or study design (retrospective/prospective) are given.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. This concept (experts establishing ground truth) is more relevant for diagnostic imaging AI studies or studies where subjective interpretation is foundational. For an in vitro diagnostic assay, "ground truth" would typically refer to a gold standard measurement or clinically established normal/abnormal ranges, not expert consensus interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided. Adjudication methods are typically used in studies where multiple human readers are evaluating data, often for diagnostic imaging, and is not relevant to this type of in vitro diagnostic assay submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not provided. This is an in vitro diagnostic assay, not an AI-powered diagnostic imaging system that would involve human readers. Therefore, an MRMC study is not applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not explicitly detailed in the way it would be for an AI algorithm. The Tandem-MP Ostase Assay is an immunoenzymetric assay, which is a laboratory test. Its performance is inherent in its chemical and biological reactions and subsequent colorimetric measurement, not in an "algorithm" in the AI sense. The entire assay itself is "standalone" in that it produces a quantitative result from a serum sample without human interpretive assistance for the result itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For an in vitro diagnostic assay like this, "ground truth" would typically refer to:

    • Reference Method/Predicate Device Comparison: The text implies comparison to the predicate device (Tandem-R Ostase Immunoradiometric Assay) is the basis for demonstrating equivalence in performance. This predicate device would serve as a de-facto "ground truth" for showing the new device measures the same analyte accurately.
    • Clinical Correlation: The device is used as an aid in managing postmenopausal osteoporosis and Paget's disease. The ultimate "ground truth" for clinical utility would be disease progression, response to treatment, or bone mineral density measurements, but the device itself directly measures sALP, which is an indicator of osteoblastic activity. The document doesn't detail how the correlation with clinical outcomes was established for this specific device (Tandem-MP Ostase), but it relies on the predicate's established clinical utility.

    Specific details about how "ground truth" was established for testing the new device are not explicitly provided beyond the substantial equivalence claim.

    8. The sample size for the training set

    This information is not provided. As an in vitro diagnostic assay, it doesn't typically have a "training set" in the machine learning sense. The development and optimization of such assays involve reagent formulation, antibody selection, and protocol refinement, which are different from supervised learning with a data set.

    9. How the ground truth for the training set was established

    This information is not applicable as there is no "training set" in the AI/machine learning sense for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K964764

    Validate with FDA (Live)

    Device Name
    OPUS BONE ALP
    Date Cleared
    1997-01-24

    (58 days)

    Product Code
    Regulation Number
    862.1050
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OPUS Bone ALP is an in vitro fluorogenic enzyme immunoassay (ELISA) for the quantitative measurement of bone alkaline phosphatase (bone ALP) in human serum as an aid in the management of patients with diagnosed Paget's disease. OPUS Bone ALP is intended for use with the OPUS analyzers.

    Device Description

    OPUS Bone ALP is a set of reagents intended to be used together with the OPUS immunoassay analyzers for the quantitative measurement of bone alkaline phosphatase (bone ALP) in human serum.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Behring Diagnostics Inc. OPUS® Bone ALP 510(k) Notification, focusing on acceptance criteria and study details:

    Executive Summary:

    The provided document describes a 510(k) notification for the OPUS® Bone ALP assay, a fluorogenic enzyme immunoassay for the quantitative measurement of bone alkaline phosphatase (bone ALP) in human serum, intended as an aid in the management of patients with diagnosed Paget's disease. The study provided focuses on establishing substantial equivalence to a legally marketed device (Hybritech Tandem-R Ostase) through performance characteristics like precision, recovery, and correlation with the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    PrecisionAcceptable intra-assay and inter-assay variability for a diagnostic immunoassay.Intra-assay %CV: 2.7% to 5.23% (3 control levels, 20 replicates each)
    Inter-assay %CV: 4.57% to 7.06% (3 control levels, 20 replicates over 5 days)
    RecoveryAcceptable percentage recovery indicating accurate measurement across different analyte concentrations.98% to 103% (4 dilutions of elevated bone AP sample into normal human serum)
    Accuracy by Correlation (vs. predicate device)High correlation coefficient (ideally >0.95), slope close to 1, and y-intercept close to 0 when compared to a legally marketed device.Correlation coefficient (r): 0.992
    y-intercept: 0.77
    Slope: 1.06

    Note: The document does not explicitly state numerical acceptance criteria for each metric. The "implied" criteria are based on standard expectations for diagnostic immunoassay performance for 510(k) submissions demonstrating substantial equivalence. The reported performance values fall within generally accepted ranges for such assays.

    Detailed Study Information:

    The provided document focuses on analytical performance studies for substantial equivalence. It does not describe a study related to clinical acceptance criteria in the sense of a diagnostic outcome (e.g., sensitivity, specificity for disease detection/management). The "acceptance criteria" here refer to analytical performance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Accuracy by Correlation): 50 serum samples
    • Data Provenance: The document does not explicitly state the country of origin or if the data was retrospective or prospective. It implies the data was collected for the purpose of the 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    Not applicable. For this type of analytical performance study (especially correlation against a predicate device), "ground truth" as established by expert consensus is not typically used. The accuracy is determined by comparison to an existing, legally marketed device of the same intended use.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this study is an analytical performance comparison, not a clinical study requiring adjudication of diagnostic outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is an immunoassay, not an imaging device or AI-assisted diagnostic tool that would involve human readers or MRMC studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in a way. The performance data presented (precision, recovery, and correlation) represents the standalone performance of the OPUS Bone ALP assay system. Since it's an automated immunoassay, it inherently operates as an "algorithm only" system (in the sense of the assay chemistry and instrument processing samples without direct human interpretive intervention beyond loading/unloading and reviewing automated results).

    7. The Type of Ground Truth Used

    • For Precision and Recovery: The "ground truth" is inherent to the manufacturing and characterization of the control materials and spiked samples.
    • For Accuracy by Correlation: The "ground truth" is effectively the results obtained from the Hybritech Tandem-R Ostase, the legally marketed predicate device to which the OPUS Bone ALP assay is claiming substantial equivalence.

    8. The Sample Size for the Training Set

    The document does not describe a "training set" in the context of machine learning or AI. This is an immunoassay kit. The development of reagents and optimization would involve internal validation, but a distinct "training set" in the AI sense is not applicable here.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of an AI/machine learning model described in this 510(k) notification.

    Ask a Question

    Ask a specific question about this device

    K Number
    K961573

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-08-19

    (118 days)

    Product Code
    Regulation Number
    862.1050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    an indicator of osteoblastic activity which is intended to be used as an aid in the management of postmenopausal osteoporosis and Paget's disease

    Device Description

    Tandem®-R Ostase® is an in vitro device for the quantitative measurement of skeletal alkaline phosphatase (sALP) in human serum. The assay is a solid-phase, two-site immunoradiometric assay. Samples containing sALP are reacted with a plastic bead (solid phase) that is coated with a monoclonal antibody directed toward a specific site on the sALP molecule and, simultaneously, with a second radiolabeled monoclonal antibody directed toward a different antigenic site on the sALP molecule. Following formation of the solidphase/sALP/radiolabeled antibody sandwich, the bead is washed to remove unbound labeled antibody. The radioactivity bound to the solid phase is measured in a gamma counter. The amount of radioactivity measured is directly proportional to the concentration of sALP in the test sample, which is determined from a standard curve. The standard curve is based on the concurrent testing of Tandem-R Ostase calibrators ranging from 0 to 120 µg sALP/L.

    AI/ML Overview

    This looks like a 510(k) premarket notification for a medical device, which typically focuses on demonstrating substantial equivalence to a predicate device rather than comprehensive clinical trials with strict acceptance criteria found in PMA applications. As such, the provided text does not contain the detailed information you're requesting regarding quantitative performance metrics, study designs, sample sizes, expert involvement, and ground truth establishment in the traditional sense of a clinical validation study.

    Here's an analysis based on the provided text, highlighting what can be extracted and what information is missing:

    The core of this submission is to demonstrate "substantial equivalence" of the Tandem-R Ostase assay to a previously cleared predicate device (K930810), specifically regarding an expanded indication for use. This means the focus is less on proving absolute performance against a set acceptability standard, but rather showing that it performs as well as or identically to the predicate, and that the new indication does not raise new safety or effectiveness concerns.


    1. A table of acceptance criteria and the reported device performance

    Based on the provided text, specific quantitative acceptance criteria and detailed device performance metrics (e.g., sensitivity, specificity, accuracy against a gold standard) are not explicitly stated. The primary "acceptance criterion" for this 510(k) is demonstrating substantial equivalence to the predicate device and that the expanded indication poses no new safety or effectiveness issues.

    Acceptance Criteria (Inferred from 510(k) context)Reported Device Performance (as stated in text)
    Technological Characteristics: Identical to predicate device (K930810)"technological characteristics that are identical to those of the predicate device. The components, manufacture, specifications, and procedure for the Tandem-R Ostase assay remain unchanged from those of the previously-cleared device (#K930810)."
    Intended Use (Analyte & Specimen Matrix): Unchanged from predicate"The intended use of Tandem-R Ostase remains unchanged from the predicate with respect to the analyte being measured (skeletal alkaline phosphatase) and the specimen matrix (human serum)."
    Safety & Effectiveness with new indications (Postmenopausal Osteoporosis & Paget's Disease management): No new issues raised."It has been demonstrated in this premarket notification that Paget's disease of bone and postmenopausal osteoporosis are both bone disorders in which patients undergoing therapy can be managed with the aid of Tandem-R Ostase." "No new issues of safety or effectiveness are raised by the change in indication for use. The Tandem-R Ostase assay remains both safe and effective when used as indicated in the product labeling."
    Clinical Value for new indications: Clinically useful for managing patients with postmenopausal osteoporosis and Paget's disease."This premarket notification includes clinical data demonstrating that the Tandem-R Ostase assay provides the clinician with information that is of value in the management of patients with postmenopausal osteoporosis and Paget's disease."

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified. The text mentions "clinical data" but does not quantify the number of patients or samples included in this data.
    • Data Provenance: Retrospective/Prospective: Not specified. Country of origin: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable/Not specified. Since this is an in vitro diagnostic (IVD) assay measuring a biomarker (sALP), the "ground truth" would typically relate to accurately measuring sALP concentrations and correlating these with clinical conditions. The text implies the clinical utility of the assay for managing patients, rather than a diagnostic decision that requires expert interpretation of images or complex data. Therefore, there's no mention of experts establishing a ground truth in the way one would for diagnostic imaging.

    4. Adjudication method for the test set

    • Not applicable/Not specified. Adjudication methods (like 2+1, 3+1) are typically used when subjective interpretations or diagnoses from multiple experts need to be reconciled, such as in imaging studies. For an IVD measuring a quantitative biomarker, this would not be a standard requirement. The "clinical data" mentioned would likely involve correlations between sALP levels and patient management decisions or disease progression, rather than adjudicating diagnostic calls.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is an in vitro diagnostic device, not an AI-assisted diagnostic tool that involves human readers interpreting output. Therefore, an MRMC study is not relevant to this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Partially applicable, but not framed as "standalone performance" in the AI sense. The Tandem-R Ostase assay itself is a standalone laboratory test. Its performance (e.g., analytical accuracy, precision, linearity) is inherent to the assay and doesn't involve a human-in-the-loop for its result generation. The "clinical data" would have been used to demonstrate that the results from this standalone assay are valuable for clinicians in patient management. The text confirms the assay's procedure is "unchanged" from the predicate, implying its standalone analytical performance is equivalent.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Inferred Clinical Outcomes/Management Decisions. The text states that the "clinical data demonstrating that the Tandem-R Ostase assay provides the clinician with information that is of value in the management of patients with postmenopausal osteoporosis and Paget's disease." This suggests the ground truth was related to established clinical diagnoses of these conditions and how the sALP levels (measured by the assay) correlated with patient management strategies or clinical outcomes, rather than a single "ground truth" diagnosis established by pathology or expert consensus for each case. The "value in management" is the key here.

    8. The sample size for the training set

    • Not applicable/Not specified. This is not a machine learning or AI device that requires a training set in the conventional sense. The "training" for the assay itself would refer to its development and optimization, which isn't described in terms of a data set.

    9. How the ground truth for the training set was established

    • Not applicable. As above, this is not an AI/ML device that uses training data with established ground truth. The "ground truth" for the assay's development would be analytical standards and known concentrations during its chemical and biological development.

    In summary: The provided text is a 510(k) summary focused on demonstrating substantial equivalence and expanding an indication for use for an in vitro diagnostic test. It emphasizes that the device's technological characteristics and core intended use (analyte, matrix) are unchanged. The "study" mentioned refers to collecting "clinical data" to support the expanded indications for managing postmenopausal osteoporosis and Paget's disease, but it does not provide the granular detail typically associated with clinical validation studies for novel devices or AI algorithms. The lack of specific quantitative performance metrics, detailed sample sizes, and expert involvement for ground truth adjudication is typical for such a submission where the primary argument rests on equivalence to an already cleared device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1