Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K242598

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-14

    (76 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DEFINE System with the non-invasive applicators employs RF energy for various applications:

    The DEFINE System with the Cheek and Chin Applicators is indicated for the temporary relief of minor muscle aches and pain, temporary relief of muscle spasm, and temporary improvement of local blood circulation.

    The DEFINE System with the FORMA Applicator is intended for the relief of minor muscle aches and pain. relief of muscle spasm. and for the temporary improvement of local blood circulation.

    The DEFINE System with the MiniFX Applicator is intended for the relief of minor muscle aches and pain, relief of muscle spasm, for the temporary improvement of local blood circulation and for the temporary reduction in the appearance of cellulite.

    The DEFINE System with the Morpheus8 Applicator is intended for use in dermatologic procedures where coagulation/contraction of soft tissue or hemostasis is needed.

    At higher energy levels greater than 62 mJ/pin, the use of the Morpheus8 Applicator is limited to Skin Types I-IV.

    Device Description

    The DEFINE device is designed to deliver non-thermal RF energy to the skin and subdermal fat. The RF energy is transmitted to the Applicators and delivered to the treatment area throughout the set of RF electrodes positioned on the treatment units. The invested RF energy is transformed into local heating of the treatment area. Local heating of the skin and underlying tissue layers temporarily ease the symptoms of muscle aches, pain & spasm. It also contributes to the temporary improvement of local blood circulation.

    The DEFINE device platform comprises a AC/DC power supply unit, main controller, two RF generators, distributor card and user interface including a touch screen. The system platform contains four designated connectors for the following applicators:

    • Cheek Applicator (hands-free) ●
    • Chin Applicator (hands-free) ●
    • FORMA Applicator ●
    • MiniFX ●
    • MORPHEUS8 Applicator with the resurfacing, 12 and 24 pin tip heads ●

    The DEFINE device RF Applicators are connected to the console via a cable. The delivery of the RF Energy is controlled by a "Start"/"Stop" button on the LCD screen or by a foot pedal. The DEFINE device incorporates a treatment deactivation button. Pressing the deactivation button by either the patient or caregiver would immediately halt the treatment and switch the device into a Pause mode until the operator re-enables it.

    AI/ML Overview

    The provided text is a 510(k) summary for the InMode Ltd. DEFINE System. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific performance acceptance criteria through clinical studies. Therefore, much of the requested information regarding acceptance criteria, specific performance metrics, sample sizes, expert involvement, and ground truth establishment for a study proving the device meets acceptance criteria is not present in this type of regulatory submission.

    The 510(k) process primarily relies on showing that a new device is as safe and effective as a legally marketed predicate device. This is often achieved through:

    • Comparison of technological characteristics: Demonstrating that the new device has the same fundamental technology, intended use, and similar design as the predicate.
    • Non-clinical performance data (bench testing): Verifying that the device meets safety and performance standards (e.g., electrical safety, electromagnetic compatibility, RF energy output).
    • Absence of clinical performance data for substantial equivalence: In many cases, if technological characteristics are sufficiently similar and non-clinical testing confirms equivalence, new clinical data is not required.

    Based on the provided text, here's what can be extracted and what is not available:

    1. Table of Acceptance Criteria and Reported Device Performance

    The concept of specific "acceptance criteria" and "reported device performance" as typically seen in a clinical trial or performance study (e.g., sensitivity, specificity, accuracy for a diagnostic device) is not applicable or stated in this 510(k) summary for the DEFINE System.

    Instead, the submission demonstrates compliance with:

    • Safety standards: IEC 60601-1-2, ANSI AAMI ES60601-1, IEC 60601-2-2 (for RF surgical equipment), IEC 60601-1-6 (usability), ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), IEC 62304 (software life cycle), ISO 14971 (risk management).
    • Equivalence in intended use and technological characteristics to predicate devices.

    The "performance" is primarily demonstrated through bench testing for electrical safety, EMD safety, RF output (IEC 60601-2-2), and software validation. These are conformity assessments to established standards rather than performance against a clinical acceptance threshold for diagnostic or therapeutic efficacy outcomes.

    Acceptance Criteria (Implied by 510(k))Reported Device Performance (as stated in submission)
    Compliance with IEC 60601-1 (Electrical Safety)The modified device was tested for compliance withIEC 60601-1.
    Compliance with IEC 60601-1-2 (EMD Safety)The modified device was tested for compliance with IEC 60601-1-2.
    Compliance with IEC 60601-2-2 (RF Specific)The modified device was tested for compliance with IEC 60601-2-2.
    Compliance with IEC 62304 (Software)Validation of the device software was performed in accordance with clause 14 of IEC 60601-1 (third edition) Software requirements, IEC to 62304:2006/A1:2016.
    Biocompatibility (Cytotoxicity) Compliance with ISO 10993-5Compliance demonstrated (listed as a standard the device complies with).
    Biocompatibility (Irritation/Sensitization) Compliance with ISO 10993-10Compliance demonstrated (listed as a standard the device complies with).
    Risk Management Compliance with ISO 14971Compliance demonstrated (listed as a standard the device complies with).
    Equivalence of RF Maximum Output Power (Cheek & Chin Applicators)Predicate: 50 W; DEFINE System: idem (50 W).
    Equivalence of Energy Output Levels (Cheek & Chin Applicators)Predicate: Cheek Up to 60, Chin Up to 40; DEFINE System: idem.
    Equivalence of Frequency (Cheek & Chin Applicators)Predicate: 1 MHz; DEFINE System: idem.
    Equivalence of RF Pulse Width (Cheek & Chin Applicators)Predicate: 2 seconds; DEFINE System: idem.

    Note: The submission explicitly states "Clinical Performance Data: Not Applicable" (page 7 of 14). This signifies that the decision for substantial equivalence was based solely on non-clinical (bench) testing and comparison to predicate devices, without new human clinical studies for performance metrics.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable. No clinical test set data is provided or referenced, as the submission states "Clinical Performance Data: Not Applicable." The "test set" for non-clinical performance refers to the device and its components undergoing bench testing according to various standards. The document does not specify the number of units tested.
    • Data Provenance: Not applicable for clinical data. For non-clinical data, it's typically performed by the manufacturer or a certified testing lab. The country of origin of the data is implied to be within controlled engineering/testing environments, likely at the manufacturer's location (Israel) or a contracted lab. The data is implicitly "prospective" in the sense that the tests were conducted specifically for this submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable. There was no clinical test set for which ground truth needed to be established by experts.

    4. Adjudication Method for the Test Set

    • Not applicable. There was no clinical test set requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not applicable. No clinical studies were conducted or referenced for this 510(k) submission. Therefore, no MRMC study, human reader improvement effect size, or AI assistance claims are present.

    6. Standalone (Algorithm Only) Performance Study

    • Not applicable. This device is an electrosurgical cutting and coagulation device, not an AI or algorithm-only device in the context of diagnostic or interpretive performance. Its performance is tied to its physical mechanisms and safety parameters validated through bench testing, not an algorithm's standalone accuracy. Software validation (IEC 62304) ensures the software functions as intended and is safe, not that it performs a diagnostic or prescriptive task independently.

    7. Type of Ground Truth Used

    • For non-clinical testing: The "ground truth" is adherence to established engineering standards (e.g., IEC, ANSI, ISO). For example, electrical safety is "ground true" if the device passes the specified tests in IEC 60601-1.
    • For clinical performance claims: Not applicable, as no clinical performance data was provided.

    8. Sample Size for the Training Set

    • Not applicable. This is not an AI/ML device that requires a training set in the typical sense (e.g., for image classification). The "software validation" mentioned relates to the software controlling the device's operation, not to a learned model.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, for the same reasons as point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1