    The Cure Catheter Insertion Kit is an intermittent urinary catheter kit indicated for the purpose of bladder drainage for males and females. The urinary catheter kit comes in a variety of configurations and sizes packaged sterile for single-use.

    Device Description

    The Cure Catheter Insertion Kits (K1, K2, K2-90, K3), Cure Hydrophilic Catheter Kit (HMxxUK), Cure Pocket Catheter Kit (MxxUK), Cure Catheter Closed System Kit (CSxx)) contain sterile products used during intermittent urinary catheterization. Two types of kit are provided, one where a urinary catheter is included and one without a catheter. The components allow users to select one or more of the products to help prevent contamination of the environment and the user and to maintain a no-touch technique during the insertion of the catheter. Catheters are inserted through the urethra and indicated for the purpose of bladder drainage.

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for a medical device called the "Cure Catheter Insertion Kit" and related products. It outlines the regulatory classification, intended use, and a comparison to a predicate device.

    However, it does not contain information about specific acceptance criteria or a study that objectively proves the device meets those criteria in terms of performance metrics like accuracy, sensitivity, or specificity for a diagnostic or screening device.

    The "Performance Data" section lists various validation tests for general device characteristics, such as:

    • Sterilization validation: Conforms to AAMI/ISO 11135-1:2014/AMD 1:2018 and ISO 10993-7:2008/AMD:2019.
    • Biocompatibility testing: Conforms to ISO 10993-1:2018 and FDA Guidance "Use of International Standard ISO 10993-1".
    • Sterile packaging: Conforms to ISO 11607-1:2019 and ISO 11607-2:2019.
    • Real-time aged shelf-life testing: Conforms to ISO 11607-1:2006 with justification to version 2019.
    • Packaging integrity testing: Conforms to ASTM F2096-11 (2019).
    • Urinary catheter testing: Conforms to ISO 20696:2018.

    These are standard engineering and safety tests for medical devices, ensuring they are safe, sterile, and perform their basic mechanical function. They are not clinical performance studies to assess diagnostic accuracy or the direct clinical efficacy of a software-driven device.

    Therefore, the specific information requested in the prompt, such as acceptance criteria in terms of diagnostic performance (e.g., sensitivity, specificity), study design details (sample size, data provenance, expert adjudication, MRMC, standalone performance, ground truth establishment), is not available in this 510(k) clearance letter.

    This document is for a urological catheter and accessories, which are physical medical devices, not an AI or software-based diagnostic tool. The performance data listed refers to the manufacturing, material, and sterility aspects, which are crucial for this type of device, but do not involve the types of criteria and studies you've asked about for AI/diagnostic devices.