Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K241324

    Validate with FDA (Live)

    Date Cleared
    2024-10-23

    (166 days)

    Product Code
    Regulation Number
    866.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LifeScale AST system is a multiplexed in vitro diagnostic test that uses a microfluidic sensor and resonant frequency to calculate organism concentration and/or mass distribution for quantitative antimicrobial susceptibility testing (AST). Testing is performed directly on blood cultures signaled as positive by a continuous monitoring blood culture system and confirmed by Gram stain. The LifeScale AST system does not provide organism identification and is not indicated for use with polymicrobial samples. Interpretive results (Susceptible/Intermediate/Susceptible-dose dependent/Resistant) are provided for specific drug/organism combinations. Results are intended to be used in conjunction with other clinical and laboratory findings. Standard laboratory protocols for processing positive blood cultures should be followed to ensure availability of isolates for supplemental testing as needed. Additionally, subculture of positive blood culture is necessary for the susceptibility testing of organisms present in polymicrobial samples, for testing antimicrobial agents and species not indicated for testing with the device and for epidemiologic testing and for recovery of organisms present in microbial samples.

    The LifeScale Gram Negative Kit (LSGN) is intended for use with the LifeScale AST system for in vitro testing of positive blood culture samples confirmed by Gram stain as containing gram-negative bacilli for the antimicrobial agents and specific target organisms identified below:

    1. Ampicillin: Escherichia coli
    2. Aztreonam: Escherichia coli, Klebsiella pneumoniae, Klebsiella aerogenes, Klebsiella oxytoca
    3. Cefazolin: Escherichia coli, Klebsiella pneumoniae, Klebsiella variicola
    4. Ceftazidime: Acinetobacter spp. (other than Acinetobacter ursingii), Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella variicola, Klebsiella pneumoniae, Pseudomonas aeruginosa
    5. Ertapenem: Escherichia coli, Klebsiella aerogenes, Klebsiella pneumoniae, Klebsiella oxytoca
    6. Trimethoprim-Sulfamethoxazole: Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella variicola
    7. Amikacin: Acinetobacter spp., Escherichia coli, Klebsiella pneumoniae, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella variicola, Pseudomonas aeruqinosa
    8. Cefepime: Escherichia coli, Klebsiella pneumoniae, Klebsiella oxytoca, Pseudomonas aeruginosa
    9. Ceftazidime-avibactam: Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca
    10. Gentamicin: Escherichia coli, Klebsiella pneumoniae, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella variicola, Pseudomonas aeruginosa
    11. Levofloxacin: Escherichia coli, Klebsiella pneumoniae, Klebsiella derogenes, Klebsiella oxytoca, Pseudomonas aeruginosa
    12. Meropenem: Acinetobacter spp., Escherichia coli, Klebsiella pneumoniae, Klebsiella oxytoca, Pseudomonas aeruginosa
    13. Meropenem-vaborbactam: Escherichia coli, Klebsiella pneumoniae, Klebsiella aerogenes, Klebsiella oxytoca
    14. Piperacillin-tazobactam: Acinetobacter spp., Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa
    Device Description

    The Affinity Biosensors LifeScale Gram Negative Kit (LSGN) is a semi-automated instrument system for antimicrobial susceptibility testing (AST) directly from positive blood cultures for which the Gram stain shows gram-negative bacilli. The system uses a microfluidic sensor that detects organisms in suspension and measures differences in cell mass between bacterial suspensions incubated in the presence and absence of antibiotic. Minimum inhibitory concentrations (MICs) are determined from data obtained during sample measurement including organism concentration and/or cell mass distributions of individual organisms. The system automatically interprets the measurements to determine MIC values and interpretive results (susceptible, intermediate, or resistant) based on FDA-defined or recognized breakpoints. The organism identification determined using a platform FDA-cleared for use with positive blood culture samples is entered by the user. If the organism identification has not been entered or if the sample has not been confirmed as monomicrobial, the system provides a preliminary report that indicates that organism identification or monomicrobial status is pending. The device Instructions for Use indicates that the preliminary laboratory report should not be reported to the healthcare provider. The final report is provided to the healthcare provider when the organism identification is entered into the system and the culture is confirmed to be monomicrobial samples should not be tested with the LifeScale LSGN Kit. Preliminary results are available in most cases within four hours from initiation of the assay.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study details for the LifeScale Gram Negative Kit (LSGN) with the LifeScale AST system, based on the provided FDA 510(k) submission:

    1. Table of Acceptance Criteria and Reported Device Performance (Clinical Performance Data)

    The acceptance criteria for clinical performance are implicitly demonstrated through the achieved Essential Agreement (EA), Category Agreement (CA), Very Major Discrepancy (VMJ), Major Discrepancy (MAJ), and Minor Discrepancy (MIN) rates. While specific pass/fail thresholds for each of these are not explicitly stated as "acceptance criteria" in a single table, the summary of performance data (Table 3) and subsequent detailed descriptions for each antimicrobial agent confirm the performance was considered acceptable, with a few specified limitations. Generally, for most AST devices, high EA and CA (typically >90-95%) and low VMJ/MAJ rates (typically <3%) are desired.

    For reproducibility, an acceptance criterion of "≥95%" agreement for EA was used for stability, and acceptable "Best-case (≥95%) and Worst-case (>89%) reproducibility" was stated for the reproducibility study itself.

    For inoculum density, the study aimed to show that organism concentration does not impact performance and that the device terminates tests with insufficient growth.

    For media equivalence, an EA agreement of "≥95% for the various blood culture media compared to LifeScale AST mode MICs from the BD BACTEC Standard Aerobic Media" was the acceptance criterion.

    For interfering substances, "EA ≥ 95%" was acceptable.

    Below is a summarized table of clinical performance for the additional claimed antimicrobial/organism combinations and the performance for the removal of limitations, as these were the focus of the submission. The table combines various performance metrics from Tables 3 and 12, focusing on the overall performance for each drug/organism group. Specific criteria are often tied to regulatory guidelines (e.g., CLSI, FDA), which this document notes were used.

    Antimicrobial-Organism Combination (Type of Claim)Total Evaluated (n)Essential Agreement (EA%)Category Agreement (CA%)VMJ (%)MAJ (%)MIN (%)Acceptance Criteria (Implicit)Reported PerformanceNotes
    New Claims (Table 3):
    Amikacin-Acinetobacter spp.77100.0%97.4%0.00%0.00%2.60%High EA/CA, Low VMJ/MAJMet
    Amikacin-E. coli, K. aerogenes, K. oxytoca, K. pneumoniae, K. variicola48097.3%95.8%0.00%0.75%3.54%High EA/CA, Low VMJ/MAJMet
    Amikacin-Pseudomonas aeruginosa5998.3%94.9%0.00%0.00%5.08%High EA/CA, Low VMJ/MAJMet
    Cefepime-E. coli, K. aerogenes, K. oxytoca, K. pneumoniae44592.6%95.7%0.00%1.07%3.60%High EA/CA, Low VMJ/MAJMet
    Cefepime-Pseudomonas aeruginosa10193.1%84.2%10.34%18.06%0.00%High EA/CA, Low VMJ/MAJNot MetLimitation proposed due to high VMJ/MAJ at 4 µg/mL.
    Ceftazidime-avibactam-E. coli, K. aerogenes, K. oxytoca30399.0%99.0%0.00%1.17%0.00%High EA/CA, Low VMJ/MAJMet
    Gentamicin-E. coli, K. aerogenes, K. oxytoca, K. pneumoniae, K. variicola48098.8%97.7%0.79%0.57%1.67%High EA/CA, Low VMJ/MAJMet
    Gentamicin-Pseudomonas aeruginosa5994.9%93.2%0.00%0.00%6.78%High EA/CA, Low VMJ/MAJMet
    Levofloxacin-E. coli, K. aerogenes, K. oxytoca, K. pneumoniae43798.2%96.6%0.00%0.00%3.43%High EA/CA, Low VMJ/MAJMet
    Levofloxacin-Pseudomonas aeruginosa10196.0%88.1%0.00%0.00%11.88%High EA/CA, Low VMJ/MAJMetAll categorical errors were minor.
    Meropenem-Acinetobacter spp.7896.2%97.4%0.00%0.00%2.56%High EA/CA, Low VMJ/MAJMet
    Meropenem-E. coli, K. oxytoca, K. pneumoniae39291.6%96.7%3.25%0.38%2.04%High EA/CA, Low VMJ/MAJNot MetLimitation proposed for K. oxytoca when MIC is 0.5 µg/mL due to high VMJ.
    Meropenem-Pseudomonas aeruginosa5996.6%89.8%0.00%0.00%10.17%High EA/CA, Low VMJ/MAJMetEA of evaluable results was 94.9%.
    Meropenem-vaborbactam-E. coli, K. aerogenes, K. oxytoca, K. pneumoniae44295.0%94.6%0.00%0.55%4.98%High EA/CA, Low VMJ/MAJMet
    Piperacillin-tazobactam-Acinetobacter spp.8992.1%94.4%0.00%0.00%5.62%High EA/CA, Low VMJ/MAJMet
    Piperacillin-tazobactam-E. coli, K. pneumoniae39190.8%92.1%1.52%1.58%6.38%High EA/CA, Low VMJ/MAJNot MetLimitation proposed for K. pneumoniae when MIC is 16 µg/mL due to unacceptable CA.
    Piperacillin-tazobactam-Pseudomonas aeruginosa18594.1%93.5%1.49%0.89%5.41%High EA/CA, Low VMJ/MAJMet
    K211815 Limitation Removals (Table 12):
    Ertapenem-E. coli (Off-line)9490.4%90.4%0.00%1.35%8.51%High EA/CA, Low VMJ/MAJMet (combined data)Removed limitation.
    Ertapenem-E. coli (On-line)10897.2%96.3%0.00%2.08%2.78%High EA/CA, Low VMJ/MAJMet (combined data)
    Aztreonam-K. pneumoniae16090.0%94.4%1.32%2.41%3.75%High EA/CA, Low VMJ/MAJMet (combined data)Removed limitation.
    Ceftazidime-K. pneumoniae16096.3%96.9%1.27%2.53%1.25%High EA/CA, Low VMJ/MAJMet (combined data)Removed limitation.
    Ceftazidime-Acinetobacter spp.11894.9%98.3%0.00%0.00%1.69%High EA/CA, Low VMJ/MAJMet (combined data)Revised limitation to A. ursingii only.
    Cefazolin-E. coli20397.0%87.7%1.71%1.37%10.83%High EA/CA, Low VMJ/MAJMet (combined data)Removed limitation.
    Ertapenem-K. pneumoniae15897.5%97.5%1.41%0.00%1.90%High EA/CA, Low VMJ/MAJMet (combined data)Removed limitation.

    2. Sample Size Used for the Test Set and Data Provenance

    The test set consisted of both prospective clinical blood cultures (PBCs) and contrived blood cultures.

    • Clinical Samples: 986 total clinical tests were initiated (Table 2). The specific breakdown by drug/organism combination for clinical samples is embedded within the detailed performance sections (e.g., "465 clinical (75.5%)" for Amikacin, indicating 465 clinical samples for Amikacin).
      • Provenance: Collected at 6 US Clinical sites.
      • Retrospective/Prospective: Prospective.
    • Contrived Samples: A total of 3307 analytical tests were initiated, which includes contrived samples (Table 2). For individual drug/organism combinations, the number of challenge samples (e.g., "151 challenge (24.5%)" for Amikacin) contributes to the overall number of samples evaluated.
      • Provenance: Prepared from frozen isolates supplied by Affinity Biosensors or contemporary isolates collected by the laboratory.
      • Retrospective/Prospective: Primarily prospective (prepared and tested during the study).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number of experts or their qualifications for establishing ground truth. However, it indicates:

    • Ground Truth Method: The reference method for Antimicrobial Susceptibility Testing (AST) was Broth Microdilution (BMD), performed according to CLSI guidance (CLSI M07).
    • Organism Identification: Organism ID for prospective samples was performed using an FDA-cleared direct from positive blood culture ID system, and confirmed by MALDI (matrix-assisted laser desorption/ionization). If there was a discordant identification, MALDI was considered the final identification.
    • Reference Testing Location: Reference testing was performed at two trial sites that received isolates from the PBC purity panel.
    • This suggests that trained laboratory personnel with expertise in clinical microbiology and adherence to CLSI standards established ground truth.

    4. Adjudication Method for the Test Set

    The document states:

    • "Each sample submitted for BMD testing was assigned a unique Trial ID, and LifeScale results were kept blinded to prevent bias."
    • Performance was evaluated by comparing quantitative (MIC) and qualitative (S/I/SDD/R) AST results generated by the LifeScale AST System with those of the reference BMD.
    • No explicit "adjudication method" (like 2+1 or 3+1 expert review) for resolving discrepancies between the device and ground truth is described. The comparison appears to be a direct assessment against the BMD reference. Discrepancies (VMJ, MAJ, MIN) are reported, but a formal adjudication process for these discrepancies by additional experts is not detailed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study focuses on the performance of the automated device (LifeScale AST system) compared to a laboratory reference method (BMD), rather than comparing human reader performance with and without AI assistance. This device is an automated AST system, not an AI-assisted diagnostic tool for human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    Yes, the study primarily assessed standalone performance of the LifeScale AST system. The system automatically interprets measurements to determine MIC values and interpretive results. While a user enters the organism identification, the core AST determination is an automated function of the device, making this a standalone performance assessment in the context of AST. The phrase "algorithm only without human-in-the loop performance" applies to the AST result generation itself.

    7. The Type of Ground Truth Used

    The primary ground truth used was FDA-recognized reference Broth Microdilution (BMD) method, following CLSI guidance (CLSI M07). For organism identification, MALDI was the definitive method.

    8. The Sample Size for the Training Set

    The document does not provide information on the sample size for the training set. This submission is a 510(k) for an addition of claims to an already cleared device (K211815). The focus is on the performance evaluation for the new claims and removal of limitations, implying that the underlying algorithms were likely developed and validated prior to this specific submission.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established, as it pertains to the development of the device's algorithms. As mentioned above, this document focuses on the performance of the device for new claims, not its initial development.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1