Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K241372

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-07-09

    (55 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

    Device Description

    The AdvisorTM HD Grid X Mapping Catheter, Sensor EnabledTM, is a sterile, single use, irrigated, high-density mapping catheter with a 7.5F shaft and an 8F distal shaft deflectable section. It is available in a D-F bi-directional curve model that is deflected using the actuator located on the catheter handle. The catheter working length is 110 cm. The device consists of a paddle-shaped distal tip with 16 electrodes, two paddle magnetic sensors, two distal shaft ring electrodes, two shaft magnetic sensors, polymer braided shaft, handle, fluid lumen extension with a luer, and an electrical connector. The catheter also has an insertion tool intended to compress and guide the distal paddle into, and withdraw from, the hemostasis valve of an introducer sheath. The catheter is intended to be used with the EnSiteTM X EP System and other accessories, including the connecting cable and commercially available irrigation pumps.

    AI/ML Overview

    This document is a 510(k) summary for the Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™. It states that the device is substantially equivalent to a predicate device and includes a summary of non-clinical testing.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly provide a table of acceptance criteria alongside specific quantitative device performance metrics. Instead, it states:

    Acceptance Criteria (General Statement)Reported Device Performance (General Statement)
    Bench design verification activities were performed and met their respective acceptance criteria.Testing included dimensional, visual, mechanical integrity, simulated use, electrical, packaging, and shelf-life testing to assess substantial equivalence. Additionally, biocompatibility testing was conducted per the voluntary consensus standards under ISO 10993 and met all the biocompatibility endpoints for a medical device category for an external communicating medical device with limited (≤24 hour) circulating blood contact.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes "bench design verification activities" and "biocompatibility testing." These are typically conducted in a lab environment using manufactured samples of the device.

    • Sample Size for Test Set: Not explicitly stated in terms of number of devices or data points, but implies standard engineering and biocompatibility test sample sizes.
    • Data Provenance: The tests are "bench design verification activities" and "biocompatibility testing." This typically refers to retrospective testing on manufactured devices, not prospective clinical studies with human or animal subjects. The origin is likely the manufacturer (Abbott Medical, Plymouth, MN, USA).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. Bench and biocompatibility testing typically rely on standardized protocols and measurements, rather than expert consensus on "ground truth" in the way clinical studies might.

    4. Adjudication Method for the Test Set

    This information is not provided. Adjudication methods like 2+1 or 3+1 are relevant for clinical trials involving human interpretation of data, which is not the type of testing described here.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document explicitly states "Non-Clinical Testing Summary" and focuses on bench and biocompatibility tests. It does not mention any clinical studies, MRMC studies, or human reader performance.

    6. Standalone (Algorithm Only) Performance Study

    No. The device is a physical catheter, not an AI algorithm. Therefore, a standalone algorithm performance study is not applicable.

    7. Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The ground truth for the non-clinical testing would be the physical and chemical specifications and performance standards outlined in the design and verification protocols for the device components and overall system, as well as the established safety criteria in ISO 10993 for biocompatibility. This is not "expert consensus," "pathology," or "outcomes data" in the clinical sense.

    8. Sample Size for the Training Set

    This information is not applicable. The device is a physical medical instrument, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1