510(k) Data Aggregation

    K Number
    K250113

    Validate with FDA (Live)

    Date Cleared
    2025-03-13

    (56 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fibers are indicated for use in general surgical applications for incision, vaporization, ablation, hemostasis or coagulation of soft tissue. It is also indicated for use in open or closed endoscopic applications where incision, excision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors or lesions, tissue vaporization, hemostasis and or coagulation may be indicated. The neoLaser Laser Surgery Fibers have a wavelength range of 450 nm to 2100 nm, can be used in contact mode and are indicated for use in general surgery, urology, gastroenterology, dermatology, vascular surgery, neurosurgery, plastic surgery, ENT and endovenous occlusion of the greater saphenous vein in the patient with superficial vein reflux and laser assisted lipolysis with a cleared compatible laser marketed for the mentioned intended uses and using an SMA 905 connector.

    Device Description

    The neoLaser Laser Surgery Fibers are sterile, single use, laser delivery devices intended for medical applications in various fields of laser surgery. They are intended to deliver energy to soft tissue in contact and non-contact mode during various surgical applications, including via endoscopes and cytoscopes. The fibers are designed to transmit energy from the laser system to the treatment site, as well as to transmit a low power laser aiming beam to assist in the visibilty of the target tissue.

    AI/ML Overview

    The provided text is a 510(k) summary for the neoLaser Laser Surgery Fibers. It describes a medical device and its substantial equivalence to a predicate device, BeaMed Laser Surgery Fibers (K232769). This type of submission relies on demonstrating that the new device is as safe and effective as a legally marketed predicate device, rather than requiring extensive clinical trials to prove efficacy from scratch.

    Therefore, the document does not outline acceptance criteria or a study proving the device meets those criteria in the context of an AI/algorithm-based medical device performance study. Instead, it focuses on demonstrating equivalence through comparison of technical characteristics and performance bench testing.

    Here's why the requested information cannot be extracted from this document:

    • The device is a physical medical instrument (laser surgery fibers), not an AI/algorithm-based device. The questions about MRMC studies, human reader improvement with AI, standalone algorithm performance, and ground truth establishment for training/test sets are relevant to AI/ML medical devices, not simple surgical tools.
    • The 510(k) pathway for this device is based on "substantial equivalence" to a predicate device. This means the manufacturer proves the new device is as safe and effective as an already cleared device through comparison of specifications, materials, and bench testing, not through new clinical performance studies against defined acceptance criteria for a novel technology.
    • The document explicitly states: "Animal or clinical studies: None." This confirms that performance studies in the way you've outlined for AI/ML models (with test sets, ground truth, expert readers, etc.) were not conducted or required for this 510(k) submission.

    However, I can extract information regarding the device's characteristics and the types of "performance" assessments that were deemed sufficient for this 510(k) clearance, even though they don't align with the typical AI/ML study framework you've described.


    Based on the provided document, here's an attempt to address the prompt's points, reinterpreting them for a physical medical device clearance based on substantial equivalence:

    1. A table of acceptance criteria and the reported device performance

    The "acceptance criteria" for a 510(k) submission for a physical device like this are primarily demonstrated through substantially equivalent technical specifications and material properties to a predicate device, along with successful bench testing to ensure safety and functionality. The document presents a comparative table rather than an acceptance criteria table.

    SpecificationSubject Device (neoLaser Laser Surgery Fibers)Predicate Device (BeaMed Laser Surgery Fibers)"Performance" / Outcome (Demonstrated Equivalence)
    Indications for UseGeneral surgical applications (incision, vaporization, ablation, hemostasis, coagulation of soft tissue, endoscopic applications, etc.). Wavelength: 450nm-2100nm. Contact/non-contact mode. Used in general surgery, urology, gastroenterology, gynecology, dermatology, vascular surgery, neurosurgery, plastic surgery, ENT, endovenous occlusion, laser assisted lipolysis. SMA 905 connector.Identical wording for Indications for Use.Met: Substantially equivalent indications for use.
    Fiber core materialSilica quartz glassSilica quartz glassMet: Identical material.
    Fiber cladding materialsilica quartz glass cladding and/or Hardcladsilica quartz glass cladding and/or HardcladMet: Identical material.
    Buffersilicone buffersilicone bufferMet: Identical material.
    Jacket materialsilicone / ETFE /polymide /nylon / acrylate / PVDF/ PU (with color additives)silicone / ETFE /polymide /nylon / acrylate / PVDF/ PU (with color additives)Met: Identical materials.
    Maximal temperatureNylon jacket: 176°F (80°C); ETFE jacket: 302°F (150°C)Nylon jacket: 176°F (80°C); ETFE jacket: 302°F (150°C)Met: Identical temperature tolerances.
    Minimal bend radius51 x core diameter (short term); 121 x core diameter (long term)51 x core diameter (short term); 121 x core diameter (long term)Met: Identical bend radius specifications.
    Maximal laser powerSuitable for laser power up to 100WSuitable for laser power up to 100WMet: Identical power handling.
    Shelf life5 years5 yearsMet: Identical shelf life.
    Numerical aperture range0.37 (for quartz/hardclad fibers); 0.22 (for quartz/quartz)0.37 (for quartz/hardclad fibers); 0.22 (for quartz/quartz)Met: Identical numerical aperture range.
    Core diameter range200 µm to 1000 µm200 µm to 1000 µmMet: Identical core diameter range.
    Outer diameter range240 µm to 1800 µm typically240 µm to 1800 µm typicallyMet: Identical outer diameter range.
    Distal fiber tip typesFlat, conical, ball, spherical or bended bare fibers, round or conical capsFlat, conical, ball, spherical or bended bare fibers, round or conical capsMet: Identical tip configurations.
    Fiber tip outer diameter range1.0mm to 1.8mm1.0mm to 1.8mmMet: Identical tip diameter range.
    Length range6.5 feet (2 meters) to 9.9 feet (3 meters)6.5 feet (2 meters) to 9.9 feet (3 meters)Met: Identical length range.
    Emission angle rangeStraight, forward directed cone or 60° cone angle from fiber axis, diffuse emission, and radial 360°Straight, forward directed cone or 60° cone angle from fiber axis, diffuse emission, and radial emission of 360°Met: Identical emission patterns.
    Single/multiple useSingle useSingle useMet: Identical use designation.
    SterilitySterile by EOSterile by EOMet: Identical sterilization method.
    PackagingSterile, double pouchedSterile, double pouchedMet: Identical packaging.
    Proximal endSMA 905 connectorSMA 905 connectorMet: Identical connector type.

    Performance Bench Testing included in the submission (serving as "proof"):

    • Packaging seal strength
    • Dye penetration test
    • Packaging microbial barrier
    • Shelf-life validation summary
    • Sterilization process validation plan
    • Sterilization process validation protocol
    • Sterilization process validation report
    • Sterility Assurance Level (SAL)
    • EO-ECH residuals report
    • LAL Endotoxin test
    • Bioburden bare fibers test / Bioburden capped fibers tests
    • Biocompatibility tests protocols & reports
    • Thermal safety validation

    The success of these tests (implicitly, meeting pre-defined thresholds for each) serves as the "reported device performance" that proves the device's safety and functionality for a 510(k) submission based on substantial equivalence.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This is for a physical device, and the submission explicitly states "Animal or clinical studies: None." The "tests" mentioned are bench tests on the device itself or its packaging/materials, not performance studies with human data. Therefore, concepts like sample size for a test set of patient data or data provenance (country, retrospective/prospective) are not relevant here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No "ground truth" was established in the context of an AI/ML algorithm's performance against human expert interpretations or pathology. The evaluation is based on engineering and material specifications, and physical bench testing results.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There was no test set requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI/ML algorithm that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an AI/ML algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" for this device's clearance is its conformity to engineering specifications, material standards, and successful completion of bench tests (e.g., sterilization efficacy, biocompatibility).

    8. The sample size for the training set

    Not applicable. There is no AI/ML training set for this physical device.

    9. How the ground truth for the training set was established

    Not applicable. There is no AI/ML training set or associated ground truth.