    Intended Use

    The PIVO™ Pro Needle-free Blood Collection Device attaches to a peripheral IV catheter system for use to obtain venous blood specimens into a vacuum tube, blood culture bottle, or syringe from adult and pediatric patients, including those with difficult intravenous access who may have small, fragile, and/or non-palpable veins. For blood culture, the device can be used at initial placement of the peripheral IV catheter.

    Device Description

    The PIVOTM Pro Needle-free Blood Collection Device is a sterile, single use needle-free collection device that attaches to a peripheral intravascular (IV) catheter system for use to obtain venous blood specimens. The device is comprised of an inner flow tube with a slider, proximal flexible tube with female Luer, outer housing and winged clip on the distal end. The winged clip attaches to the catheter system The female Luer attaches to an evacuated tube holder or syringe. The inner flow tube is then advanced to collect a blood sample. Once complete, the inner flow tube is retracted, and the device is removed from the IV. The device is also suitable for use for blood culture collection at initial placement of peripheral IV catheters. The device is available in three sizes: 20GA, 22GA and 24GA IV compatible. The device is compatible with the corresponding IV gauge and larger IV catheters. The PIVOTM Pro Needle-free Blood Collection Device is compatible with peripheral IV catheters and extension sets with NearPortTM IV Access.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a medical device called the "PIVO™ Pro Needle-free Blood Collection Device." This document details the device's indications for use, technological characteristics, and comparison to a predicate device, along with a summary of performance tests. However, it does not describe an AI/ML powered device or a study involving human experts for ground truth establishment, MRMC studies, or training/test set sample sizes in the context of AI/ML.

    The document primarily focuses on establishing substantial equivalence for the device itself (a needle-free blood collection device), specifically for the addition of blood culture collection and an extended shelf life. The performance tests mentioned (e.g., Insertion Test, Blood Leak Test, Package Integrity) are related to the physical and functional aspects of the hardware device, not the performance of an AI/ML algorithm.

    Therefore, I cannot extract the requested information about acceptance criteria for an AI/ML powered device, the study proving its performance (in terms of AI metrics), sample sizes for AI test/training sets, expert involvement in ground truth, or MRMC studies, as this information is not present in the provided text.

    The text does state:

    • "Clinical studies are not required to demonstrate substantial equivalence to the predicate device."
    • "Summary of Performance tests completed on the subject devices were limited to those tests required Performance to support a determination of substantial equivalence to the predicate device."
    • "Per design control requirements specified in 21 CFR 820.30, the subject device met all predetermined acceptance criteria for the above-listed performance test, demonstrating substantial equivalence to the predicate device."

    This indicates that internal performance tests were conducted against pre-determined acceptance criteria, but these are for the physical device's functionality and safety, not for an AI/ML algorithm's diagnostic or predictive performance.