Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243693

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-12-23

    (24 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NEO Pedicle Screw System™

    The NEO Pedicle Screw System™, when used as a posterior pedicle screw system, is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion. The system is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (scoliosis, kyphosis), tumor, pseudarthrosis, and/or failed previous fusion.

    When used in conjunction with BonOs® Inject Cement, the NEO Pedicle Screw System™ is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. NEO Pedicle Screws augmented with BonOs® Inject Cement are for use with 5 mm to 8 mm screw diameters at spinal levels (thoracic and lumbar T1 to L5) where the structural integrity of the spine is not severely compromised.

    BonOs® Inject

    BonOs® Inject bone cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using vertebroplasty or balloon kyphoplasty procedure.

    When used in conjunction with NEO Pedicle Screw System™ BonOs® Inject is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. NEO Pedicle Screws augmented with BonOs® Inject Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.

    Device Description
    1. BonOs® Iniect

    BonOs® Iniect is a radiopaque, iniectable bone cement for use in spine surgery like percutaneous vertebral augmentation during vertebroplasty or kyphoplasty. It is a twocomponent system consisting of a powder and a liguid. Methylmethacrylate polymer is the primary constituent of the powder component. Zirconium dioxide is added as radiopacifier. Methylmethacrylate monomer is the primary constituent of the liquid component. Mixing the two separate sterile components, initially an injectable paste is produced which can be transferred into a syringe and which then can be injected under slight pressure into the vertebral body. After curing of the bone cement by exothermic polymerization it stabilizes the vertebral lesions and vertebral compression fractures.

    1. NEO Pedicle Screw System™

    The NEO Pedicle Screw System™ has been optimized and consists of 27 pedicle and 8 iliac screws, rods and connectors which are available in different sizes. The system includes a reduced optimized trav of instruments providing options for open and MIS surgical techniques. The implants are all delivered sterile and ready to use. The relevant instruments are mainly delivered as single use, disposable and delivered sterile, just a few optional instruments are reusable and delivered non-sterile.

    All the system components are made of materials compliant with ASTM and/or ISO standards. Screws and rods are made of titanium-alloy (Ti-6AI-4VELI) and comply with ISO 5832-3 or ASTM-F136. The screws are made out of a titanium alloy and delivered pre-mounted to a screw extender including a tissue dilator and sterile. The rods are made out of a titanium alloy or cobalt chrome alloy CoCrMo (ISO 5832-12, ASTM F1537) and delivered sterile, Connectors are made out of titanium alloy and delivered sterile.

    The pedicle screws are offered in diameters of 4.5 – 8.0 mm and lengths of 25 - 55 mm. Illiac screws are offered in diameters of 8.0 mm and 10.0 mm and lengths of 70 – 100 mm. Three different types of rods are available: pre-bent, straight or special bent rod for S1/L5. All rods have a diameter of 5.5 mm. Pre-bent rods are offered in lengths of 35 – 100 mm, straight rods in lengths from 30 - 500 mm and the special-bent rod in either 30 or 40 mm length.

    The NEO Pedicle Screw System™ provides immobilization and stabilization of spinal seqments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine. The system can be used via an open or minimally invasive posterior approach. lliac screws shall be placed via Sacral Alar Iliac (SAI) and/or Posterior Superior Illiac Spine (PSIS) approach. The size and form of the device is adjusted to the morphology of the body and the surgical technique.

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter and a 510(k) summary for a medical device called the "NEO Pedicle Screw System™" and "BonOs® Inject Bone Cement." This document details the administrative information, device description, indications for use, and a summary of performance data to support substantial equivalence to a predicate device.

    However, the document specifies that "No clinical studies were conducted." This means that the information requested regarding acceptance criteria for a study proving device performance (especially for AI/algorithm-based devices), sample size for test and training sets, expert involvement, adjudication methods, MRMC studies, standalone algorithm performance, and ground truth establishment cannot be provided from this document.

    The performance data mentioned in the document are limited to non-clinical testing, primarily focusing on mechanical performance, biological evaluation, and MRI compatibility. These types of tests are used to demonstrate that changes to the device (like new components or size extensions) do not negatively impact its existing, cleared performance characteristics, and that it is substantially equivalent to the predicate device.

    Therefore, I cannot populate the requested table or answer the specific questions related to the acceptance criteria and study that proves the device meets the acceptance criteria in the context of an AI/algorithm-driven device's performance study, as this submission is for a physical orthopedic implant system, not a software or AI product requiring such a study.

    Here's what can be inferred from the document regarding "acceptance criteria" and "proof" in the context of this device:

    The "acceptance criteria" for the modifications introduced in this 510(k) submission are implicitly met if the device demonstrates substantial equivalence to its predicate device. This is primarily demonstrated through:

    • Mechanical Performance: Meeting the requirements for pedicle screw spinal systems according to relevant ASTM and ISO standards (e.g., ASTM F1717 for dynamic compression bending). The document states: "Non-clinical performance testing demonstrate that the NEO Pedicle Screw System™ meets the requirements for Pedicle screw spinal systems according to Spinal System 510(k)s Guidance for Industry and FDA Staff Document issued on: May 3, 2004 and is substantially equivalent to its predicate devices."
    • Biological Evaluation: Compliance with ISO 10993-1 series for the new components.
    • MRI Compatibility: Meeting standards like ASTM F2052-15, ASTM F2182-11a, and ASTM F2119-07 (R2013).
    • Functionality: Internal validation confirming new components function as intended.

    Table of Acceptance Criteria (Implicit for a Mechanical Device) and Reported Performance:

    Acceptance Criteria CategorySpecific Acceptance Criteria (Implicit from Submission)Reported Device Performance (as stated in the document)
    Mechanical PerformanceMeets requirements of relevant standards (e.g., ASTM F1717 for dynamic compression bending for spinal systems)."Finite element analysis and engineering rationales were performed to justify that the subject components do not represent a new worst case for mechanical performance. Additionally, the new axial connector was included in a construct tested according to ASTM F1717 for dynamic compression bending." "Mechanical testing also confirmed that the conical press-fit connection between the ball attachment and the 10.0 mm diameter iliac screw thread is sufficiently strong through measurement of the torque resistance of the assembled screw." "Results of these evaluations demonstrate substantially equivalent mechanical performance."
    BiocompatibilityNew components comply with ISO 10993-1 series."Biological evaluation has been performed for the new 10.0 mm diameter iliac screws of the modified NEO Pedicle Screw System™ in accordance with the ISO 10993-1 series."
    Cleaning & SterilizationValidation performed for new components."A new cleaning and sterilization validation has been performed for the new 10.0 mm diameter iliac screws."
    MRI CompatibilityEvaluated against ASTM F2052-15, F2182-11a, F2119-07 (R2013)."The new system components have been evaluated in regards to Magnetic field interactions ASTM F2052-15, MRI-related heating ASTM F2182-11a and Artifacts ASTM F2119-07 (R2013). Non-clinical testing demonstrated that the entire family of the NEO Pedicle Screw System™ is MRI conditional. The previous conclusions and MR conditional K222256 remain valid."
    Functional Performance (Internal)New axial connectors function as intended."An internal validation of the new axial connectors of the modified NEO Pedicle Screw System™ was performed and confirmed they function as intended."
    Overall EquivalenceDemonstrated substantial equivalence to the predicate device in terms of intended use, material, design, and function."Based on the indications for use, technological characteristics, and comparison to predicate device, the subject modified BonOs® Inject Bone Cement, NEO Pedicle Screw System™ has been shown to be substantially equivalent to the currently cleared predicate device."

    Regarding the specific questions about an AI/algorithm performance study:

    1. Sample size for the test set and data provenance: Not applicable. No clinical or external algorithm performance test set was used.
    2. Number of experts used to establish the ground truth and qualifications: Not applicable.
    3. Adjudication method for the test set: Not applicable.
    4. MRMC comparative effectiveness study: Not applicable.
    5. Standalone (algorithm only without human-in-the-loop performance) study: Not applicable.
    6. Type of ground truth used: Not applicable. Ground truth for a physical device is often demonstrated by meeting design specifications and performance under defined test conditions (e.g., material strength and fatigue testing).
    7. Sample size for the training set: Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document is for a traditional mechanical medical device, not an AI/ML powered one. Therefore, the specific criteria and study details requested, which are pertinent to AI/ML device evaluations, are not present in this submission. The "proof" for this device's acceptance rests on its demonstration of substantial equivalence to a predicate device through non-clinical laboratory testing.

    Ask a Question

    Ask a specific question about this device