    Intended Use

    3M™ Attest™ Super Rapid Readout Biological Indicator 1493:
    Use the 3M™ Attest™ Super Rapid Readout Biological Indicator 1493 in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater, the 3M™ Attest™ Auto-reader 490H having software version 4.0.0 or greater, or the 3M™ Attest™ Mini Auto-reader 490M to qualify or monitor the following steam sterilization cycles:
    Cycle Type: Gravity Displacement, Exposure Temperature: 250°F (121°C), Exposure Time: 30 minutes
    Cycle Type: Dynamic-air-removal (pre-vacuum and SFPP), Exposure Temperature: 250°F (121°C), Exposure Time: 15, 20, 30, or 35 minutes

    3M™ Attest™ Super Rapid Steam Gravity Clear Challenge Pack 1493PCDG: Use the 3M™ Attest™ Super Rapid Steam Gravity Clear Challenge Pack 1493PCDG in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater, the 3M™ Attest™ Auto-reader 490H having software version 4.0.0 or greater, or the 3M™ Attest™ Mini Auto-reader 490M to qualify or monitor the following steam sterilization cycles:
    Cycle Type: Gravity Displacement, Exposure Temperature: 250°F (121°C), Exposure Time: 30 minutes

    3M™ Attest™ Auto reader 490:
    The 3M™ Attest™ Auto reader 490 is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295 and 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491, 1492V, and 1493 at 60°C for a final fluorescent result at 24 minutes.

    3M™ Attest™ Auto-reader 490H:
    The 3M™ Attest™ Auto-reader 490H is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295, and 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491, 1492V, and 1493 at 60°C for a final fluorescent result at 24 minutes.

    3M™ Attest™ Mini Auto-reader 490M:
    The 3M™ Attest™ Mini Auto-reader 490M is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295 and 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491, 1492V and 1493, at 60°C for a final fluorescent result at 24 minutes.

    Device Description

    The 3MTM Attest™ Super Rapid Readout Biological Indicator (BI) 1493 is a self-contained biological indicator specifically designed for rapid and reliable qualification testing and routine monitoring of 121°C (250°F) dynamic-air-removal and gravity-displacement steam sterilization processes in healthcare facilities. The 1493 BI is used in conjunction with a 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater, or a 3MTM Attest™ Auto-reader 490H having software version 4.0.0 or greater, or a 3M™ Attest™ Mini Auto-reader 490M. The 1493 BI is a single-use device composed of a plastic vial containing a spore carrier and media ampoule, enclosed with a color-coded cap. On each 1493 BI cap is a chemical process indicator that changes color from pink to light brown or darker when exposed to steam. The presence of fluorescence within the specified incubation time of the 1493 BI in the 490 Autoreader, the 490H Auto-reader, or the 490M Mini Auto-reader indicates a steam sterilization process failure.

    The 3M™ Attest™ Super Rapid Steam Gravity Clear Challenge Pack 1493PCDG is specifically designed to qualify and monitor gravity steam sterilization processes at 250°F (121°C) in healthcare facilities. The 1493PCDG Challenge Pack consists of a clear plastic shell with an opening to the cavity containing the monitoring products, all covered by a foil lid. The 1493PCDG is a single-use device. Each 1493PCDG Challenge Pack contains a 3M™ Attest™ Super Rapid Readout Biological Indicator (BI) 1493 and a 3M™ Attest™ Steam Chemical Integrator (CI) (Type 5 (Category i5) Integrating Indicator as categorized by ISO 11140-1:2014). The 3M™ Attest™ Steam Chemical Integrator offers an immediate ACCEPT or REJECT reading. A chemical process indicator is present on the cap of the 1493 BI and is visible through the clear plastic shell of the challenge pack. The process indicator on the BI cap turns color from pink to light brown or darker upon exposure to steam. 3M™ Attest™ 1493 BI controls are provided with the Challenge Pack.

    The 1493 BI is specifically designed for rapid and reliable qualification testing and routine monitoring of steam sterilization process when used in conjunction with the 3MM Attest™ Auto-reader 490 having software version 4.0.0 or greater, a 3M™ Attest™ Auto-reader 490H having software version 4.0.0 or greater, or a 3M™ Attest™ Mini Auto-reader 490M. The presence of fluorescence within the specified incubation time of the 1493 BI in the 490. 490H or 490M Auto-reader indicates a steam sterilization process failure.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the 3M™ Attest™ Super Rapid Readout Biological Indicator 1493 and related devices, as extracted from the provided text:

    1. Acceptance Criteria and Reported Device Performance

    Biological Indicator (1493 BI)

    Test PerformedDevice DescriptionApplicable StandardsPurposeAcceptance CriteriaResults
    Population1493 BIFDA Guidance¹, ISO 11138-1:2017²To evaluate the total viable spore count≥ 10⁶ sporesAcceptance criteria met
    Positive Control Test1493 BIFDA Guidance¹, ISO 11138-1:2017², and ISO 11138-3:2017³To evaluate performance of BI without steam exposureAll BIs must be fluorescent positive at 24 minutes and visual pH color change positive (yellow) at 168 hoursAcceptance criteria met
    D-value for 121°C (250°F) cycles1493 BIFDA Guidance¹, ISO 11138-1:2017², and ISO 11138-3:2017³To evaluate the resistance characteristics of the BID-value must be ≥ 1.5 minutesAcceptance criteria met
    Z-value for 121°C (250°F) cycles1493 BIFDA Guidance¹, ISO 11138-1:2017², and ISO 11138-3:2017³Not explicitly stated but implied to be part of characterizing resistanceZ-value must be ≥ 10°CAcceptance criteria met
    Survival for 121°C (250°F) cycles1493 BIFDA Guidance¹, ISO 11138-1:2017², and ISO 11138-3:2017³To evaluate the resistance characteristics of the BIMeets the requirements for Calculated survival time* or 5 minutes, whichever is longer (*ISO 11138-1:2017)Acceptance criteria met
    Kill for 121°C (250°F) cycles1493 BIFDA Guidance¹, ISO 11138-1:2017², and ISO 11138-3:2017³Not explicitly stated but implied to be part of characterizing resistanceMeets the requirements for Calculated kill time* (*ISO 11138-1:2017)Acceptance criteria met
    Component Inhibition Studies1493 BIFDA Guidance¹, ISO 11138-1:2017², and ISO 11138-3:2017³To evaluate the effects of carrier and packaging materials on the resistance characteristics of the BIComponents have no impact on the recovery of 10-100 organismsAcceptance criteria met
    Holding Time Assessment for 121°C (250°F) cycles1493 BIFDA Guidance¹, ISO 11138-1:2017², and ISO 11138-3:2017³To evaluate the effect of the labeled holding time on the D-valueD-value must be within +/- 20% of the initial D-value calculated per ISO 11138-1:2017Acceptance criteria met
    Reduced Incubation Time for 121°C (250°F) cycles1493 BIFDA Guidance¹, ISO 11138-1:2017², and ISO 11138-8:2021⁴To validate the reduction in incubation time from 7 days (visual color change) to 24 minutes (fluorescent readout)Meets FDA's requirements for Reduced Incubation Time with ≥ 97% alignment with the conventional incubation timeAcceptance criteria met
    Attest 1493 Simulated Use1493 BIFDA Guidance¹Verification of performance in claimed cyclesThe following readout times: • Fluorescent result in 24 minutesAcceptance criteria met
    490 and 490M Auto-reader Performance1493 BIFDA Guidance¹, ISO 11138-1:2017²Verify equivalent performance of the 1493BI with the 3M™ Attest 490 and 490M Auto-readersBIs must be fluorescent positive in 24 minutes and visual pH color change (growth) positive at 168 hoursAcceptance criteria met

    Challenge Pack (1493PCDG)

    Test PerformedDevice DescriptionApplicable StandardsPurposeAcceptance CriteriaResults
    Resistance of 1493PCDG Challenge Pack compared to AAMI 16 Towel PCD in claimed cycle1493PCDGFDA Guidance¹ and ANSI/AAMI ST79:2017, Comprehensive guide to steam sterilization and sterility assurance in health care facilitiesDemonstrate the performance of the 1493PCDG Challenge Pack is equivalent to the performance of the AAMI 16 Towel PCD in the claimed cycleIndicators contained in the 1493PCDG Challenge Pack must demonstrate equivalent resistance as compared to the indicators contained in the AAMI 16 Towel PCD in the claimed cycleAcceptance criteria met
    Resistance of 1493PCDG Challenge Pack compared to standalone indicators in claimed cycle1493PCDGFDA Guidance¹Demonstrate the 1493PCDG Challenge Pack provides a greater challenge than the standalone indicators in the claimed cycleIndicators contained in the 1493PCDG Challenge Pack must demonstrate greater resistance compared to the standalone indicators in the claimed cycleAcceptance criteria met
    1493PCDG Peel Force (unsterilized)1493PCDGASTM F88 / F88M-15: Standard Test Method for Seal Strength of Flexible Barrier and ASTM F2824 – 10: Standard Test Method for Mechanical Seal Strength Testing for Round Cups and Bowl Containers with Flexible Peelable LidsDemonstrate acceptable peel force of Clear Challenge Pack 1493PCDG heat seal foil lid from plastic shell≥2.75lbF and ≤10lbFAcceptance criteria met
    1493PCDG Peel Force (pre and post-sterilization)1493PCDGASTM F88 / F88M-15: Standard Test Method for Seal Strength of Flexible Barrier and ASTM F2824 – 10: Standard Test Method for Mechanical Seal Strength Testing for Round Cups and Bowel Containers with Flexible Peelable LidsDemonstrate acceptable peel force of Clear Challenge Pack 1493PCDG heat seal foil lid from plastic shell, pre and post sterilization≥2.75lbF and ≤10lbFAcceptance criteria met

    2. Sample Size Used for the Test Set and Data Provenance

    • Biological Indicator (1493 BI):
      • Population, Positive Control Test, D-value, Z-value, Survival, Kill, Component Inhibition Studies, Reduced Incubation Time, Simulated Use: 3 lots of 1493 BIs were tested.
      • Holding Time Assessment: 2 lots of 1493 BIs were tested.
      • 490 and 490M Auto-reader Performance: 1 lot of 1493 BIs was evaluated.
    • Challenge Pack (1493PCDG):
      • Resistance compared to AAMI 16 Towel PCD: 3 lots of 1493PCDG (containing 3 lots of 1493 BIs and 3 lots of 1243R CIs) were tested.
      • Resistance compared to standalone indicators: 3 lots of 1493PCDG (containing 3 lots of 1493 BIs and 3 lots of 1243R CIs) were tested.
      • Peel Force: 3 lots of 1493PCDG were tested for both unsterilized and pre/post sterilization conditions.

    The data provenance is not explicitly stated in terms of country of origin or whether it's retrospective or prospective. However, tests were conducted according to recognized international (ISO) and US (FDA, AAMI, ASTM) standards, suggesting controlled laboratory environments.

    3. Number of Experts and Qualifications for Ground Truth

    Not applicable. The "device" in question (biological indicators and challenge packs for sterilization) directly measures biological viability and physical characteristics. The ground truth for these types of devices is established through direct laboratory measurements, microbiological assays (like spore counts and viability after exposure), and physical property testing, rather than expert interpretation of images or clinical data. Therefore, there are no "experts" in the sense of medical professionals establishing a ground truth for a test set.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the studies involve direct measurement and laboratory testing against predefined criteria in recognized standards, not subjective assessment requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic imaging devices where human readers interpret results, sometimes with AI assistance. The devices described here (biological indicators and auto-readers) are designed for direct, objective measurement of sterilization efficacy, not for human interpretation that would be enhanced by AI assistance.

    6. Standalone Performance Study

    Yes, standalone performance studies were done. All the tests listed in the tables (Population, D-value, Survival, Kill, Reduced Incubation Time, Simulated Use, etc., for the BI, and Resistance, Peel Force for the Challenge Pack) represent the standalone performance of the devices as they are intended to function without human interpretation. For example, the auto-readers automatically provide a fluorescent result, which is the direct output of the system.

    7. Type of Ground Truth Used

    • Biological Indicator (1493 BI): The ground truth is primarily established through direct microbiological testing (e.g., enumeration of viable spores, determination of D-values, survival, and kill times through controlled exposure to steam under defined conditions) and physical/chemical measurements (e.g., fluorescent or pH color change as an indicator of spore viability/growth). This is based on established sterilization science principles and standards.
    • Challenge Pack (1493PCDG): The ground truth for the challenge pack's resistance is established by comparing its performance to recognized reference standards (e.g., AAMI 16 Towel PCD) and to standalone indicators, all under controlled sterilization conditions. Physical integrity is assessed via mechanical property testing (peel force).

    8. Sample Size for the Training Set

    Not applicable. The devices described are not AI/machine learning devices that require a separate "training set" of data to learn patterns or make predictions. They are physical and chemical indicators that operate based on biological and material science principles.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for these devices.