510(k) Data Aggregation

    K Number
    K251196

    Validate with FDA (Live)

    Date Cleared
    2025-04-21

    (4 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of composite brackets and buttons is to be bonded to the teeth to provide orthodontic movement of natural teeth.

    The main indication for use of orthodontic brackets and buttons is a malocclusion, such as misaligned teeth, overcrowding, spacing issues, and bite irregularities.

    Device Description

    Composite orthodontic brackets and buttons present a clear and aesthetically pleasing choice for orthodontic patients undergoing orthodontic treatment. The Composite Bracket and Button is engineered to offer simple and aesthetic devices that address both minor and complex misalignments in patients.

    Composite orthodontic brackets and buttons are classified as Class II medical devices. In accordance with ISO 10993-1:2018, based on the nature of body contact, orthodontic brackets and buttons are categorized as having mucosal contact, as they are affixed to the teeth and maintain prolonged contact with the oral mucosa. Furthermore, the contact duration is categorized as "permanent exposure" (=more than thirty days), as brackets and buttons are expected to remain in place for extended periods during orthodontic treatment.

    Utilized by orthodontic professionals, these brackets and buttons serve as integral components - together with adhesives, wires, bands, elastics and other attachments in the alignment and correction of dental malocclusions and misalignments. The duration of the treatment depends on the degree of misalignment and is between 6 months and a maximum of 2,5 years.

    Brackets and buttons are strategically placed and affixed to the surface of the teeth using dental adhesive. This bonding process ensures that the brackets and buttons remain securely attached to the teeth throughout the orthodontic treatment. Then Orthodontic arch wires are pushed into the slot of the bracket. The slot acts as an aid to hold and guide the orthodontic wires that - in conjunction with bands and other orthodontic accessories - exert gentle forces to the teeth, gradually moving them into their desired position. These direct and indirect orthodontic bonding techniques with orthodontic adhesives is thoroughly documented in orthodontic literature. These additional products required for orthodontic treatment are chosen by the orthodontist as required and are hence not part of this 510(k) application. The removal of the brackets/buttons follow standard procedures that do not require additional training.

    Composite Orthodontic Products are only to be used by Orthodontists/Dentists. As educated specialists they know exactly how to use brackets and buttons. They have received training in order to analyse and decide with which other products they are best compatible (wires, adhesives, elastics and instruments). This 510(k) application is only for composite orthodontic brackets and buttons.

    AI/ML Overview

    It's important to note that the provided document is an FDA 510(k) clearance letter for "Composite Orthodontic Brackets and Buttons" (ECC Premium Line). While it confirms substantial equivalence to a predicate device, it does not include detailed study results in the format requested, particularly for acceptance criteria and specific performance metrics for AI/ML devices. This document describes a physical medical device, not an AI/ML software device.

    Therefore, many of the requested fields, especially those related to AI/ML specific studies (like MRMC, standalone performance, training set details, and ground truth for AI), cannot be extracted from this document, as they are not applicable to the device described.

    However, I can provide information based on the presented text for the relevant sections:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that a series of bench tests were completed to show that the medical devices function as intended and meet specified performance characteristics. It also states they withstand stresses and do not pose a risk. However, precise numerical acceptance criteria and their corresponding reported performance values are not detailed in this 510(k) summary. The document generally asserts that the device "meets the specified performance characteristics" and "functions as intended."

    Test NameAcceptance Criteria (Not Explicitly Stated in Document)Reported Device Performance (Not Explicitly Stated with Specific Values in Document)
    Dimensional Test(Implied: Conformance to ISO 27020:2019)(Implied: Conforms to ISO 27020:2019)
    Hardness Test(Not specified)(Implied: Meets specified characteristics)
    Shear Test(Not specified)(Implied: Meets specified characteristics, withstands stresses)
    Friction Force Test(Not specified)(Implied: Meets specified characteristics)
    Gate Pull Test(Not specified)(Implied: Meets specified characteristics)
    Torque Test(Not specified)(Implied: Meets specified characteristics)
    Wear Abrasion Test(Not specified)(Implied: Meets specified characteristics, withstands stresses)
    Water Absorption Test(Not specified)(Implied: Meets specified characteristics)
    Discoloration Test(Not specified)(Implied: Meets specified characteristics)

    2. Sample Size Used for the Test Set and Data Provenance

    The document is for a physical medical device (orthodontic brackets and buttons), not a software/AI device that typically uses "test sets" of data. The "tests" performed are bench tests on the physical product. The sample sizes for these bench tests are not specified in the provided text. The concept of "data provenance" (country of origin, retrospective/prospective) is not applicable here as these are laboratory bench tests, not data-driven studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable as the device is a physical medical device. Ground truth, in the context of expert consensus, refers to the labeling of medical images or data by experts, typically for AI/ML model training or evaluation. The performance evaluation here is based on physical material properties and mechanical function through bench testing.

    4. Adjudication Method for the Test Set

    This question is not applicable as the device is a physical medical device. Adjudication methods (like 2+1, 3+1) are used for resolving disagreements among human readers/experts when establishing ground truth for AI/ML datasets.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This question is not applicable. An MRMC study is relevant for evaluating the clinical effectiveness of AI/ML-assisted diagnostic or decision-support systems by comparing human performance with and without AI assistance. This document describes a physical orthodontic device; therefore, no such study was conducted or reported.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. Standalone performance refers to the evaluation of an AI/ML algorithm's performance independent of human intervention. Since this is a physical medical device, there is no algorithm or AI component to evaluate in a standalone manner.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the bench tests conducted, the "ground truth" implicitly refers to the engineering specifications and standards (e.g., ISO 27020:2019 for dimensional tests) against which the physical device's performance is measured. There is no biological "ground truth" like pathology or outcomes data generated for these non-clinical tests.

    8. The Sample Size for the Training Set

    This question is not applicable. A "training set" is used for developing AI/ML models. This device is a physical product, not an AI/ML software.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable. As there is no AI/ML model, there is no training set and thus no ground truth for a training set.