Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K252524

    Validate with FDA (Live)

    Date Cleared
    2025-09-09

    (29 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Quadsense and Quadsense Pro Systems are indicated for any medical condition in which primary Total Knee Arthroplasty (TKA) would be indicated, and the patella is resurfaced.

    For use as a comparative tool to measure patellofemoral joint force. The Quadsense Sensor and Quadsense Pro Sensor are sterile, for single patient use.

    The devices do not make a diagnosis and are not intended to replace a surgeon's clinical judgement.

    Device Description

    The Quadsense and Quadsense Pro Systems have been designed to allow surgeons to assess relative quadricep forces at the patellofemoral joint during total knee surgery.

    The Quadsense and Quadsense Pro Systems includes a sensor that attaches to the patella, adjustment shims that attach to the sensor, a control puck, and a touchscreen Panel PC with Quadsense software. The patella sensor can take readings of Medial, Lateral, Superior, and Inferior loads during range of motion trials. The sensor records data that is displayed on the Panel PC in real time. The Panel PC remains outside of the sterile field, and the software workflow can be controlled by the surgeon using the control puck, within the sterile field. The Panel PC collects and temporarily stores sensor reading data so the surgeon can view the sensor readings of force through the patellofemoral joint during different stages of the surgical procedure.

    The sensors are designed to attach to the patella at the start of the surgical procedure, following a conservative resection of the patella. This allows the surgeon to take a sensor reading of the patient's force through the patellofemoral joint during range of motion trials, prior to the resurfacing of the femur and the tibia. The surgeon can then take a reading of the force through the patellofemoral joint with the trial femoral and tibial components in situ. This allows the surgeon to assess how the new position of the femoral and tibial components has impacted the patient's force through the patellofemoral joint by comparing the reference reading taken prior to resurfacing to the reading(s) taken following trial implant insertion.

    The sensor is supplied with several shims with different thicknesses and angles. By changing the thickness, rotation angulation parameters of the shim attached to the sensor during the surgical procedure the surgeon can assess how changes in patella thickness or angulation will impact the force through the patellofemoral joint.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1