    Intended Use

    The intended use of the Electric Wheelchair (Robooter E80), Model name: BBR-E80-01 and BBR-E80-02 is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.

    Device Description

    The Electric Wheelchair (Robooter E80), Models name: BBR-E80-01,BBR-E80-02, is an indoor/outdoor, battery- operated, 2-wheel drive (rear-wheel drive) powered wheelchair.

    It consists of four modules: seat system, control system, braking system, and drive system. The user sits in the wheelchair seat and uses the control system.

    The control pad positioned on the right armrest, user can turn the wheelchair on, control the speed, and direct the movement.

    The braking system employs an electromagnetic brake, when release the controller rocker, the electromagnetic brakes will be actuated, and the electric wheelchair will stop in several seconds. Electromagnetic brake will not take effect immediately, it will take effect after the wheel rotates for 1/2 cycle.

    The wheelchair is powered by a 24V DC,12Ah or 20Ah rechargeable lithium-ion battery charged by an offboard lithium-ion battery charger. The wheelchair is driven by two DC motors.

    The Electric Wheelchair (Robooter E80), Models name: BBR-E80-01,BBR-E80-02, contains Bluetooth 4.1 BLE technology. The device can be controlled by the controller rocker or remote control by a smartphone app via Bluetooth 4.1 Low Energy (BLE) wireless communication interface. The smartphone app is used to drive the chair remotely. For safety, controller rocker control is priority over the remote control by design. The smartphone app can also view the battery's status, adjust the speed gear level and lock/unlock the unattended device.

    The wheelchair can be folded/expanded manually. The Left and right handrails (joystick) can be interchange.