    Intended Use

    Visalys® Bulk Fill: Loss of tooth substance due to caries, dental wear or developmental disorders. Techniques: Class I and II restorations, non-tunneled class III restorations, filling in deciduous teeth, core buildups, cavity floor elevation, repair of direct restorations after appropriate pretreatment and splinting / locking of teeth.

    Visalys® Bulk Flow: Loss of tooth substance due to caries, fractures, dental wear or developmental disorders. Techniques: Class I and II restorations, non-tunneled class III restorations, filling in decidious teeth, core build-ups, cavity lining, cavity floor elevation, blocking out undercuts, repair of direct restorations after appropriate pretreatment, extended fissure sealing and splinting / locking of teeth.

    Device Description

    Visalys® Bulk Fill is a light-cured, nano-hybrid filling composite that can be modeled and is available in a universal shade. For adhesive cementation, a dentin-enamel bond is used. The material is available in screw syringes and can be applied in layers up to 4 mm thick. The filling material contains approx. 82% inorganic fillers by weight. The radiopacity is 2.5 mm aluminium*. The color stability meets the requirements of DIN EN ISO 4049. Visalys® Bulk Fill is thus suitable for both posterior teeth and also for fillings without tunnel preparation in the anterior tooth area, particularly class III and V.

    Visalys® Bulk Flow is a light-cured, flowable nano-hybrid filling composite and is available in a universal shade. For adhesive cementation, a dentin-enamel bond is used. The material is available in syringes and can be applied in layers up to 4 mm thick. The filling material contains approx. 72% inorganic fillers by weight. The radiopacity is 2.0 mm aluminium*. The color stability meets the requirements of DIN EN ISO 4049. Visalys® Bulk Flow is thus suitable for both posterior teeth and also for fillings without tunnel preparation in the anterior tooth area, particularly class III and V. Capping is not necessary.

    AI/ML Overview

    The provided text focuses on the substantial equivalence of dental restorative materials (Visalys® Bulk Fill and Visalys® Bulk Flow) to a predicate device. It does not describe a device that uses AI or machine learning, nor does it detail acceptance criteria related to typical AI/ML performance metrics such as sensitivity, specificity, or AUC.

    Therefore, I cannot extract the requested information regarding acceptance criteria and study details for an AI/ML device from this document. The document describes traditional performance testing for dental materials, primarily focusing on physical and chemical properties and biocompatibility.

    Here's an overview of the information present in the document that aligns with the context of a medical device submission, but not specifically for an AI/ML device:

    1. Table of acceptance criteria and reported device performance:

    The document lists performance tests conducted in accordance with various ISO standards for "Polymer-based restorative materials" (ISO 4049). It states that a "comparison of the parameters in respect to mechanical properties was performed. The results demonstrate the substantial equivalence to the predicate device." However, specific numerical acceptance criteria (e.g., "flexural strength > X MPa") and the reported numerical performance values for the subject devices are not explicitly stated in the provided text. It only mentions that the results demonstrate substantial equivalence.

    Acceptance Criteria CategoryStandard UsedReported Device Performance
    Sensitivity to lightISO 4049 / 7.9Results demonstrate substantial equivalence to the predicate device.
    Depth of cureISO 4049 / 7.10Results demonstrate substantial equivalence to the predicate device.
    Flexural strengthISO 4049 / 7.11Results demonstrate substantial equivalence to the predicate device.
    Elastic modulusISO 4049 / 7.11Results demonstrate substantial equivalence to the predicate device.
    Water sorptionISO 4049 / 7.12Results demonstrate substantial equivalence to the predicate device.
    SolubilityISO 4049 / 7.12Results demonstrate substantial equivalence to the predicate device.
    Shade of restorative materialsISO 4049 / 7.13 & ISO 7491 / ANSI ADA Spec. 80Results demonstrate substantial equivalence to the predicate device.
    Color stability after irradiation and water sorptionISO 4049 / 7.13 & ISO 7491 / ANSI ADA Spec. 80Results demonstrate substantial equivalence to the predicate device.
    Radio-opacityISO 4049 / 7.14 & ISO 13116Results demonstrate substantial equivalence to the predicate device.
    Volume shrinkageISO 17304Results demonstrate substantial equivalence to the predicate device.
    Shear bond strength to dentinISO 29022Results demonstrate substantial equivalence to the predicate device.
    Shear bond strength to restorative materialsISO 29022Results demonstrate substantial equivalence to the predicate device.
    Duration of light curingInternal test methodResults demonstrate substantial equivalence to the predicate device.
    Extrusion force (Bulk Flow only)Internal test methodResults demonstrate substantial equivalence to the predicate device.
    Consistency (Bulk Fill only)Internal test methodResults demonstrate substantial equivalence to the predicate device.
    High shear viscosity (Bulk Flow only)Internal test methodResults demonstrate substantial equivalence to the predicate device.
    Low shear viscosity (Bulk Flow only)Internal test methodsResults demonstrate substantial equivalence to the predicate device.
    Universal colorIn-vitro testResults demonstrate substantial equivalence to the predicate device.
    Biocompatibility (Cytotoxicity, Sensitization, Genotoxicity, Toxicological risk assessment, Chemical characterization)ISO 10993-5, ISO 10993-10, ISO 10993-3, ISO 10993-17, ISO 10993-18, ISO 7405Comparable to other materials on the market, no toxicological risks.

    2. Sample size used for the test set and the data provenance: Not applicable. The tests are bench tests of material properties, not clinical studies with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for material properties is established by standardized physical/chemical testing, not expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, this is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No, this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for the performance testing is based on established physical and chemical measurements defined by international ISO standards for dental materials.

    8. The sample size for the training set: Not applicable, as this is not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable, as this is not an AI/ML device.