    K Number
    K991320

    Validate with FDA (Live)

    Date Cleared
    1999-11-10

    (205 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MedComp® Duo-Coat Catheter is indicated for use in attaining short term vascular access for hemodialysis and apheresis. It may be inserted percutaneously and is ideally placed in the internal jugular vein of an adult patient. Alternate insertion sites are the subclavian or femoral vein.

    Device Description

    The MedComp® DUO-COAT Catheter Double Lumen combines the MEDCOMP Duo-Flow Catheter with a dual purpose surface coating. Bound heparin reduces thrombus formation on the catheter outer surface and at arterial and venous ports, in 4-hour invitro-studies'. The round catheter design and lubricious surface may aid in catheter insertion by reducing friction on the outer surface. The fixed circular tube within a tube allows for a 360 arterial uptake and provides consistent flows from 350 to 400 ml/min. The catheter lumen is fabricated from a soft, thermo-sensitive, polyurethane material that is rigid upon insertion, but which becomes soft upon reaching body temperature to reduce vessel trauma. The external lumen surface incorporates heparin to reduce the risk of thrombus. The catheter hub is molded from soft pliable polyurethane to increase patient comfort. The soft radiopaque tip reduces vessel trauma and aids in confirmation of tip placement. Extensions are color coded with a red luer and clamp for the arterial lumen and a blue luer and clamp for the venous lumen for easy identification.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the MedComp® DUO-COAT Catheter Double Lumen, based on the provided text:

    Important Note: This submission is a 510(k) Pre-market Notification, which primarily focuses on demonstrating substantial equivalence to a predicate device. Therefore, the "acceptance criteria" and "study" described are often comparisons to the predicate and performance tests to ensure safety and effectiveness, rather than extensive clinical trials with human subjects as might be seen for novel devices.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Performance ClaimReported Device Performance
    Predicate EquivalenceSubstantially equivalent to Duo-Flow Double Lumen Catheter (K974236).
    Thrombus Formation Reduction (Outer Surface & Ports)Bound heparin reduces thrombus formation on the catheter outer surface and at arterial and venous ports, in 4-hour in-vitro-studies. Hemodynamic testing under strenuous simulation testing demonstrated that the surface modification reduced the risk of thrombosis. (Results from comparative evaluation against uncoated control catheters after simulated dialysis testing with heparinized bovine blood.)
    Insertion Aid / Reduced FrictionThe round catheter design and lubricious surface may aid in catheter insertion by reducing friction on the outer surface. Comparison of treated and untreated surfaces showed a marked reduction in drag on the surface after application of the lubricious heparin.
    Consistent Flow Rates (Arterial Uptake)The fixed circular tube within a tube allows for a 360° arterial uptake and provides consistent flows from 350 to 400 ml/min.
    Material Properties (Thermo-sensitive)Fabricated from a soft, thermo-sensitive, polyurethane material that is rigid upon insertion, but which becomes soft upon reaching body temperature to reduce vessel trauma.
    BiocompatibilityTesting conducted on the uncoated catheter, similar to that conducted on the predicate device, confirmed that the presence of the lubricious heparin did not affect the safety of the catheter.
    Mechanical IntegrityTesting conducted on the uncoated catheter, similar to that conducted on the predicate device, confirmed that the presence of the lubricious heparin did not affect the safety of the catheter.
    Patient ComfortThe catheter hub is molded from soft pliable polyurethane to increase patient comfort. (This is a design feature, not directly a tested performance metric in the provided text).
    Tip Placement Confirmation / Vessel TraumaThe soft radiopaque tip reduces vessel trauma and aids in confirmation of tip placement. (Design features).
    Safety (General)Confirmed that the presence of the lubricious heparin did not affect the safety of the catheter (via biocompatibility screens and mechanical integrity tests).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated as a number of devices or subjects in the provided text. The testing mentioned refers to "comparative evaluation against uncoated control catheters" and "hemodynamic testing under strenuous simulation testing." It implies in vitro testing, not human trials.
    • Data Provenance:
      • Country of Origin: Not specified, but generally assumed to be within the US or compliant with US regulatory standards for a US FDA submission.
      • Retrospective or Prospective: The reported tests ("4-hour in-vitro-studies", "hemodynamic testing under strenuous simulation testing," "comparative evaluation") are presented as prospective in vitro bench tests conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the text. The evaluation method described is primarily in vitro laboratory testing and comparison to an uncoated control, not an expert-driven ground truth assessment for a diagnostic or imaging device.


    4. Adjudication Method for the Test Set

    None. The testing described is in vitro laboratory performance evaluation (e.g., thrombus formation, drag reduction, consistent flow rates, biocompatibility, mechanical integrity) against a control, not a subjective assessment requiring adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is typically performed for diagnostic imaging or AI devices where human readers interpret cases. This device is a medical catheter, and its evaluation focuses on physical and biological performance characteristics, not diagnostic interpretation. Therefore, there is no mention of human readers or effect size of AI assistance.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not Applicable. This is a physical medical device (catheter), not an AI algorithm. Its performance is inherent to its design and materials, not learned from data.


    7. The Type of Ground Truth Used

    The "ground truth" for the performance claims appears to be established through:

    • Quantitative In Vitro Measurements: Direct measurement of parameters like flow rates (350-400 ml/min), thrombus formation (reduction in 4-hour studies), and drag reduction (marked reduction).
    • Comparative Assessment: Performance compared against "uncoated control catheters" for thrombus formation and drag.
    • Predicate Device Performance: Substantial equivalence claims often benchmark against the known safety and effectiveness of the predicate.
    • Biocompatibility Standard Tests: Implied adherence to established biocompatibility screens.
    • Mechanical Integrity Tests: Implied adherence to established tests for device durability and function.

    8. The Sample Size for the Training Set

    Not Applicable. This device is a physical medical device, not an AI/ML algorithm that requires a training set.


    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. As this is not an AI/ML device, there is no training set and therefore no ground truth established for it.