Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K993337

    Validate with FDA (Live)

    Date Cleared
    2000-05-16

    (224 days)

    Product Code
    Regulation Number
    892.1170
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The bone density estimates from the Apollo™ bone densitometer can be compared to gender and ethnic matched reference populations as an aid in determining the presence of bone disease other than age-related bone loss. T-Score, % Young Reference, Z-Score, % Age Matched, and long and short term change values are provided. These reference populations include Caucasian, Afro-American, Hispanic, and Asian men and women.

    The Apollo bone density estimates can be used as an aid to physicians in determining fracture risk.

    Device Description

    The Apollo™ Multi-Racial Reference Population capability compares patient values to average values for people without bone related disease, who have the same gender and ethnic background as the patient. Its purpose is to aid the physician in determining the presence of bone disease other than age related bone loss. Both age matched and young reference comparisons are made. T-Score, Z-Score, % Young Reference, % Age Matched, and a graphical representation are presented. This reference population also includes Fracture Risk assessment based on the World Health Organization (WHO) criteria. In general, this means that patients with T- Scores between +1 to -1 are considered to be normal with a low risk of fracture. Patients with T-Scores from -1 to -2.5 have low bone mass and an increased risk of fracture. Likewise, patients with T- Scores below -2.5 are considered to be osteoporotic and have a high risk of fracture. The Apollo™ scans the heel using the industry standard DXA pencil beam technology to assess bone density of the Os Calcis. A water bath is NOT used; and gels are not necessary. The dose to the patient is 0.2 mRem and the scatter/leakage radiation is <0.1 mRem/hour at 3 feet.

    AI/ML Overview

    This 510(k) summary for the Apollo™ Multi-Racial Reference Population does not contain the detailed study information typically found in a clinical trial report. The document primarily focuses on establishing substantial equivalence to previously cleared devices rather than presenting a standalone performance study with specific acceptance criteria and results.

    However, based on the information provided, we can infer some details and highlight what is not present:

    1. Table of Acceptance Criteria and Reported Device Performance

    This information is not explicitly stated in the provided 510(k) summary. The document focuses on establishing comparability to existing reference populations rather than setting novel performance thresholds for accuracy or diagnostic efficacy. The "Safety and Effectiveness" section merely states: "This Apollo Multi-Racial Reference Population is comparable to reference population capabilities in use with other bone densitometers in the industry. No new safety or effectiveness issues are raised with this capability."

    Therefore, a table with specific acceptance criteria (e.g., accuracy, precision) and corresponding reported performance metrics cannot be constructed from this document.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the 510(k) summary. The document describes the purpose of the reference population (comparing patient values to average values for people without bone-related disease, of the same gender and ethnic background) but does not detail the specific population used to create or test this "Multi-Racial Reference Population." It only states that the reference populations include "Caucasian, Afro-American, Hispanic, and Asian men and women." No sample sizes, geographical origin, or whether the data was retrospective or prospective is mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the 510(k) summary. Given that this is a reference population for a bone densitometer, the "ground truth" would likely relate to the bone density measurements themselves and the health status of the individuals in the reference population. However, the exact methodology and expert involvement in establishing this "ground truth" for the reference data are not described.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not provided in the 510(k) summary. An adjudication method would typically be relevant for studies involving human interpretation or subjective assessments. For a reference population based on objective bone density measurements, this might not be directly applicable in the same way it would be for a diagnostic image analysis algorithm. However, any process for selecting or classifying the individuals within the reference population for inclusion or exclusion would involve some form of consensus or criteria, which is not detailed here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable and therefore not provided in this 510(k) summary. The Apollo™ Multi-Racial Reference Population is a reference database used by a bone densitometer (DXA pencil beam technology) to interpret bone density values (T-scores, Z-scores). It is not an AI-powered diagnostic algorithm that assists human readers in interpreting images or making a diagnosis in a comparative effectiveness study. Therefore, an MRMC study and effect size related to AI assistance would not be part of this submission. The device is a database, not an AI model.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable in the typical sense of an algorithm's standalone performance. The "device" here is a reference population database. Its "performance" is implicitly tied to the accuracy and representativeness of the data it contains, which then allows the interpretation of bone density measurements from the Apollo™ densitometer. The 510(k) focuses on the comparability of this reference population to others in the industry, implying its utility is recognized rather than presenting a novel standalone performance study for the database itself.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The document mentions that the reference population compares patient values to "average values for people without bone related disease." This implies that the "ground truth" for the individuals included in the reference population is their lack of bone-related disease, and their actual bone density measurements (from DXA). The method for determining "without bone related disease" is not specified but would likely involve clinical assessments, medical history, and potentially other diagnostic tests, all contributing to a clinical "truth" of health status.

    8. The Sample Size for the Training Set

    This information is not provided in the 510(k) summary. The document describes a "reference population," which implies a dataset used to establish normal ranges. This would be analogous to a "training" or "development" dataset if it were an AI model, but for a reference population, it's the fundamental data itself. The size of this population for each gender and ethnic group (Caucasian, Afro-American, Hispanic, and Asian men and women) is not disclosed.

    9. How the Ground Truth for the Training Set Was Established

    The document states the reference population includes "average values for people without bone related disease, who have the same gender and ethnic background as the patient." Therefore, the implicitly established "ground truth" for the individuals whose data forms this reference population is that they were identified as healthy (i.e., "without bone related disease") and their demographics (gender, ethnicity) were recorded. The specific diagnostic criteria or medical evaluations used to confirm the absence of bone-related disease for inclusion in this reference data are not detailed in this summary.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1