Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K250583

    Validate with FDA (Live)

    Date Cleared
    2025-06-20

    (113 days)

    Product Code
    Regulation Number
    868.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AdheResp Smart Breath-actuated Mesh Nebulizer is a system to aerosolize liquid medications for inhalation for patients ranging from pediatrics equal to or greater than 5 years old to adults who can coordinate breathing.

    The environment of use are Hospital/institutional settings, home care use, schools, and long-term care facilities.

    Device Description

    The AdheResp Smart Breath-actuated Mesh Nebulizer consists of a main unit, medication reservoir, mouthpiece, key tag and charging cable, or power adapter. The medication reservoir could be further separated into a medication container and an aerosol chamber.

    Principle of Operation: There are two key features in the AdheResp Smart Breath-actuated Mesh Nebulizer, which are vibrating mesh nebulization and breath-actuation functions. The AdheResp Smart Breath-actuated Mesh Nebulizer is an active mesh nebulizer. Active mesh nebulizers operate by electrically activating a piezoelectric (PZT) ceramic element that rapidly contracts and expands when applied a voltage to it. The vibrations of the PZT are transmitted to the mesh membrane, which is in contact with the liquid medication loaded in the medication container. This vibration pushes the liquid or suspension through the fine conical holes of the mesh, forming aerosol droplets for the user to inhale.

    AI/ML Overview

    This document is a 510(k) clearance letter for the AdheResp Smart Breath-actuated Mesh Nebulizer. It outlines the device's characteristics and its equivalence to predicate devices, particularly focusing on performance data related to aerosol delivery.

    Here’s the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Important Note: The provided text is a 510(k) summary and clearance letter. It summarizes findings and states that acceptance criteria were met, but it does not provide the detailed study protocols, raw data, or the explicit acceptance criteria values themselves in a direct, quantifiable manner for all aspects. Instead, it presents comparative performance data and lists categories of testing. Therefore, some information will be inferred from the context or noted as 'not explicitly stated' if the document does not provide it.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state quantitative acceptance criteria for each parameter in a "limit" vs. "result" format. Instead, it presents comparative performance data against a predicate device and states that the subject device "met its acceptance criteria" and that its performance is "similar." For the purpose of this table, I will present the reported performance of the Subject Device and mention the comparison to the Predicate. The implicit acceptance criterion for the comparative performance data is that the subject device's performance is demonstrably non-inferior or similar to the predicate.

    Acceptance Criterion (Implicit)Reported Device Performance (Subject Device)
    Particle Size (MMAD): Similar to predicate device15 LPM: Albuterol 3.98 ±0.18 µm, Ipratropium 3.984 ±0.11 µm, USP Cromolyn 4.07 ±0.14 µm (Predicate generally higher MMAD)
    30 LPM: Albuterol 3.48 ±0.15 µm, Ipratropium 3.39 ±0.21 µm, USP Cromolyn 3.58 ±0.15 µm (Predicate generally higher MMAD)
    Geometric Standard Deviation (GSD): Similar to predicate15 LPM: Albuterol 1.82 ±0.03, Ipratropium 1.94 ±0.10, USP Cromolyn 1.88 ±0.02 (Predicate generally higher GSD)
    30 LPM: Albuterol 1.89 ±0.09, Ipratropium 1.98 ±0.16, USP Cromolyn 1.90 ±0.07 (Predicate generally higher GSD)
    Total Delivered Dose (mg): Similar to predicate15 LPM: Albuterol 4.25 ±0.36, Ipratropium 0.45 ±0.07, USP Cromolyn 16.23 ±1.35 (Subject generally slightly higher or similar to Predicate)
    30 LPM: Albuterol 3.83 ±0.29, Ipratropium 0.40 ±0.04, USP Cromolyn 14.65 ±1.09 (Subject generally slightly higher or similar to Predicate)
    Respirable Dose <5µm [mg]: Similar to predicate15 LPM: Albuterol 2.71 ±0.32, Ipratropium 0.28 ±0.03, USP Cromolyn 10 ±0.85 (Subject generally higher than Predicate)
    30 LPM: Albuterol 2.67 ±0.21, Ipratropium 0.27 ±0.03, USP Cromolyn 9.95 ±0.71 (Subject generally higher than Predicate)
    Coarse Particle Dose <4.7µm [mg]: Similar to predicate15 LPM: Albuterol 1.47 ±0.10, Ipratropium 0.15 ±0.02, USP Cromolyn 5.85 ±0.65 (Subject generally lower or similar to Predicate)
    30 LPM: Albuterol 1.05 ±0.28, Ipratropium 0.11 ±0.01, USP Cromolyn 4.55 ±0.72 (Subject generally lower or similar to Predicate)
    Fine Particle Dose <4.7µm [mg]: Similar to predicate15 LPM: Albuterol 2.56 ±0.30, Ipratropium 0.27 ±0.03, USP Cromolyn 9.41 ±0.65 (Subject generally higher than Predicate)
    30 LPM: Albuterol 2.54 ±0.21, Ipratropium 0.26 ±0.03, USP Cromolyn 9.46 ±0.48 (Subject generally higher than Predicate)
    Ultra-Fine Particle Dose <1µm [mg]: Similar to predicate15 LPM: Albuterol LOD, Ipratropium 0.01 ±0.00, USP Cromolyn LOD (Similar to Predicate)
    30 LPM: Albuterol 0.21 ±0.06, Ipratropium 0.02 ±0.01, USP Cromolyn 0.58 ±0.05 (Similar to Predicate)
    Nebulization Rate: Must meet specified rate>0.25 ml/min (Similar to predicate)
    Inter- and Intra-sample Variability: Must meet criteriaMet acceptance criteria (no specific values given)
    Minimum Breath Actuation Tests: Must meet criteriaMet acceptance criteria (no specific values given)
    Biocompatibility (ISO 10993 & ISO 18562): Must pass testsApplicable ISO 10993 and ISO 18652 testing was performed and passed.
    Reprocessing, Sterility, and Shelf-Life: Must pass testsCleaning Validation, Aging, Simulated Life (cleaning, aging, drop) were performed and passed. (implicitly met acceptance criteria related to these tests)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • For the comparative performance testing (Breath Actuated Particle Characterization): "Testing was performed with 3 samples of each device in triplicate with the 3 drugs." This implies $3 \text{ devices} \times 3 \text{ replicates} = 9$ runs per drug per flow rate for both the subject and predicate devices.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The study is non-clinical bench testing ("Bench testing," "Non-Clinical Testing Summary"), meaning it's a controlled lab environment, not patient data. It is a prospective study in the sense that the tests were conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not applicable. This device is a nebulizer, and the tests described are bench tests (e.g., particle size, dosage, flow rates). These are physical measurements, not interpretations of medical images or patient conditions that would require expert human readers to establish ground truth.
    • Qualifications of Experts: N/A, as noted above.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. This is not a study involving human interpretation of data, where adjudication of disagreements would be necessary. The performance data is derived from direct and repeatable physical measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No. An MRMC study is typically performed for AI/CADe (Computer-Assisted Detection/Diagnosis) devices that assist human readers in interpreting medical images. This device is a nebulizer, a drug delivery system, and its performance is evaluated through bench testing.

    • Effect Size of Human Readers' Improvement: Not applicable, as no MRMC study was performed.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance: Yes, in essence. The "AdheResp Smart Breath-actuated Mesh Nebulizer" itself has an algorithm for "Breath Actuation" that "Measures breath and averages the cycle, then delivers aerosol during the designated period." The "Breathe Actuated Particle Characterization Performance Comparison" is a standalone evaluation of the device's physical output based on this internal algorithm responding to simulated breathing patterns (15 LPM and 30 LPM). There is no "human-in-the-loop" for the output of the device itself.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for this device's performance is established by standardized laboratory measurements using traceable equipment and protocols (e.g., measuring particle size, drug quantity, flow rates) under controlled conditions. This is considered an objective physical truth, not a subjective interpretation. The comparison is made against a legally marketed predicate device's performance.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This device is a physical medical device (nebulizer) with an embedded algorithm, not a machine learning/AI model that requires a "training set" of data in the typical sense for image recognition or diagnostic purposes. The device's "algorithm" is for controlling the nebulization based on breath patterns, which would be developed through engineering and control system design, not data-driven machine learning.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set Establishment: Not applicable, as there is no "training set" in the context of a machine learning model. The "ground truth" for the device's operational parameters (like breath-actuation logic or aerosol delivery range) would be based on established engineering principles, physiological understanding of human breathing, and clinical requirements for drug delivery, validated through rigorous bench testing and potentially human factors testing, though the latter is not detailed in this summary.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1