510(k) Data Aggregation

    K Number
    K243522

    Validate with FDA (Live)

    Date Cleared
    2025-05-30

    (197 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ViVi® Toga Premium is a component of the ViVi® System and is intended to be worn by surgical personnel to provide a barrier between the operating environment and the surgical personnel in order to help protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.

    Device Description

    The ViVi® Toga Premium is a one-piece head and body cover that is worn by healthcare professionals. The ViVi® Toga Premium is a component of the ViVi® Surgical Helmet System and worn over the air-exchanging surgical helmet. The ViVi® Toga Premium is used with the ViVi® Helmet or the ViVi® Helmet HPL (High Power LED) with Light Emitting Diode (LED) light to provide a barrier between the operating environment and the surgical personnel to protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.

    ViVi® Toga Premium is designed and has been tested to meet the applicable AAMI PB70 standards for level 4 compliance. The AAMI standard does not cover apparel for the head, face, and eyes. Therefore, the hood area and the lens are exempt from classification under this standard.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the ViVI® Toga Premium is for a Class II surgical apparel device. This document primarily details the substantial equivalence of the ViVi® Toga Premium to a predicate device based on non-clinical performance data. It explicitly states that "No clinical evaluation is necessary for this device."

    Therefore, this document does not describe a study involving an AI/Machine Learning (ML) algorithm with a test set, ground truth, expert adjudication, or MRMC studies. The device is a physical product (surgical toga/gown), and its "performance" is related to its physical properties and barrier capabilities, not diagnostic or predictive accuracy.

    Given the information provided, it's not possible to fulfill the request for acceptance criteria and study details related to an AI/ML device. The "study" mentioned in the document refers to non-clinical laboratory testing of material properties.

    However, I can extract the acceptance criteria and the "reported device performance" based on the provided document as it relates to this specific physical device.


    Acceptance Criteria and Reported Device Performance for ViVi® Toga Premium (Surgical Apparel)

    Based on the provided FDA 510(k) clearance letter for the ViVi® Toga Premium, the acceptance criteria and reported device performance relate to its physical properties and barrier capabilities, as it is a surgical apparel device and not an AI/ML device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the standards and tests performed, with "PASS" indicating successful compliance.

    Property or CharacteristicTest Method / StandardAcceptance Criteria (Implied)Reported Device Performance (ViVi® Toga Premium)
    Flammability of clothing textiles16 CFR 1610Class 1 compliantClass 1 compliant PASS
    Biocompatibility: CytotoxicityISO 10993-5CompliantCompliant PASS
    Biocompatibility: SensitizationISO 10993-10CompliantCompliant PASS
    Biocompatibility: IrritationISO 10993-23CompliantCompliant PASS
    Tear ResistanceASTM D5587CompliantCompliant PASS
    Tensile StrengthASTM D5034CompliantCompliant PASS
    Seam StrengthASTM F88/F88MCompliantCompliant PASS
    Laser ResistanceISO 11810:2015CompliantCompliant PASS
    Linting (Lint and other particles generation in the dry state)ISO 9073-10:2005CompliantCompliant PASS
    Water Vapor Transmission rateASTM E96/E96M-24ACompliantCompliant PASS
    Barrier Performance (Liquid)ANSI/AAMI PB70:2012; ASTM F1671/F1671MLevel 4 compliantCompliant Level 4 PASS
    Sterilization Assurance Level (SAL)Not explicitly detailed by test method but stated as requirement10⁻⁶10⁻⁶

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes (number of garments or material samples) used for each individual non-clinical test. The testing is reported as "tests were conducted and completed," indicating standard laboratory evaluations.
    • Data Provenance: Not applicable in the context of clinical data. The tests are material and performance tests conducted in a laboratory setting. No geographical origin of "data" in the sense of patient data is mentioned, as this is a physical product. The manufacturer is THI Total Healthcare Innovation GmbH, based in Austria. The predicate device manufacturer is Zimmer Surgical, Inc.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not applicable. This device is surgical apparel, and its "performance" is assessed through standardized physical and biological material testing, not through expert human interpretation of data like images or clinical outcomes. There is no "ground truth" in the diagnostic AI sense.

    4. Adjudication Method for the Test Set

    • Not applicable. As described above, there is no expert adjudication process for this type of device. Performance is determined by meeting pre-defined regulatory standards and test method specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No. An MRMC study is a type of clinical study used to evaluate the performance of diagnostic devices (often imaging-based AI) by comparing the accuracy of multiple human readers with and without AI assistance across multiple cases. This does not apply to a surgical apparel device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study was Done

    • Not applicable. This is not an AI algorithm.

    7. The Type of Ground Truth Used

    • Not applicable in the context of AI/ML. The "ground truth" for this device's performance is derived from established international and national standards (e.g., ISO, ASTM, ANSI/AAMI, CFR), which define objective physical and chemical properties and test methodologies. Meeting these standards serves as the "truth" for device functionality.

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI/ML device requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. This is not an AI/ML device and does not have a training set or associated ground truth.