    Intended Use

    The ABBOTT Advisor One-Step Pregnancy Test is a self-performing immunoassay designed for the qualitative determination of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy.

    The ABBOTT Advisor One-Step Pregnancy Test is an over-the counter in vitro test which can be used by the consumer to detect pregnancy as early as the first day of her missed period, using a direct urine stream sampling method. Some positive results appear as soon as three minutes, however, all results are confirmed when the Test Timer turns pink/red. The Test Timer will turn pink/red approximately 5 minutes after the urine has been added to the test device.

    Device Description

    The ABBOTT Advisor One-Step Pregnancy Test is an elongated device composed of two pieces of molded plastic which contain the internal test strip. The three openings on the device are the Urine Well, the Result Window and the Test Timer. The Result Window and the Test Timer are protected with a clear seal to prevent potential contamination caused by splashing urine.

    The ABBOTT self-performing immunoassay for hCG in urine is based on the principle of a two site immunometric assay. Urine is added to the Urine Well, where it migrates through the Absorbent Pad to the Conjugate Pad portion of the test strip. As the urine moves through the Coniugate Pad, it mobilizes the mouse monoclonal anti-alpha hCG antibody coated colloidal particles. Any hCG present in the urine sample combines with the anti-alpha hCG antibody on the colloidal particle.

    The urine, now containing an hCG-antibody-colloid complex and unbound colloid, continues to move up the test strip to the Result Window which contains immobilized polyclonal anti-BhCG on a vertical (Patient) bar and anti-mouse/anti-βhCG on a horizontal (Control) bar. The color generated is proportional to the concentration of hCG in the urine sample. Anti-hCG (mouse monoclonal) coated colloidal particles bind to the anti-mouse antibody immobilized on the Control Bar which indicates the assay is performing properly. A positive sign (+) indicates the presence of hCG. Low hCG concentration samples may display a "T" result, which is also interpreted as positive. The absence of detectable hCG will result in a negative sign (-). The assay is complete when the "Test Timer" indicator changes to pink or red, approximately three minutes after the start of the assay.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from Predicate Device)Reported Device Performance (ABBOTT Advisor One-Step Pregnancy Test)
    Consumer Use Accuracy: 98%98% (146 out of 149 subjects correctly interpreted results)
    Laboratory Use Accuracy: 100%100% (50/50 hCG-positive and 50/50 hCG-negative specimens correctly identified)
    Time to Assay Result: 3 minutes (some positives seen in 1 minute)5 minutes (some positives seen in 3 minutes)
    End of Assay Color Change: Yes (Do not wait for signal before reading results)Yes (Wait to read results until Test Timer changes color)

    Note: The acceptance criteria for "Time to Assay Result" and "End of Assay Color Change" are based on the predicate device's characteristics, as the Abbott device is aiming for substantial equivalence. The Abbott device has a modified time-to-result requirement compared to the predicate.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Laboratory Accuracy Test Set:

      • Sample Size: 100 specimens (50 hCG-positive, 50 hCG-negative).
      • Data Provenance: Clinical urine specimens. The document doesn't specify the country of origin but states the testing was performed by "Advanced Care Products." Given the context of a US 510(k) submission, it's highly likely to be US-based. Retrospective.
    • Consumer Testing Test Set:

      • Sample Size: 149 female volunteers.
      • Data Provenance: Volunteers recruited from a "central New Jersey location." Prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Laboratory Accuracy Test Set: The document does not explicitly state the number or qualifications of experts used to establish the ground truth.

      • Ground Truth Confirmation:
        • hCG-positive samples were confirmed using the "Hybritech ICON pregnancy test."
        • hCG-negative specimens were confirmed using the "FactPLUS® Cup and Dropper Pregnancy Test."
      • It's implied that these reference tests provided the ground truth, likely interpreted by trained laboratory personnel, but no explicit "expert" count or qualification is given.
    • Consumer Testing Test Set: The ground truth for this test was the actual presence or absence of hCG in the volunteers' urine, as determined by laboratory methods (implied through the "correct results" phrasing). The consumer's interpretation was compared against this objective ground truth. No specific "experts" were used to establish the ground truth for the volunteers beyond standardized laboratory tests.

    4. Adjudication Method for the Test Set

    • Laboratory Accuracy Test Set: Not explicitly stated. The confirmation with reference tests (Hybritech ICON and FactPLUS Cup and Dropper) suggests a direct comparison rather than an adjudication process between multiple readers of the Abbott device.
    • Consumer Testing Test Set: Not applicable. The consumers themselves were the "readers," and their interpretation was compared against a laboratory-determined ground truth. There was no adjudication among consumer interpretations reported.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not explicitly described. The studies focused on:

    1. Device performance against ground truth in a lab setting.
    2. Consumer ability to correctly use and interpret the device's results.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Yes, the "Laboratory Accuracy testing" essentially describes standalone performance, as the device was tested with known hCG-positive and hCG-negative samples, and the results were read by trained personnel against objective ground truth. The consumer testing also evaluates the device's standalone performance in the hands of the intended user, even though a human is "interpreting" the result.

    7. The Type of Ground Truth Used

    • Laboratory Accuracy Test Set: Reference standard tests (Hybritech ICON pregnancy test for positive, FactPLUS® Cup and Dropper Pregnancy Test for negative). This is considered a form of "reference test consensus" or "established reference method."
    • Consumer Testing Test Set: The actual presence or absence of hCG in the urine, determined by objective laboratory testing (implied by the comparison to "correct results").

    8. The Sample Size for the Training Set

    The document does not describe a separate training set or machine learning algorithm. This device is an immunoassay, not an AI/ML-driven device. Therefore, the concept of a "training set" in the context of machine learning is not applicable here. The device's "training" or development would have involved optimizing its chemical components and design characteristics.

    9. How the Ground Truth for the Training Set Was Established

    As this is not an AI/ML device, the concept of a "training set" and its ground truth in that context is not applicable. The device's performance is based on its chemical and physical design.