510(k) Data Aggregation

    K Number
    K253822

    Validate with FDA (Live)

    Date Cleared
    2026-03-27

    (119 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GPA (hexafocon A) Rigid Gas Permeable Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 0.75 diopters.

    The GPA-AT (hexafocon A) Rigid Gas Permeable Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 6.00 diopters.

    The GPA-M, GPA-PN and GPA-PD (hexafocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 0.75 diopters.

    The GPA-PDT, GPA-MT, and GPA-PNT (hexafocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 6.00 diopters and are presbyopic requiring add power of up to +4.00 diopters.

    The GPA-K (hexafocon A) Rigid Gas Permeable Contact Lens may be prescribed in otherwise non-diseased eyes that require a rigid gas permeable lens for the management of irregular corneal conditions such as; keratoconus, pellucid marginal degeneration or following penetrating keratoplasty or refractive (e.g. LASIK) surgery.

    Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

    Device Description

    The GPA (hexafocon A) Rigid Gas Permeable Contact Lenses are lathe manufactured from the lens blank material hexafocon A, which is blue-tinted for visibility and may include an ultraviolet (UV) light absorber. The contact lens material name hexafocon A has been adopted by the United States Adopted Names (USAN) Council.

    The GPA (hexafocon A) Rigid Gas Permeable Contact Lenses are a family of rigid gas permeable contact lenses available in the following design configurations:

    ModelDescription
    GPA-ARigid gas permeable contact lens with double aspherical design on both the front and back surfaces.
    GPA-KRigid gas permeable contact lens with double aspherical design on both the front and back surfaces, specifically designed for keratoconus.
    GPA-MProgressive multifocal rigid gas permeable contact lens with double aspherical design on both the front and back surfaces.
    GPA-PNMultifocal rigid gas permeable contact lens with an aspherical back surface and a concentric multi-zone front surface where the center zone is for near vision, specifically designed for presbyopia correction.
    GPA-PDMulti-focal rigid gas permeable contact lens with an aspherical back surface and a concentric multi-zone front surface where the center zone is for distance vision, specifically designed for presbyopia correction.
    GPA-ATRigid gas permeable contact lens with double toric design on both the front and back surfaces.
    GPA-MTProgressive multifocal rigid gas permeable contact lens with a toric back surface and an aspherical front surface.
    GPA-PNTMultifocal rigid gas permeable contact lens with a toric back surface and a concentric multi-zone front surface where the center zone is for near vision, specifically designed for correcting presbyopia in patients with large corneal astigmatism.
    GPA-PDTMulti-focal rigid gas permeable contact lens with an aspherical back surface and a concentric multi-zone front surface where the center zone is for distance vision, specifically designed for correcting presbyopia in patients with large corneal astigmatism.

    The GPA (hexafocon A) Rigid Gas Permeable Contact Lenses are in the following lens parameters:

    ParameterAbbr.Range (mm)Step (mm)Tolerance
    GPA-AGPA-KGPA-MGPA-PNGPA-PD
    BC radius /mmBC6.50~9.504.50~7.506.50~9.506.50~9.506.50~9.500.05± 0.05 mm
    Total diameter /mmTD8.50~11.507.00~11.008.00~12.008.00~12.008.00~12.000.10± 0.10mm
    Spherical power /DSPH-20.00~+20.00-20.00~+20.00-20.00~0.00-20.00~+6.00-20.00~+6.000.25± 0.12 (0 to </= 5D)± 0.18 (5 to </= 10.0D)± 0.25 (10 to </= 15D)± 0.37 (15 to </= 20D)± 0.50 (over 20D)
    Additional power /DAddN/AN/A-5.00~+5.000~+4.000~+4.000.25± 0.38D
    Center thickness /mmCT0.11~0.600.11~0.600.11~0.600.11~0.600.11~0.60-± 0.02 mm
    Optional back toric design-GPA-ATN/AGPA-MTGPA-PNTGPA-PDT--
    Optional back toric diopters /DTOR0.00~-6.00N/A0.00~-6.000.00~-6.000.00~-6.000.25± 0.25 (0 to </= 2D)± 0.37 (2 to </= 4D)± 0.50 (over 4D)
    Surface Appearance-Lenses should be clear with no surface defect

    The physical properties of the GPA (hexafocon A) Rigid Gas Permeable Contact Lenses are as follows:

    Hexafocon A Rigid Gas Permeable Contact Lens
    Refractive Index: 1.415
    Modulus (MPa): 1269
    Hardness (Shore D): 81
    Specific Gravity: 1.27
    Oxygen Permeability (Dk): 100 x 10⁻¹¹(cm²/sec) (ml O₂/ml x mm Hg @ 35°C)
    Color Additive(s): D&C Green #6

    GPA (hexafocon A) Rigid Gas Permeable Contact Lenses are distributed in a dry state, without immersion in solution. The primary packaging consists of a polycarbonate plastic screw-top vial equipped with a lens clamp to secure the product during transit. All accompanying documentation—including the case, packing slip, and invoice—identifies lens specifications: base curve, dioptric power, cylindrical power, cylindrical axis, additional power, diameter, central thickness, color, and lot number.

    AI/ML Overview

    N/A