Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Peripheral vessel and Dermatology.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

    Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode, MV-Flow Mode.

    Device Description

    V5 Diagnostic Ultrasound System; H5 Diagnostic Ultrasound System; XV5 Diagnostic Ultrasound System: XH5 Diagnostic Ultrasound System; V4 Diagnostic Ultrasound System; H4 Diagnostic Ultrasound System: XV4 Diagnostic Ultrasound System; XH4 Diagnostic Ultrasound System are a general purpose, mobile, software controlled, diagnostic ultrasound system. Their function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, MV-Flow mode, Multi-Image mode(Dual, Quad), 3D/4D mode. The V5/H5/XV5/XH5, V4/H4/XV4/XH4 diagnostic ultrasound system also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The V5/H5/XV5/XH5, V4/H4/XV4/XH4 diagnostic ultrasound system have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and supporting study details for each AI-powered feature of the V5/H5/XV5/XH5 and V4/H4/XV4/XH4 Diagnostic Ultrasound Systems, as provided in the document:


    HeartAssist (Fetal and Adult)

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureModalityAcceptance CriteriaReported Performance
    View RecognitionFetalAverage recognition accuracy ≥ 89%93.21%
    AdultAverage recognition accuracy ≥ 84%98.31%
    SegmentationFetalAverage Dice-score ≥ 0.80.88
    AdultAverage Dice-score ≥ 0.90.93
    Size MeasurementFetalError rate of area measured value ≤ 8%≤ 8%
    (Error Rate)FetalError rate of angle measured value ≤ 4%≤ 4%
    FetalError rate of circumference measured value ≤ 11%≤ 11%
    FetalError rate of diameter measured value ≤ 11%≤ 11%
    Size MeasurementAdultPCC value (B-mode) ≥ 0.8Passed (PCC ≥ 0.8)
    (Pearson Correlation)AdultPCC value (M-mode) ≥ 0.8Passed (PCC ≥ 0.8)
    AdultPCC value (Doppler-mode) ≥ 0.8Passed (PCC ≥ 0.8)

    2. Sample Sizes and Data Provenance:

    • Test Set (Fetal):

      • Number of individuals: 80
      • Number of static images: 280
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from five hospitals.
      • Country of Origin: Not explicitly stated, but clinical experts are from multiple countries including US and Korea, suggesting data could be from these regions.
    • Test Set (Adult):

      • Number of individuals: 30
      • Number of static images: 540
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from five hospitals.
      • Country of Origin: Not explicitly stated, but clinical experts are from multiple countries including US and Korea, suggesting data could be from these regions.

    3. Number of Experts and Qualifications for Ground Truth:

    • Fetal:

      • 3 participating experts:
        • 1 obstetrician with >20 years of experience (in fetal cardiology)
        • 2 sonographers with >10 years of experience (in fetal cardiology)
      • Supervised by: 1 obstetrician with >25 years of experience.
    • Adult:

      • 4 experts:
        • 2 cardiologists with at least 10 years of experience
        • 2 sonographers with at least 10 years of experience.

    4. Adjudication Method:

    • Fetal: All acquired images were first classified into correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn. The entire process was supervised by another obstetrician.
    • Adult: Experts manually traced the contours of the heart and the signal outline on the images. (Implicitly, this suggests an expert consensus or expert-defined ground truth, but a specific adjudication method like "2+1" is not detailed.)

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned for evaluating human readers with and without AI assistance for HeartAssist. The study focused on standalone performance metrics against ground truth.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported performance metrics (recognition accuracy, Dice-score, error rates, PCC values) are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-defined outlines. For fetal images, experts classified views and manually drew anatomy areas. For adult images, cardiologists and sonographers manually traced contours. The PCC for adult size measurement was calculated against "the cardiologist's measurements."

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for training, tuning and validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • Not explicitly stated, but it can be inferred that a similar expert-based process was used for generating ground truth for training data, as noted for the validation data: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn..."

    BiometryAssist (Fetal)

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance CriteriaReported Performance
    SegmentationAverage Dice-score ≥ 0.80.919
    Size Measurement (Circumference)Error rate ≤ 8%≤ 8%
    Size Measurement (Distance)Error rate ≤ 4%≤ 4%
    Size Measurement (NT)Error rate ≤ 1mm≤ 1mm

    2. Sample Sizes and Data Provenance:

    • Test Set:
      • Number of individuals: 77
      • Number of static images: 320
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from two hospitals.
      • Country of Origin: Americans and Koreans, suggesting data could be from these regions.

    3. Number of Experts and Qualifications for Ground Truth:

    • 3 participating experts:
      • 1 obstetrician with >20 years of experience (in fetal cardiology)
      • 2 sonographers with >10 years of experience (in fetal cardiology)
    • Supervised by: 1 obstetrician with >25 years of experience.

    4. Adjudication Method:

    • All acquired images were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each image. The entire process was supervised by another obstetrician.

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported performance metrics (Dice-score, error rates) are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-defined outlines. Experts classified views and manually drew anatomy areas.

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for training, tuning and validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • It's inferred that a similar expert-based process was used as for validation data: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn..."

    ViewAssist (Fetal)

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance CriteriaReported Performance
    View RecognitionAverage accuracy ≥ 89%94.26%
    Anatomy Annotation (Segmentation)Average Dice-score ≥ 0.80.885

    2. Sample Sizes and Data Provenance:

    • Test Set:
      • Number of individuals: 77
      • Number of static images: 680
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from two hospitals.
      • Country of Origin: Americans and Koreans, suggesting data could be from these regions.

    3. Number of Experts and Qualifications for Ground Truth:

    • 3 participating experts:
      • 1 obstetrician with >20 years of experience (in fetal cardiology)
      • 2 sonographers with >10 years of experience (in fetal cardiology)
    • Supervised by: 1 obstetrician with >25 years of experience.

    4. Adjudication Method:

    • All acquired images were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each image. The entire process was supervised by another obstetrician.

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported performance metrics (recognition accuracy, Dice-score) are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-defined outlines. Experts classified views and manually drew anatomy areas.

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for training, tuning and validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • It's inferred that a similar expert-based process was used as for validation data: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn..."

    UterineAssist

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance CriteriaReported Performance
    Segmentation (Uterus)Average Dice-score not explicitly stated, but 96% is reported.96%
    Segmentation (Endometrium)Average Dice-score not explicitly stated, but 92% is reported.92%
    Feature Points Extraction (Uterus)Errors ≤ 5.8 mm≤ 5.8 mm
    Feature Points Extraction (Endometrium)Errors ≤ 4.3 mm≤ 4.3 mm
    Size MeasurementErrors ≤ 2.0 mm≤ 2.0 mm

    2. Sample Sizes and Data Provenance:

    • Test Set (Segmentation):

      • Number of static images: 450 sagittal uterus images and 150 transverse uterus images (total 600 images).
      • Number of individuals: 60 contributed to the validation dataset (for segmentation AND feature points/size measurement)
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from three hospitals.
      • Country of Origin: All Koreans.
    • Test Set (Feature Points Extraction & Size Measurement):

      • Number of static images: 48 sagittal and 44 transverse plane images of the uterus (total 92 images).
      • Number of individuals: 60 individuals contributed to the validation dataset (same as segmentation).
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from three hospitals.
      • Country of Origin: All Koreans.

    3. Number of Experts and Qualifications for Ground Truth:

    • 3 participating OB/GYN experts with >10 years' experience.

    4. Adjudication Method:

    • Segmentation of the ground truth was generated by three participating OB/GYN experts. The image set was divided into three subsets, and each expert drew ground truths for one subset. Ground truths drawn by one expert were cross-checked by the other two experts. Any images not meeting inclusion/exclusion criteria were excluded.

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported performance metrics (Dice-score, error rates) are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-drawn contours and measurements.

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for test/training validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • It's inferred that a similar expert-based process was used as for validation data: "Segmentation of the ground truth was generated by three participating OB/GYN experts with more than 10 years' experience." This process for ground truth establishment is also applicable to training data generation.

    NerveTrack (Nerve Detection)

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance CriteriaReported Performance
    DetectionAverage accuracy from 10 image sequence not explicitly stated as an acceptance criterion, but 89.91% is reported. A detection is considered correct if Dice coefficient is ≥ 0.5.Average accuracy from 10 image sequence: 89.91% (95% CI: 86.51, 93.35)
    SpeedNot explicitly stated as an acceptance criterion, but 3.98 fps is reported.Average speed (fps): 3.98 (95% CI: 3.98, 3.99)

    2. Sample Sizes and Data Provenance:

    • Test Set:
      • Number of individuals: 22
      • Number of images: 3,999 (extracted from 2D sequences, with each individual contributing at least ten images per sequence).
      • Provenance: Prospective data collected in clinical practice from eight hospitals.
      • Country of Origin: Koreans.

    3. Number of Experts and Qualifications for Ground Truth:

    • 3 participating experts:
      • 1 anesthesiologist with >10 years of experience in pain management (for drawing GT).
      • "Other doctors" with >10 years of experience (for GT verification).
      • Doctors who scanned the ultrasound were directly involved in GT construction.

    4. Adjudication Method:

    • Manual drawing of nerve areas by an anesthesiologist. For verification of GT, "other doctors" checked every frame. If they disagreed, corrections were made to finalize the GT.

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported accuracy and speed are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-annotated rectangular regions for nerve locations.

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for training, tuning and validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • It's inferred that a similar expert-based process was used as for validation data: "The GT data were built by three participating experts. Nerve areas in all acquired images for training, tuning and validation were manually drawn by an anesthesiologist... For verification of GT, other doctors with more than 10 years of experience checked every frame... corrections were made to make the final GT."

    NerveTrack (Nerve Segmentation)

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance CriteriaReported Performance
    SegmentationAverage accuracy from nine image sequences not explicitly stated as an acceptance criterion, but 98.30% is reported. A segmentation is considered correct if Dice coefficient is ≥ 0.5.Average accuracy from nine image sequences: 98.30% (95% CI: 95.43, 100)
    SpeedNot explicitly stated as an acceptance criterion, but 3.98 fps is reported.Average speed (fps): 3.98 (95% CI: 3.98, 3.98)

    2. Sample Sizes and Data Provenance:

    • Test Set:
      • Number of individuals: 20
      • Number of images: 1,675 (extracted from 2D sequences, with each individual contributing at least ten images per sequence).
      • Provenance: Prospective data collected in clinical practice from ten hospitals.
      • Country of Origin: Koreans.

    3. Number of Experts and Qualifications for Ground Truth:

    • 3 participating experts:
      • 1 anesthesiologist with >10 years of experience in pain management (for drawing GT).
      • "Other doctors" with >10 years of experience (for GT verification).
      • Doctors who scanned the ultrasound were directly involved in GT construction.

    4. Adjudication Method:

    • Manual drawing of nerve areas (contours) by an anesthesiologist. For verification of GT, "other doctors" checked every frame. If they disagreed on nerve and other organ contours, corrections were made to finalize the GT.

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported accuracy and speed are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-annotated contours for nerve regions.

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for training, tuning and validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • It's inferred that a similar expert-based process was used as for validation data: "The GT data were built by three participating experts. Nerve areas in all acquired images for training, tuning and validation were manually drawn by an anesthesiologist... For verification of GT, other doctors with more than 10 years of experience checked every frame... corrections were made to make the final GT."
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1