510(k) Data Aggregation

    K Number
    K251867

    Validate with FDA (Live)

    Date Cleared
    2025-09-19

    (94 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V has been designed to be used with a video system center, light source, documentation equipment, monitor, endo therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.

    Device Description

    The EVIS EXERA III DUODENOVIDEOSCOPE TJF-Q190V has been designed to be used with a video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum. The TJF-Q190V is compatible with Olympus system "Video System Center OLYMPUS CV-190 and XENON LIGHT SOURCE OLYMPUS CLV-190 (K112680). TJF-Q190V is also compatible with Olympus system "EVIS X1 Video System Center OLYMPUS CV-1500" (K222584).

    The subject device consists of a flexible insertion section, control section and endoscope connector section with equipped charge-coupled device (CCD) chip which delivers images.

    The light from the light source travels through the light guide to the light guide lens at the distal end. The light source can offer both white light for normal observation and narrow band imaging (NBI). The CCD chip transduces the incident light from the objective lens to electrical signal. The video processor transduces the electrical signal to video signal.

    There is an instrument channel located inside of the flexible insertion section. EndoTherapy accessories can be inserted through the instrument channel. A forceps elevator is located at the distal end of the insertion section to elevate endo therapy accessories for endoscopic treatment.

    A sterile, single-use distal cover (MAJ-2315) has been designed to be attached to the OLYMPUS TJF-Q190V to cover the distal end of the insertion tube and fit around the forceps elevator. MAJ-2315 is to be discarded after clinical use. MAJ-2315 and TJF-Q190V were previously cleared under 510(k)s K193182, K202661, K220587, and K250701.

    The subject device has the same technological characteristics and similar design as the predicate device.