    Intended Use

    Surgify Halo is a sterile surgical burr indciated for shaping and removal of hard tissue and bone in Neurosurgical, Spinal, ENT, and general surgical procedures

    Device Description

    The Surgify Halo is a sterile-packaged, single-use rotary cutting device (a burr) made from high-grade metallic materials, designed to cut bone and other hard tissues selectively while minimizing chattering. The device is similar to conventional surgical cutting burrs but includes a ring mechanism in the burr head to reduce chattering. Like traditional burrs, the device is designed to be compatible with commonly marketed, high-speed, surgical drill systems.

    The Surgify Halo comprises:

    • Shank: Transfers rotary motion from the drill motor to the burr head.
    • Locking Features: Located at the distal end of the shank, these features securely attach the burr to the high-speed surgical drill.
    • Burr Head: Shaped with two cutting edges for cutting of bone and hard tissue.
    • Safety Mechanism: A self-centering ring and spring mechanism installed in the burr head. The moving ring, larger than the burr head, can adopt two positions:
      • First Position: (Hard Tissue Mode) The ring protrudes, preventing the cutting edges from engaging with the work material.
      • Second Position: (Soft Tissue Mode) The ring extends beyond the cutting edge, shielding it to minimize accidental soft tissues and an exposed position for cutting hard surfaces like bone.
    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Surgify Halo) and does not contain information about an AI/ML component or a study proving its performance against acceptance criteria in the context of AI/ML. The device is a surgical burr, which is a physical cutting tool, not a software or AI-driven diagnostic or therapeutic device.

    Therefore, I cannot extract the requested information regarding acceptance criteria, study details for AI/ML performance, sample sizes, expert involvement, or ground truth establishment related to AI/ML.

    The "Performance Data" section describes functional testing for a physical device (durability, chattering, compatibility with drill systems) and sterilization/passivation validation, along with a table of "Design Verification Tests" and their result that "All tests fulfilled Design input requirements". This kind of performance data is standard for mechanical surgical tools and is not related to the performance of an AI/ML algorithm.