510(k) Data Aggregation

    K Number
    K251906

    Validate with FDA (Live)

    Device Name
    Z1 Hip System
    Manufacturer
    Date Cleared
    2025-07-18

    (28 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Z1 Hip System is intended for total or hemi hip arthroplasty and is indicated for the following conditions:

    • Advanced wear of the joint due to degenerative, post-traumatic or rheumatic diseases.
    • Failed previous hip surgery including joint reconstruction (osteotomy), arthrodesis, hemi-arthroplasty or total hip replacement (THR).
    • Acute traumatic fracture of the femoral head or neck.
    • Avascular necrosis of the femoral head.

    Z1 Hip System is for cementless use only.

    Device Description

    The Z1 Hip System consists of femoral hip stems intended for use in total or hemi hip arthroplasty. The stem is designed for cementless implantation in the proximal femur and mates with compatible femoral heads and adapters for use in total or hemi hip arthroplasty through a 12/14 male taper connection. The stems are manufactured from titanium alloy Ti-6Al-4V and have a wedge-shaped design, with a proximal-to-distal taper. Apart from the highly polished femoral neck region and the taper, the entire surface of the stem is grit-blasted, proximally sprayed with a Ti-6Al-4V titanium alloy plasma coating, and followed by a hydroxyapatite (HA) overcoat. Offered in multiple sizes and neck lengths, the stems are available in standard, high offset, and coxa vara offsets and as collared or collarless stems in each offset to accommodate various patient anatomies. The hip stems are provided sterile and are for single use only. System-specific instrumentation is available to prepare the femur for implantation of the Z1 Hip System femoral stems. The stems are compatible with previously cleared Zimmer Biomet modular femoral heads and/or acetabular components.

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter for the Z1 Hip System, which is a medical device for hip arthroplasty. The document details the administrative aspects of the clearance, the device description, intended use, and a summary of non-clinical testing.

    However, the question asks for information related to a "study that proves the device meets the acceptance criteria," specifically in the context of an AI/Software as a Medical Device (SaMD). The Z1 Hip System is a physical orthopedic implant, not an AI/SaMD. The document clearly states:

    • "No clinical testing was conducted." (Page 6)
    • The tests performed are "Distal Fatigue Testing per ISO 7206-4", "Proximal Fatigue Testing per ISO 7206-6", "Evaluation of Compatibility with Femoral Heads", and "Evaluation of Range of Motion" (Page 6). These are all mechanical and performance tests for a physical implant, not studies involving AI algorithms, expert reads, or clinical outcomes data as would be relevant for an AI/SaMD.

    Therefore, I cannot provide the requested information as the provided text does not contain any details about acceptance criteria or a study design relevant to an AI/SaMD. The information requested (multi-reader multi-case studies, ground truth establishment, sample size for training sets, etc.) is specific to the validation of AI algorithms, which is not applicable to the Z1 Hip System described in this 510(k) clearance.