    K Number
    K231927

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-03-22

    (266 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitation of intact parathyroid hormone in human serum and plasma for the differential diagnosis of hypercalcemia and hypocalcemia. This assay can be used intraoperatively. The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers

    Device Description

    Both the Elecsys PTH and Elecsys PTH STAT immunoassays make use of a sandwich test principle using a biotinylated monoclonal PTH-specific antibody labeled with a ruthenium complex. Both the Elecsys PTH and Elecsys PTH STAT immunoassays are intended for the in vitro quantitative determination of intact parathyroid hormone in human serum and plasma for the differential diagnosis of hypercalcemia and hypocalcemia. These assays can be used intraoperatively. They are intended for use on the cobas e immunoassay analyzers.

    Results are determined via a calibration curve which is instrument-specifically generated by a two-point calibration and a master curve provided via the reagent barcode.

    The reagent working solutions include: Rack Pack (kit placed on the analyzer)

    • M Streptavidin coated microparticles (transparent cap), 1 bottle, 6.5 mL: Streptavidin . coated microparticles 0.72 mg/mL; preservative.
    • . R1 Anti PTH Ab~biotin (gray cap), 1 bottle, 7 mL: Biotinylated monoclonal anti PTH antibody (mouse) 2.3 mg/L; phosphate buffer 100 mmol/L, pH 7.0; preservative.
    • . R2 Anti PTH Ab~Ru(bpy) (black cap), 1 bottle, 7 mL: Monoclonal anti PTH antibody (mouse) labeled with ruthenium complex 2.0 mg/L; phosphate buffer 100 mmol/L, pH 7.0; preservative.
    AI/ML Overview

    Unfortunately, the provided text does not describe a study involving an AI/Machine Learning powered medical device, nor does it contain information about typical acceptance criteria for such devices (e.g., sensitivity, specificity, AUC).

    Instead, the document details a 510(k) premarket notification for Elecsys PTH and Elecsys PTH STAT, which are immunoassays for quantitative determination of intact parathyroid hormone. These are laboratory diagnostic tests, not AI-powered devices.

    The document focuses on demonstrating substantial equivalence to a previously cleared predicate device (K070709) by:

    • Comparing technological characteristics (Table 1 and Table 2).
    • Providing non-clinical performance evaluations such as precision, analytical sensitivity (Limit of Blank, Limit of Detection, Limit of Quantitation), linearity, high dose hook effect, interference studies, and sample matrix comparison.
    • Presenting a method comparison to the predicate devices.
    • Reporting on stability and a reference range study.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and study that proves an AI device meets acceptance criteria based on this input. The information required for your questions (e.g., sample size for test set, data provenance, number of experts for ground truth, MRMC study, standalone performance, training set details) are specific to AI/ML device evaluations and are not present in this document about an immunoassay.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232791

    Validate with FDA (Live)

    Date Cleared
    2024-03-01

    (172 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Intact PTH assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems. It is indicated to aid in the differential diagnosis of hyperparathyroidism, hypoparathyroidism, or hypercalcemia of malignancy and can be used intraoperatively. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.

    Device Description

    The Access Intact PTH assay is a sandwich immunoenzymatic assay. The Access Intact PTH assay consists of the reagent pack and calibrators. Other items needed to run the assay include substrate and wash buffers. The Access intact PTH assay reagent pack, Access intact PTH assay calibrators, along with the Access wash buffer and substrate are designed for use on the Dxl 9000 Access Immunoassay Analyzers in a clinical laboratory setting.

    AI/ML Overview

    The provided text describes the performance validation of the Beckman Coulter Access Intact PTH assay on the Dxl 9000 Access Immunoassay Analyzer. This is a biomedical measurement device, specifically an in vitro diagnostic (IVD) rather than an AI-powered system as might be initially implied by the format of the request. Therefore, some of the requested information regarding AI-specific criteria (like number of experts for ground truth, adjudication methods, MRMC studies, training set details) is not applicable to this type of device.

    Here's an analysis of the acceptance criteria and study results, focusing on the relevant aspects for an IVD device:

    Acceptance Criteria and Reported Device Performance

    The study primarily focuses on demonstrating the performance characteristics of the Access Intact PTH assay run on the Dxl 9000 system, compared to its predicate device (Access Intact PTH on the Access 2 Immunoassay System). The "acceptance criteria" here are performance targets related to analytical precision, linearity, and detection/quantitation limits, based on standard laboratory guidelines (CLSI documents).

    Table of Acceptance Criteria and Reported Device Performance:

    Performance MetricAcceptance Criteria (Design Criteria)Reported Device Performance (Routine Mode)Reported Device Performance (Intraoperative Mode)
    Imprecision (Within-Laboratory)≤ 0.96 pg/mL (0.10 pmol/L) SD at concentrations ≤ 12 pg/mL (1.3 pmol/L) ≤ 8.0% CV at concentrations > 12 pg/mL (1.3 pmol/L)SD: Sample 1: 0.46 pg/mL (0.05 pmol/L) %CV: Sample 1: 5.9%, Sample 2: 6.5%, Sample 3: 4.3%, Sample 4: 3.7%, Sample 5: 2.7%SD: Sample 1: 0.23 pg/mL (0.02 pmol/L) %CV: Sample 1: 13.2% (exceeds 8%), Sample 2: 5.0%, Sample 3: 4.8%, Sample 4: 4.2%, Sample 5: 4.0%, Sample 6: 4.3%, Sample 7: 3.3%
    Linearity(Implicitly: demonstrate linearity across measuring interval)Demonstrated linearity across measuring interval (1.7 - 3500 pg/mL)Demonstrated linearity across measuring interval (1.7 - 3500 pg/mL)
    Limit of Blank (LoB)≤ 1.0 pg/mL (≤ 0.11 pmol/L)0.5 pg/mL (0.05 pmol/L)0.5 pg/mL (0.05 pmol/L)
    Limit of Detection (LoD)≤ 1.0 pg/mL (≤ 0.11 pmol/L)0.7 pg/mL (0.08 pmol/L)0.8 pg/mL (0.08 pmol/L)
    Limit of Quantitation (LoQ) (≤ 20% within-lab CV)≤ 1.7 pg/mL (≤ 0.18 pmol/L)0.7 pg/mL (0.08 pmol/L)0.8 pg/mL (0.08 pmol/L)
    Method Comparison (vs. Predicate)(Implicitly: demonstrate substantial equivalence with predicate)Expected slopes near 1.0 and intercepts near 0,
    high correlation (R near 1.0)Slope: 0.97 (95% CI: 0.96-0.98) Intercept: -0.23 (95% CI: -0.33 to -0.12) Correlation (R): 1.00Slope: 1.02 (95% CI: 1.00-1.04) Intercept: -0.081 (95% CI: -0.36 to 0.20) Correlation (R): 0.99

    Notes on Performance:

    • For Imprecision, one sample in the "Intraoperative Mode" (Sample 1, 1.7 pg/mL) shows a %CV of 13.2%, which exceeds the general criteria of "≤ 8.0% CV at concentrations > 12 pg/mL (1.3 pmol/L)" and also "≤ 0.96 pg/mL (0.10 pmol/L) SD at concentrations ≤ 12 pg/mL (1.3 pmol/L)". However, given the actual concentration and the stated SD for this sample (0.23 pg/mL), it might be acceptable for very low concentrations where relative variability can be higher, or it might fall into a specific acceptance sub-criterion not fully detailed here. The SD of 0.23 pg/mL is still well below the 0.96 pg/mL SD target for concentrations ≤ 12 pg/mL.
    • The correlation coefficients (R) of 1.00 and 0.99 for method comparison indicate excellent agreement with the predicate device. The slopes and intercepts also demonstrate good agreement, supporting the claim of substantial equivalence.

    Study Details:

    As this is a 510(k) submission for an in vitro diagnostic immunoassay system, the concept of AI-driven image analysis or clinical decision support is not applicable. Therefore, many of the questions related to AI-specific validation (e.g., number of experts, adjudication, MRMC studies, AI training sets) are not relevant to this document. The "ground truth" for an IVD refers to the true concentration of an analyte as measured by a reference method or established through rigorous characterization of control materials.

    1. Sample sizes used for the test set and data provenance:

      • Imprecision Study:
        • Routine Mode: 86 replicates for most samples (one sample with 84 replicates).
        • Intraoperative Mode: 88 replicates for all samples.
        • Data Provenance: Not explicitly stated, but typically these studies are conducted in a controlled laboratory setting (prospective data collection from prepared samples/controls). No country of origin for the data is specified, but given the manufacturer (Beckman Coulter, Inc., Chaska, MN, USA), it's highly likely to be US-based or from their global R&D facilities.
      • Method Comparison Study:
        • Routine Mode: N=144 patient samples.
        • Intraoperative Mode: N=116 patient samples.
        • Data Provenance: Not explicitly stated regarding country of origin or retrospective/prospective. These are typically patient plasma/serum samples collected prospectively for the study or from biobanks.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. This is an immunoassay device for quantitative determination of a hormone. The "ground truth" for these tests is based on the chemical measurement itself, validated against reference materials or established methods, not subjective expert interpretation of images or clinical data.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not Applicable. See point 2.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable. This is an automated immunoassay system, not an AI-assisted diagnostic tool requiring human reader interpretation studies.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This entire validation represents the "standalone" performance of the analytical instrument and assay kit. The Dxl 9000 Access Immunoassay Analyzer functions largely as an automated "algorithm" in that it processes samples and produces quantitative results without direct human interpretive input during the measurement process. Human oversight and quality control are part of its real-world use.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this immunoassay is likely established through:
        • Reference Materials/Calibrators: These are materials with precisely known concentrations of intact PTH, used to calibrate the instrument and verify accuracy.
        • Predicate Device Comparison: The "Method Comparison" study uses the results from the legally marketed predicate device (Access Intact PTH on the Access 2 Immunoassay System) as the comparative reference, implicitly relying on the established accuracy of that predicate.
        • Analytical Standards: Performance metrics (LoB, LoD, LoQ, linearity) are assessed against internationally recognized analytical standards, such as those from CLSI (Clinical and Laboratory Standards Institute), which define the statistical and experimental methods for confirming these basic analytical performance parameters.
    7. The sample size for the training set:

      • Not Applicable. This is an immunoassay device, not an AI/machine learning model that requires a "training set" in the conventional sense. The "training" of the device involves calibration using specified calibrator materials and internal quality control procedures.
    8. How the ground truth for the training set was established:

      • Not Applicable. See point 7.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221197

    Validate with FDA (Live)

    Date Cleared
    2023-09-05

    (498 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VITROS Immunodiagnostic Products Intact PTH II Reagent Pack quantitatively measures intact parathyroid hormone (iPTH) in human serum and plasma (K2-EDTA, lithium heparin) using the automated VITROS 3600 Immunodiagnostic System.

    Intact PTH is indicated to aid in the diagnosis of hyperparathyroidism, differential diagnosis of hypocalcemia or hypercalcemia and for intraoperative measurement of iPTH levels.

    Device Description

    The VITROS Immunodiagnostic Products Intact PTH II assay is performed using the VITROS Intact PTH II Reagent Pack and the VITROS Intact PTH II Calibrators on the VITROS 3600 Immunodiagnostic System.

    VITROS Intact PTH II Reagent Pack contains:

    1 reagent pack containing:

    • 100 coated wells (biotinylated anti-PTH antibody, 2ug/ml). ●
    • 7.4 mL assay reagent as buffer with bovine gamma globulin, bovine serum albumin, and antimicrobial agent.
    • 7.4 mL conjugate reagent (HRP-mouse monoclonal anti-PTH, 6 ug/mL) in buffer with bovine serum albumin and antimicrobial agent).
    AI/ML Overview

    The provided text is a 510(k) premarket notification for Ortho-Clinical Diagnostics' VITROS Immunodiagnostic Products Intact PTH II Reagent Pack, a device for quantitatively measuring intact parathyroid hormone (iPTH). While it details various performance characteristics, it does not describe an AI/ML-based device and therefore does not contain information on acceptance criteria for such a device, nor an AI/ML specific study.

    The document focuses on the analytical and clinical performance of an in vitro diagnostic (IVD) assay, comparing it to a predicate device (Roche Elecsys PTH assay). The studies described are typical for IVD assays: precision, linearity, detection limits, analytical specificity (interference and cross-reactivity), and method comparison against a commercially available PTH assay and a clinical study for intra-operative use.

    Therefore, I cannot extract the requested information regarding AI/ML device acceptance criteria and study details from the provided text. The questions below are specifically tailored for AI/ML device evaluations.

    However, I can provide the available information relevant to the IVD device's performance studies:


    Based on the provided document, here is the information available for the VITROS Immunodiagnostic Products Intact PTH II Reagent Pack, noting that it is an IVD assay, not an AI/ML device:

    1. A table of acceptance criteria and the reported device performance

    The document does not formally present a table of "acceptance criteria" against which performance is measured in the same way an AI/ML device would demonstrate specified accuracy metrics. Instead, it presents various performance characteristics and states that results "met the acceptance criteria" for certain comparisons without detailing those criteria explicitly in a table.

    However, key performance parameters and their reported values are listed below:

    Performance ParameterReported Device Performance
    Measuring (Reportable) Range6.8–5000 pg/mL (0.7–530 pmol/L)
    Limit of Blank (LoB)0.8 pg/mL
    Limit of Detection (LoD)1.2 pg/mL
    Limit of Quantitation (LoQ)1.2 pg/mL (at 20% CV)
    Precision (Repeatability)CV% between 0.9% and 1.3% across different concentration levels
    Precision (Within Lab)CV% between 1.5% and 3.0% across different concentration levels
    Method Comparison (vs. Predicate)
    Slope (VITROS 3600 vs. Comparative Method)1.01
    Correlation Coefficient (r)0.991
    Intercept0.3 (pg/mL) / 0.0 (pmol/L)
    Matrix Comparison (Serum vs. Plasma)Slopes between 0.988 and 1.011; Correlation Coefficients between 0.973 and 0.996 (for K2-EDTA, Li-Hep, Na-Hep plasma vs. serum)
    Intra-operative Clinical Study Agreement
    Primary Endpoint Positive Agreement100% (29/29)
    Primary Endpoint Negative Agreement100% (3/3)
    Primary Endpoint Overall Agreement100% (32/32)

    2. Sample size used for the test set and the data provenance

    • Precision/Reproducibility: Patient pools tested with 80 observations (measurements) over 20 days for each concentration level. Data provenance not specified (country of origin, retrospective/prospective implied as prospective measurement in lab).
    • Method Comparison: 206 patient (serum) samples. Data provenance not specified.
    • Intra-operative Clinical Study: 32 subjects (sets of specimens) qualified for primary endpoints. Data provenance not specified (likely prospective for the study measurements, but subjects were undergoing surgery, so possibly a mix of retrospective patient selection and prospective sample collection).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This device is an IVD assay, not an imaging/AI device requiring expert interpretation for ground truth. Ground truth for an IVD is established by standardized reference methods or the performance of a predicate device/comparator assay.
    • For the intra-operative clinical study, the "success" criterion for surgery (50% or greater drop in PTH level) is based on the result from the comparator assay (hospital's assay) and the investigational device, rather than expert judgment on the device's output.

    4. Adjudication method for the test set

    • Not applicable as this is an IVD assay measuring a biomarker concentration, not an AI/ML device relying on human interpretation or adjudication of output.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable as this is an IVD assay, not an AI/ML imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The device is a standalone IVD assay. Its performance (precision, linearity, detection limits, analytical specificity, method comparison) is evaluated directly, without human interpretation of its output in the sense of an AI/ML diagnosis. The intra-operative study compares its quantitative results to another quantitative assay.

    7. The type of ground truth used

    • Analytical Performance (Precision, Linearity, Detection Limits): Established through laboratory testing using controlled samples (e.g., patient pools, dilutions, blanks) according to CLSI (Clinical and Laboratory Standards Institute) protocols (EP05, EP06, EP17). The "ground truth" here is the expected value of the controlled sample or the known characteristics of the samples used.
    • Analytical Specificity (Interference, Cross-Reactivity): Known concentrations of interferents or cross-reactants added to samples with known PTH concentrations. Ground truth is the unbiased PTH value.
    • Method Comparison: Comparison against a "commercially available PTH assay" (presumably the predicate device, although it's not explicitly named as such in this section, it's implied by "comparative method") as the reference.
    • Intra-operative Clinical Study: The "success" criterion (50% or greater drop in PTH) was determined by comparing the device's results to the results from the hospital's (comparator) assay for the same patient samples. The "ground truth" for surgical success was defined by this 50% drop as measured by the comparator, and the study assessed concordance.

    8. The sample size for the training set

    • Not applicable. This is an IVD assay, not an AI/ML algorithm that requires a "training set." The system is a reagent pack and instrument system, not a learned model.

    9. How the ground truth for the training set was established

    • Not applicable (no training set for an IVD assay).
    Ask a Question

    Ask a specific question about this device

    K Number
    K190702

    Validate with FDA (Live)

    Date Cleared
    2019-08-30

    (165 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use.
    Lumipulse G whole PTH is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the quantitative measurement of PTH (1-84) in human serum and plasma on the LUMIPULSE G System. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism.

    Device Description

    The Lumipulse G whole PTH Immunoreaction Cartridges consists of 3 × 14 tests. Each kit contains the following:

    1. Antibody-Coated Particle Solution (Liquid when used, 200 µL/Immunoreaction Cartridge) Contains 200 µg/mL anti-PTH polyclonal antibodies (goated ferrite particles, protein stabilizers (bovine and goat) and chemical stabilizers in MES buffer. This solution contains gelatin and turns into gel at 15 °C or lower. Preservative: ProClin 300
    2. Enzyme-Labeled Antibody Solution (Liquid, 120 µL/Immunoreaction Cartridge) Contains 0.2 µq/mL alkaline phosphatase (ALP: calf)-labeled anti-PTH polyclonal antibody (qoat), protein stabilizers (bovine) and chemical stabilizers in MES buffer. Preservative: ProClin 300
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Lumipulse G whole PTH device, based on the provided FDA 510(k) summary:

    Device: Lumipulse G whole PTH (Fujirebio Diagnostics, Inc.)
    Intended Use: Quantitative measurement of PTH (1-84) in human serum and plasma on the LUMIPULSE G System for differential diagnosis of hypercalcemia and hypocalcemia.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes various analytical performance characteristics. For each characteristic, the reported performance is presented. Acceptance criteria are generally implied by the positive outcomes of the studies and their adherence to CLSI guidelines. For example, for precision, a certain percentage of Coefficient of Variation (%CV) is considered acceptable.

    Acceptance Criteria CategorySpecific Metric / StudyReported Device Performance (Lumipulse G whole PTH)
    Analytical Precision20-Day Precision (Total %CV)≤ 4% (e.g., Control Level 1: 3%, Serum 7: 3%)
    Lot-to-Lot Precision (Inter-lot %CV)≤ 4% (e.g., Control Level 1: 4%, Serum 4: 3%)
    Site-to-Site Precision (Total %CV)≤ 6.7% (e.g., Serum 1: 6.3%, Serum 2: 6.7%)
    Linearity/Reportable RangeLinearity Range (Correlation R²)1.4 pg/mL - 2190.3 pg/mL (R²=0.9984)
    Measuring Range4.0 pg/mL - 1800.0 pg/mL
    Detection LimitsLimit of Blank (LoB)0.0 pg/mL
    Limit of Detection (LoD)0.295 pg/mL
    Limit of Quantitation (LoQ)2.128 pg/mL (defined as interassay CV of 10%)
    Analytical Specificity/Cross-ReactivityCross-Reactivity with specified substances (e.g., Calcitonin, PTH fragments)< 0.002% (e.g., Calcitonin: < 0.001%, PTH (7-84): < 0.002%) - no cross-reactivity observed
    Interfering SubstancesInterference from endogenous & therapeutic substances (Average Interference)≤ 10% (e.g., Conjugated Bilirubin 44 mg/dL, Hemoglobin 510 mg/mL, Acetaminophen 22 mg/dL) - found not to interfere
    High Dose Hook EffectHigh Dose Hook EffectNot observed for spiked serum up to 80,000 pg/mL
    Specimen StabilitySpecimen Storage ConditionsSerum 2-10°C up to 24 hrs; Plasma 2-10°C up to 7 days; Serum/Plasma -20°C up to 3 months. Specimens on-board LUMIPULSE G System within 3 hours. Max 1 freeze/thaw cycle.
    Matrix ComparisonEquivalence between matrices (SST, K2EDTA, Lithium Heparin, Sodium Heparin vs. Red Top Serum) - Pearson Correlation Coefficient0.9980 to 0.9995 (indicating strong correlation and equivalency)
    Method ComparisonComparison to Predicate Device (LIAISON® 1-84 PTH assay) - Correlation Coefficient (r)0.9808 (Intercept: -0.5351, Slope: 0.9909, Average Bias: -6.271 pg/mL)

    2. Sample Size and Data Provenance (for test set/performance studies)

    The document refers to various samples for different analytical studies:

    • 20-Day Precision: A panel of seven native human serum samples (n=80 replicates for each sample). The study was conducted at an internal laboratory. Origin of samples is "native human serum" but specific country/retrospective/prospective nature is not specified beyond that.
    • Lot-to-Lot Precision: A panel of four native human serum samples and two control levels (n=32 replicates per lot, 96 total replicates per sample/control across three lots). Internal laboratory. "Native human serum" for samples.
    • Site-to-Site Precision: A panel of five native serum samples (n=combined data for site-to-site analysis). Internal laboratory and two additional sites. "Native serum" for samples.
    • Linearity: High and low sample pools created using patient serum samples that contained naturally expressed whole PTH. Specific number of samples not detailed, but the linearity range was derived from these pools.
    • Detection Limit (LoB, LoD, LoQ): Four blank and four low-level serum specimens (LoB/LoD: 60 determinations for each panel per lot); five low-level serum (LoQ: on 2 systems with 2 lots across several days).
    • Analytical Specificity/Cross Reactivity: A panel of serum samples with naturally occurring whole PTH (n=4 panel members) for cross-reactivity with 10 different substances.
    • Interfering Substances: Human serum specimen pools with whole PTH concentrations of approximately 35.1, 81.5, and 235.9 pg/mL.
    • Matrix Comparison: Seventy-one (71) matched sets of serum and plasma samples.
    • Method Comparison: A total of 275 matched human serum samples.
    • Expected Values/Reference Range: 147 serum specimens from an "apparently healthy adult population (22-72 years old)."

    Data Provenance Remarks:

    • All studies mentioned use "native human serum" or "patient serum samples," indicating biological samples.
    • Studies were conducted at "internal laboratory" and "two additional sites," suggesting multi-site evaluation for some parameters.
    • The document does not explicitly state geographical origin (e.g., country) or whether samples were retrospective or prospectively collected for most analytical studies. The "apparently healthy adult population" for expected values implies a specific collection for that purpose.

    3. Number of Experts and Qualifications for Ground Truth

    • The document is for an in vitro diagnostic (IVD) device that measures a biomarker (PTH). The ground truth for such devices is typically established by reference methods, defined concentrations, or clinically characterized samples.
    • Therefore, no "experts" (e.g., radiologists) are explicitly mentioned or required to establish a ground truth for diagnostic imaging interpretation, as this is not an imaging device.
    • The ground truth for the analytical studies comes from:
      • Known concentrations: For calibrators, controls, and spiked samples (e.g., linearity, detection limits, cross-reactivity, interfering substances).
      • Clinical characterization: For "native human serum" samples, their PTH levels are measured.
      • International standards: Traceability of calibration is to the 1st international standard for Parathyroid Hormone 1-84 (code: 95/646) provided by the NIBSC.

    4. Adjudication Method for the Test Set

    • Not applicable. This section typically applies to diagnostic imaging studies where expert readers adjudicate findings. For an IVD device measuring a biomarker, the "ground truth" is determined by the analytical measurement process itself using established methods, standards, and controls.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. An MRMC study is relevant for diagnostic imaging AI, where multiple human readers interpret cases with and without AI assistance. This device is an in vitro diagnostic for quantitative measurement of a biomarker and does not involve human readers interpreting images.

    6. Standalone Performance Study

    • Yes, for the device itself. All the analytical performance studies (precision, linearity, detection limits, specificity, interference, specimen stability, matrix comparison) described are studies of the Lumipulse G whole PTH device operating in a standalone mode (i.e., the algorithm/assay system without human-in-the-loop performance influencing the measurement output directly).
    • The Method Comparison study directly compares the standalone performance of the Lumipulse G whole PTH to a legally marketed predicate device (LIAISON® 1-84 PTH Assay), which also measures PTH quantitatively without human interpretation within the measurement process.

    7. Type of Ground Truth Used

    The ground truth for the performance studies is established by:

    • Reference Materials/International Standards: Calibration is traceable to the 1st international standard for Parathyroid Hormone 1-84 (code: 95/646) provided by the National Institute for Biological Standards and Controls (NIBSC).
    • Spiked Samples/Known Concentrations: For studies like linearity, detection limits, cross-reactivity, interfering substances, where specific amounts of analyte or interfering substances are added to samples.
    • Clinically Characterized Patient Samples: "Native human serum" or "patient serum samples" are used, with their PTH levels being the "truth" for evaluation against expected performance or comparison with other methods.
    • Predicate Device Comparison: For the method comparison study, the results from the legally marketed predicate device (LIAISON® 1-84 PTH Assay) serve as a comparative "ground truth" or reference for evaluating substantial equivalence.

    8. Sample Size for the Training Set

    • The document does not provide information on a specific "training set" sample size. This submission is for an IVD assay, not typically a machine learning or AI algorithm that undergoes explicit "training" as defined in the context of imaging AI.
    • For IVD assays, method development and optimization phases internally use numerous samples, but these are generally considered part of the product development, not a formal "training set" like in AI. The studies presented here are for validation and verification of the developed assay.

    9. How Ground Truth for the Training Set Was Established

    • Not applicable in the context of an IVD assay as presented. As explained above, there isn't a "training set" in the common AI/ML sense.
    • Ground truth for assay development and optimization (analogous to "training" in a broad sense) would rely on the same principles as the validation studies: known concentrations, reference materials, and clinically characterized samples.
    Ask a Question

    Ask a specific question about this device

    K Number
    K163658

    Validate with FDA (Live)

    Date Cleared
    2017-05-23

    (151 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA Centaur® Intact Parathyroid Hormone (PTH) assay is for in vitro diagnostic use in the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma using the ADVIA Centaur XP system. This assay is intended to be used to aid in the differential diagnosis of hyperparathyroidism, or disorders of calcium metabolism. This assay can be used intra-operatively.

    Device Description

    The ADVIA Centaur® Intact Parathyroid Hormone (PTH) assay is a chemiluminescence sandwich immunoassay. It consists of a Lite Reagent containing acridinium ester-labeled mouse monoclonal anti-human PTH antibody and a Solid Phase Reagent containing biotinylated mouse monoclonal anti-human PTH antibody bound to streptavidin-coated paramagnetic particles. The assay also includes Low and High Calibrators which are reconstituted synthetic peptide in buffered saline with human EDTA plasma, surfactants, and preservatives.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ADVIA Centaur Intact Parathyroid Hormone (PTH) Assay, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally not explicitly stated as quantitative thresholds in this type of 510(k) summary for all tests (e.g., "must achieve X% recovery"). Instead, the document demonstrates that the device meets performance expectations through studies that align with recognized standards. Where explicit criteria or comparative results are given, they are listed below.

    Performance CharacteristicAcceptance Criteria (Implied/Explicit)Reported Device Performance
    PrecisionDemonstrated precision according to CLSI EP5-A2 (within-run and within-lab CVs at various concentrations). The CV values are presented as meeting acceptable laboratory practice.Repeatability (Within-Run) CV: Range from 0.81% to 5.16%Within-Lab (Total Precision) CV: Range from 1.62% to 6.78% (See Table in section 10.1 for detailed values at different concentrations).
    LinearityDemonstrated linearity across the assay range with acceptable deviation from linear fit, according to CLSI EP06-A. Explicit: Bias from the linear fit estimate was <10%.Linear Range: 6.3 to 2000 pg/mLDeviation from Linear Fit: Ranged from -4% to 4% (See Table in section 10.2).
    Dilution RecoveryDemonstrated acceptable recovery percentages for diluted high-concentration samples.Recovery (%): 101.6%, 104.2%, 102.6% for three different spiked concentrations. Mean recovery of 102.8%.
    Method ComparisonDemonstrated good correlation with the predicate device across the assay range using patient samples, according to CLSI EP9-A3.Regression Equation: ADVIA Centaur PTH = 1.02(iPTH) – 2.18 pg/mLCorrelation Coefficient (r): 0.99 (indicating strong correlation).
    Matrix ComparisonNo significant difference between various tube types (serum, SST, EDTA, lithium heparin, sodium heparin) compared to dipotassium EDTA plasma, according to CLSI EP9-A3.Correlation Coefficients (r): 0.996 to 0.998 for various matrix comparisons against Dipotassium EDTA plasma. Regression equations showed close to 1.0 slope and small intercepts (e.g., Serum = 0.99 EDTA – 1.85).
    Reference IntervalsEstablished interval based on healthy donors consistent with clinical utility, according to CLSI EP28-A3c. The ranges are presented as acceptable for diagnostic aid.EDTA Plasma: 18.4 - 80.1 pg/mL (95% CI)Serum: 18.5 – 88.0 pg/mL (95% CI)
    Detection LimitEstablished LoB, LoD, and LoQ according to CLSI EP17-A2. The values are presented as meeting analytical sensitivity requirements. Explicit: LoQ defined as lowest concentration at a total CV of 20%.LoB: 1.5 pg/mLLoD: 3.2 pg/mLLoQ: 4.6 pg/mL
    InterferenceNo significant interference (< 10% effect) from common endogenous substances and drugs at specified concentrations, according to CLSI EP07-A2.Endogenous Substances & Drugs: All tested substances showed interference < 10% (range -7.5% to 6.9%).HAMA: Difference from expected values ranged from -11.5% to 3.2% (for 60 pg/mL PTH) and -3.8% to 2.4% (for 340 pg/mL PTH).
    Cross-ReactivityMinimal cross-reactivity with structurally similar compounds, presented as acceptable for specificity.Most cross-reactants showed < 0.001% cross-reactivity. Significant cross-reactivity observed with PTH (7-84) at 37.36% in normal patient samples and 17.5033% in blank.
    High-Dose Hook EffectNo paradoxical decrease in RLU up to a specified high concentration, confirming reliable measurement at high PTH levels.Patient samples with intact PTH levels as high as 100,000 pg/mL are reported as > 2000 pg/mL (no hook effect within this range).
    Intra-Operative Use (Clinical Study)Miami Criterion: A 50% or greater drop in PTH level from baseline to 10 minutes post-excision. The device should show high agreement with a clinical PTH test cleared by the FDA for intra-operative use.Positive Agreement: 100% (29/29 successful surgeries)Overall Agreement: 96.7% (29/30). One discrepant sample explained by the local device failing to meet the 50% drop at 10 min, while the ADVIA Centaur PTH assay did.

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision: Six EDTA plasma samples and three levels of controls.
    • Linearity: 11 serially diluted samples.
    • Dilution Recovery: 3 spiked samples (low sample pool spiked to ~3,000, ~6,000, and ~9,000 pg/mL).
    • Method Comparison: 349 EDTA plasma patient samples.
    • Matrix Comparison: 56 matched specimens drawn in different tube types (serum red top, SST, EDTA, lithium heparin, and sodium heparin).
    • Reference Intervals: 142 paired serum and EDTA plasma samples from apparently healthy donors.
    • Detection Limit: Not explicitly stated beyond "CLSI protocol EP17-A2," which typically involves multiple replicates of blank and low-concentration samples.
    • Interference: Two EDTA sample pools (at ~75 pg/mL and ~600 pg/mL PTH) for endogenous substances/drugs. Six human plasma samples for HAMA.
    • Cross-Reactivity: One normal human EDTA plasma sample and blank assay diluent for each test compound.
    • High-Dose Hook Effect: Patient samples as high as 100,000 pg/mL.
    • Intra-Operative Clinical Study: Sets of specimens from 30 subjects.

    Data Provenance (Country of origin / Retrospective or Prospective):
    The document does not explicitly state the country of origin for the samples used in the performance or clinical studies. The method comparison and clinical study are likely prospective in nature (collecting new data for the device evaluation), given the testing against a predicate or a comparator device. The reference interval study is explicitly from "apparently healthy donors," implying a prospective collection or at least a specific selection criteria. The other studies typically involve controlled laboratory conditions or spiked samples rather than broad patient cohorts for provenance purposes.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For this type of in vitro diagnostic (IVD) assay, "ground truth" is typically established by reference methods, established predicate devices, or pathology results, rather than by individual human experts evaluating diagnostic images or clinical cases.

    • For the analytical performance studies (Precision, Linearity, etc.): Ground truth is established by the known concentrations of calibrators, controls, spiked samples, or by comparison to a legally marketed predicate device (ADVIA Centaur iPTH assay) which is itself a validated IVD. No specific number or qualification of experts is mentioned for establishing these analytical "ground truths."
    • For the Clinical Study (Intra-operative use): The "ground truth" for surgical success was defined by the Miami Criterion (50% or greater drop in PTH level). The comparison was made against "the assays used by participating surgeons/sites (comparator devices)." This implies that the 'truth' was the outcome determined by the established clinical practice and the results of a clinically used (FDA-cleared) PTH assay at the sites. No external experts were used to establish this ground truth, but rather comparison to existing clinical standards.

    4. Adjudication Method for the Test Set

    Not applicable in the conventional sense of image or case adjudication. For the clinical study, the "Miami Criterion" served as the objective rule for determining surgical success, which was then compared between the investigational and comparator devices. There was no explicit adjudication process by multiple human readers for a single "ground truth" determination.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study, as typically understood for imaging devices involving multiple human readers, was not conducted. This is an IVD assay, not an imaging device. The clinical study focused on the agreement of the device's quantitative output with existing clinical practice based on a well-defined criterion (Miami Criterion), not on how human readers' diagnostic accuracy improves with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, this entire submission revolves around the standalone performance of the ADVIA Centaur Intact Parathyroid Hormone (PTH) Assay. It's an automated in vitro diagnostic device, designed to produce quantitative results independently for clinical interpretation. The clinical study for intra-operative use also evaluated the device's ability to meet the Miami Criterion, comparing its raw output to that of other clinical devices, not its assistance to a human in a diagnostic workflow.

    7. The Type of Ground Truth Used

    • Analytical Studies (Precision, Linearity, Dilution Recovery, Detection Limit): Ground truth is based on known concentrations of standards and controls, or scientifically established principles of measurement procedure performance.
    • Method Comparison & Matrix Comparison: Ground truth is the results obtained from a legally marketed predicate device (ADVIA Centaur Intact Parathyroid Hormone (iPTH) assay) or a comparison against a "gold standard" or accepted methodology.
    • Reference Intervals: Ground truth is derived from results obtained from apparently healthy donors which define a normal range.
    • Clinical Study (Intra-operative): Ground truth for surgical success was the Miami Criterion (a predefined clinical outcome threshold based on PTH levels), as determined by a separate, FDA-cleared clinical PTH test device used at the sites.

    8. The Sample Size for the Training Set

    This document does not specify a "training set" in the context of an AI/ML algorithm. This is a traditional IVD immunoassay. Therefore, there's no machine learning model that requires a distinct training dataset for its development that would be reported in this manner. The development and optimization of the assay reagents and parameters would be internal to the manufacturer's R&D process.

    9. How the Ground Truth for the Training Set Was Established

    As explained above, there is no "training set" or "ground truth for the training set" in the context of AI/ML for this device. The development of the assay's reagents and methodologies would be based on biochemical principles, antigen-antibody interactions, and optimization studies performed by the manufacturer, rather than by training a model on labeled data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161158

    Validate with FDA (Live)

    Date Cleared
    2017-01-31

    (281 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS-iSYS Intact PTH+ assay is an in vitro diagnostic device intended for the quantitative determination of intact PTH in human serum or plasma on the IDS-iSYS Multi-Discipline Automated System. Results are to be used in the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism.

    Device Description

    The IDS-iSYS Intact PTH" assay is based on chemiluminescence technology. 100 uL of patient sample is incubated with a biotinylated polyclonal anti-PTH (39-84) antibody and an acridinium labelled PTH (13-34) antibody. Streptavidin labelled magnetic particles are added prior to a second incubation step. The magnetic particles are captured using a magnet and a wash step performed to remove any unbound analyte. Trigger reagents are added; the resulting light emitted by the acridinium label is directly proportional to the concentration of Intact PTH in the original sample.

    IDS-iSYS Intact PTH reagent kit consists of one (1) Immunoassay Cartridge, two (2) vials each of 2 concentration Calibrator levels and a mini-CD containing the Instructions For Use (IFU), CRY files and Certificate of Analysis.

    IDS-iSYS Intact PTHN Cartridge, sufficient for 100 tests, consists of reagents provided in individual compartment within a plastic container called the Cartridge. The IDS-iSYS Intact PTHN Cartridge contains the following ready to use reagents:

    • · Magnetic particles coated with streptavidin in a phosphate buffer containing sodium azide as preservative (<0.1%), 1 bottle.
    • Goat polyclonal antibody against 13-34 PTH labelled with an acridinium ester derivative, in buffer containing goat serum with sodium azide as preservative (<0.1%), 1 bottle.
    • Goat polyclonal antibody against 39-84 PTH 1abelled with biotin, in buffer containing bovine and goat proteins with sodium azide as preservative (<0.1%), 1 bottle.

    IDS-iSYS Intact PTH Calibrators are included in the reagent kit.

    • · Calibrator A: Two vials of lyophilized porcine serum matrix buffer containing low level PTH and sodium azide as preservative >1% (w/w%).
    • · Calibrator B: Two vials of lyophilized porcine serum matrix buffer containing high level PTH and sodium azide as preservative >1% (w/w%).

    The submission is due to a new source of antibody for the assay.

    AI/ML Overview

    This document describes the performance characteristics of the IDS-iSYS Intact PTHN assay, an in vitro diagnostic device for quantitative determination of intact PTH.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct numerical targets but are implicitly defined by the reported performance and the comparison to the predicate device. The information below summarizes the performance characteristics provided, which in a regulatory context, demonstrate the device meets its intended use and is substantially equivalent to the predicate.

    Performance CharacteristicPredicate Device (K103325) PerformanceCandidate Device (IDS-iSYS Intact PTHN) PerformanceImplied Acceptance Criteria (Demonstrated)
    Method Comparison (against a commercially available method)N/An=312 samples; Slope: 1.02 (95% CI: 0.99 to 1.04); Intercept: -0.7 pg/mL (95% CI: -5.4 to 4.1); Correlation coefficient (r): 1.00Demonstrate strong correlation and agreement with a commercially available quantitative chemiluminescence Intact PTH assay (e.g., r close to 1, slope close to 1, intercept close to 0) as per CLSI EP-9A2 guidelines.
    Sample Matrix Comparison (vs. K2 EDTA plasma)N/A
    - Serum (Red Top)N/An=52; Slope: 0.94 (95% CI: 0.92 to 0.97); Intercept: 2.55 (95% CI: 0.86 to 3.16); r: 1.00Demonstrate acceptable agreement across different sample matrices (serum, lithium heparin plasma) compared to the control (K2 EDTA plasma) as per CLSI EP9-A3 guidelines. Slopes close to 1 and intercepts close to 0 with good correlation coefficients.
    - Serum (SST)N/An=52; Slope: 0.93 (95% CI: 0.91 to 0.96); Intercept: 2.38 (95% CI: 1.25 to 3.15); r: 1.00
    - Lithium HeparinN/An=52; Slope: 0.98 (95% CI: 0.95 to 0.99); Intercept: 0.42 (95% CI: -0.43 to 1.63); r: 1.00
    Reference Interval11.5 to 78.4 pg/mL10.3 to 80.5 pg/mLEstablish a 95% reference interval for a healthy population, consistent with clinical expectations and CLSI C28-A3 guidelines. The new range should be comparable to the predicate.
    Sensitivity
    - Limit of Blank (LoB)1.2 pg/mL0.9 pg/mLDemonstrate low limits of blank, detection, and quantitation, indicating the ability to detect very low concentrations of PTH, as per CLSI EP17-A guidelines. Should be comparable or better than the predicate.
    - Limit of Detection (LoD)2.5 pg/mL2.3 pg/mL
    - Limit of Quantitation (LoQ)4.5 pg/mL4.5 pg/mL
    Precision
    - Within-run CV%1.1% to 6.3%1.5% to 9.9%Demonstrate consistent and reproducible results across different concentration levels, within acceptable ranges for clinical diagnostic assays, as per CLSI EP-5A2 guidelines. Total CV should be within acceptable limits (e.g., <10-15% depending on concentration).
    - Total CV%4.1% to 8.2%1.8% to 9.9%
    Linearity
    - K2 EDTA Plasma$y = 1.002 x – 4.748; R^2 = 1.00$$y = 0.96 x − 0.1; R^2 = 1.00$Demonstrate linearity across the claimed measurement range (5 to 3500 pg/mL) with a regression coefficient ($R^2$) close to 1 and minimal deviation from expected values, as per CLSI EP-6A guidelines.
    - SerumN/A (Only K2 EDTA provided for predicate)$y = 1.02x – 0.2; R^2 = 1.00$
    High Dose Hook EffectNot observed up to 95,000 pg/mLNot observed up to 100,000 pg/mLNo significant hook effect should be observed within a clinically relevant and extended high concentration range to prevent falsely low results in patients with very high PTH levels. The demonstrated range should be comparable or wider than the predicate.
    Interferences (≤±10% bias)
    - Bilirubin, conjugatedN/A22 mg/dLNo significant interference (≤±10% bias) from common endogenous and exogenous interfering substances at clinically relevant concentrations, as per CLSI EP7-A2 guidelines. The tested concentrations should be comparable or improved compared to the predicate device.
    - Bilirubin, unconjugated20 mg/dL40 mg/dL
    - Biotin300 nmol/L300 nmol/L
    - CholesterolN/A395 mg/dL
    - HAMA1000 ng/mL1000 ng/mL
    - Hemoglobin250 mg/dL500 mg/dL
    - Rheumatoid Factor1500 IU/mL1836 IU/mL
    - Total ProteinN/A10 g/dL
    - Triglycerides3000 mg/dL3000 mg/dL
    - AcetaminophenN/A200 µg/mL
    - CarbamazepineN/A200 µg/mL
    - IbuprofenN/A500 µg/mL
    Cross-Reactivity (vs. PTH (1-84))
    - PTH (1-84)100%100%Demonstrate high specificity to intact PTH (1-84) and minimal cross-reactivity with PTH fragments or structurally similar proteins, improving upon the predicate where possible, as per CLSI EP7-A2 guidelines.
    - PTH (7-84)60%83.6%
    - PTH (1-34)0.5%<0.01%
    - PTH (39-84)Not detectable<0.01%
    - PTH (53-84)2%<0.01%
    - PTH (39-68)N/A<0.01%
    - PTH (44-68)2%<0.01%
    - Human Calcitonin4%<0.01%
    - CTX-1 (β CrossLaps)2%<0.01%
    - Osteocalcin2%<0.01%

    2. Sample Sizes and Data Provenance for Test Sets

    • Method Comparison:

      • Sample Size: 312 serum samples (291 native, 21 spiked).
      • Data Provenance: Not explicitly stated but clinical samples are generally considered prospective or retrospective from a clinical laboratory setting. The use of "commercially available quantitative chemiluminescence Intact PTH assay" as the comparator suggests real-world clinical samples.
    • Sample Matrix Comparison:

      • Sample Size: 52 matched samples (45 native, 7 spiked).
      • Data Provenance: Not explicitly stated, but clinical samples are typically used for this type of evaluation.
    • Reference Interval:

      • Sample Size: 129 individuals (67 males, 62 females; 21 to 89 years of age).
      • Data Provenance: K2 EDTA plasma samples collected from individuals with normal calcium, phosphate, and TSH values from the north, central and south regions of the United States. This is prospective or a well-characterized retrospective collection.
    • Sensitivity (LoB, LoD, LoQ):

      • Sample Size: 60 blanks and 10 low-level serum samples.
      • Data Provenance: Lab-prepared samples and low-level serum samples, likely from internal or commercially acquired biobanks.
    • Precision:

      • Sample Size: 7 samples (tested multiple times, N=80 for each sample).
      • Data Provenance: Lab-prepared controls and potentially pooled clinical samples ("Serum Sample," "EDTA Plasma Sample").
    • Linearity:

      • Sample Size: Not explicitly stated as a single number, but involved diluting high patient samples with low patient samples. Each sample was measured in 4 replicates.
      • Data Provenance: High and low intact PTH human K2 EDTA plasma and serum samples.
    • High Dose Hook Effect:

      • Sample Size: 3 samples (1 EDTA Plasma, 2 Serum) per spiked concentration.
      • Data Provenance: Spiked samples, likely from a laboratory setting using clinical matrices.
    • Interferences:

      • Sample Size: Two K2-EDTA samples (low and high Intact PTH levels) and two serum samples (different Intact PTH concentrations).
      • Data Provenance: Spiked samples, using clinical matrices (K2-EDTA and serum).
    • Cross-Reactivity:

      • Sample Size: A serum sample and a zero calibrator matrix.
      • Data Provenance: Spiked samples, using clinical serum matrix.

    3. Number of Experts and Qualifications for Ground Truth

    This device is an in vitro diagnostic assay, where the "ground truth" for performance characteristics such as sensitivity, precision, linearity, and interference is established by:

    • Reference Methods: Comparison against established, commercially available, and often FDA-cleared or gold-standard assays (e.g., the predicate device or a "commercially available quantitative chemiluminescence Intact PTH assay" for method comparison).
    • Analytical Standards/Spikes: Precisely known concentrations of analyte or interfering substances are used to create "spiked" samples, where the "true" concentration is known by design.
    • Clinical Laboratory Standard Institute (CLSI) Guidelines: Adherence to these guidelines (EP-9A2, EP9-A3, C28-A3, EP17-A, EP-5A2, EP-6A, EP7-A2) dictates the experimental design and statistical analysis which are considered "best practice" for establishing performance characteristics in IVDs.

    There are no "experts" in the sense of physicians or radiologists directly assessing images or patient data to establish ground truth for this type of analytical device performance. The "ground truth" is analytical and derived from established laboratory practices and reference methods.

    4. Adjudication Method for the Test Set

    Not applicable for this type of in vitro diagnostic device study. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies, particularly for diagnostic imaging or subjective clinical assessments, where consensus among multiple expert readers is needed to establish a ground truth for a disease state. For IVD analytical performance, the ground truth is either the value from a reference method or a gravimetrically/volumetrically determined "true" concentration.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is primarily relevant for diagnostic imaging AI algorithms where multiple human readers interpret cases with and without AI assistance. This document describes an in vitro diagnostic assay for a biochemical marker, not an imaging device or an AI application that assists human interpretation.

    6. Standalone Performance

    Yes, the studies described are standalone performance assessments of the IDS-iSYS Intact PTHN assay. The device is an automated system for measuring PTH levels, and the performance characteristics (method comparison, precision, linearity, etc.) directly evaluate the algorithm/assay's ability to accurately quantify PTH in patient samples. There is no human-in-the-loop component being evaluated in these tests; the device's output is directly compared to established analytical standards or reference methods.

    7. Type of Ground Truth Used

    • Reference Method Results: For method comparison, the results from a "commercially available quantitative chemiluminescence Intact PTH assay" served as the ground truth comparator.
    • Assigned Values of Controls/Standards: For precision, linearity, sensitivity, interference, and cross-reactivity studies, the ground truth was based on:
      • Known concentrations: Spiked samples with precisely known concentrations of analyte or interfering substances.
      • Expected values: Derived from dilution series or preparation of controls.
      • Blank matrices: For limit of blank determination.
    • Clinical Criteria: For the reference interval study, the ground truth for "normal" individuals was based on clinical criteria (normal calcium, phosphate, and TSH values).

    8. Sample Size for the Training Set

    The document does not specify a "training set" in the context of machine learning. This is an analytical medical device, not an AI/ML algorithm that is "trained" on data. The characteristics and parameters of the assay (e.g., antibodies, reagent concentrations, calibration curve algorithm) are developed and optimized during the product development phase, which is analogous to "training" in AI, but it's not described in terms of a distinct dataset.

    9. How Ground Truth for the Training Set Was Established

    As explained above, there isn't a "training set" in the common AI/ML sense. The "ground truth" for the development and optimization of the assay's chemical and instrumental parameters typically comes from:

    • Biochemical principles: Understanding of antigen-antibody interactions, chemiluminescence, and PTH biology.
    • Calibration materials: Use of international standards or highly characterized primary calibrators with assigned values.
    • Internal validation studies: Extensive testing during R&D to optimize reagent formulations, incubation times, and instrumental settings to achieve desired performance characteristics against known samples and reference methods.
    • Iterative development and testing: Performance data from pilot studies or earlier assay versions would inform adjustments to achieve acceptable analytical performance compared to the established medical consensus for PTH measurement.
    Ask a Question

    Ask a specific question about this device

    K Number
    K150879

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-06-22

    (82 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DiaSorin LIAISON® 1-84 PTH assay is an in vitro chemiluminescent immunoassay (CLIA) intended for the quantitative determination of parathyroid hormone (1-84) in human serum and EDTA plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia resulting from disorders of calcium metabolism.

    The test has to be performed on the LIAISON® Analyzer.

    The DiaSorin LIAISON® 1-84 PTH Control Set is intended for in vitro diagnostic use as assayed quality control samples to monitor the accuracy and precision of the LIAISON® 1-84 PTH Assay.

    The DiaSorin LIAISON® 1-84 PTH Calibration Verifiers are assayed quality control materials intended for in vitro diagnostic use in the quantitative verification of calibration and reportable range of the LIAISON® 1-84 PTH Assay.

    Device Description

    The LIAISON® 1-84 PTH Assay is a modified two-step, two-site sandwich assay that uses two polyclonal antibodies for capture and detection of the 1-84 PTH molecule. Results are determined by a 2 point calibration conversion of the master curve to a working curve. The light signal is measured by a photomultiplier as relative light units (RLU) and is proportional to the concentration of 1-84 PTH present in the calibrators, controls or samples.

    LIAISON® 1-84 PTH Control set contains;

    • 2 levels controls containing human plasma spiked with 1-84 PTH, and preservatives: 4 vials each level; lyophilized
      The target concentration for control level 1 is 30 pg/mL. The target concentration for control Level 2 is 260 pg/mL.
      The range of concentrations of each control is reported on the certificate of analysis provided with each LIAISON® 1-84 PTH Control set.

    LIAISON® 1-84 PTH Calibration Verifier set contains:

    • 4 levels containing human plasma spiked with 1-84 PTH, with preservative, 1 vial each level, lyophilized.
      The target concentration for cal verifier A is 10 pg/mL. The target concentration for cal verifier B is 80 pg/mL. The target concentration for cal verifier C is 400 pg/mL. The target concentration for cal verifier D is 1450 pg/mL.
      The range of concentrations of each calibration verifier is reported on the certificate of analysis provided with each LIAISON® 1-84 PTH Calibration Verifier set.
    AI/ML Overview

    This document describes the performance of the DiaSorin LIAISON® 1-84 PTH Assay, LIAISON® 1-84 PTH Control Set, and LIAISON® 1-84 PTH Calibration Verifiers, concluding they are substantially equivalent to their predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in the provided text as numerical targets for each performance characteristic. Instead, the study demonstrates performance by comparing the LIAISON® 1-84 PTH Assay with its predicate device and established CLSI guidelines. The performance characteristics for the LIAISON® 1-84 PTH Assay are summarized below:

    Performance CharacteristicReported Device Performance (LIAISON® 1-84 PTH Assay)
    Method Comparison- n: 193 patient samples (comparative to Scantibodies Laboratory Inc. Whole PTH™ (1-84) Specific IRMA assay).- Passing & Bablok Regression (R): 0.9812 (indicates strong correlation).- Slope: 0.9810 (95% CI: 0.9497 to 1.0204) (close to 1).- Intercept: -2.23 pg/mL (95% CI: -5.29 to -0.61) (close to 0).
    Sample Matrix Comparison- n: 188 matched patient sets of EDTA plasma and serum samples.- Passing & Bablok Regression (R²): 0.9981 (between serum and EDTA plasma).- Slope: 1.0481 (95% CI: 1.03 – 1.08).- Intercept: 0.14 pg/mL (95% CI: -0.48 – 0.69). Comparisons indicate good agreement between sample types.
    Reference Range- EDTA Plasma: 5.72 - 45.4 pg/mL (Median: 25.0 pg/mL, N=125).- Serum: 5.68 - 47.8 pg/mL (Median: 25.2 pg/mL, N=124).- Established from healthy adults (21-70 years) from mixed ethnic backgrounds in the U.S. with normal relevant health markers.
    Precision- Kit Control 1 (Mean 30.2 pg/mL): Total %CV 5.0% (N=160 replicates).- Kit Control 2 (Mean 305.6 pg/mL): Total %CV 4.6% (N=160 replicates).- EDTA PTH-S1 (Mean 11.8 pg/mL): Total %CV 10.4% (N=160 replicates).- EDTA PTH-S7 (Mean 1743.9 pg/mL): Total %CV 5.8% (N=160 replicates).- Calibration Verifier A (Mean 13.0 pg/mL): TOTAL (Within-lot) %CV 8.2% (N=80 replicate results).- Calibration Verifier D (Mean 1487 pg/mL): TOTAL (Within-lot) %CV 5.7% (N=80 replicate results).
    Linearity- Serum: Observed 1-84 PTH = 0.9992x + 0.0835; R² = 0.9976.- EDTA plasma: Observed 1-84 PTH = 0.9679x - 3.328; R² = 0.9971. (Very high R² values indicate strong linearity across the tested range). If an AI component is involved, please describe what the AI is (e.g. classification, segmentation) and what it is trying to predict
    High Dose Hook EffectNo high dose hook effect observed for 1-84 PTH concentrations measured up to 60,000 pg/mL.
    RecoveryMean Recovery across various spiked EDTA plasma samples was 95%.
    Analytical Specificity- Very low % Cross Reactivity for all tested PTH fragments (e.g., 7-84 PTH: 0.00105%, 1-34 PTH: 0.00005%) and structurally similar proteins.
    InterferenceNo significant interference (≥ ±10%) observed for a wide range of endogenous substances (e.g., Hemoglobin up to 500 mg/dL, Bilirubin up to 40 mg/dL, Triglycerides up to 3,000 mg/dL) and exogenous substances (e.g., Acetaminophen up to 20 mg/dL, Vitamin D2/D3 up to 240 ng/mL, Biotin up to 1 ug/mL) at two PTH levels (40 and 70 pg/mL).
    Limit of Blank (LOB)≤ 0.5 pg/mL
    Limit of Detection (LOD)1.0 pg/mL
    Limit of Quantitation (LOQ)4.0 pg/mL
    StabilityReagent Integral: 28 days (open vial on system/at 2-8°C). Calibrators/Controls: 8 weeks (open vial reconstituted and frozen at -20°C). Calibration curve: 7 days. Calibration Verifiers: 2 weeks (open vial reconstituted and frozen at -20°C).
    Measuring Range4 - 1800 pg/mL
    TraceabilityCalibrators, Controls, and Calibration Verifiers referenced to an in-house standard preparation containing synthetic human PTH (1-84).

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Method Comparison: 193 patient samples. The provenance is implied to be from patient samples used for laboratory testing, but specific country of origin or whether they were retrospective/prospective is not stated.
    • Sample Matrix Comparison: 188 matched patient sets of EDTA plasma and serum samples. Data provenance is not explicitly stated as retrospective or prospective, nor is the country of origin.
    • Reference Range: 125 EDTA plasma samples (91 females, 34 males) and 124 serum samples (90 females, 34 males) collected from "apparently healthy adults; 21 - 70 years of age, from mixed ethnic backgrounds (32.0% dark-skinned, 67.2% lightskinned and 0.8% unknown) with normal 25 OH Vitamin D, TSH, Total Calcium, Phosphorus, Magnesium, Creatinine and Alkaline Phosphatase levels from the northern and southern regions of the U.S." This indicates the data is from the United States and is prospective in nature, as samples were collected from healthy individuals to establish ranges.
    • Precision:
      • Coded panel (7 frozen EDTA plasma samples): 160 replicate results per sample (2 replicates per run, 2 runs per day for 20 operating days).
      • Kit Controls (2 levels): 160 replicate results per control (2 replicates per run, 2 runs per day for 20 operating days).
      • Calibration Verifiers (4 levels): 80 replicate results per calibration verifier (2 replicates per run, 2 runs per day for 20 operating days).
      • Data provenance not specified beyond "prepared by DiaSorin Inc."
    • Linearity: Not explicitly stated, but "one sample pool of each type: serum and EDTA plasma were diluted and analyzed."
    • High Dose Hook Effect: Three serum and three EDTA plasma samples were used, spiked with 1-84 PTH. Measured in triplicate.
    • Recovery Study: Five high concentration EDTA plasma samples and five low concentration EDTA plasma samples were analyzed.
    • Analytical Specificity Cross-Reactivity Studies: Not specified, but involved spiking LIAISON® 1-84 PTH Specimen Diluent with various substances.
    • Interference Studies: Not specified, but involved spiking EDTA plasma at two PTH levels (40 and 70 pg/mL) with various substances.
    • Limit of Blank, Limit of Detection, Limit of Quantitation: Not specified, but determined using CLSI EP17-A2.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For in vitro diagnostic (IVD) assays, the "ground truth" for patient samples is typically established by the reference method (either a validated predicate device or a clinical reference standard) and the inherent biological characteristics of the samples. This is not a study involving human readers or interpretation of images. Therefore:

    • Number of Experts: Not applicable in the context of interpretation, as the device measures a quantitative analyte. The analytical performance is compared against a predicate device and established clinical guidelines.
    • Qualifications of Experts: Not applicable in the context of interpretation. However, the studies were conducted according to CLSI guidelines, implying professionals with expertise in clinical laboratory testing and assay validation.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study requiring adjudication of interpretations (e.g., 2+1, 3+1). The performance is based on quantitative measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes an in vitro diagnostic assay, not a medical imaging or AI-assisted diagnostic device that involves human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented are for the standalone performance of the LIAISON® 1-84 PTH Assay on the LIAISON® Analyzer. This is a fully automated immunoassay system; there is no "human-in-the-loop" once the samples are loaded and the assay is initiated. The reported performance characteristics (e.g., precision, linearity, accuracy against predicate) represent the algorithm's performance.

    7. The Type of Ground Truth Used

    The ground truth for the performance studies was established through:

    • Reference Method Comparison: Comparison against the Scantibodies Laboratory, Inc. Whole PTH™ (1-84) Specific immunoradiometric (IRMA) assay (predicate device) for patient sample results.
    • Spiked Samples: For linearity, high-dose hook effect, recovery, analytical specificity, and interference studies, known concentrations of PTH or interfering substances were added to samples, with the expectation that the assay would accurately measure these known concentrations.
    • Healthy Donor Population: For reference range establishment, samples from apparently healthy individuals with normal levels of related biomarkers formed the basis for defined reference intervals.
    • Internal Standards: The calibrators, controls, and calibration verifiers are traceable to "an in-house standard preparation containing synthetic human PTH (1-84)," serving as an internal ground truth for the assay's measurements.

    8. The Sample Size for the Training Set

    This document describes the validation of an IVD assay, not an AI/ML device that requires a distinct "training set" in the machine learning sense. The assay works based on established biochemical principles and does not learn from data in the same way an AI algorithm does. Therefore, a "training set" in that context is not applicable.

    However, if one were to consider the data used to develop and optimize the assay parameters (prior to the validation studies reported here), that would be considered the equivalent of a "training set." The document does not specify the sample size for this developmental phase.

    9. How the Ground Truth for the Training Set Was Established

    As noted in point 8, the concept of a "training set" for this type of IVD, where performance is based on chemical-immunological reactions rather than machine learning, is not directly applicable. The "ground truth" during the development and optimization of such an assay would typically be established through:

    • Reference materials: Highly characterized synthetic PTH or purified natural PTH standards with known concentrations.
    • Validated methods: Existing, well-established reference methods for PTH measurement.
    • Clinical samples: Use of patient samples with confirmed clinical states (e.g., hyperparathyroidism, hypoparathyroidism) to ensure the assay's output correlates with physiological conditions.
    • Protocols following good laboratory practices and quality system regulations to ensure the accuracy and reliability of these references.

    The document implicitly refers to these by stating the LIAISON® 1-84 PTH Calibrators, Controls, and Calibration Verifiers are "referenced to an in-house standard preparation containing synthetic human PTH (1-84)." This "in-house standard" would be the foundational "ground truth" for the assay's quantitative measurements.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133601

    Validate with FDA (Live)

    Date Cleared
    2014-02-05

    (72 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA® Centaur Intact Parathyroid (iPTH) assay is for in vitro diagnostic use in the quantitative determination of intact parathyroid hormone (iPTH) in EDTA plasma or serum using the ADVIA Centaur and ADVIA Centaur XP systems. This assay is intended to be used to aid in the differential diagnosis of hyperparathyroidism and hypoparathyroidism.

    Device Description

    The ADVIA Centaur iPTH assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses constant amounts of an antihuman PTH antibody in the Lite Reagent and an antihuman PTH antibody in the Solid Phase Reagent. The first antibody is a polyclonal goat antihuman PTH (N-terminal 1-34) antibody labeled with acridinium ester. The second antibody is a biotinylated polyclonal goat antihuman PTH (39-84 region) antibody that is preformed to streptavidin coated paramagnetic latex particles in the Solid Phase reagent.

    AI/ML Overview

    The provided text describes the ADVIA® Centaur Intact Parathyroid Hormone (iPTH) Assay, a device intended for the quantitative determination of iPTH in EDTA plasma or serum to aid in the differential diagnosis of hyperparathyroidism and hypoparathyroidism. The document focuses on demonstrating the substantial equivalence of a modified version of this assay to its predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly list "acceptance criteria" in a separate table for each performance characteristic with pass/fail outcomes. Instead, it describes various performance studies and presents their results, implying that these results met internal criteria for demonstrating substantial equivalence. Based on the described tests, here's a reconstructed table of implicit acceptance criteria and the reported performance:

    Performance CharacteristicImplicit Acceptance Criteria (based on study design and intent)Reported Device Performance
    PrecisionCVs within acceptable clinical limits; comparable to predicate.Pool 1: 16.1 pg/mL, 7.7% Total CV Control 1: 38.7 pg/mL, 4.7% Total CV Pool 2: 62.8 pg/mL, 3.7% Total CV Control 2: 185 pg/mL, 3.2% Total CV Control 3: 663 pg/mL, 2.8% Total CV Pool 3: 839 pg/mL, 3.9% Total CV Pool 4: 1698 pg/mL, 3.2% Total CV
    Detection Limits (LoB, LoD, LoQ)LoB, LoD, LoQ established and clinically acceptable. Compare favorably to predicate (implied, though not directly shown).LoB: 0.1 pg/mL LoD: 2.7 pg/mL LoQ: 6.3 pg/mL
    Linearity / Assay RangeLinear response across the claimed assay range.Regression: y = 0.98x - 4.98, R = 1.00 Assayed range: 4.4 to 2021 pg/mL (claimed range 6.3 - 1900 pg/mL)
    High Dose Hook EffectNo high dose hook effect observed.No high dose hook was observed for either reagent lot (tested up to 144,000 pg/mL).
    Interfering SubstancesBias ≤ 10% effect at specified interferent levels.Hemoglobin (500mg/dL): 4% and -1% bias Triglycerides (3000mg/dL): 7% and -4% bias Conjugated Bilirubin (40mg/dL): 7% and 0% bias Unconjugated Bilirubin (40mg/dL): 7% and 8% bias Biotin (1000ng/mL): 8% and 6% bias
    Cross-reacting SubstancesLow to no cross-reactivity with related substances, except for clinically expected interactions.PTH 7-84: ~51% cross-reactivity (noted as an exception, implying this level is understood or acceptable) All other listed cross-reactants (PTH 1-34, 39-68, 39-84, 44-68, 53-84, Calcitonin, Beta-Cross Laps, Osteocalcin): showed 0.0% to 0.2% cross-reactivity (very low).
    Method ComparisonGood correlation with the predicate device (R-value close to 1, slope near 1, intercept near 0).EDTA Plasma: y = 0.98x + 10.6, R = 1.00 (N=106, range 11.8 - 1862 pg/mL) Serum: y = 1.00x + 5.0, R = 1.00 (N=105, range 9.8 - 1868 pg/mL)
    Expected Values (Reference Range Verification)Percentage of samples falling within the predicate's reference range (12-85 pg/mL) should be high (e.g., ≥95%).Plasma (Lot 1 & 2): 100.0% within range Serum (Lot 1): 97.5% within range Serum (Lot 2): 95.0% within range
    StabilityReagent stability confirmed for claimed periods.On-board stability: 28 days (with 14-day calibration interval). Reagent stable until printed date at 2-8°C.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Precision: 7 samples (4 EDTA plasma pools, 3 commercial control materials). No specific provenance (country of origin) mentioned, but likely clinical samples and commercial controls from within Siemens' testing facilities or sourced according to industry standards.
    • Detection Limits (LoB, LoD, LoQ): A single blank (Multi-Diluent 11) and 5 test samples (spiked/diluted plasma samples). Each sample tested n=6 x 2 instruments x 5 days (total n=60 per reagent lot for 2 reagent lots). Provenance not specified.
    • Linearity / Assay Range: 12 test concentrations (pools). Provenance not specified.
    • High Dose Hook Effect: One hook sample. Provenance not specified.
    • Interfering Substances: Two patient plasma pools. Provenance not specified.
    • Cross-reacting Substances: A plasma sample with endogenous iPTH and an assay-specific diluent. Provenance not specified.
    • Method Comparison: 106 unaltered native matched serum and plasma samples. No specific provenance (country of origin) or retrospective/prospective status explicitly stated. Clinical samples are typically retrospective or prospectively collected for validation studies.
    • Expected Values (Reference Range Verification): 40 plasma and 40 serum samples. Obtained from "apparently healthy individuals" (implying human clinical samples). Provenance not specified.

    All studies appear to be retrospective in nature, as they involve testing existing samples or prepared samples under controlled conditions to evaluate device performance characteristics.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    For this type of in vitro diagnostic device (immunoassay), ground truth is not typically established by human expert review in the same way it would be for imaging diagnostics.

    • Ground truth for quantitative assays like this one is generally established through:
      • Reference Methods: Highly accurate and precise laboratory methods (e.g., mass spectrometry). While not explicitly stated for all measurements, this is the gold standard for establishing true analyte concentrations.
      • Predicate Device: For method comparison, the predicate device's results serve as the comparison "truth" against which the modified device is evaluated.
      • Defined Concentrations: For studies like linearity, precision, and detection limits, samples are often prepared with known, spiked, or characterized concentrations.

    There is no mention of human "experts" establishing ground truth for individual test samples. The "experts" involved are implied to be the laboratory personnel, scientists, and statisticians who designed and conducted the studies, and who interpret the results against established statistical and clinical criteria (e.g., CLSI guidelines).

    4. Adjudication Method for the Test Set

    Not applicable for this type of quantitative diagnostic assay. Adjudication methods like 2+1 or 3+1 are typically used in imaging studies where multiple human readers assess findings and resolve discrepancies. Here, the "truth" for evaluation is based on quantitative measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is a laboratory immunoassay, not an AI-assisted diagnostic imaging device that involves human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, this is a standalone device. The performance characteristics (precision, linearity, detection limits, interference, cross-reactivity, method comparison) are all evaluations of the assay itself, demonstrating its analytical performance without any human intervention beyond standard laboratory procedures for running the test. It's a fully automated system for quantitative determination.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The ground truth used in these studies is primarily based on:

    • Reference standards/known concentrations: For linearity, precision, and detection limit studies, samples are often prepared with known concentrations or characterized against highly accurate reference methods.
    • Values from the predicate device: For method comparison, the results obtained from the predicate device serve as the comparative "ground truth."
    • Clinical status for reference ranges: Normal healthy individuals are used to verify the expected reference range.

    8. The Sample Size for the Training Set

    Not applicable in the traditional sense of machine learning. This device is an immunoassay, not a machine learning algorithm that requires a "training set" to learn a model. Its performance is based on the chemical and immunological reactions of the assay components.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the reasons stated above (not an ML-based device requiring a training set). The assay's performance is driven by its reagent formulation and instrument design, which are developed and validated through iterative testing and process control during manufacturing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132515

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-11-08

    (88 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIAISON® N-TACT® PTH Gen II is an in vitro chemiluminescent immunoassay (CLIA) intended for the quantitative determination of intact human parathyroid hormone in serum, EDTA and Lithium Heparin plasma samples. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism. The test is to be performed on the LIAISON® XL Analyzer.

    The LIAISON® N-TACT® PTH Gen II Control Set is intended for use as assayed quality control samples to monitor the accuracy and precision of the DiaSorin LIAISON® N-TACT® PTH Gen II assay.

    The LIAISON® N-TACT® PTH Gen II Calibration Verifiers are assayed quality control materials intended for the quantitative verification of calibration and reportable range of the LIAISON® N-TACT® PTH Gen II assay.

    Device Description

    The LIAISON® N-TACT® PTH Gen II assay is a modified two-step, two-site sandwich assay that uses two goat polyclonal antibodies for capture and detection of intact PTH. Results are determined by a 2 point calibration conversion of the master curve to a working curve. The light signal is measured by a photomultiplier as relative light units (RLU) and is proportional to the concentration of intact PTH present in the calibrators, controls or samples.

    LIAISON® N-TACT® PTH Gen II Control set contains:
    2 levels controls containing 80% human plasma spiked with 1-84 PTH, and preservatives; 4 vials each level; lyophilized
    The target concentration for control level 1 is 20 pg/mL. The target concentration for control Level 2 is 300 pg/mL.
    The range of concentrations of each control is reported on the certificate of analysis provided with each LIAISON® N-TACT® PTH Gen II Control set.

    LIAISON® N-TACT® PTH Gen II Calibration Verifier set contains:
    4 levels containing 80% human plasma spiked with 1-84 PTH, with preservative, . 1 vial each level, lyophilized
    The target concentration for cal verifier A is 10 pg/mL. The target concentration for cal verifier B is 150 pg/mL. The target concentration for cal verifier C is 650 pg/mL. The target concentration for cal verifier D is 1600 pg/mL.
    The range of concentrations of each calibration verifier is reported on the certificate of analysis provided with each LIAISON® N-TACT® PTH Gen II Calibration Verifier set.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the LIAISON® N-TACT® PTH Gen II device based on the provided 510(k) summary:

    Acceptance Criteria and Reported Device Performance

    The 510(k) summary for the LIAISON® N-TACT® PTH Gen II primarily demonstrates substantial equivalence to a predicate device. As such, the "acceptance criteria" are generally implied by the performance of the predicate device and the demonstration that the new device performs comparably or better, meeting established clinical laboratory guidelines. Specific numeric acceptance criteria are not explicitly stated in a "PASS/FAIL" format for each performance characteristic, but rather the study results are presented to show satisfactory performance.

    Here's a table summarizing the performance characteristics and their reported results, which implicitly serve as the demonstration of meeting acceptance:

    Performance CharacteristicAcceptance Criteria (Implied / Predicate Performance)Reported Device Performance (LIAISON® N-TACT® PTH Gen II)
    Method Comparison (vs. Predicate)Substantial equivalence to Siemens ADVIA® CENTAUR INTACT iPTH (K020217)n=198 Slope: 1.010 (95% CI: 0.99 to 1.03) Intercept: -1.5851 pg/mL (95% CI: -3.11 to -0.44) Correlation coefficient (r): 0.9953
    Measuring RangeComparable to predicate (2.5 - 1900 pg/mL)3 - 1900 pg/mL
    Sample Matrix EquivalenceEquivalent results across EDTA plasma, serum, SST serum, Lithium Heparin plasmaEDTA plasma vs. Serum: Slope 0.97, Int. -2.45, R² 0.9986 EDTA plasma vs. SST Serum: Slope 1.01, Int. -2.25, R² 0.9996 EDTA plasma vs. Lithium Heparin: Slope 0.98, Int. -0.01, R² 0.9991
    Reference RangeEstablished and clinically appropriate14.5 - 87.1 pg/mL (n=125 healthy adults from US)
    PrecisionDemonstrated by low %CVs across various PTH levels (following CLSI EP5-A2)Total %CVs across lots (selected examples): 19.3 pg/mL: 3.3% 250 pg/mL: 3.5% 12.6 pg/mL: 4.2% 1477 pg/mL: 2.8%
    LinearityLinear response across the assay range (following CLSI EP6-A)R² for various matrices: Serum: 0.9982, SST Serum: 0.9987, EDTA plasma: 0.9983, Lithium Heparin plasma: 0.9992 (all close to 1, indicating linearity)
    High Dose Hook EffectNo hook effect within a specified rangeNo hook effect observed up to 1,000,000 pg/mL of PTH
    RecoveryAcceptable percentage recovery valuesMean Recovery: 97% (range 93%-103% across various spiked samples)
    Analytical Specificity (Cross-Reactivity)Minimal cross-reactivity with related substancesPTH (7-84): 53% Other PTH fragments, Calcitonin, C-Telopeptide, Osteocalcin: < 0.01%
    Interference Studies (Endogenous)No significant interference with common endogenous substances (≤ ±10%)No significant interference at specified concentrations (e.g., Hemoglobin 500 mg/dL, Bilirubin 40 mg/dL, Triglycerides 3,000 mg/dL, etc.)
    Interference Studies (Exogenous)No significant interference with common exogenous substances (≤ ±10%)No significant interference at specified concentrations (e.g., Acetaminophen 0.2 mg/mL, Ibuprofen 0.5 mg/mL, various bisphosphonates, vitamins, calcium/magnesium/aluminum salts)
    Limit of Blank (LoB)Low LoB value< 0.5 pg/mL
    Limit of Detection (LoD)Low LoD value5 pg/mL
    Limit of Quantitation (LoQ)Low LoQ value3.0 pg/mL
    Stability (Reagent Integral on system)Acceptable on-board stability56 days
    Stability (Calibration curve)Acceptable calibration interval28 days

    Study Details for LIAISON® N-TACT® PTH Gen II

    1. Sample sizes used for the test set and data provenance:

      • Method Comparison: 198 patient samples. The provenance is not explicitly stated as country of origin, but the reference range study mentions "northern and southern regions of the U.S.", suggesting US-based samples for some studies. The study is retrospective, utilizing existing patient samples.
      • Sample Matrix Comparison: 65 matched patient sets (EDTA plasma, serum, SST serum, and Lithium Heparin plasma). Data provenance is not explicitly stated beyond being "patient sets." Retrospective.
      • Reference Range: 125 apparently healthy adults aged 21-70 years from mixed ethnic backgrounds (32.5% dark-skinned, 66.7% light-skinned, 0.8% unknown) from the northern and southern regions of the U.S. This is a prospective or retrospective collection of samples for establishing a reference interval.
      • Precision: 7 frozen EDTA plasma samples (coded panel) and 2 lots of LIAISON® N-TACT® PTH Gen II controls (2 levels). Samples were tested on two lots of reagents. This is a controlled lab study, not involving patient data per se beyond the initial source of the plasma samples.
      • Linearity: One sample pool of each type: serum, SST serum, EDTA plasma, and Lithium Heparin plasma. These were diluted. Controlled lab study.
      • High Dose Hook Effect: A zero sample spiked with PTH. Controlled lab study.
      • Recovery Study: 5 high concentration EDTA plasma samples and 5 low concentration EDTA plasma samples. Controlled lab study, not directly patient data.
      • Analytical Specificity (Cross-Reactivity & Interference): Samples spiked with specific substances. These are controlled lab studies.
      • Limit of Blank, Limit of Detection, Limit of Quantitation: Not explicitly stated, usually involves multiple replicates of blank and low-concentration samples. Controlled lab study.
    2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

      • This device is an in vitro diagnostic (IVD) immunoassay, not an image-based AI device. The "ground truth" is established through analytical methods and comparison to established, cleared diagnostic devices (predicate device) and clinical standards, rather than expert human interpretation.
      • For the method comparison, the "ground truth" for the test set is the result from the predicate device, Siemens ADVIA® CENTAUR INTACT Parathyroid Hormone (iPTH) Assay, which is an FDA-cleared device.
      • For other analytical performance studies (precision, linearity, recovery, etc.), the ground truth relies on carefully prepared samples with known concentrations or expected behaviors, measured against established analytical principles and validated laboratory methods. There are no "experts" establishing image-based ground truth here.
    3. Adjudication method for the test set:

      • Not applicable as this is an IVD immunoassay, not a diagnostic imaging AI where human adjudication of ambiguous cases is typically required. The comparison is objective, numerical data between two analytical methods or against defined analytical targets.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is an IVD assay, not an AI-powered diagnostic imaging tool that assists human readers. The performance is assessed against a predicate device and analytical standards.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • The LIAISON® N-TACT® PTH Gen II assay is a standalone device in the sense that it provides quantitative results for PTH concentration. It operates autonomously on the LIAISON® XL Analyzer without direct human interpretation of the result generation process. Interpretation of the PTH level in a clinical context (e.g., diagnosis of hyper/hypocalcemia) still involves human clinicians, but the device provides the raw measurement itself without human input into the measurement.
    6. The type of ground truth used:

      • For Method Comparison: Results from the predicate device (Siemens ADVIA® CENTAUR INTACT PTH (iPTH) Assay, K020217).
      • For other analytical studies (Precision, Linearity, Recovery, etc.): Typically based on known concentrations of analytes in spiked samples or reference materials, or expected analytical behaviors according to established CLSI guidelines (e.g., EP5-A2 for Precision, EP6-A for Linearity, EP7-A2 for Interference, EP17-A2 for Detection Capability). For the reference range, it's derived from a healthy population cohort carefully selected using specific clinical criteria.
    7. The sample size for the training set:

      • This document describes performance characteristics for an IVD kit, not an AI/Machine Learning model. Therefore, there isn't a "training set" in the sense of data used to train an algorithm. The development of the assay itself would involve internal optimization and validation, but these stages are not typically referred to as "training sets" in the context of conventional IVDs in 510(k) summaries.
    8. How the ground truth for the training set was established:

      • As there is no "training set" in the AI/ML context, this question is not applicable. The assay's analytical characteristics are developed and verified through standard laboratory practices and comparison to reference methods, not through an iterative learning process with labeled data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K121981

    Validate with FDA (Live)

    Date Cleared
    2013-03-21

    (258 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA® Centaur Intact Parathyroid (iPTH) assay is for in vitro diagnostic use in the quantitative determination of intact parathyroid hormone (iPTH) in EDTA plasma or serum using the ADVIA Centaur XP system. This assay is intended to be used to aid in the differential diagnosis of hyperparathyroidism and hypoparathyroidism.

    Device Description

    The ADVIA Centaur iPTH assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses constant amounts of an antihuman PTH antibody in the Lite Reagent and an antihuman PTH antibody in the Solid Phase Reagent. The first antibody is a polyclonal goat antihuman PTH (N-terminal 1-34) antibody labeled with acridinium ester. The second antibody is a biotinylated polyclonal goat antihuman PTH (39-84 region) antibody that is preformed to streptavidin coated paramagnetic latex particles in the Solid Phase reagent.

    The ADVIA Centaur iPTH reagent kit contains the following:
    . ReadyPack® primary reagent pack containing ADVIA Centaur Lite and Solid Phase Reagent)
    ADVIA Centaur iPTH Master Curve card .

    Materials Required but Not Provided
    iPTH Calibrator .
    Optional Reagents
    ADVIA Centaur Multi-Diluent 11 .
    iPTH 1, 2, 3 quality control material .
    iPTH Master Curve Material .

    AI/ML Overview

    The Siemens ADVIA® Centaur Intact Parathyroid (iPTH) Assay is a quantitative in vitro diagnostic device used to measure intact parathyroid hormone (iPTH) in EDTA plasma or serum to aid in the differential diagnosis of hyperparathyroidism and hypoparathyroidism. The device's performance was evaluated through studies on precision, interfering substances, cross-reactivity, and method comparison, demonstrating substantial equivalence to the predicate device, the Abbott Architect Intact PTH assay.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria were implicitly established by demonstrating comparability to the predicate device and meeting common analytical performance standards for in vitro diagnostic assays. The specific acceptance thresholds for each performance characteristic (e.g., %CV for precision, % bias for interference) are not explicitly stated as numbered criteria but are inferred from the reported results and the conclusion of substantial equivalence.

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
    PrecisionWithin-run and Total Imprecision should be acceptable for clinical use and comparable to predicate device performance.Within Run %CV: Ranged from 2.6% (Medical Decision Pool 5, 1802.9 pg/mL and Control 3, 691.0 pg/mL) to 6.0% (Low Calibrator, 33.0 pg/mL).Total %CV: Ranged from 3.2% (Medical Decision Pool 5, 1802.9 pg/mL) to 9.2% (Patient Pool, 16.7 pg/mL). Overall, coefficients of variation were below 10%, indicating good precision.
    Interfering SubstancesBias due to interferents (hemoglobin, bilirubin, triglycerides, biotin) should be < 10%.Hemoglobin (up to 500 mg/dL): Bias ranged from -9.59% to -0.22%. At 200 mg/dL, bias ranged from -8.26% to -2.03%. All within < 10% effect.Bilirubin (unconjugated up to 40 mg/dL, conjugated up to 40 mg/dL): Bias ranged from -5.49% to 6.33%. All within < 10% effect.Triglycerides (up to 3000 mg/dL): Bias ranged from -3.75% to 7.14%. All within < 10% effect.Biotin (up to 0.10 mg/dL): Bias ranged from -3.35% to 7.54%. All within < 10% effect.
    Cross-reactivityNegligible cross-reactivity with related substances, or reported if significant.PTH 1-34, PTH 39-68, PTH 39-84, PTH 44-68, PTH 53-84, Calcitonin, Beta-Cross Laps, Osteocalcin: Generally 0.00% cross-reactivity or negligible (e.g., -0.02% to 0.17%).PTH 7-84 fragment: Showed significant cross-reactivity of ~51% (50.74% in Multi-Dil 11 and 51.39% in Plasma Sample). This was noted for reporting in labeling.
    Method ComparisonGood correlation with the predicate device (Abbott Architect Intact PTH assay), indicating substantial equivalence in results.Correlation (y = ADVIA Centaur iPTH, x = Predicate): Based on 177 EDTA plasma samples, the regression equation was y = 1.03x - 3.32 with an R-value of 0.991 (using Least Squares Linear regression) and a mean (median) bias of -2.0% (-1.5%). This demonstrates strong correlation.
    Matrix EquivalencyEDTA plasma and serum results should be highly correlated, indicating interchangeability.Correlation (y = Serum, x = EDTA plasma): Based on 79 paired samples, the regression equation was 0.98x + 3.72 with an R-value of 0.992 (using Weighted Linear regression). This indicates that the two matrices are equivalent.
    Expected ValuesReference ranges for plasma and serum should be established and comparable to clinically accepted values and predicate device.Plasma: 13.8 to 85.0 pg/mL (1.46 to 9.01 pmol/L) for 95% of healthy donors.Serum: 12.4 to 76.8 pg/mL (1.31 to 8.14 pmol/L) for 95% of healthy donors. These values were established to confirm comparability to the predicate device and clinical norms.

    2. Sample Size and Data Provenance

    • Precision:
      • One patient EDTA plasma pool, five Medical Decision Pools (iPTH-spiked human EDTA plasma), two levels of calibrators, and three levels of commercial control materials.
      • Each sample was assayed in 2 replicates per run, 2 runs per day for 20 days.
      • Total of 80 replicates per sample.
      • Provenance: Not explicitly stated, but typically these studies are performed by the manufacturer, implying internal data collection. The use of "human EDTA plasma pools" suggests human-derived samples. These are likely prospective studies conducted for the purpose of regulatory submission.
    • Interfering Substances:
      • Three levels of PTH were evaluated for each interferent (Hemoglobin, Bilirubin, Triglycerides, Biotin).
      • Provenance: Not explicitly stated, likely internal, prospective.
    • Cross-reactivity:
      • One normal EDTA plasma sample and an assay-specific Multi-diluent 11 (buffered goat serum with no analyte).
      • Provenance: Not explicitly stated, likely internal, prospective.
    • Method Comparison:
      • Test Set Sample Size: 177 EDTA plasma samples.
      • Data Provenance: Not explicitly stated, but typically collected internally for comparison against a predicate device, likely prospective for the purpose of the study.
    • Serum Plasma Matrix Equivalency:
      • Test Set Sample Size: 79 matched EDTA plasma and serum samples.
      • Data Provenance: Not explicitly stated, likely internal, prospective.
    • Expected Values:
      • Matched serum and plasma samples from apparently healthy donors. Specific number not stated but usually involves a statistically significant cohort (e.g., >100 individuals) to establish a 95% reference range.
      • Data Provenance: Not explicitly stated, but typically involves collecting samples from a healthy population, likely prospective.

    3. Number of Experts and Qualifications

    This submission pertains to an in vitro diagnostic assay, which relies on quantitative measurements rather than expert interpretation of images or clinical findings. Therefore, the concept of "experts" to establish ground truth in the context of diagnostic imaging or clinical assessment is not applicable here. The ground truth is based on the chemical and immunological properties of the assay and comparison to a legally marketed predicate device.

    4. Adjudication Method

    Not applicable for an in vitro diagnostic assay like this. Adjudication methods are typically used in studies involving human interpretation or subjective assessments (e.g., by multiple readers).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is an in vitro diagnostic assay, not a device requiring human reader interpretation or assistance.

    6. Standalone Performance

    Yes, the studies described are for the standalone performance of the ADVIA Centaur iPTH assay (algorithm only, without human-in-the-loop performance in the sense of interpretative assistance). The performance metrics (precision, interference, cross-reactivity, method comparison) directly evaluate the assay's ability to accurately and reliably measure iPTH concentrations.

    7. Type of Ground Truth Used

    The ground truth for this device is established primarily by:

    • Quantitative Chemical Measurement: The assay directly measures the concentration of intact parathyroid hormone using a chemiluminescent immunoassay. The "truth" is the actual concentration of iPTH in the samples.
    • Comparison to a Predicate Device: For method comparison, the results from the legally marketed Abbott Architect Intact PTH assay (K063232) served as the reference or "ground truth" to establish substantial equivalence.
    • Known Reference Standards/Spiked Samples: For precision and interference studies, the ground truth is often established by using clinically relevant pools with known or spiked analyte concentrations.
    • Clinically Established Reference Ranges: Normal physiological ranges for iPTH from healthy donors are used to establish expected values.

    8. Sample Size for the Training Set

    The provided summary does not explicitly mention a "training set" in the context of machine learning or AI models. This device is a traditional immunoassay, which does not typically involve distinct training and test sets in the same way an AI/ML device would. The development (optimization and calibration) of the assay itself would involve numerous samples, but these are part of the assay's development and validation, not a formal "training set" as described in AI. The "testing" samples mentioned above are for performance validation for regulatory submission.

    9. How the Ground Truth for the Training Set was Established

    As noted above, the concept of a formal "training set" with established ground truth is not directly applicable to this traditional immunoassay. The development and optimization of such assays involve:

    • Biochemical assays and calibration curves: These define the relationship between the signal detected (Relative Light Units) and the analyte concentration (iPTH). This is established using known iPTH standards.
    • Cross-validation and optimization: During development, various reagents and parameters are tested and optimized using a wide range of samples to ensure robust and accurate performance across the assay's range.
    • Reference materials and spiked samples: These are used to ensure the assay can accurately quantify iPTH across different concentrations and in the presence of potential interferences.