Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K241164

    Validate with FDA (Live)

    Date Cleared
    2024-09-06

    (133 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IB3D™ PL Spinal System is indicated for use in lumbar spinal fusion procedures for patients diagnosed with Degenerative Disc Disease (DDD) at 1 or 2 contiguous levels from L2 to S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. When used for these indications, the IB3D™ PL Spinal System is intended for use with supplemental fixation systems cleared for use in the lumbar spine.

    Additionally, the IB3D™ PL Spinal System can be used to provide anterior column support in patients diagnosed with degenerative scoliosis as an adjunct to pedicle screw fixation.

    All patients should be skeletally mature and have had at least 6 months of nonoperative treatment. The IB3D™ PL Spinal System is intended to be used with autogenous bone and/or allograft bone graft comprised of cancellous and/or corticocancellous bone graft, and/or demineralized allograft bone marrow aspirate when the subject device is used as an adjunct to fusion. These implants may be implanted via an open or a minimally invasive posterior or transforaminal approach. When implanting via posterior approach (PLIF), a minimum of two implants is required per spinal level.

    Device Description

    The IB3D™ PL Spinal System implants are inter-somatic spacers manufactured by additive manufacturing (Direct Laser Metal Sintering) from Titanium alloy Ti-6Al-4V ELI powder, according to ASTM F3001 and ASTM F136.

    The IB3D™ PL Spinal System implants are intended for insertion between two adjacent vertebrae by a posterior or a transforaminal approach on the lumbar spine only.

    The subject IB3D™ PL Spinal System interbody devices are available in a variety of heights and lordosis angles for treatment of lumbar interbody fusion procedure. The implant is designed with a large hollow region in the center to house autograft or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate. The design incorporates hexalock macro-rough surface on the superior and inferior surfaces of the device along with angular teeth to prevent expulsion from the interbody space.

    AI/ML Overview

    This is a medical device submission, not an AI/ML device submission. Therefore, it does not contain information about acceptance criteria, test sets, ground truth, or training sets typical for AI/ML performance evaluation.

    The provided document describes the IB3D™ PL Spinal System, an intervertebral body fusion device. The acceptance criteria and supporting studies for this type of device focus on mechanical performance, biocompatibility, and manufacturing quality, not diagnostic accuracy or AI algorithm performance.

    Here's a breakdown of the relevant information from the document:

    1. A table of acceptance criteria and the reported device performance:

    The document mentions several non-clinical tests performed to support substantial equivalence. These tests serve as the basis for demonstrating the device meets certain performance criteria. However, explicit "acceptance criteria" presented in a table format with corresponding "reported device performance" values are not detailed in this summary.

    Non-clinical tests performed in support of substantial equivalence:

    Test NameStandard/Method
    Mechanical Testing
    Static and Dynamic Axial CompressionASTM E2077
    Compression Shear TestingASTM E2077
    Subsidence TestingASTM F2267
    Impaction TestingISO 23089-2 (recommended)
    Particulate and wear analysisASTM F1877

    The summary states that these tests were performed on "worst-case constructs" of the IB3D™ PL Spinal System. The implication is that the device met the performance requirements of these standards, demonstrating substantial equivalence to its predicates.

    2. Sample size used for the test set and the data provenance:

    For mechanical and material tests of this nature, "sample size" typically refers to the number of test articles (implants) subjected to testing. This information is not specified in the provided 510(k) summary. The document does not describe patient data (e.g., country of origin, retrospective/prospective) because no clinical testing was used to support the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as the submission is for a medical device (intervertebral body fusion device) and does not involve diagnostic interpretation or AI algorithm evaluation requiring human experts to establish ground truth from medical images or clinical data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable for the same reasons as point 3.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable as this is not an AI-assisted device. The submission explicitly states: "No clinical testing was used in order to support this submission."

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This information is not applicable as this is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the non-clinical tests, the "ground truth" is defined by the specified test standards (ASTM, ISO). Meeting the criteria outlined in these standards for mechanical strength, fatigue, wear, and biocompatibility constitutes the "ground truth" for proving the device's performance characteristics and safety.

    8. The sample size for the training set:

    This information is not applicable as this is a medical device, not an AI/ML product that requires a training set.

    9. How the ground truth for the training set was established:

    This information is not applicable as this is not an AI/ML product.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1