510(k) Data Aggregation

    K Number
    K243222

    Validate with FDA (Live)

    Date Cleared
    2024-11-06

    (30 days)

    Product Code
    Regulation Number
    866.4001
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BIOFIRE FILMARRAY Pneumonia Panel (BIOFIRE Pneumonia Panel) is a multiplexed nucleic acid test intended for use with BIOFIRE FILMARRAY 2.0 (BIOFIRE 2.0) or BIOFIRE FILMARRAY TORCH (BIOFIRE TORCH) systems for the simultaneous detection of multiple respiratory viral and bacterial nucleic acids, as well as select antimicrobial resistance genes, in sputum-like speciorated sputum, or endotracheal aspirates) or bronchoalveolar lavage (BAL)-like specimens (BAL) obtained from individuals suspected of lower respiratory tract infection.

    The following bacteria are reported semi-quantitatively with bins representing approximately 10^4, 10^5, or ≥10°7 genomic copies of bacterial nucleic acid per milliliter (copies/mL) of specimen, to aid in estimating relative abundance of nucleic acid from these common bacteria within a specimen:

    Bacteria reported with bins of 10^4, 10^5, 10^6, or ≥10^7 copies/mL

    • · Acinetobacter calcoaceticus-baumannii complex
    • · Klebsiella oxytoca
    • · Serratia marcescens
    • · Enterobacter cloacae complex
    • Klebsiella pneumoniae group
    • · Staphylococcus aureus
    • · Escherichia coli
    • · Moraxella catarrhalis
    • · Streptococcus agalactiae
    • Haemophilus influenzae
    • · Proteus spp.
    • · Streptococcus pneumoniae
    • Klebsiella aerogenes
    • Pseudomonas aeruginosa
    • · Streptococcus pyogenes

    The following atypical bacteria, viruses, and antimicrobial resistance genes are reported qualitatively:

    Atypical Bacteria

    • Chlamydia pneumoniae
    • · Legionella pneumophila
    • Mycoplasma pneumoniae

    Viruses

    • · Adenovirus
    • Human rhinovirus/enterovirus
    • · Parainfluenza virus
    • · Coronavirus
    • · Influenza A virus
    • Respiratory syncytial virus

    • Human metapneumovirus

    • Influenza B virus
      Antimicrobial Resistance Genes

    • · CTX-M

    • IMP

    • КРС

    • NDM

    • OXA-48-like

    • VIM

    • · mecA/C and MREJ (MRSA)

    The detection and identification of specific viral and bacterial nucleic acids, as well as the estimation of relative abundance of nucleic acid from common bacterial analytes, within specimens collected from individuals exhibiting signs and/or symptoms of a respiratory infection, aids in the diagnosis of lower respiratory infection with other clinical and epidemiological information. The results of this test should not be used as for diagnosis, treatment, or other patient management decisions.

    Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, pathogens below the limit of detection, or in the case of bacterial analytes, present at levels below the lowest reported 10^4 copies/mL bin. Detection of analytes does not rule out co-infection with other organisms; the agent(s) detected by the BIOFIRE Pneumonia Panel may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible lower respiratory tract infection.

    Detection of bacterial nucleic acid may be indicative of colonizing or normal respiratory flora and may not indicate the causative agent of pneumonia. Semi-quantitative Bin (copies/mL) results generated by the BIOFIRE Pneumonia Panel are not equivalent to CFU/mL and do not consistently correlate with the quantity of bacterial analytes compared to CFU/mL. For specimens with multiple bacteria detected, the relative abundance of nucleic acids (copies/mL) may not correlate with the relative abundance of bacteria as determined by culture (CFU/mL). Clinical correlation is advised to determine significance of semi-quantitative Bin (copies/mL) for clinical management.

    The antimicrobial resistance gene detected may or may not be associated with the agent(s) responsible for disease. Negative results for these antimicrobial resistance gene assays do not indicate susceptibility to corresponding classes of antimicrobials, as multiple mechanisms of antimicrobial resistance exist.

    Antimicrobial resistance can occur via multiple mechanisms. A "Not Detected" result for a genetic marker of antimicrobial resistance does not indicate susceptibility to associated antimicrobial drugs or drug classes. A "Detected" result for a genetic marker of antimicrobial resistance cannot be definitively linked to the microorganism(s) detected. Culture is required to obtain isolates for antimicrobial susceptibility testing, and BIOFIRE Pneumonia Panel results should be used in conjunction with culture results for determination of bacterial susceptibility or resistance.

    Due to the genetic similarity between human rhinovirus and enterovirus, the test cannot reliably differentiate them. A positive Rhinovirus/Enterovirus result should be followed up using an alternate method (e.g., cell culture or sequence analysis) if differentiation is required.

    Culture is required to identify pathogens not detected by the BIOFIRE Pneumonia Panel, to further speciate analytes in genus, complex, or group results if desired, to identify bacterial pathogens present below the 10°4 copies/mL bin if desired, and for antimicrobial susceptibility testing.

    BIOFIRE FILMARRAY Pneumonia Panel plus:

    The BIOFIRE FILMARRAY Pneumonia Panel plus (BIOFIRE Pneumonia Panel plus) is a multiplexed nucleic acid test intended for use with BIOFIRE FILMARRAY 2.0 (BIOFIRE 2.0) or BIOFIRE FILMARRAY TORCH (BIOFIRE TORCH) systems for the simultaneous detection and identification of nucleic acids from Middle East respiratory syndrome coronavirus (MERS-CoV) and multiple respiratory viral and bacterial nucleic acids, as well as select antimicrobial resistance genes, in sputum-like specimens (induced or expectorated sputum, or endotracheal aspirates) or bronchoalveolar lavage (BAL)-like specimens (BAL or mini-BAL) obtained from individuals meeting MERS-CoV clinical and/or epidemiological criteria.

    Testing with BIOFIRE Pneumonia Panel plus should not be performed unless the patient meets clinical and/or epidemiologic criteria for testing suspected MERS-CoV specimens. Thical signs and symptoms assocated with MERS-CoV infection, contact with a probable or confirmed MERS-CoV case, history of travel to geographic locations where MERS-CoV cases were detected, or other epidemiological links for which MERS-CoV testing may be indicated.

    The following bacteria are reported semi-quantitatively with bins representing approximately 10^4, 10^5, or ≥10°7 genomic copies of bacterial nucleic acid per milliliter (copies/mL) of specimen, to aid in estimating relative abundance of nucleic acid from these common bacteria within a specimen:

    Bacteria reported with bins of 10^4, 10^5, 10^6, or ≥10^7 copies/mL

    • Acinetobacter calcoaceticus-baumannii complex
    • Enterobacter cloacae complex
    • Escherichia coli
    • Haemophilus influenzae
    • Klebsiella aerogenes
    • · Klebsiella oxytoca
    • · Klebsiella pneumoniae group
    • Moraxella catarrhalis
    • Proteus spp.
    • Pseudomonas aeruginosa
    • · Serratia marcescens
    • Staphylococcus aureus
    • Streptococcus agalactiae
    • · Streptococcus pneumoniae
    • · Streptococcus pyogenes

    The following atypical bacteria, viruses, and antimicrobial resistance genes are reported qualitatively: Atypical Bacteria

    • Chlamydia pneumoniae
    • · Legionella pneumophila
    • Mycoplasma pneumoniae

    Viruses

    • · Middle East respiratory syndrome coronavirus (MERS-CoV)
    • Adenovirus
    • Coronavirus
    • Human metapneumovirus
    • Human rhinovirus/enterovirus
    • · Influenza A virus
    • Influenza B virus
    • Parainfluenza virus
    • · Respiratory syncytial virus

    Antimicrobial Resistance Genes

    • CTX-M

    • IMP

    • · KPC

    • NDM

    • OXA-48-like

    • VIM

    • · mecA/C and MREJ (MRSA)

    The detection and identification of specific viral and bacterial nucleic acids from MERS-CoV and other respiratory pathogens, as well as the estimation of relative abundance of nucleic acid from common bacterial analytes, within specimens collected from individuals meeting MERS-CoV clinical and/or epidemiological criteria aids in the differential diagnosis of MERS-CoV infection, if used in conjunction with other clinical and epidemiological information in accordance with the guidelines provided by the appropriate public health authorities.

    BIOFIRE Pneumonia Panel plus MERS-CoV positive results are for the presumptive identification of MERS-CoV. The definitive identification of MERS-CoV requires additional testing and confirmation procedures in consultation with the appropriate public health authorities (e.g., local or state public health departments, etc.) for whom reporting is necessary. The diagnosis of MERS-CoV infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence in addition to the identification of MERS-CoV.

    BIOFIRE Pneumonia Panel plus MERS-CoV negative results, even in the context of a BIOFIRE Pneumonia Panel plus positive result for one or more of the common respiratory pathogens, do not preclude MERS-CoV infection and should not be used as the sole basis for patient management decisions. The levels of MERS-CoV that would be present in sputum-like or BAL-like specimens from individuals with early infection and from asymptomatic MERS-CoV carriers are not well understood. A negative BIOFIRE Pneumonia Panel plus MERS-CoV result in an asymptomatic individual does not rule out the possibility of future illness and does not demonstrate that the individual is not infectious.

    Viral culture should not be attempted on specimens with positive BIOFIRE Pneumonia Panel plus results for MERS-CoV unless a BSL 3 facility is available to receive and culture specimens.

    Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, pathogens below the limit of detection, or in the case of bacterial analytes, present at levels below the lowest reported 10^4 copies/mL bin. Detection of analytes does not rule out co-infection with other organisms; the agent(s) detected by the BIOFIRE Pneumonia Panel plus may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible lower respiratory tract infection.

    Detection of bacterial nucleic acid may be indicative of colonizing or normal respiratory flora and may not indicate the causative agent of pneumonia. Semi-quantitative Bin (copies/mL) results generated by the BIOFIRE Pneumonia Panel plus are not equivalent to CFU/mL and do not consistently correlate with the quantity of bacterial analytes compared to CFU/mL. For specimens with multiple bacteria detected, the relative abundance of nucleic acids (copies/mL) may not correlate with the relative abundance of bacteria as determined by culture (CFU/mL). Clinical correlation is advised to determine significance of semi-quantitative Bin (copies/mL) for clinical management.

    The antimicrobial resistance gene detected may or may not be associated with the agent(s) responsible for disease. Negative results for these antimicrobial resistance gene assays do not indicate susceptibility to corresponding classes of antimicrobials, as multiple mechanisms of antimicrobial resistance exist.

    Antimicrobial resistance can occur via multiple mechanisms. A "Not Detected" result for a genetic marker of antimicrobial resistance does not indicate susceptibility to associated antimicrobial drugs or drug classes. A "Detected" result for a genetic marker of antimicrobial resistance cannot be definitively linked to the microorganism(s) detected. Culture is required to obtain isolates for antimicrobial susceptibility testing, and BIOFIRE Pneumonia Panel plus results should be used in conjunction with culture results for determination of bacterial susceptibility or resistance.

    Due to the genetic similarity between human rhinovirus and enterovirus, the test cannot reliably differentiate them. A positive Rhinovirus/Enterovirus result should be followed up using an alternate method (e.g., cell culture or sequence analysis) if differentiation is required.

    Culture is required to identify pathogens not detected by the BIOFIRE Pneumonia Panel plus, to further speciate analytes in genus, complex, or group results if desired, to identify bacterial pathogens present below the 10°4 copies/mL bin if desired, and for antimicrobial susceptibility testing.

    Device Description

    The BIOFIRE® FILMARRAY® Pneumonia Panel and BIOFIRE® FILMARRAY® Pneumonia Panel plus use nested, multiplex reverse transcription polymerase chain reaction (PCR), followed by melting curve analysis for the detection of select organisms and antimicrobial resistance (AMR) genes in sputum-like (induced and expectorated sputum as well as endotracheal aspirate, ETA) and bronchoalveolar lavage (BAL)-like (BAL and mini-BAL) specimens. The panels allow for the identification of specific bacteria, atypical bacteria, viruses, and AMR genes as indicated in Table 1. The BIOFIRE Pneumonia Panel and BIOFIRE Pneumonia Panel plus pouches are identical, but the BIOFIRE Pneumonia Panel plus includes reporting of Middle East Respiratory Syndrome Coronavirus (MERS-CoV), which is not included in the BIOFIRE Pneumonia Panel. Reporting of MERS-CoV is controlled through software masking of the MERS-CoV result for the BIOFIRE Pneumonia Panel.

    The BIOFIRE Pneumonia Panels are compatible with bioMérieux's PCR-based in vitro diagnostic BIOFIRE® FILMARRAY® 2.0 and BIOFIRE® FILMARRAY® TORCH Systems for infectious disease testing. Specific software module (i.e. BIOFIRE Pneumonia Panel Pouch Module Software) are used to perform BIOFIRE Pneumonia Panels testing on these systems.

    AI/ML Overview

    This document refers to a 510(k) premarket notification for a medical device (BIOFIRE FILMARRAY Pneumonia Panel and BIOFIRE FILMARRAY Pneumonia Panel plus). This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a full de novo study with strict acceptance criteria and performance validation against a test set. The document clearly states that the submission is for software updates to mitigate false positive Coronavirus and CTX-M results and that "Reanalysis of the performance data with the modified pouch module software did not result in an overall change of the study conclusions or performance claims for non-clinical/analytical studies."

    Therefore, the information typically requested in your prompt regarding acceptance criteria, study details, sample sizes, expert ground truth establishment, MRMC studies, and standalone performance might not be explicitly detailed in this type of FDA submission as it would be for a de novo marketing authorization. However, I can extract what is available and clarify what is not.

    Based on the provided text, here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state formal "acceptance criteria" in a quantitative table format as might be seen for a new device submission. Instead, the focus is on the impact of the software update on existing performance. The relevant performance change mentioned is:

    Performance MetricPrevious Performance (without software update)Reported Performance (with software update)
    Clinical specificity/NPA of Coronavirus assay in BAL specimens98.4%98.7%
    Clinical specificity/NPA of Coronavirus assay in Sputum specimens99.3%99.5%

    The document implies that these updated specificities are acceptable because they represent an improvement in mitigating false positives and "did not result in an overall change of the study conclusions or performance claims for non-clinical/analytical studies."

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions a "Clinical Prospective Study" for which the Coronavirus specificity numbers are reported. It does not provide the exact sample size for this specific study, nor does it explicitly state the country of origin. It indicates that the reanalysis of existing performance data was done.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. As this is a molecular diagnostic test, ground truth would typically be established by highly sensitive and specific laboratory methods (e.g., PCR, sequencing, culture) rather than expert human interpretation in the way radiologists interpret images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study is not applicable here as this is a molecular diagnostic device, not an AI-assisted diagnostic imaging device that involves human reader interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the assay itself. The document implicitly discusses the "standalone" performance of the BIOFIRE FILMARRAY Pneumonia Panel and Panel Plus, which is a molecular diagnostic test. The reported specificities are a measure of this standalone performance. The software update is an internal modification to the assay's interpretation logic.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The document implies that the ground truth for the clinical performance evaluations was established through highly sensitive and specific methods for pathogen detection, as is standard clinical laboratory practice for molecular diagnostics. It does not explicitly state the specific ground truth methods but mentions that "Culture is required to obtain isolates for antimicrobial susceptibility testing, and BIOFIRE Pneumonia Panel results should be used in conjunction with culture results for determination of bacterial susceptibility or resistance." This suggests that culture and other definitive laboratory tests would be part of the ground truth establishment, particularly for bacterial analytes and antimicrobial resistance genes.

    8. The sample size for the training set

    This document describes a software update to an already cleared device. It does not provide details about a "training set" in the context of machine learning model development. The software update appears to be a rule-based or algorithmic adjustment to optimize melting curve analysis and mitigate cross-reactivity with human genomic DNA, rather than a re-training of a complex AI model. The modification was driven by "routine post-market monitoring and complaint investigations combined with concurrent findings from an internal product development study."

    9. How the ground truth for the training set was established

    As there's no mention of a traditional "training set" in the context of an AI model, this information is not provided. The "ground truth" that informed the software change was likely observations of false positives in clinical samples, identified through investigations and potentially confirmed by orthogonal testing or characterization of the offending interactions (cross-reactivity with hgDNA).