510(k) Data Aggregation

    K Number
    K243047

    Validate with FDA (Live)

    Date Cleared
    2025-05-28

    (243 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zoom 7X Catheter, with the Zoom Aspiration Tubing or Zoom POD Aspiration Tubing, and the Zoom Aspiration Pump (or equivalent vacuum pump), is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of last known well.

    Patients who are ineligible for intravenous thrombolytic drug therapy or who have not responded to thrombolytic drug therapy are candidates for treatment.

    The Zoom Aspiration Tubing and the Zoom POD Aspiration Tubing are intended to connect the Zoom (7X, 71, 55, 45, 35) Catheter and the TracStar LDP Large Distal Platform, the Zoom 88 Large Distal Platform, or the Zoom 88 Large Distal Platform Support to the Zoom Canister of the Zoom Aspiration Pump (or equivalent vacuum pump) and to allow the user to control the fluid flow.

    Device Description

    The Zoom™ 7X Catheter is a single lumen, braid and coil reinforced, variable stiffness catheter that facilitates removal of thrombus/clot from the neurovasculature when connected to a vacuum source, such as the Zoom Aspiration Pump, using the Zoom Aspiration Tubing or the Zoom POD Aspiration Tubing.

    The Zoom 7X Catheter is comprised of a hollow cylindrical tube which is bonded to a standard luer fitting. The wall of the tube is constructed using a combination of metal coils/braids and medical grade polymers.

    The distal section of the Zoom 7X Catheter has a hydrophilic coating to enhance tracking through the vasculature. The beveled distal tip allows for atraumatic tracking past vessel branches during insertion. A radiopaque marker provides the user with visual confirmation of the distal tip location under fluoroscopy.

    The Zoom 7X Catheter is packaged with an accessory Rotating Hemostasis Valve (RHV). The RHV is intended to be attached to the proximal hub of the catheter and used to control hemostasis during use with other devices. The Zoom 7X Catheter is also packaged with two (2) introducer sheath accessories. The introducer is intended to assist with insertion of the Zoom 7X Catheter into the RHV or the guide catheter.

    The Zoom Aspiration Tubing and the Zoom POD Aspiration Tubing are comprised of a hollow cylindrical tube which is bonded to a standard luer fitting that connects to the Zoom 7X Catheter and a slip fit connector that connects to the canister on the aspiration pump. The Zoom Aspiration Tubing is made of common medical grade polymers.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided FDA 510(k) clearance letter:

    1. Acceptance Criteria and Reported Device Performance:

    Test AttributeAcceptance CriterionReported Device Performance
    Zoom 7X Catheter
    Delivery, Compatibility, and Retraction (Trackability)The catheter shall be able to be delivered, deployed, and retracted per the IFU within a simulated neurological model without incurring any damage to the catheter.Pass
    Compatibility with Other Devices (External)The catheter shall be able to be delivered through the minimum introducer sheath or guide catheter size indicated in the product labeling.Pass
    Guidewire CompatibilityThe catheter shall be able to be delivered over the maximum size guidewire indicated in the product labeling.Pass
    Microcatheter / Intermediate Catheter CompatibilityThe catheter shall be able to accommodate a microcatheter/intermediate catheter up to the maximum size indicated in the product labeling.Pass
    Tip FlexibilityThe catheter distal tip flexibility shall be comparable to the predicate.Pass
    Visual InspectionThe catheter shall meet visual inspection criteria. The printing on the strain relief must be legible.Pass
    Dimensional (ID, Distal OD, Tip Length)All defined catheter dimensions are within the specified tolerances.Pass
    Catheter Bond StrengthThe catheter shall have sufficient bond strength to remain intact throughout a procedure.Pass
    Dynamic BurstThe catheter must withstand pressure testing under dynamic flow conditions.Pass
    Static BurstThe catheter shall meet criteria for static burst pressure testing.Pass
    Catheter Torque StrengthThe catheter shall not be damaged when rotated at least two (2) full rotations (720 degrees) with the tip constrained.Pass
    Kink ResistanceThere shall be no kinking of the catheter shaft around respective clinically relevant minimum bend radii in distal tip, medial and proximal locations.Pass
    FlexibilityThe catheter needs to have acceptable flexure values for tracking in the vasculature.Pass
    Luer CompatibilityDevice and accessories shall be compatible with standard syringe luer fittings per ISO 80369-7.Pass
    Accessory CompatibilityDevice shall be compatible with an RHV.Pass
    Coating - ParticulateThe amount of particulate matter generated during simulated use testing shall be determined and comparable to competitive products.Pass
    Coating – Lubricity, Durability, and IntegrityCoating must be lubricious with a specified average pull force. There were no coating anomalies or significant wear observed post simulated use.Pass
    Clot RetrievalThe device shall be able to aspirate a variety of clot types in a range of vessel diameters.Pass
    Zoom POD Aspiration Tubing (new component)
    Visual InspectionThe Zoom POD Aspiration Tubing shall meet visual inspection criteria.Pass
    Dimensional (Working Length)The working length is within the specified tolerances.Pass
    Vacuum Force at Catheter TipThe vacuum force delivered by the aspiration tubing to the tip of the catheter should be comparable to the vacuum force delivered by the predicate aspiration tubing.Pass
    Connector CompatibilityThe aspiration tubing connectors shall securely connect to the pump canister lid and standard luer fittings.Pass
    Lumen Collapse TestThe tubing lumen shall not collapse under vacuum.Pass
    Flow Control FunctionalityThe flow control mechanism shall allow users to start and stop flow multiple times when the connected pump is running at maximum vacuum.Pass
    Freedom From LeakageThe vacuum pressure delivered at the tip of the aspiration tubing shall be consistent with the pressure generated by the pump.Pass
    Tensile StrengthThe bonds between the tubing and connectors shall be sufficiently strong to ensure the tubing remains intact during use.Pass
    Clot Filter Functionality (Zoom POD only)The clot filter should be able to be opened and closed without causing leak.Pass
    Introducer Sheath Accessory (new component biocompatibility tests)
    CytotoxicityTest article extracts must yield grade 2 or lower.Pass, Non-cytotoxic
    SensitizationTest article extracts must yield grade 0.Pass, Non-sensitizing
    Intracutaneous ReactivityTest article extracts must yield a difference between the test extract overall mean score and corresponding control overall mean score of 1.0 or less.Pass, Non-irritating
    Acute Systemic ToxicityNone of the animals treated with the test article extracts must show a significantly greater reactivity than animals treated with the control article.Pass, Non-toxic (acute systemic)
    Material-Mediated PyrogenicityTest article extract must yield <0.5°C rise in individual animal temperature.Pass, Non-pyrogenic
    Indirect HemolysisTest article extract must result in a hemolytic index of <2%.Pass, Non-hemolytic

    2. Sample Size Used for the Test Set and Data Provenance:

    The document primarily refers to non-clinical bench testing to demonstrate substantial equivalence. It does not provide specific numerical sample sizes for each bench test conducted on the subject devices (Zoom 7X Catheter, Zoom Aspiration Tubing, Zoom POD Aspiration Tubing, and the Introducer Sheath accessories). The tests were conducted on physical devices in simulated environments.

    The data provenance is from bench testing and laboratory testing performed by the manufacturer, Imperative Care, Inc. There is no indication of country of origin for the data beyond the manufacturer's location in Campbell, California. The study is a pre-market regulatory submission, and the data is generated for this specific purpose rather than being retrospective or prospective patient data.


    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. The clearance letter describes bench testing for physical device performance and biocompatibility testing. This type of evaluation does not involve establishing ground truth from human expert interpretations of medical images or patient outcomes. The "ground truth" for these tests is defined by established engineering and biological safety standards (e.g., ISO, ASTM).


    4. Adjudication Method for the Test Set:

    Not applicable. As described above, no human expert adjudication of a test set (e.g., for image interpretation or clinical diagnosis) was performed. The "adjudication" in this context refers to the comparison of test results against predefined acceptance criteria for physical, mechanical, and biological properties.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No. The document explicitly states:

    • "Substantial equivalence was established based on non-clinical testing presented above."
    • "Clinical Study: Substantial equivalence was established based on non-clinical testing presented above."
    • "Animal Study: Substantial equivalence was established based on non-clinical testing presented above."

    This indicates that no MRMC study or any clinical study was conducted. Therefore, there is no effect size reported for human readers improving with AI vs. without AI assistance. The device is a physical medical device (catheter and aspiration tubing system), not an AI-powered diagnostic or assistive tool.


    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    No. This question applies to AI/software as a medical device. The Zoom 7X Catheter and its associated aspiration tubing are physical medical devices. There is no mention of an algorithm or AI component in the device's function.


    7. Type of Ground Truth Used:

    For the device performance tests (Tables 3 and 4), the "ground truth" is defined by:

    • Established engineering standards: e.g., ISO 10555-1 for catheter performance, ISO 80369-7 for Luer compatibility.
    • Manufacturer's internal specifications: The acceptance criteria (e.g., specific dimensions, torque strength, burst pressure) are derived from the device's design requirements and comparison to the predicate device.
    • Simulated neurological models: Used for trackability and clot retrieval tests.

    For the biocompatibility tests (Table 5), the "ground truth" is based on:

    • International standards for biological evaluation: e.g., ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11, ISO 10993-23.
    • ASTM standards: e.g., ASTM F756-17 for indirect hemolysis.
    • FDA guidance documents.

    8. Sample Size for the Training Set:

    Not applicable. This device is a physical medical device cleared based on substantial equivalence to a predicate through bench and biocompatibility testing, not through machine learning or AI model development that requires a training set.


    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as no training set was used for this device.