Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K994164

    Validate with FDA (Live)

    Date Cleared
    2000-04-13

    (126 days)

    Product Code
    Regulation Number
    862.3870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OnTrak TesTstik 2 for Cocaine/THC is an in vitro diagnostic test intended for professional use for the qualitative detection of cocaine metabolite and cannabinoids in urine. The OnTrak TesTstik 2 for Cocaine/THC cutoff levels (300 ng/ml for cocaine and 50 ng/ml for THC) are based on the Federal Mandatory Guidelines. OnTrak TesTstik 2 for Cocaine/THC is not intended for over-the-counter sale.

    OnTrak TesTstik provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmation method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

    Device Description

    The OnTrak TesTstik 2 Cocaine/THC assay described in this submission is an in vitro test intended for professional use in the qualitative detection of cocaine metabolite and cannabinoids in urine at or above a cutoff concentration of 300 ng/ml (cocaine) and 50 ng/ml (THC).

    The TesTstik assays are based on the principle of microparticle capture inhibition. The test relies on the competition between drug, which may be present in the urine being tested, and drug conjugate immobilized on a membrane.

    When the TesTstik is immersed in the urine sample, some of the sample is absorbed into the TesTstik sample pad. The absorbed sample travels through a reagent strip contained in the device by capillary action. In the reagent strip, the sample rehydrates and mobilizes antibody-coated blue microparticles. The microparticle-urine suspension continues to migrate through the reagent strip and comes in contact with the immobilized drug conjugate. In the absence of drug in the urine, the antibody-coated microparticles bind to the drug conjugates and blue bands are formed at the result window ("negative" sign).

    When drug is present in the specimen, it binds to the antibody-coated microparticles. If sufficient drug is present, the micro-particles are inhibited from binding the drug conjugate and no blue band is formed at the result window. A positive sample caused the membrane to remain white.

    An additional antibody/antigen reaction occurs at the "TEST VALID" area. The "TEST VALID" blue band forms when antibodies, which are imbedded in the reagent membrane, bind to the antigen on the blue microparticles. The presence of the "TEST VALID" band indicates that the test has completed, the reagents in the TEST VALID area are viable, and the results are ready to interpret.

    AI/ML Overview

    This K994164 submission describes a new device, OnTrak TesTstik™ 2 for Cocaine/THC, and claims substantial equivalence to two predicate devices, OnTrak TesTstik for Cocaine and OnTrak TesTstik for THC (both cleared under K973075).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    ItemAcceptance Criteria (New Device)Reported Device Performance (New Device) - CocaineReported Device Performance (New Device) - THC
    MethodologyCompetitive microparticle capture inhibitionCompetitive microparticle capture inhibitionCompetitive microparticle capture inhibition
    MeasurementQualitativeQualitativeQualitative
    Sample TypeUrineUrineUrine
    Endpoint readColorColorColor
    Cutoff300 ng/ml (cocaine) 50 ng/ml (THC)300 ng/ml (cocaine)50 ng/ml (THC)
    Reagent (active ingredients)• Blue dyed microparticles coated with mouse monoclonal anti-benzoylecgonine antibody and anti-cannabinoid antibody. • Drug conjugates immobilized on a membrane. • Mouse monoclonal anti-BSA antibody immobilized on membrane.• Blue dyed microparticles coated with mouse monoclonal anti-benzoylecgonine antibody and anti-cannabinoid antibody. • Drug conjugates immobilized on a membrane. • Mouse monoclonal anti-BSA antibody immobilized on membrane.• Blue dyed microparticles coated with mouse monoclonal anti-benzoylecgonine antibody and anti-cannabinoid antibody. • Drug conjugates immobilized on a membrane. • Mouse monoclonal anti-BSA antibody immobilized on membrane.
    Performance: Precision>95% confidence at 150% cutoff>95% confidence at 150% cutoff (Same as predicate)>95% confidence at 150% cutoff (Same as predicate)
    Performance: Accuracy (Cocaine)Clinical correlation for all positive (including near cutoff samples) and negative specimens demonstrated as acceptable agreement to GC/MS and predicate.94% agreement for all positive (including near cutoff samples) and negative specimens (Cocaine) compared to GC/MS and predicate (automated immunoassay & GC/MS for negatives).N/A (This row is for Cocaine performance)
    Performance: Accuracy (THC)Clinical correlation for all positive (including near cutoff samples) and negative specimens demonstrated as acceptable agreement to GC/MS and predicate.N/A (This row is for THC performance)95% agreement for all positive (including near cutoff samples) and negative specimens (THC) compared to GC/MS and predicate (automated immunoassay & GC/MS for negatives).

    2. Sample Size Used for the Test Set and Data Provenance

    • Cocaine:

      • Positive Specimens: 70 clinical specimens, including near cutoff specimens.
      • Negative Specimens: 100 urine specimens from a clinical laboratory (screened negative by automated immunoassay), with 15% confirmed negative by GC/MS.
      • Data Provenance: Clinical specimens obtained from a clinical laboratory; the country of origin is not specified but implicitly in the USA, where the submitter is located. The data is retrospective as it refers to "clinical specimens screened" and "obtained from a clinical laboratory."
    • THC:

      • Positive Specimens: 70 clinical specimens, including near cutoff specimens.
      • Negative Specimens: 100 urine specimens from a clinical laboratory (screened negative by automated immunoassay), with 15% confirmed negative by GC/MS.
      • Data Provenance: Clinical specimens obtained from a clinical laboratory; the country of origin is not specified but implicitly in the USA. The data is retrospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not specify the number or qualifications of experts directly establishing the ground truth. Instead, it relies on established laboratory methods:

    • For positive specimens: Initially screened by an automated immunoassay, and then confirmed positive by GC/MS. GC/MS (Gas Chromatography/Mass Spectrometry) is a highly accurate and widely accepted "gold standard" for drug confirmation in toxicology.
    • For negative specimens: Screened negative by automated immunoassays, and a portion (15%) confirmed negative by GC/MS.

    Therefore, the ground truth is established by a combination of automated immunoassay screening and confirmatory GC/MS, rather than a human expert panel.

    4. Adjudication Method for the Test Set

    There is no mention of an adjudication method involving human consensus for the test set results. The comparison is made directly against the "ground truth" established by laboratory methods (GC/MS for confirmation).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study was performed or mentioned. This type of study (human readers with/without AI assistance) is not relevant for this in-vitro diagnostic device, which is an automated or semi-automated test reading designed for professional use. The performance evaluation focuses on the device's accuracy against laboratory standards.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone performance study was done. The document details the "OnTrak TesTstik 2 for Cocaine/THC was evaluated using clinical specimens" and directly reports its agreement percentages (94% for Cocaine, 95% for THC) against the ground truth established by GC/MS and predicate comparison. This evaluation is of the device's intrinsic performance without human interpretation or assistance beyond loading the sample.

    7. The Type of Ground Truth Used

    The primary type of ground truth used is GC/MS (Gas Chromatography/Mass Spectrometry), which is considered a definitive confirmatory method in toxicology. This is supplemented by initial screening with an automated immunoassay.

    8. The Sample Size for the Training Set

    The document does not provide information about a "training set" or its size. In a 510(k) submission for a diagnostic test like this, the focus is on the performance of the finalized device (the "test set" in your query context) rather than detailing developmental training data. This implies a traditional assay development workflow rather than a machine learning model that would require explicit training and validation sets.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is described, the method for establishing its ground truth is not applicable/mentioned.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1