The LYNX◊ COBLATION◊ Laryngeal Wand is indicated for ablation, resection, and coagulation of soft tissues and hemostasis of blood vessels for otorhinolaryngology (ENT) procedures of the larynx and trachea.
The LYNX™ COBLATION™ Laryngeal Wand is intended for ablation, resection, and coagulation of soft tissues and hemostasis of blood vessels in otorhinolaryngology (ENT) procedures.
It is intended for procedures using a conductive media, such as normal saline or Ringer's lactate. It is intended to be used with qualified controllers.
The LYNX COBLATION Laryngeal Wand is a single-use, disposable, bipolar, radio frequency, electrosurgical device intended for ablation, resection and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) procedures. The device is intended to work with Qualified Controllers (WEREWOLF COBLATION SYSTEM (WEREWOLF ENT and WEREWOLF +). There are no changes to the WEREWOLF Controllers or Irrigation pump for use with LYNX Wand.
The LYNX Wand consists of three integrated lines built into the handpiece that allow the wand to work as intended:
- An integrated suction line allows connection to suction source within the operating environment.
- An integrated saline tube set designed to be compatible with the WEREWOLF flow control module allows connection with saline source within the operating environment.
- An integrated cable compatible with the WEREWOLF controller.
The device operates in Coblation (resection/ablation) and Coagulation (coagulation/hemostasis) modes by using their biopolar configuration to create energy field using a conductive irrigating solution.