Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K252665

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-20

    (56 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    brAIn™ Shoulder Positioning is intended to be used as an information tool to assist in the preoperative surgical planning and visualization of a primary total shoulder replacement.

    Device Description

    The brAIn™ Shoulder Positioning software is a cloud-based application intended for shoulder surgeons. The software does not perform surgical planning but provides tools to assist the surgeon with planning primary anatomic and reverse total shoulder replacement surgeries using FX Shoulder Solutions implants. The software is accessible via a web-based interface, where the user is prompted to upload their patient's shoulder CT-scan (DICOM series) accompanied with their information in a dedicated interface. The software automatically segments (using machine learning) and performs measurements on the scapula and humerus anatomy contained in the DICOM series. These segmentations serve as a foundation for the surgeon's manual planning, which is performed using an interactive 3D viewer that allows for soft tissue visualization. The surgeon positions the glenoid and humerus implants manually within this same 3D interface using a dedicated manipulation panel. The changes in shoulder anatomy resultant from the implants are relayed in a post-position interface that displays information related to distalization and lateralization. The software outputs a planning multimodal summary that includes textual information (patient information, pre- and post-op measurements) and visual information (screen captures of the shoulder pre- and post-implantation).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the brAIn™ Shoulder Positioning device, based on the provided FDA 510(k) clearance letter:


    Table of Acceptance Criteria and Reported Device Performance

    Feature/MetricAcceptance CriteriaReported Device Performance
    Segmentation PerformanceMean Dice Similarity Coefficient (DSC) $\geq$ 0.95Met acceptance criteria that the segmentation performance meets the acceptance criteria. The validation criterion was a Dice Similarity Coefficient (DSC) coefficient of 0.95 or higher, demonstrating that the segmentation produced by the model after post-processing closely matches the ground truth.
    Shoulder Side DetectionCorrect shoulder side (right or left) in DICOM imagesAll performance tests for Shoulder Side Detection validation were successfully completed with no deviations, confirming compliance with the required performance standards.
    Measurement Accuracy (Angles)$\leq$ 1° for angle measurementAll performance tests for Measurement Accuracy Validation were successfully completed with no deviations, confirming compliance with the required performance standards.
    Measurement Accuracy (Distances)$\leq$ 1 mm for distance measurementAll performance tests for Measurement Accuracy Validation were successfully completed with no deviations, confirming compliance with the required performance standards.
    Measurement Accuracy (3D Subluxation)$\leq$ 1% for 3D subluxationAll performance tests for Measurement Accuracy Validation were successfully completed with no deviations, confirming compliance with the required performance standards.
    Landmark PerformanceMean distance $\leq$ 3 mm (compared to final positions adjusted by experts)All performance tests for landmark validation were successfully completed with no deviations, confirming compliance with the required performance standards; achieving accuracy similar to manual positioning.
    Streaming StabilityNo performance degradation (frames per second, jitter, packet loss) with simultaneous multiple usersAll performance tests for the streaming stability were successfully completed with no deviations, confirming compliance with the required performance standards.
    Ruler PerformancePrecision of one millimeter for linear (Euclidean) distance between two user-selected points on the scapula’s unreamed 3D mesh.All performance tests for the ruler tool accuracy were successfully completed with no deviations, confirming compliance with the required performance standards.

    Study Details

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size for Test Set: 173 samples.
      • Data Provenance: Retrospective, with a split based on patient gender, shoulder side, and geographical region of origin.
        • Geographical Origin (Test Set):
          • Left shoulder: 58.2% Europe (46), 41.8% USA (33)
          • Right shoulder: 56.4% Europe (53), 43.6% USA (41)
        • The data corresponds to patients that underwent total shoulder arthroplasty with an FX Shoulder implant, with diversity in gender, imaging equipment, institutions, and study year. The image acquisition protocol was standard for this type of procedure.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

      • The document does not explicitly state the number of experts used.
      • Qualifications: "Medical professionals" are mentioned for creating manual segmentation labels. For landmark performance, "experts" adjusted final positions, but their specific qualifications are not detailed beyond "medical professionals." For shoulder side detection, a "Clinical Solutions Specialist" performed a manual assessment.
    3. Adjudication Method for the Test Set:

      • The document does not specify an explicit adjudication method such as 2+1 or 3+1 for establishing ground truth from multiple experts. It mentions labels created "manually by medical professionals" and "final positions adjusted by experts," implying a consensus or single-expert approach, but no detailed adjudication process is described.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No, a multi-reader multi-case (MRMC) comparative effectiveness study evaluating human reader improvement with AI vs. without AI assistance was not explicitly mentioned or described in the provided information. The studies primarily focus on the standalone performance of the AI for various tasks.
    5. Standalone (Algorithm Only) Performance Study:

      • Yes, a standalone performance study was conducted. The "Segmentation Performance Testing," "Shoulder Side Detection performance testing," "Measurement Accuracy performance testing," "Landmark Performance Testing," "Streaming Stability Testing," and "Ruler Performance Testing" sections all describe the evaluation of the brAIn™ Shoulder Positioning software's algorithmic performance against established ground truths or benchmarks, without explicit human-in-the-loop interaction as part of the primary evaluation metrics.
    6. Type of Ground Truth Used:

      • Segmentation: Manual segmentation performed by medical professionals.
      • Shoulder Side Detection: Manual assessment performed by a Clinical Solutions Specialist.
      • Measurement Accuracy: Reported accuracy of the predicate device (for comparison when editing positions) and theoretical distances calculated from spatial coordinates (for ruler tool).
      • Landmark Performance: Final positions adjusted by experts.
    7. Sample Size for the Training Set:

      • Sample Size for Training Set: 335 samples (corresponding to 65.9% of the total dataset).
    8. How the Ground Truth for the Training Set Was Established:

      • The labels (ground truth) for both the training and testing sets were created "manually by medical professionals."
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1