510(k) Data Aggregation

    K Number
    K250785

    Validate with FDA (Live)

    Date Cleared
    2025-05-06

    (53 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Single-use Flexible Ureteroscope is designed for use with Vathin Display Units, endotherapy accessories and other ancillary devices for the endoscopy and endoscopic surgery within urinary tract and kidney in adults.

    Device Description

    The Single-use Flexible Ureteroscope is intended to be used with Vathin Digital Video Monitor. During diagnosis and treatment, the Single-use Flexible Ureteroscope is inserted into the ureter or into the kidney or renal pelvis through the ureter, and the image sensor (CMOS) at the end of the Single-use Flexible Ureteroscope converts the received mucosal reflected light signals into electrical signals, transmitted to the Digital Video Monitor through the cable, the Digital Video Monitor receives the image signal from the endoscope and processes it, converts it into an image signal that can be displayed on the display screen, and finally presents it on the screen of the display.

    The Single-use Flexible Ureteroscope is provided sterile (sterilized by EO) and intended to be single-use. It is for use in professional Healthcare Facility Environment.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and its associated summary for the "Single-use Flexible Ureteroscope" does not contain information regarding a study with specific acceptance criteria and detailed device performance metrics in the way one might expect for an AI-enabled device or a diagnostic accuracy study. This document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance data.

    Therefore, the requested information cannot be fully extracted from the provided text. I will provide the acceptance criteria as implied by the non-clinical tests conducted to prove substantial equivalence, and note where specific information is missing based on your request.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria for each bench test, nor does it provide specific numerical performance results. Instead, it states that the device "complies with" or "met its design specifications" or "has been validated according to" relevant standards.

    Acceptance Criterion (Implied)Reported Device Performance
    Biocompatibility:
    No CytotoxicityNo Cytotoxicity
    No Irritation to SkinNo Irritation to Skin
    No significant sensitizationNo significant evidence of sensitization
    No Pyrogenicity(Test performed)
    No Acute Systemic Toxicity(Test performed)
    Electrical Safety:
    Compliance with IEC 60601-1Complies with IEC 60601-1
    Compliance with IEC 60601-2-18Complies with IEC 60601-2-18
    Electromagnetic Compatibility (EMC):
    Compliance with IEC 60601-1-2Complies with IEC 60601-1-2
    Compliance with IEC 60601-4-2Complies with IEC 60601-4-2
    Sterilization:
    Validation according to ISO 11135Validated according to ISO 11135
    Shelf Life & Packaging:
    Packaging integrity validationValidated
    Product performance after accelerated aging per ASTM F1980-21Carried out and presumably passed
    Product performance after simulated transportation per ASTM D4169-22Carried out and presumably passed
    Mechanical Performance:
    Compliance with ISO 8600-1Performed in accordance with
    Compliance with ISO 8600-4Performed in accordance with
    Optical Performance:
    Field of view (120°)(Tested to meet design specifications)
    Direction of view (0°)(Tested to meet design specifications)
    Depth of field(Tested to meet design specifications)
    Geometric distortion(Tested to meet design specifications)
    Image intensity uniformity(Tested to meet design specifications)
    Color performance(Tested to meet design specifications)
    Signal-To-Noise Ratio(Tested to meet design specifications)
    Dynamic Range(Tested to meet design specifications)
    Photobiological Safety:
    Compliance with IEC 62471:2006Tested according to IEC 62471:2006

    Missing Information Details:

    1. Sample size used for the test set and the data provenance: This information is not provided in the document. The studies were non-clinical bench tests and biocompatibility tests, not clinical trials with patient data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This is not applicable as the studies were non-clinical bench tests and material safety tests, not studies requiring expert interpretation of clinical images or data.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for non-clinical bench tests.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document describes a medical instrument, not an AI-enabled device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For non-clinical tests, the "ground truth" is typically defined by the specified performance limits or standards outlined in the applicable international standards (e.g., ISO, IEC, ASTM). For biocompatibility, it's the lack of adverse biological reactions.
    7. The sample size for the training set: Not applicable, as this is not an AI/machine learning device.
    8. How the ground truth for the training set was established: Not applicable, as this is not an AI/machine learning device.

    In summary, the provided FDA document details the non-clinical testing performed to demonstrate the safety and effectiveness of a physical medical device (a single-use flexible ureteroscope) for substantial equivalence to a predicate device. It does not contain information related to diagnostic accuracy studies, AI performance, or clinical trials with human readers or patient data.