Search Results
Found 412 results
510(k) Data Aggregation
(213 days)
Ask a specific question about this device
(29 days)
Ask a specific question about this device
(188 days)
The SleepRes PAP system is indicated for the treatment of obstructive sleep apnea (OSA) in patients weighing more than 66 lbs. (30 kg.). The SleepRes PAP system is intended for home, hospital / institutional and sleep center settings.
The SleepRes PAP system consists of a firmware-controlled console, power cable and power adapter, air filter, water tank, and heated breathing tube. The PAP console is composed of a control system, motor system, and heating system. The control system includes the main board, buttons, screen, and sensor. The motor system includes the motor, fan blade, motor outer shell, and silence box. The heating system includes a heating plate and temperature sensor. The PAP console of the subject device does not include foam.
The SleepRes PAP system also includes the following components of PAP therapy systems:
- Water tank
- Air filter
- Breathing Tube
The SleepRes PAP system offers several modes:
- CPAP
- Provides continuous positive airway pressure at the prescribed pressure
- APAP
- Functions to titrate the pressure up (and down as well) to balance reducing or eliminating sleep events while using the lowest effective pressure. Pressure limits are pre-set by the prescribing clinician.
- Kairos Positive Airway Pressure (KPAP)
- This algorithm reduces airway pressure during both inspiration and during most of expiration, only returning pressure to the prescribed treatment level towards the end of expiration.
Connectivity to a Cloud platform via Wi-Fi connection.
The provided FDA 510(k) clearance letter and summary for the SleepRes PAP System (K251770) offer some details about acceptance criteria and the study that proves the device meets these criteria. However, it's important to note that a 510(k) summary is a high-level overview and does not typically contain all the granular details of a full study report.
Here's an attempt to extract and present the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The 510(k) summary broadly asserts that the device met performance and safety requirements. Specific, quantifiable acceptance criteria with corresponding performance outcomes are not explicitly listed in a detailed format within the provided document, beyond the comparison table with the predicate.
| Acceptance Criteria (Inferred from Predicate Comparison & Testing) | Reported Device Performance (SleepRes PAP System) |
|---|---|
| Pressure Range | 4 to 20 cmH2O |
| Pressure Accuracy (Static) | +/- 0.5 cmH2O |
| Pressure Accuracy (Dynamic) | +/- 1.0 -1.5 cmH2O from 10-20 BPM |
| Flow Range | 10-60 lpm |
| Flow Accuracy | +/- 2 lpm or 10% of reading |
| Operating Temperature Range | 5℃ to 32℃ (41˚F to 95˚F) |
| Storage Temperature Range | -20℃ to 55℃ (-13˚F to 158˚F) |
| Atmospheric Pressure Range | 700 to 1060 hPa (0-3000m / 0-9880ft) |
| Noise / Sound | <35 dB(A) (Note: Higher than predicate's 27 dB(A), addressed by labeling mitigation) |
| Therapy Algorithm Effectiveness (Comparative Sleep Studies) | Substantially equivalent to predicate in disordered breathing events (AHI3a), sleep quality, and sleep architecture. |
| Biocompatibility | Met ISO 10993 and ISO 18562 series requirements, including post-accelerated aging tests. |
| Electrical Safety and EMC | Complied with IEC 60601-1 and related collateral standards. |
| Software Verification and Validation | Conducted as per FDA guidances. |
| Human Factors Validation | Conducted as per FDA guidances. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 50 OSA patients.
- Data Provenance: The document states "Each patient underwent a full night polysomnogram." It does not specify the country of origin of the data. The study involved patients "all of whom were on standard APAP in the home," implying they were existing patients.
- Retrospective or Prospective: Implied to be prospective for the clinical validation part, as it describes a protocol where patients "underwent a full night polysomnogram" with different treatment conditions (CPAP, APAP, and APAP + KPAP) specifically for the study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The document does not specify the number or qualifications of experts used to establish the ground truth for the test set (polysomnogram scoring). It only mentions "PSG sleep studies."
4. Adjudication Method for the Test Set
The document does not specify the adjudication method for the test set (PSG scoring).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- Was an MRMC study done? No, an MRMC study was not explicitly described. The clinical study focused on patient outcomes comparing devices/algorithms, not on human readers' performance with or without AI assistance.
- Effect size of human readers improvement: Not applicable, as this was not an MRMC study assessing human reader performance. The study compared the device's therapy algorithms directly through patient outcomes.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
While the clinical study evaluated the device's algorithms (APAP and KPAP) directly through patient polysomnography, it's not described as a "standalone algorithm-only" performance evaluation in the context of interpretation often seen in AI/ML applications. Rather, it evaluated the therapy delivered by the device to patients. The outcomes (AHI3a, sleep quality, sleep architecture) are direct measures of the device's therapeutic effect.
7. The Type of Ground Truth Used
- Type of Ground Truth: Polysomnography (PSG) results, specifically metrics like AHI3a (Apnea-Hypopnea Index with 3% desaturation or arousal criterion), sleep quality, and measures of sleep architecture. These are clinical outcomes obtained from detailed sleep studies.
8. The Sample Size for the Training Set
The document does not specify the sample size for the training set. It only describes the clinical validation study's test set.
9. How the Ground Truth for the Training Set Was Established
Since no training set details are provided, the method for establishing its ground truth is not mentioned.
Ask a specific question about this device
(189 days)
Personalized Therapy Comfort Settings ("PTCS") is an optional software accessory indicated for use by healthcare professionals and patients, providing recommended non-prescription comfort settings options to support patients when using their compatible Resmed PAP therapy device. The outputs provided by PTCS are optional recommendations and are not required to use the therapy device. It is intended for use in a home or clinical environment.
The subject device, Personalized Therapy Comfort Settings (PTCS), is an AI-enabled software accessory intended for use with compatible ResMed positive airway pressure (PAP) therapy devices. PTCS provides optional, non-prescription comfort setting (e.g., ramp, EPR and AutoSet Response) recommendations within the cleared configuration limits of the connected device. It is designed to support patients in personalizing comfort settings during onboarding and early therapy adaptation.
PTCS functions as a stand-alone, machine learning (ML)-based algorithm that generates individualized comfort setting recommendations based on patient-specific input data. The software outputs a limited set of predefined configuration options that can be reviewed by the patient or healthcare provider. PTCS does not automatically adjust device settings, modify prescribed therapy parameters, create new settings, or alter configuration limits. The recommendations are optional and may be accepted, disregarded, or further adjusted according to personal preference or clinical guidance. PTCS is integrated into compatible ResMed products through an application programming interface (API) and has no direct user interface or end-user access.
The provided FDA 510(k) clearance letter and summary for the Personalized Therapy Comfort Settings (PTCS) device contains only limited information regarding specific acceptance criteria and the detailed study that proves the device meets these criteria. Much of the information requested in your prompt is not explicitly stated in the provided text.
Based on the available text, here's what can be extracted and what remains unknown:
Acceptance Criteria and Reported Device Performance
The document does not explicitly provide a table of acceptance criteria with specific numerical targets. It broadly states that "PTCS met all acceptance criteria and passed testing requirements," and that "Analytical validation demonstrated robust sensitivity, specificity, and generalizability." It also mentions "measurable improvements in patient therapy engagement, including increased nightly usage and additional usage days, while maintaining comparable residual AHI and mask leak."
Without specific numerical acceptance criteria, the table below provides what can be inferred.
| Acceptance Criteria (Inferred from text) | Reported Device Performance |
|---|---|
| Reliable, safe, and effective AI algorithm performance for intended use. | "PTCS met all acceptance criteria and passed testing requirements." "Analytical validation demonstrated robust sensitivity, specificity, and generalizability, thereby establishing safety and effectiveness without prospective clinical study." |
| Consistent and reliable model performance across large, representative datasets. | "Analytical validation confirmed consistent and reliable model performance across large, representative datasets, with consistent treatment-effect results and no identified safety concerns." |
| Measurable improvements in patient therapy engagement. | "Retrospective comparisons of PTCS recommended settings to default configurations showed measurable improvements in patient therapy engagement, including increased nightly usage and additional usage days." |
| Comparable residual AHI (Apnea-Hypopnea Index). | "maintaining comparable residual AHI" |
| Comparable mask leak. | "maintaining comparable residual AHI and mask leak." |
| User-comprehensible decision support without impacting therapy delivery or patient safety. | "Usability evaluations with both clinical and non-clinical users confirmed clear comprehension of PTCS recommendations, appropriate application of judgment, and no critical tasks or use-related risks." |
Study Details
1. Sample sized used for the test set and the data provenance:
* Sample Size: Not specified. The document only mentions "large, representative datasets."
* Data Provenance: "Retrospective, de-identified clinical data representative of the intended population." The country of origin is not specified.
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
* Not specified. The document does not detail how ground truth for the test set was established.
3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
* Not specified.
4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
* No MRMC comparative effectiveness study involving human readers is mentioned. The study referenced "Usability evaluations with both clinical and non-clinical users confirmed clear comprehension of PTCS recommendations, appropriate application of judgment, and no critical tasks or use-related risks," but this is not a comparative effectiveness study measuring human reader improvement with AI assistance.
5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
* Yes, a standalone performance evaluation was done. The document states: "PTCS functions as a stand-alone, machine learning (ML)-based algorithm that generates individualized comfort setting recommendations based on patient-specific input data." And "Analytical validation demonstrated robust sensitivity, specificity, and generalizability, thereby establishing safety and effectiveness without prospective clinical study." This refers to the algorithm's performance independent of human input during the core recommendation generation.
6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
* Not explicitly stated for the "ground truth" of the comfort settings. However, for evaluating the impact of the recommendations, "outcomes data" were used, specifically "patient therapy engagement, including increased nightly usage and additional usage days, while maintaining comparable residual AHI and mask leak."
7. The sample size for the training set:
* Not specified. The document mentions "large, representative datasets" for model performance validation but does not differentiate between training and test sets in terms of size.
8. How the ground truth for the training set was established:
* Not specified. The document only mentions that the model is "ML-based" and that "Analytical validation demonstrated robust sensitivity, specificity, and generalizability." It does not provide details on how the training data was curated or how its ground truth was established.
Ask a specific question about this device
(210 days)
A-Model
The F&P Nova Nasal Mask is intended to be used by adults aged 22 years and older weighing ≥ 66 lb (30 ks) who have been prescribed non-invasive positive airway pressure therapy such as CPAP or Bi-Level by a physician. The Nova Nasal mask is intended for single-patient use in the home.
SLA-Model
The F&P Nova Nasal mask is intended to be used by adults aged 22 years and older weighing ≥ 66 lb (30 kg) who have been prescribed non-invasive positive airway pressure therapy such as CPAP or Bi-Level by a physician. The Nova Nasal mask is intended for single-patient use in the home and for multiple patient use in the hospital or other clinical setting where proper disinfection of the device can occur between patient uses.
The F&P Nova Nasal Mask is a non-invasive, Positive Airway Pressure (PAP) therapy nasal mask that features an over-the-nose cushion that seals the patient's nose, held in place by adjustable headgear straps. The mask is designed to aid in the delivery of PAP by providing an interface between the flow generator and tubing, and the patient. The F&P Nova Nasal Mask is a prescription-only device, provided in a non-sterile state.
The Nova Nasal Mask range is available in three cushion sizes – Small, Medium, and Large.
The mask has two models: A-Model and Sleeplab (SL) A-Model. Both models are identical except for their Indications for Use, Operating Environment, and Reusability. The A-Model is intended to be single-patient use in the home, while the SL A-Model is intended to be used on multiple patients in a hospital or other clinical setting where proper disinfection of the device can occur between patient uses.
I am sorry, but based on the provided FDA 510(k) Clearance Letter and the accompanying 510(k) Summary, there is no information provided regarding acceptance criteria for an AI/ML powered medical device, nor any study details that would prove such a device meets these criteria.
The document describes the clearance of a physical medical device, the F&P Nova Nasal Mask, which is a non-invasive positive airway pressure (PAP) therapy nasal mask. The performance data section lists standard engineering and biocompatibility tests typically performed on such devices (e.g., cleaning validation, leak, CO2 rebreathing, resistance to flow, vibration and noise, dead space analysis, human factors, mechanical integrity, shelf-life, and various ISO standards related to masks and biological evaluation).
Therefore, I cannot fulfill your request to describe acceptance criteria and a study for an AI-powered device using this document. The information you are asking for (e.g., sample sizes for test/training sets, data provenance, number/qualifications of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth) are specifically relevant to the validation of AI/ML algorithms in medical devices, which this document does not cover.
Ask a specific question about this device
(86 days)
The myAir app is indicated for patients:
- prescribed with a compatible ResMed Air11 platform device to simulate therapy prior to using their device with their prescribed settings, and to configure their settings to support therapy. It is an optional software accessory to allow patients to acclimate to and operate their therapy device.
- prescribed a NightOwl wearable device to provide the user interface to operate the connected device and aid in the home sleep testing process.
The device is intended for home and hospital use for:
- new and existing patients of ResMed Air10 and Air11 PAP therapy devices and
- new users who are prescribed a compatible NightOwl home sleep test (HST).
myAir is a companion mobile medical application ("app") that serves as a patient self-monitoring, therapy usage tracking, and engagement platform for patients prescribed with compatible ResMed PAP therapy devices and the NightOwl home sleep test (HST) sensors. The app allows the patient to connect via Bluetooth to a compatible hardware device for control of their prescribed PAP or HST device and to allow self-tracking of device usage data. myAir can also be used as a communication pathway using the Bluetooth connection with the compatible device in order to send or receive data. Analysis of patient diagnostic data or display of diagnostic results are not in scope of myAir features.
The myAir subject device for this premarket submission adds device software functions related to starting and stopping therapy, and control of comfort settings (collectively referred to as "device control"). The addition of these device software functions within the myAir subject device is a modification to the K241216 myAir predicate device, in addition to sustaining previously cleared medical device functions.
It appears the provided document is an FDA Clearance Letter for the "myAir" device, specifically a 510(k) summary. This type of document primarily focuses on establishing substantial equivalence to a predicate device, rather than detailing the full scope of acceptance criteria and the comprehensive study results typically found in a clinical study report or a more detailed technical submission.
Based on the provided text, the device "myAir" is a companion mobile medical application. The submission is for a modified device, adding "device software functions related to starting and stopping therapy, and control of comfort settings (collectively referred to as 'device control')".
Therefore, the study detailed is primarily non-clinical verification and validation testing of software functions introduced with these modifications. The document states:
"Non-clinical verification and validation testing completed for device software functions under review introduced with the modifications to the myAir app demonstrated that the device met all intended performance requirements. Testing included:
- Software verification and validation, including non-functional requirements and end-to-end functional testing."
Given this, I cannot provide detailed information about acceptance criteria and study results for clinical performance or human-in-the-loop studies (MRMC, effect size, etc.), as such studies are not described in this 510(k) summary for a software modification. The focus is on the software's functional performance and adherence to guidance documents for software and cybersecurity.
Here's what can be extracted from the provided text according to your request, with a clear indication where information is not available in the document:
Acceptance Criteria and Device Performance for "myAir" (K250624)
The "myAir" device is a mobile application that received FDA 510(k) clearance (K250624) as a modified device, primarily adding "device control functions" (starting/stopping therapy, controlling comfort settings) for compatible ResMed Air11 platform devices. The acceptance criteria and performance detailed below pertain specifically to the non-clinical verification and validation of these new software functions.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Acceptance Criteria (Inferred from document) | Reported Device Performance (As stated/inferred from document) |
|---|---|---|
| Software Functionality | Device control functions (starting/stopping therapy, comfort settings) operate as intended. | "device met all intended performance requirements." "Software verification and validation, including non-functional requirements and end-to-end functional testing." |
| Interoperability | myAir app successfully connects and operates with compatible ResMed Air11 platform devices. | "Device Interoperability and V&V: Identical [to predicate] - ResMed Air11 platform devices and NightOwl HST." "myAir app must be connected to a compatible AS11 or HST device to operate device functions as intended." |
| Compliance with FDA Guidance | Adherence to relevant FDA guidance for medical device software and cybersecurity. | "The following FDA guidance were conformed to: Content of Premarket Submissions for Device Software Functions: June 2023; Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions: September 2023." |
| Safety and Effectiveness | The additions to the device do not raise new questions of safety or effectiveness. | "The differences [from predicate] do not raise any new questions of safety or effectiveness; and It is as safe and effective as the predicate device." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not specified. The document mentions "Software verification and validation, including non-functional requirements and end-to-end functional testing." This typically involves a range of test cases, but the specific number is not disclosed in this summary.
- Data Provenance: The document describes "non-clinical verification and validation testing." This implies internal testing conducted by the manufacturer (ResMed Corp, USA). It does not involve patient data or clinical datasets. Therefore, provenance (country, retrospective/prospective) is not applicable in the typical sense of a clinical study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Not Applicable/Not specified. For non-clinical software verification and validation, "ground truth" is typically defined by functional requirements specifications and design documents. The testing validates that the software performs according to these pre-defined specifications. Expert human readers or adjudicators are not typically involved in establishing ground truth for this type of testing.
4. Adjudication Method for the Test Set
- Not Applicable/None stated. As the testing is non-clinical software verification, adjudication methods like 2+1 or 3+1 (common in image interpretation studies) are not relevant here. The success or failure of a test case is determined by whether the software performs as specified.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No. An MRMC study was not conducted as part of this submission, based on the provided document. The submission pertains to software functionality and its substantial equivalence, not to comparative effectiveness for human readers.
- Effect Size of Human Readers Improvement with AI vs. Without AI Assistance: Not applicable, as no MRMC study or human reader performance evaluation was described. The "myAir" app is an accessory for device control and patient engagement, not an AI diagnostic tool assisting human readers.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Yes, implicitly. The "non-clinical verification and validation testing" is a form of standalone testing for the software's functional performance. The software is tested to operate its intended device control functions (starting/stopping therapy, comfort settings) without direct human intervention in the execution of those specific software commands once initiated. However, this is not "standalone performance" in the sense of a diagnostic algorithm's accuracy, sensitivity, or specificity. It's about the software performing its programmed tasks correctly.
7. The Type of Ground Truth Used
- Functional Requirements and Design Specifications. For this type of software verification, the "ground truth" is established by the detailed specifications of how the software functions are designed to operate. Test cases are then executed to confirm that the actual software behavior matches these expected behaviors. No clinical ground truth (e.g., pathology, outcomes data, expert consensus on patient conditions) is mentioned as being used for the specific software modifications under review.
8. The Sample Size for the Training Set
- Not Applicable/Not disclosed. Since this is software functional verification and validation for a non-AI/ML application (it controls a medical device and provides user interface), there is no "training set" in the sense of data used to train a machine learning model. The software's logic is developed through traditional programming methods, not data training.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. See point 8. No training set for an AI/ML model is indicated, so no ground truth establishment process for a training set is relevant or described.
Ask a specific question about this device
(238 days)
The Mask is intended to provide an interface for Continuous Positive Airway Pressure (CPAP) or bi-level therapy. The Mask is intended for single-patient reuse in the home and multi-patient multi-use in the hospital environment.
The Mask is to be used by adults weighing ≥ 30 kg whom positive airway pressure (CPAP or bi-level system) has been prescribed.
WiZARD 520 Full Face Mask's cushion and frame provide a secure interface between the mask and the patient's face and the flexible cushion improves seal. A series of vents on the elbow, serve as an exhalation vent to purge the exhaled carbon dioxide from the mask. The mask is designed to maintain a continuous airflow to minimize the amount of CO2 rebreathed by the patient. The mask features a wider visibility design that ensures a clear line of sight. It also includes a fabric frame equipped with 3D-shaped headgear for enhanced comfort and stability.
WiZARD 520 Full Face Mask is intended for connection to a positive air pressure source such as CPAP or bi-level for the treatment of Obstructive Sleep Apnea. WiZARD 520 Full Face Mask can be linked to the CPAP or bi-level system's standard 22 mm breathing tube using a swivel hose, flexible tube, and tubing connector. Various sizes (small, medium, and large) are offered to ensure adequate fit over the extended patient population.
The provided text is an FDA 510(k) clearance letter for the WiZARD 520 Full Face Mask. This document describes the substantial equivalence of the new device to a predicate device, focusing on its physical and functional characteristics. It does not describe a study proving the device meets acceptance criteria related to an AI/Machine Learning model, nor does it provide information relevant to a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or the establishment of ground truth by expert consensus.
The clearance is for a physical medical device (a CPAP mask), not a software or AI-driven diagnostic tool. Therefore, the concepts of "acceptance criteria for an AI model," "sample size for test/training sets," "number of experts for ground truth," "adjudication method," "MRMC study," or "standalone algorithm performance" are not applicable to this document.
The document discusses performance tests that are standard for a physical medical device, such as CO2 rebreathing, pressure-flow characteristics, resistance to flow, anti-asphyxia valve operating pressures, and sound levels. These are engineering and physiological performance metrics, not metrics for an AI's diagnostic accuracy or expert agreement.
Therefore, I cannot fulfill your request using the provided input, as the input material does not contain the type of information you've asked for. It's crucial to understand that a 510(k) for a physical medical device like a CPAP mask does not involve AI development or validation processes.
Ask a specific question about this device
(265 days)
AirFit F20 Mask System:
The AirFit F20 mask has two product variants:
- AirFit F20 mask is intended for single-patent reuse in the home environment.
- AirFit F20 SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment.
Both masks are intended for patients weighing more than 66lb (30kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multi-patient reprocessing and must be reprocesssed if reused between patients.
AirFit F20 NM Mask System:
The AirFit F20 Non Magnetic mask has two product variants:
- The AirFit F20 Non Magnetic variant is intended for single-patient reuse in the home environment.
- The AirFit F20 Non Magnetic SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment.
This mask is for patients weighing more than 66 lb (30 kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multi-patient reprocessing and must be reprocessed if reused between patients.
The AirFit F20 and AirFit F20 NM Mask Systems are both full-face masks intended for use with Continuous Positive Airway Pressure (CPAP) and/or Bi-Level devices. CPAP & Bi-Level devices are used for the treatment of Obstructive Sleep Apnea (OSA) and/or ventilatory support. Both treatments deliver pressurized air from a flow generator to the patient. Such delivery of pressurized air relies on the Patient Interface; commonly known as the mask. Full face masks are a common type of Patient Interface (Mask) and are known to efficaciously support the delivery of pressurized air by providing a seal around the patient's nose and mouth.
The AirFit F20 mask system has two product variants:
- AirFit F20 Mask: This is the home use variant that is intended for single patient re-use.
- AirFit F20 SLM (sleep lab mask): This is the variant that is intended for multi-patient re-use and must be reprocessed if reused between patients using the disinfection guide.
AirFit F20 (K170924) had a "For Her" variant which was cleared with aesthetic colour pops, the AirFit F20 Mask and AirFit F20 SLM also have the same "For Her" variant.
The AirFit F20 NM mask system has two product variants:
- AirFit F20 NM Mask: This is the home use variant that is intended for single patient re-use.
- AirFit F20 NM SLM (sleep lab mask): This is the variant that is intended for multi-patient re-use and must be reprocessed if reused between patients using the disinfection guide.
The AirFit F20 and AirFit F20 NM Mask Systems share a common design architecture as the predicate AirFit F20 (K170924), except that the AirFit F20 NM Mask System has non-magnetic clips.
The AirFit F20 and AirFit F20 NM mask systems are made up of the following four main component assemblies: cushion, elbow, frame, and headgear. The cushion and headgear are available in various sizes to allow for adequate mask fit over the intended patient population.
The AirFit F20 and AirFit F20 NM mask systems are prescription devices supplied non-sterile.
The provided FDA 510(k) clearance letter and associated summary pertain to the AirFit F20 Mask Systems, which are non-continuous ventilators (CPAP/Bi-Level masks). This type of device does not typically involve complex diagnostic algorithms or AI components that would necessitate the study structure outlined in your request (e.g., test sets, ground truth, expert adjudication, MRMC studies, standalone performance for AI).
Instead, the clearance is based on demonstrating substantial equivalence to a predicate device (AirFit F20, K170924) through non-clinical performance and safety testing, adherence to recognized standards, and biocompatibility evaluations.
Therefore, many of the specific questions you've asked cannot be directly answered from this document as they are not relevant to the type of device and clearance process involved.
Here's an attempt to extract the relevant information and address your questions where possible:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a "table of acceptance criteria" with corresponding "reported device performance" in a format typically used for diagnostic algorithm validation. Instead, it compares specific design parameters and performance characteristics of the new devices (AirFit F20 Mask System and AirFit F20 NM Mask System) against the predicate device (AirFit F20, K170924). The "Comments" column often indicates that the performance is "Equivalent" or "Identical," implying that the new devices meet or perform similarly to the established safety and performance profile of the predicate, which serves as the de facto acceptance benchmark.
Here's a summary derived from the comparison table (Pages 9-12 of the provided document):
| Design parameter or feature | Predicate device: AirFit F20, K170924 (Performance/Criteria based on predicate) | Modified device: AirFit F20 Mask System (Reported Performance) | Modified device: AirFit F20 NM Mask System (Reported Performance) | Acceptance Criteria (Implied) |
|---|---|---|---|---|
| Indications for Use | Specified for CPAP/bi-level, >66lb (30kg), single-patient home, multi-patient hospital use. | Equivalent to predicate, with variants for home/sleep lab use. | Equivalent to predicate, with variants for home/sleep lab use. | Equivalent to predicate |
| Intended Use | Interface for CPAP/bi-level devices. | Identical | Identical | Identical to predicate |
| FDA Product Code | BZD | BZD | BZD | Identical to predicate |
| Patient population | Weighing more than 66 lb (30 kg) | Patients weighing more than 66lb (30kg) | Patients weighing more than 66lb (30kg) | Identical to predicate |
| Environment of Use | Home or health institution | Home or health institution | Home or health institution | Identical to predicate |
| Reprocessing claims | Single patient re-use or multi-patient re-use. | Identical | Identical | Identical to predicate |
| Sterility state as provided | Non-sterile | Non-sterile | Non-sterile | Identical to predicate |
| Validated reprocessing methods | High-Level Thermal disinfection, Sterilization | High-Level Thermal disinfection, Sterilization | High-Level Thermal disinfection, Sterilization | Identical to predicate |
| Vent type | Multi-hole vent | Multi-hole vent and diffuser vent | Multi-hole vent and diffuser vent | Equivalent to predicate's safety/efficacy profile |
| PAP tubing connection point | ISO 5356-1 (22mm) | ISO 5356-1 (22mm) | ISO 5356-1 (22mm) | Identical to predicate |
| Construction material | Polymeric, Textile, magnets | Polymeric, Textile, magnets | Polymeric, Textile | Equivalent to predicate's safety/efficacy profile (NM variant removes magnets) |
| Operating pressure range (cmH2O) | 3 - 40 | 3 – 40 | 3 – 40 | Identical to predicate |
| Sizes | Cushion (3), Headgear (3), Frame (1) | Identical | Identical | Identical to predicate |
| Mask exhaust flow (Nominal) ISO 17510:2015 Annex B | Pressure (cm H2O): 3, 20, 40 -> Flow (L/min) 'Full Face' curve: 20.1, 55.5, 82.4 | Pressure (cm H2O): 3, 20, 40 -> Flow (L/min) 'Full Face' curve: 19, 54, 82 | Pressure (cm H2O): 3, 20, 40 -> Flow (L/min) 'Full Face' curve: 19, 54, 82 | Equivalent to predicate (minor differences within acceptable range for substantial equivalence) |
| CO2 rebreathing performance (normal condition) ISO 17510:2015 Annex F | Pressure (cm H2O): 3, 5, 10 -> Relative CO2 increase Non-SLM: 14.8%, 7.5%, 6.8% SLM: 11.7%, 7.3%, 2.9% | Pressure (cm H2O): 3, 5, 10 -> Relative CO2 increase Non-SLM: ≤ 11.4%, ≤ 11.4%, - SLM: 11.7%, 7.3%, 2.9% | Pressure (cm H2O): 3, 5, 10 -> Relative CO2 increase Non-SLM: ≤ 11.4%, ≤ 11.4%, - SLM: 11.7%, 7.3%, 2.9% | Equivalent (<20%) (as per ISO standard for substantial equivalence) |
| CO2 rebreathing performance (single fault condition) ISO 17510:2015 Annex F | Fault (ISO 17510:2015) -> Relative CO2 increase Non-SLM: 32.3%, 39.5% SLM: 40.5%, 52.7% | Fault (ISO 17510:2015) -> Relative CO2 increase Non-SLM: ≤ 53.1%, ≤ 53.1% SLM: 40.5%, 52.7% | Fault (ISO 17510:2015) -> Relative CO2 increase Non-SLM: ≤ 53.1%, ≤ 53.1% SLM: 40.5%, 52.7% | Equivalent (<60%) (as per ISO standard for substantial equivalence) |
| Resistance to flow (Pressure drop across mask in cmH2O) ISO 17510:2015 Annex C | @50 L/min: <0.8 cmH2O @100 L/min: <1.6 cmH2O | @50 L/min: <0.8 cmH2O @100 L/min: <1.6 cmH2O | @50 L/min: <0.8 cmH2O @100 L/min: <1.6 cmH2O | Equivalent (Identical performance) |
| Breathing Resistance (cmH2O) ISO 17510:2015 Annex E | Inspiratory: 0.44 @ 50L/min Expiratory: 0.60 @ 50L/min | Inspiratory: 0.6 @ 50L/min Expiratory: 0.80 @ 50L/min | Inspiratory: 0.6 @ 50L/min Expiratory: 0.80 @ 50L/min | Equivalent (within acceptable limits for substantial equivalence) |
| AAV Operating Pressures ISO 17510:2015 Annex D | De-activation: 1.5 cm H2O Activation: 1.0 cm H2O | De-activation: 2.0 cm H2O Activation: 1.5 cm H2O | De-activation: 2.0 cm H2O Activation: 1.5 cm H2O | Equivalent (within acceptable limits for substantial equivalence) |
| Flow generator setting on compatible ResMed CPAP and Bi-level flow generators | "Full face" | "Full face" | "Full face" | Identical to predicate |
| Sound | Sound power level: 30 dBA Sound pressure level: 23 dBA | Multi-hole vent: 30 dBA power, 22 dBA pressure. Diffuser vent: 21 dBA power, 13 dBA pressure. | Multi-hole vent: 30 dBA power, 22 dBA pressure. Diffuser vent: 21 dBA power, 13 dBA pressure. | Equivalent to predicate (Diffuser vent improves sound) |
| Operating and storage temperature | Operating: 5°C to 40°C Storage: -20°C to +60°C | Identical | Identical | Identical to predicate |
| Magnetic Clips | Number: 4, Surface Strength: <400 mT, 50mm distance: <0.5 mT | Number: 4, Surface Strength: <400 mT, 50mm distance: <0.5 mT | Number: 0, Surface Strength: N/A, 50mm distance: N/A | Equivalent (NM variant removes magnets for specific patient needs) |
2. Sample size used for the test set and the data provenance
The document states that "Non-clinical verification and validation testing completed for the AirFit F20 and AirFit F20 NM Mask Systems demonstrated that the mask systems met all intended performance requirements." This primarily refers to bench testing and in-vitro studies.
- Sample Size: Not explicitly stated as a number of "cases" or "patients" in the context of diagnostic performance. For physical device performance, testing involves multiple units of the product to ensure consistency and reliability, but specific sample sizes for tests (e.g., number of masks tested for exhaust flow, CO2 rebreathing, resistance) are not detailed in this summary.
- Data Provenance: This is not a clinical study in the typical sense. The data comes from non-clinical bench testing conducted by the manufacturer (ResMed Pty Ltd, Australia, with a US office). It is not retrospective or prospective patient data from a particular country, but rather laboratory test results.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable. The device is a medical mask system, not a diagnostic algorithm that requires expert interpretation for a "ground truth." The "ground truth" for performance is established by recognized international standards (e.g., ISO 17510:2015 for mask performance, ISO 10993 for biocompatibility) and engineering specifications. Compliance is measured against these standards, not expert consensus on medical images or diagnoses.
4. Adjudication method for the test set
This question is not applicable for the same reasons as #3. Performance is determined by objective measurements against established standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable. This clearance is for a physical medical device (CPAP/Bi-Level mask), not an AI-powered diagnostic tool. There are no "human readers" or "AI assistance" in the context described for this device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. This is not an algorithm or AI device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance is based on:
- Compliance with international standards: For physical performance (e.g., CO2 rebreathing, exhaust flow, resistance, AAV pressures, sound) as defined by ISO 17510:2015 and other relevant ISO standards.
- Biocompatibility testing: According to ISO 10993 series and ISO 18562 series.
- Mechanical integrity testing: To ensure durability after various environmental stressors (home cleaning, transport, storage, operation, free fall, reprocessing).
- Cleaning and reprocessing validation: Against standards like ISO 17664-1, ISO 17664-2, ST98, and AAMI TIR12.
Essentially, the ground truth is objective measurement against established engineering and safety standards.
8. The sample size for the training set
This question is not applicable. There is no "training set" as this is not a machine learning or AI algorithm.
9. How the ground truth for the training set was established
This question is not applicable for the same reasons as #8.
Ask a specific question about this device
(112 days)
Respiration Data Management Software is installed in a computer to perform patient management by treatment data viewing, treatment data reporting and remotely adjusting the prescribed compatible BMC CPAP and BPAP devices' settings. It is intended for healthcare professionals in healthcare facilities use only.
The Respiration Data Management Software is a softwareonly device, which allows physicians and other clinical staff to transmit, analyze and review respiration data from the compatible CPAP and BPAP devices (non-life support systems), thus to assist the users in patient management and follow-up work. This also includes remote modification of the treatment parameters of the compatible devices. It can also store patient data, upload data to the cloud platform, generate and print reports.
The provided text is a 510(k) Summary for the "Respiration Data Management Software." It highlights the device's indications for use and its substantial equivalence to a predicate device, but it does not contain acceptance criteria for device performance or a study demonstrating that the device meets such criteria.
The document focuses on:
- Software Verification and Validation: Stating that it was conducted according to FDA guidance.
- Cybersecurity: Stating that it was considered and evaluated according to FDA guidance.
- Comparison to Predicate Device: Detailing similarities and differences in functionality, compatible devices, and communication mediums, and asserting that differences do not raise new questions of safety or effectiveness.
Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets them because this information is not present in the provided text. The document is a regulatory submission demonstrating substantial equivalence, not a detailed performance study.
Ask a specific question about this device
(84 days)
The Nasal Pillow Mask is intended for patients (> 30 kg) who have been prescribed CPAP/VPAP therapy in home, hospital, or institutional environments. This device is intended to be use under the direction of a physician.
The Nasal Pillow Mask is patient interface device intended for use with CPAP/VPAP therapy systems. The device provides non-invasive means of delivering positive airway pressure to the user by creating a sealed interface using nasal pillows that fit directly into the patients' nostrils. The Nasal Pillow Mask is made of silicone material and comes in 3 different sizes (S, M & L). The nasal mask comes with standard hose fitting intended to be used with standard CPAP/VPAP system. It is intended to be use on patients (> 30 kg) who have been prescribed CPAP/VPAP therapy at home, hospital, or institutional environments. The complete mask set incudes headgear, frame, tube, and nasal pillows (S, M, & L sizes).
This document is a 510(k) summary for a Nasal Pillow Mask, indicating that the device is substantially equivalent to a previously cleared predicate device. As such, it highlights the lack of a new study to demonstrate acceptance criteria because the device is deemed identical to its predicate in all relevant aspects.
Here's the breakdown of why the requested information cannot be fully provided based solely on the input, and what can be extracted:
1. A table of acceptance criteria and the reported device performance
The document explicitly states: "The subject device is identical to predicate device except for manufacturer name, brand name, and resulting labeling updates. The indications for use, design, principles of operation, and materials are all identical to the predicate device already cleared by FDA. Therefore, no new non-clinical testing is required."
This means that for the current 510(k) submission (K243225), no new acceptance criteria or reported device performance data was generated or provided for the subject device. The acceptance criteria and performance data would have been established and demonstrated for the predicate device (K120920) in its original submission.
The table in the document provides specifications for Deadspace Volume, Resistance, Therapy pressure, Noise level, Environmental Conditions, Weight, and Size Flow, which are common technical specifications for such devices. These are not explicitly stated as "acceptance criteria" but rather as "Specifications" that are "Same" as the predicate.
| Specification Item | Subject Device Performance (Reported as "Same" as Predicate) |
|---|---|
| Deadspace Volume | S size: 90.8M size: 94.3L size: 95 |
| Resistance 50 L/m | (Value not numerically given, stated as "Same") |
| Resistance 100 L/m | S size: 6.87M size: 5.93L size: 5.3 |
| Therapy pressure | 5~20cmH2O (Reported as 10cmH2O) |
| Noise level | S size: 35.0 |
| Environmental Conditions | Storage and Transportation: 15°C ~ 25°C, up to 95% non-condensing Storage relative humidity: <85% |
| Weight | 69.9g |
| Size Flow (l/min) @ 5 cmH2O | S: 24.94, M: 24.85, L: 25.28 |
| Size Flow (l/min) @ 10 cmH2O | S: 35.66, M: 35.61, L: 35.79 |
| Size Flow (l/min) @ 15 cmH2O | S: 44.76, M: 45.07, L: 44.71 |
| Size Flow (l/min) @ 20 cmH2O | S: 52.46, M: 52.21, L: 51.99 |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- No new sample size or test set was used for the subject device as "no new non-clinical testing is required." Testing relevant to these specifications would have been performed for the predicate device.
- Data provenance: Not applicable for the subject device's 510(k) as no new studies were conducted. The origin of the predicate device's data is not provided in this document.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. No expert ground truth establishment was conducted for the subject device's 510(k), as it leveraged the equivalence to a predicate device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. No new test set or expert adjudication was performed for the subject device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This device is a physical medical device (nasal pillow mask), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study or assessment of AI assistance is irrelevant and was not performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No. This device is a physical medical device and does not involve an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable for the subject device's 510(k). The substantial equivalence determination is based on the device being identical to the predicate in design, materials, and intended use, not on new clinical performance data or ground truth establishment for the subject device. The specifications listed are objective physical measurements.
8. The sample size for the training set
- Not applicable. This is a physical medical device, not an AI/machine learning model that requires a training set.
9. How the ground truth for the training set was established
- Not applicable. As above, no training set was used.
In summary of the provided document:
The 510(k) submission for the Nasal Pillow Mask (K243225) is based on the premise of substantial equivalence to a predicate device (Hsiner Nasal Pillow Mask, K120920). The manufacturer asserts that the subject device is identical to the predicate device in terms of indications for use, design, principles of operation, and materials. Therefore, no new performance studies, acceptance criteria demonstration, or clinical data were generated for the subject device. The FDA's clearance is based on the established safety and effectiveness of the predicate device, which the subject device is claimed to replicate.
Ask a specific question about this device