Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K251877

    Validate with FDA (Live)

    Date Cleared
    2025-08-15

    (58 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JMS CAVEO A.V. Fistula Needle Set is intended for temporary cannulation to vascular access for extracorporeal blood treatment for hemodialysis. This device is intended for single use only. The anti-needlestick safety feature aids in prevention of needle stick injuries when removing and discarding the needle after dialysis. The device also has an integrated safety mechanism that is designed to automatically generate a partial occlusion of the internal fluid path and trigger the hemodialysis machine to alarm and shut off if a complete dislodgement of the venous needle from the arm inadvertently occurs. In vitro testing supports that this feature triggers the hemodialysis machine to alarm and shut off.

    Device Description

    The subject device is the JMS CAVEO A.V. Fistula Needle Set (CAVEO) with an anti-needlestick safety feature. The Caveo is predicted to protect patients from the risks associated with venous needle dislodgement (VND) based on bench testing results. It contains an integrated stainless steel torsion spring mechanism and bottom footplate that provides an open/fluid path when the AV fistula set is fully cannulated into the access site. When the venous needle becomes completely dislodged from the patient's arm, this mechanism enables the footplate to partially occlude the blood path, generating an increased venous line pressure high enough to trigger automatic alarm and halt further blood pumping of the hemodialysis machine. In vitro testing supports that this feature triggers the hemodialysis machine to alarm and shut off. Based on bench testing results, this may significantly reduce patient blood loss in the event of a complete VND. The Caveo has a pre-attached anti-stick needle guard for prevention of needlestick injury at the time of needle withdraw after completion of a hemodialysis procedure.

    In vitro performance testing using dialysis machine Fresenius 2008K supports the function of the Caveo VND feature with a venous pressure limit set to 200mmHg symmetric mode, a maximum dialyzer membrane surface area of 2.5 m2, minimum blood flow rate of 200 mL/min, maximum ultrafiltration rate of 4000 mL/hour, and simulated treatment duration of 8 hours. If different machine and/or setting are used, before introducing the device, refer to Directions for Use.

    AI/ML Overview

    This document describes the FDA 510(k) clearance for the JMS CAVEO A.V. Fistula Needle Set. This device is a physical medical device, specifically a needle set for hemodialysis, and does not involve Artificial Intelligence (AI). Therefore, many of the requested criteria related to AI/software performance, ground truth establishment, expert adjudication, MRMC studies, and training datasets are not applicable.

    The document primarily focuses on bench testing (in vitro performance) and a simulated clinical usability study to demonstrate device safety and effectiveness.

    Here's a breakdown based on the provided text:

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily demonstrated through various performance tests, with "Passed" as the acceptance. The document doesn't explicitly state numerical acceptance thresholds for all tests (e.g., how much "Force to Depress the Footplate" is acceptable), but it implies successful completion. For some, like needle penetration resistance and retraction lock strength, numerical criteria are provided.

    Acceptance Criterion (Test)Reported Device Performance
    Needle Penetration Resistance
    14G≤ 40g (Predicate: ≤ 40g)
    15G≤ 35g (Predicate: ≤ 35g)
    16G≤ 30g (Predicate: ≤ 30g)
    17G≤ 30g (Predicate: ≤ 30g)
    Needle Retention Strength> 6.0kgf (Predicate: > 6.0kgf)
    Needle Surface (Visually)No dented/damaged needle (Predicate: No dented/damaged needle)
    Product LeakNo air bubble should appear when subjected to air pressure 0.40 kgf/cm2 and immersed in water. (Predicate: Same)
    Needle Retraction Final Lock Strength≤ 2.0kgf (Predicate: ≤ 2.0kgf)
    Connector (Air tightness, Luer fit)Passed (ISO 80369-7 compliant)
    Connection Strength (Tube to Connector/Joint, Tube to Pivot Valve Core)> 6.0kgf (Predicate: > 3.0kgf for Tube to Connector/Joint, > 6.0kgf for Tube to Hub)
    Leakage by Pressure Decay (Female Luer Lock)Passed
    Positive Pressure Liquid Leakage (Female Luer Lock)Passed
    Sub-atmospheric Pressure Air Leakage (Female Luer Lock)Passed
    Stress Cracking (Female Luer Lock)Passed
    Resistance to Separation from Axial Load (Female Luer Lock)Passed
    Resistance to Separation from Unscrewing (Female Luer Lock)Passed
    Resistance to Overriding (Female Luer Lock)Passed
    Tube to Connector Pull Test (Female Luer Lock)Passed
    Luer Lock Cover Open Torque Test (Female Luer Lock)Passed
    Testing Activation of the Sharps Injury Protection FeaturePassed
    Needle Pushback Strength TestPassed
    Needle Guard Detachment Strength TestPassed
    Appearance Check (Caveo)Passed
    Cover Pull with Hub (Caveo)Passed
    Air Leak Test (Caveo)Passed
    Positive Pressure Leak Test (Caveo)Passed
    Negative Pressure Leak Test (Caveo)Passed
    Needle Guard Retraction Final Lock Test (Caveo)Passed
    Tube to Hub Pull Test (Caveo)Passed
    Cannula to Hub Tensile Test (Caveo)Passed
    Dimensional Analysis of Footplate to Pivot Valve Core (Caveo)Passed
    TPE Front & Back Ends Internal Diameter (Y-axis) Measurements (Caveo)Passed
    TPE Surface Roughness (Caveo)Passed
    Cannulation at 15 and 45-Degree Angles (Caveo)Passed
    Occlusion After Taping (Caveo)Passed
    VND Performance (Venous Needle Dislodgement)Passed
    Baseline Pressure Comparison (Caveo)Passed
    Force to Depress the Footplate (Caveo)Passed
    Mechanical Hemolysis Testing (Caveo)Passed
    Simulated Clinical Usability StudySuccessful
    Transportation TestPassed
    Human Factors TestingPassed
    Biocompatibility (Cytotoxicity, Sensitization, Irritation, Hemocompatibility, Pyrogenicity, Acute Systemic Toxicity, Subacute Toxicity, Genotoxicity)Passed (for all, per ISO 10993 standards)

    2. Sample Size for Test Set and Data Provenance

    • Test Set (Clinical Trial): 15 subjects (2 females, 13 males).
    • Data Provenance: The document does not explicitly state the country of origin. It describes the recruitment from "the general hemodialysis population," and mentions racial/ethnic demographics, but not geographic. Given the company is "JMS North America Corporation" (Hayward, CA), it is highly probable the study was conducted in the US. The study appears to be prospective as it involves recruitment and device use to confirm safety, performance, and usability.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This question is largely not applicable as the device is a physical medical device. The "ground truth" for performance is established through bench testing (objective physical measurements) and the success of the device in a simulated clinical setting. There is no mention of human experts establishing a "ground truth" for diagnostic or AI-related interpretations.

    For the simulated clinical usability study, the "ground truth" is whether the device performed as intended and was usable, as observed by clinicians/researchers during the study. The qualifications of those assessing the usability are not specified, beyond the implication that they are competent to conduct a clinical trial for hemodialysis devices.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1) are typically used in studies involving human interpretation (e.g., radiology reads) where there might be disagreement in expert opinions needing a tie-breaker. This is not applicable here as:

    • The primary "test set" involves objective performance characteristics (bench testing).
    • The clinical usability study likely involved observing successful function and user feedback, not a diagnostic interpretation needing adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. An MRMC study is relevant for evaluating the performance of AI systems or diagnostic tools where multiple human readers interpret cases, often with and without AI assistance, to see if the AI improves human performance. This device is a physical hemodialysis needle set, not an AI or diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This refers to the performance of an AI algorithm on its own. The device is a physical product. Its "standalone" performance is assessed through bench testing (e.g., VND Performance, Mechanical Hemolysis Testing, Needle Penetration Resistance). These tests evaluate the device's inherent functional characteristics independent of human interaction during the critical failure modes.

    7. The Type of Ground Truth Used

    • Bench Testing: The ground truth is based on objective physical measurements and engineering specifications (e.g., force measurements, leak tests, dimensional analyses) and functional success/failure (e.g., did the VND feature trigger the alarm?).
    • Simulated Clinical Usability Study: The ground truth is based on observed device performance during simulated use and the successful delivery of hemodialysis without impedance by the device's novel features. This is akin to outcomes data in a controlled simulated environment.

    8. The Sample Size for the Training Set

    Not applicable. The device is a physical product and does not involve AI or machine learning models that require a "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no training set for an AI model.

    In summary, the provided document details the non-clinical and limited clinical testing of a physical medical device (hemodialysis needle set). The acceptance criteria are largely met through rigorous bench testing demonstrating physical and functional robustness, and a small simulated clinical study confirming usability and safety in a controlled environment. AI-specific criteria are not relevant to this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1