510(k) Data Aggregation

    K Number
    K242622

    Validate with FDA (Live)

    Date Cleared
    2024-10-28

    (55 days)

    Product Code
    Regulation Number
    878.4850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lancet is intended for capillary blood sampling.

    Device Description

    The Sterile Lancets for Single Use is composed two components: needle and needle holder. Needle holder is plastic part that enclosed the needle. The sterile barrier is the needle holder and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The lancet is together with a lancing device to puncture the skin to obtain a drop of capillary blood from fingertip or from alternative sites. The shelf-life of the product is 5 years.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device (Sterile Lancets for Single Use). It describes the device, its intended use, a comparison to a predicate device, and the non-clinical testing performed to establish substantial equivalence.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list "acceptance criteria" with numerical targets in a formal table corresponding to performance. Instead, it refers to "Performance testing is conducted according to Inspection Specification" and states that "the test sample has undergone accelerated aging for 5.5 years" and "passed all tests." The performance attributes tested are listed.

    Here's an interpretation of the performance attributes and the implied acceptance criteria from the text:

    Acceptance Criterion (Implied)Reported Device Performance
    Appearance: Acceptable visual inspectionThe device passed tests after accelerated aging. (Implied: Appearance remained acceptable.)
    Dimension: Within specified tolerancesThe device passed tests after accelerated aging. (Implied: Dimensions remained within specified tolerances.)
    Firmness of needle and main body: Adequate mechanical integrityThe device passed tests after accelerated aging. (Implied: Firmness remained adequate.)
    Puncture Performance: Satisfactory for capillary blood samplingThe device passed tests after accelerated aging. (Implied: Puncture performance remained satisfactory.)
    Matching: Compatibility with lancing devicesThe device passed tests after accelerated aging. (Implied: Compatibility remained.)
    Launch Performance: Adequate for intended useThe device passed tests after accelerated aging. (Implied: Launch performance remained adequate.)
    Bacterial endotoxin: Below specified limitsThe device passed tests after accelerated aging. (Implied: Bacterial endotoxin levels were below limits.)
    Sterilization (SAL=10-6): AchievedSterilized by Radiation SAL=10-6. The device passed tests after accelerated aging. (Implied: Sterility was maintained.)
    Biocompatibility: Conforming to ISO 10993 standardsConforms to the requirements of ISO 10993 series standards, specifically: Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity.
    Shelf-life: 5 yearsAfter five years of accelerated aging, the product has passed all tests, so it can be guaranteed that the product can be used normally within five years.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the numerical sample size used for performance testing. It generally refers to "the test sample."
    • Data Provenance: The document indicates that the testing was conducted by Ningbo Caremed Medical Products Co., Ltd., based in China. The data would be retrospective in the sense that the tests were completed before the 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable as the device is a medical lancet and the testing involves objective physical, chemical, and biological performance tests, not subjective interpretation by experts to establish a "ground truth" like in imaging or diagnostic AI. The "ground truth" for these tests would be the established scientific and engineering standards and methods.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reasons as point 3. Adjudication methods are typically used when subjective expert opinion is involved in ground truth establishment.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This information is not applicable. An MRMC study is relevant for diagnostic devices, especially those involving human interpretation (e.g., radiologists interpreting images). This device is a sterile lancet for single-use capillary blood sampling, a physical instrument, and does not involve human readers interpreting output that would be improved by AI assistance.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance)

    This information is not applicable. This device is a physical medical instrument, not a software algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the performance testing of the lancet is based on:

    • Established scientific and engineering standards: For dimensions, firmness, puncture performance, matching, launch performance.
    • Specific ISO standards: For biocompatibility (ISO 10993 series) and sterilization (SAL of 10-6).
    • Bacterial endotoxin limits: Established regulatory or compendial limits.
    • Accelerated aging protocols: To simulate real-time shelf-life.

    8. Sample Size for the Training Set

    This information is not applicable. This device is not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this device.