510(k) Data Aggregation

    K Number
    K992297

    Validate with FDA (Live)

    Date Cleared
    1999-09-03

    (58 days)

    Product Code
    Regulation Number
    866.1620
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pasco MIC and MIC/ID panels are used for quantitatively measuring (with the exception of the Breakpoint/ID panel which provides qualitative measurement of category results) the susceptibility of rapidly growing aerobic and facultative anaerobic bacterial pathogens to a battery of antimicrobial agents and determining the biochemical identification of those organisms.

    This 510(k) notification is for the addition of Levotloxacin to Pasco panels at concentrations of 0.25-8 mcg/ml for use in determining the susceptibility of S. pneumoniae and non-pneumococcal streptococci.

    Device Description

    Varying concentrations of antimicrobial agents (usually in two-fold dilutions) are dispensed into the Pasco panels and the panels are then frozen. Panels are thawed prior to use, inoculated with the test organisms, incubated the traditional 16-24 hours and panels are then observed for visible growth or color changes as described in the package insert. The lowest concentration of each antimicrobial agent with no apparent visible growth of the test organism is recorded as the minimum inhibitory concentration (MIC). Changes in pH and production of specific metabolites from growth in biochemical substrates are interpreted as described in the package insert for conventional tubed media.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Pasco MIC and MIC/ID Panels, specifically for the inclusion of Levofloxacin:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document defines "acceptance criteria" through the lens of what is considered "acceptable" for Essential Agreement (EA), Category Agreement (CA), and error rates (Major, Very Major, Minor). The reported device performance demonstrates that these criteria were met or exceeded.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance (S. pneumoniae)Reported Device Performance (Non-pneumococcal streptococci)
    Essential Agreement (EA)Acceptable agreement (not explicitly quantified but demonstrated by 100%)100% EA on initial testing100% EA on initial testing
    Major Errors (M)No Major Errors permittedNo Major Errors observedNo Major Errors observed
    Very Major Errors (VM)No Very Major Errors permittedNo Very Major Errors observedNo Very Major Errors observed
    Minor ErrorsAcceptable (e.g., small number, within EA)No Minor Errors observed1 random Minor Error noted (within EA)
    Category Agreement (CA)Acceptable agreement (not explicitly quantified but demonstrated by 100%)100% CA100% CA
    Reproducibility100% within +/- 1 dilution100% within +/- 1 dilution (11/12 organisms on-scale)100% within +/- 1 dilution (11/12 organisms on-scale)
    QC EndpointsWithin recommended NCCLS acceptable rangeWithin recommended NCCLS acceptable range (S. pneumoniae ATCC 49619)Within recommended NCCLS acceptable range (S. pneumoniae ATCC 49619)

    2. Sample Size Used for the Test Set and Data Provenance

    • S. pneumoniae strains: 101
    • Non-pneumococcal streptococci strains: 130
    • QC organism (S. pneumoniae ATCC 49619): Tested throughout.
    • Reproducibility organisms: 12 organisms at each of two sites (totaling 24 instances of reproducibility testing).
    • Data Provenance: The study was "comparative testing...performed at two sites using CDC challenge strains and clinical isolates." This suggests a mix of standardized and real-world samples, likely prospective for the specific testing conducted for this submission. The country of origin is not explicitly stated but implied to be the US given the FDA submission context.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts used to establish the ground truth. Ground truth for antimicrobial susceptibility testing typically involves comparison to a recognized gold standard method (often broth microdilution or agar dilution) performed by trained microbiologists. The reference to "reference panel" and "CDC challenge strains" implies that the ground truth was established by highly controlled and standardized methods, likely overseen by qualified laboratory personnel.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method (such as 2+1, 3+1, none). Given the nature of antimicrobial susceptibility testing, which typically yields a definitive MIC result, a formal adjudication process between multiple readers is often not explicitly needed for individual readings but rather for resolving discrepancies between the test device and the reference method if they occur. The "reference panel" serves as the comparative gold standard.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is more common for diagnostic imaging where human interpretation varies significantly. For antimicrobial susceptibility testing, the focus is on the device's ability to accurately determine the MIC compared to a highly standardized reference method. The "human-in-the-loop" aspect is primarily the reading of the results from both the test and reference panels, but the study is not designed to measure an improvement in human reading performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device described is an "Antimicrobial Susceptibility Test Microbiology Panel," which involves physical panels inoculated with organisms and then observed for growth or color changes. While the interpretation of the lowest concentration of antimicrobial agent with no apparent growth is a critical step, the "device" itself is the physical panel and its methodology. This is not a standalone algorithm in the modern sense (e.g., an AI software reading an image). The "reading" is done by a laboratory technician based on visible changes. Therefore, the concept of "standalone algorithm performance" as separate from "human-in-the-loop" is not directly applicable in the way it is for AI/imaging devices. The device's performance is intrinsically linked to the observation and interpretation, which is typically a human task.

    7. The Type of Ground Truth Used

    The ground truth was established by a "reference panel". This typically refers to a standardized, well-accepted method for determining antimicrobial susceptibility (e.g., broth microdilution or agar dilution) performed in parallel with the test device. This "reference panel" constitutes a gold standard for comparison. Additionally, the study used "CDC challenge strains" which are well-characterized, standardized bacterial isolates with known susceptibility profiles, contributing to the robustness of the ground truth.

    8. The Sample Size for the Training Set

    The document does not specify a training set sample size. This is typical for a device based on established chemical and biological reactions, rather than a machine learning or AI-driven device. The "training" in this context refers to the development and optimization of the panel manufacturing and assay conditions, which would involve internal R&D and quality control, not a distinct "training set" in the AI sense.

    9. How the Ground Truth for the Training Set Was Established

    As there is no explicitly defined "training set" in the AI sense, the establishment of ground truth for such a set is not applicable. The development of the Pasco panels and their performance characteristics would have been guided by traditional microbiological methods and NCCLS (now CLSI) guidelines for susceptibility testing.