510(k) Data Aggregation

    K Number
    K243209

    Validate with FDA (Live)

    Device Name
    nordicMEDiVA
    Manufacturer
    Date Cleared
    2024-10-22

    (21 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The nordicMEDIVA software is an advanced visualization and processing platform with a specific focus on providing algorithms designed to analyze functional and dynamic MRI data of the brain. The software runs on a server in a networked environment and is accessed by users via a standard web browser. It can communicate with other imaging platforms that support DICOM, and process medical image data acquired through DICOM-compliant imaging devices and modalities.

    nordicMEDIVA is indicated for image analysis and visualization of functional and dynamic MRI data of the brain, presenting derived properties and parameters from the input image data in a clinically useful context.

    Device Description

    nordicMEDIVA is a software as a medical device (SaMD) for processing of MR images of the brain. Users will configure analysis pipelines, which are executed automatically when image data is received or manually by a user. The user can choose to send the results to other DICOM nodes for review or use nordicView for their review and export the results to PACS, neuro navigation systems, or other DICOM-compliant modalities. nordicMEDIVA is a server-client solution and can be installed on a local server at the customer's location or in a cloud-based setup. The software is containerized with Docker technology and operates on a GPU-enabled Linux host. This allows customers to manage the server environment themselves or use it as a Software as a Service (SaaS) hosted by NordiclmagingLab AS in the cloud. Customers can install the server on physical hardware or in their own cloud infrastructure. The device comprises a database, DICOM functionality, various APIs, a visualization engine, and medical image analysis modules. The device is not intended for long-term persistent storage of medical diagnostic data. The device incorporates rule-based algorithms for the calculation of metrics from dynamic MRI data. The device does not incorporate Al algorithms based on neural networks. The device connects to other imaging modalities, such as MR scanners, PACS, and surgical navigation systems.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the nordicMEDiVA device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Summative Usability Test: All scenarios met acceptance criteria.All scenarios from the summative usability test met the acceptance criteria completely, no new risks were found and existing risk control measures were proven to be effective.
    Diffusion and Tractography: Results were as effective as the predicate device (nordicBrainEx K163324).The results from Diffusion and Tractography were evaluated in comparison with equivalent results from the predicate device, nordicBrainEx (K163324). The results were reviewed by internal and external clinical experts and proven to be as effective as the predicate device.
    BOLD fMRI: Results were as effective as the predicate device (nordicBrainEx K163324).The results from BOLD fMRI were evaluated in comparison with equivalent results from the predicate device, nordicBrainEx (K163324). The results were reviewed by internal and external clinical experts and proven to be as effective as the predicate device.
    DSC (Dynamic Susceptibility Contrast): Lin's Concordance Correlation Coefficient (CCC) for enhancing voxels ≥ 0.8 compared to the reference device (nordicMEDiVA K241608).The results from the DSC was confirmed to be the same as the reference device nordicMEDiVA (K241608) where a Lin's Concordance Correlation Coefficient of enhancing voxels was calculated with acceptance criteria being greater than or equal to 0.8 was applied. (The text states the results were "confirmed to be the same," implying the CCC met or exceeded the 0.8 acceptance criterion, though the specific achieved value is not provided.)

    2. Sample Size and Data Provenance

    • Sample size for test set: Not explicitly stated in the provided text.
    • Data Provenance: Not explicitly stated. The text mentions "internal and external clinical experts" reviewing results, but doesn't specify the country of origin of the data or whether it was retrospective or prospective.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of experts: Not explicitly stated. The text refers to "internal and external clinical experts."
    • Qualifications of experts: Not explicitly stated, beyond being "clinical experts." Specific experience levels (e.g., "radiologist with 10 years of experience") are not provided.

    4. Adjudication Method for the Test Set

    The text does not specify an explicit adjudication method (e.g., 2+1, 3+1). It states "reviewed by internal and external clinical experts." This suggests a consensus-based review, but the exact process of reaching that consensus or resolving disagreements is not detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No, an MRMC comparative effectiveness study was not described. The studies focused on comparing the device's output to that of predicate devices and showing equivalence, rather than directly measuring human reader performance with and without AI assistance.
    • Effect size of human reader improvement: Not applicable, as an MRMC comparative effectiveness study was not performed.

    6. Standalone Performance Study (Algorithm Only)

    Yes, a form of standalone performance was implicitly done. The tests for Diffusion and Tractography, BOLD fMRI, and DSC were evaluations of the nordicMEDiVA software's output (algorithm only) compared to established predicate devices. This demonstrates algorithm-only performance in generating results equivalent to cleared devices.

    7. Type of Ground Truth Used

    The ground truth for the test set appears to be based on:

    • Predicate Device Equivalence: For Diffusion and Tractography, and BOLD fMRI, the ground truth was essentially the "equivalent results from the predicate device, nordicBrainEx (K163324)," as reviewed and confirmed by clinical experts.
    • Reference Device Equivalence/Metrics: For DSC, the ground truth was based on demonstrating "the same" results as the reference device nordicMEDiVA (K241608), quantified by a Lin's Concordance Correlation Coefficient of enhancing voxels.
    • Expert Confirmation: The final acceptance relied on confirmation by "internal and external clinical experts."

    8. Sample Size for Training Set

    The text does not provide any information regarding the sample size used for a training set. This is because the device does not incorporate AI algorithms based on neural networks. It explicitly states: "The device incorporates rule-based algorithms for the calculation of metrics from dynamic MRI data. The device does not incorporate AI algorithms based on neural networks." Therefore, there would be no "training set" in the context of deep learning models.

    9. How Ground Truth for Training Set was Established

    Given that the device uses "rule-based algorithms" and "does not incorporate AI algorithms based on neural networks," there would be no training set requiring ground truth establishment in the traditional machine learning sense. The performance is based on the correctness and validation of its rule-based calculations against established clinical principles and comparison to predicate device outputs.