Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K250475

    Validate with FDA (Live)

    Date Cleared
    2025-05-27

    (97 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Electric Wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    The product is intended only carry one person and used as a means of transportation for the old and infirm who have difficulty in moving.

    The maximum occupant mass is 100kg.

    The Electric Wheelchair is a battery powered four wheeled vehicle. It consists one Lithium battery with an off-board battery charger, frame, controller, motors, seat, back support, control device (including the battery power indicator, ON/OFF button, horn button, speed indicator, speed control button, joystick, Battery charger socket), arm supports, two rear wheels, two casters(front wheels), Foot support, anti-tip devices.

    The wheelchair can easily fold and unfold for transportation or storage.

    AI/ML Overview

    The provided text describes the FDA 510(k) clearance for an Electric Wheelchair. However, it does not include information about acceptance criteria for a device performance study in the typical sense of a clinical or analytical performance study with specific metrics like sensitivity, specificity, or accuracy.

    Instead, the "acceptance criteria" for this device are demonstrated through adherence to a comprehensive set of international standards for wheelchairs. The study that proves the device meets these criteria is a series of non-clinical tests conducted against these standards.

    Here's an analysis based on the provided document:


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Electric Wheelchair are largely defined by compliance with a comprehensive set of ISO and IEC standards related to wheelchairs. The reported device performance is demonstrated by its successful adherence to these standards.

    Acceptance Criterion (Standard)Reported Device Performance
    Biocompatibility: ISO 10993-1: 2018 (Biological Evaluation of Medical Devices)Complies with the standard.
    Static Stability: ISO 7176-1:2014 (Determination of static stability)Complies with the standard.
    Dynamic Stability: ISO 7176-2:2017 (Determination of dynamic stability of electrically powered wheelchairs)Complies with the standard.
    Brake Effectiveness: ISO 7176-3:2012 (Determination of effectiveness of brakes)Complies with the standard. (Note: Subject device's braking distances are similar to or shorter than the predicate.)
    Energy Consumption/Range: ISO 7176-4:2008 (Energy consumption for determination of theoretical distance range)Complies with the standard. (Note: Travel distances are similar to the predicate.)
    Overall Dimensions, Mass, Maneuvering Space: ISO 7176-5:2008 (Determination of overall dimensions, mass and manoeuvring space)Complies with the standard. (Note: Specific values for dimensions and mass are provided and deemed acceptable despite differences.)
    Speed, Acceleration, Deceleration: ISO 7176-6:2018 (Determination of maximum speed, acceleration and deceleration)Complies with the standard. (Note: Max speeds are specified and deemed acceptable.)
    Seating & Wheel Dimensions: ISO 7176-7:1998 (Measurement of seating and wheel dimensions)Complies with the standard.
    Static, Impact & Fatigue Strengths: ISO 7176-8:2014 (Requirements and test methods for static, impact and fatigue strengths)Complies with the standard.
    Climatic Tests: ISO 7176-9:2009 (Climatic tests for electric wheelchairs)Complies with the standard.
    Obstacle-Climbing Ability: ISO 7176-10:2008 (Determination of obstacle-climbing ability)Complies with the standard. (Note: Obstacle climbing ability is specified and deemed acceptable.)
    Test Dummies: ISO 7176-11:2012 (Test dummies)Complies with the standard.
    Coefficient of Friction: ISO 7176-13:1989 (Determination of coefficient of friction of test surfaces)Complies with the standard.
    Power and Control Systems: ISO 7176-14:2008 (Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods)Complies with the standard.
    Information Disclosure, Documentation and Labeling: ISO 7176-15:1996 (Requirements for information disclosure, documentation and labeling)Complies with the standard.
    Resistance to Ignition of Postural Support Devices: ISO 16840-10: 2021 (Wheelchair seating - Part 10: Resistance to ignition of postural support devices)Complies with the standard; flame-retardant test carried out.
    Electromagnetic Compatibility: ISO 7176-21:2009 (Electromagnetic compatibility of electrically powered wheelchairs) & IEC 60601-1-2:2014+A1:2020 (Electromagnetic disturbances)Complies with the standard.
    Batteries and Chargers: ISO 7176-25:2013 (Batteries and chargers for powered wheelchairs) & IEC 62133-2: 2017 (Safety requirements for lithium systems)Complies with the standard.
    Maximum Safe Operational Incline: (Implicit in stability standards like ISO 7176-2)6° to 10° depending on model, deemed acceptable as stability is evaluated.
    Max Loading Weight: (Implicit in strength standards like ISO 7176-8)100kg, deemed acceptable as performance validated.

    Study Details:

    1. Sample size used for the test set and the data provenance:

      • The document states "Non-clinical tests were conducted to verify that the proposed device met all design specifications." This typically implies testing of a representative sample of the manufactured device models. Specific numbers for each test (e.g., how many units were tested for static stability) are not provided.
      • Data provenance: The tests were performed by the manufacturer, Ningbo Baichen Medical Devices Co., Ltd., which is located in China. The data would be considered retrospective in the sense that the testing was completed prior to the 510(k) submission, but it's prospective in relation to the design validation process.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable in the context of this device and study type. The "ground truth" for non-clinical performance (e.g., static stability, brake effectiveness) is defined by the technical specifications and limits set forth in the international standards (ISO, IEC). The testing involves objective measurements rather than expert interpretation of data like in an image-based diagnostic study.
    3. Adjudication method for the test set:

      • This question is not applicable. Adjudication typically refers to resolving discrepancies in expert interpretations (e.g., 2+1 means 2 out of 3 experts agree, 3+1 means 3 out of 4 agree, with a fourth to resolve ties). Since these are objective physical performance tests against established numerical standards, there is no expert adjudication process. The results are either within the standard's limits or they are not.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable. This is an Electric Wheelchair, a physical medical device, not an AI software intended for interpretation by human readers. Therefore, an MRMC study is not relevant.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • This is not applicable. This is a physical medical device. While it contains an "Electronic controller" (SYC-PM30), the performance described (e.g., speeds, stability, braking) is a property of the integrated physical device, not a standalone algorithm. Software validation was performed for the control system, as mentioned under the difference D#11, but this is part of the overall device's non-clinical testing, not a standalone algorithm performance study in the context of diagnostic AI.
    6. The type of ground truth used:

      • The "ground truth" for this device's performance is established by international consensus standards (ISO and IEC). These standards define the acceptable performance parameters and test methods for electric wheelchairs to ensure their safety and effectiveness.
    7. The sample size for the training set:

      • This question is not applicable. This is a physical medical device. There is no "training set" in the context of machine learning or AI algorithms for this type of device. The design and validation are based on engineering principles and compliance with established standards.
    8. How the ground truth for the training set was established:

      • This question is not applicable, as there is no training set for this device.

    In summary, the provided document details a non-clinical performance study demonstrating substantial equivalence through rigorous adherence to internationally recognized safety and performance standards for electric wheelchairs. The specific metrics are defined by these standards, and the device's performance met these criteria through objective physical testing.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1