Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243710

    Validate with FDA (Live)

    Date Cleared
    2025-04-11

    (130 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Ureteral Access Sheath is used to establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract.

    Device Description

    The proposed device, Disposable Ureteral Access Sheath is used to establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract.

    This device is divided into five types:

    • Z type Straight joint access sheath
    • Y type negative suction Y joint access sheath
    • W type negative suction W joint access sheath
    • T type Straight joint access sheath
    • C type negative suction W joint access sheath

    This product consists of access sheath and dilator tube. The access sheath tube consists of access sheath tube body and access sheath joint. The surface of access sheath tube is coated with polyvidone (PVP). The dilator tube consists of dilator tube body and dilator tube connector. The access sheath tube body is made from nylon (PA), stainless steel and polytetrafluoroethylene (PTFE). The access sheath joint is made from polycarbonate (PC), acrylonitrile-butadiene-styrene copolymer (ABS) and silicone. The dilator tube body and dilator tube connector are made from PE.

    The proposed devices are sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of 3 years.

    AI/ML Overview

    Based on the provided FDA 510(k) clearance letter for the Disposable Ureteral Access Sheath (K243710), the device is cleared based on non-clinical performance and biocompatibility testing, demonstrating substantial equivalence to a predicate device. There is no information about a study involving human readers or AI assistance.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" alongside specific numerical "reported device performance" values for each criterion. Instead, it lists the types of performance tests conducted and states that the results "met the standard requirements" or that certain characteristics "shall not exceed" or "shall not be less than" a specified value, implying these are the acceptance criteria.

    Implied Acceptance Criteria and Performance (Based on "Comparison of Technological Characteristics" and "Non-Clinical Testing" sections):

    Acceptance Criteria (Implied)Reported Device Performance (Summary)
    Whole Device
    AppearanceMet requirements (implied by "Performance testing")
    DimensionMet requirements (implied by "Performance testing")
    Access Sheath Tube
    Liquid leakageMet requirements (implied by "Performance testing")
    Peak tensile force≥ 15 N (at tube body, and at union of tube body and joint)
    Friction coefficient≤ 0.03
    ToughnessMet requirements (implied by "Performance testing")
    Resistance To FlatteningMet requirements (implied by "Performance testing")
    Dilator Tube
    Strength of union of the dilator tube connector and dilator tubeMet requirements (implied by "Performance testing")
    Resistance to flexingMet requirements (implied by "Performance testing")
    CompatibilityMet requirements (implied by "Performance testing")
    TrafficabilityMet requirements (implied by "Performance testing")
    Dilator Tube ConnectorMet requirements (implied by "Performance testing")
    Biocompatibility
    CytotoxicityMet requirements (in accordance with ISO 10993-1)
    SensitizationMet requirements (in accordance with ISO 10993-1)
    IrritationMet requirements (in accordance with ISO 10993-1)
    Acute Systemic ToxicityMet requirements (in accordance with ISO 10993-1)
    PyrogenicityMet requirements (in accordance with ISO 10993-1)
    Sterility & Shelf-life
    Sterility Assurance Level (SAL)10^-6 (achieved via EO sterilization)
    Shelf Life3 years (maintained sterility)
    Bending ResistanceNo deformations, cracks, folding, or other defective phenomena after 5cm radius bend for 1min

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes for each non-clinical performance test. It only lists the types of tests performed.
    • Data Provenance: The testing was conducted as part of the regulatory submission by Shenzhen Trious Medical Technology Co., Ltd. The document refers to "Performance testing" and "Biocompatibility testing," which are typically done in a controlled laboratory setting. It does not mention retrospective or prospective data in the context of human studies for this device. The manufacturer is based in Shenzhen, China.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This request is not applicable to the provided document. The "ground truth" for this device's testing is based on engineering specifications, material science, and established biological safety standards (e.g., ISO and ASTM standards), not on expert clinical interpretation of medical images or data from human patients. Therefore, no human experts (e.g., radiologists) were involved in establishing "ground truth" in the way understood in diagnostic AI/image analysis studies.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods are typically used in clinical studies or expert review processes where there might be disagreements in interpretation (e.g., blinded reads, consensus reads). For non-clinical device performance testing, results are typically objective measurements against predefined specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The document explicitly states: "No clinical study is included in this submission." Therefore, no MRMC study or AI assistance evaluation was performed or submitted.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This device is a physical medical instrument (ureteral access sheath), not an AI algorithm or software. Therefore, the concept of "standalone algorithm performance" is not applicable.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is established by:

    • Engineering Specifications: Defined physical dimensions, material properties, and mechanical strengths.
    • Standardized Test Methods: Adherence to recognized international and national standards (e.g., ISO 11135, ISO 11737-1, ISO 10993-1 for biocompatibility, various ASTM standards for material properties). These standards define how performance is measured and what constitutes acceptable limits.
    • Biocompatibility Definitions: Established biological responses to materials as defined by ISO 10993-1.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical device, not an AI model or software that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1