Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240917

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-01

    (181 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Esperance 3+ Aspiration Catheter with the Wallaby Aspiration Tubing set and a compatible aspiration pump is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Esperance® 3+ Aspiration Catheter is a single-use, vascular catheter consisting of a single lumen, variable stiffness, composite catheter. The device has a tapered shaft with tapered inner diameter (ID) from 0.054'' (proximal) to 0.041'' (distal) and tapered outer diameter (OD) from 0.066'' (proximal) to 0.050'' (distal). It has five different working lengths: 120 cm, 133 cm, 145 cm, 153 cm, and 160 cm. The device is supplied as a kit with the Wallaby Aspiration Tubing Set provided with a single catheter. The distal tip of each catheter is visible under fluoroscopy and the distal shaft of each catheter is designed with an external hydrophilic coating to reduce friction during use. The proximal end of each catheter incorporates a strain relief and a standard luer adapter to facilitate the attachment of accessories. Each catheter has a semi-rigid proximal shaft which transitions into a flexible distal shaft to facilitate the advancement of the catheter in tortuous anatomy.

    The Esperance 3+ Aspiration Catheter System is a non-active, surgically invasive device intended for short term use within the neurovasculature.

    AI/ML Overview

    The document provided is a 510(k) premarket notification decision letter from the FDA for a medical device called the "Esperance 3+ Aspiration Catheter System." It primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance data (bench testing and biocompatibility).

    Based on the provided text, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document lists various performance tests and their outcomes, all indicating that the device met the acceptance criteria. The specific numerical acceptance criteria themselves are not detailed in this public FDA letter, but the conclusion for each test is consistently "All samples met the acceptance criteria" or similar.

    Test CategoryTest NameTest Method SummaryReported Device Performance/Conclusion
    BiocompatibilityCytotoxicity (ISO 10993-5)MTT- L929 Cytotoxicity Study; ISO MEM elution- L929No cytotoxic effect.
    Skin Irritation (ISO 10993-10)ISO Intracutaneous IrritationNo sensitization indicated.
    Sensitization (ISO 10993-10)ISO Guinea Pig Maximization SensitizationNo sensitization indicated.
    Systemic Toxicity (ISO 10993-11)ISO Acute Systemic Toxicity; ISO Material Mediated Rabbit Pyrogen (GLP); Systemic Toxicity - Acute Systemic InjectionNo acute systemic toxicity indicated. Esperance 3+ Aspiration Catheter is deemed non-pyrogenic. Test articles are deemed non-pyrogenic (for Introducer Sheath). RHV is deemed non-pyrogenic (for Rotating Hemostasis Valve).
    Hemocompatibility (ISO 10993-4)Hemolysis (ASTM method) Indirect Extract; ASTM Hemolysis - Direct Contact and Extract Method (GLP); Complement Activation; Thromboresistance Evaluation; Partial Thromboplastin TimeNo hemolysis indicated. Esperance 3+ Aspiration Catheter is deemed comparable to predicate (for Complement Activation, Thromboresistance Evaluation, Partial Thromboplastin Time).
    SterilizationSterility Assurance Level (SAL)Sterilization cycle verified to ensure a SAL of 10^-6 per EN ISO 11135:2014 and AAMI TIR28:2009. Validated via the half cycle method.Met the SAL of 10^-6.
    Shelf LifeAccelerated Aging (AA)Testing on devices subjected to AA process to represent 1 year of aged units. Aging studies for packaging integrity, seal strength, and device functionality.Design and performance specification requirements were met after one year shelf life. Catheters and packaging remain functional for the labeled use by date. All acceptance criteria met.
    Non-Clinical BenchDimensional VerificationDimensions of the catheter and introducer sheath are measured.All samples met the acceptance criteria.
    NavigabilityDevice tested for its ability to reach target site in an anatomical model in comparison to the predicate.All samples met the acceptance criteria.
    Clot RemovalSubject device tested for its ability to aspirate clots in an anatomical model in comparison to the predicate.All samples met the acceptance criteria.
    Kink ResistanceCatheter tested at different locations for its ability to bend to clinically relevant radii without kinking in comparison to predicate.All samples met the acceptance criteria.
    LubricityCatheter's hydrophilic coating lubricity tested by applying force to the coated section and measuring frictional force in comparison to the predicate.All samples met the acceptance criteria.
    Coating IntegrityCoated length of the catheter inspected for defects post simulated use in comparison to the predicate.All samples met the acceptance criteria.
    Torque StrengthCatheter and predicate evaluated for torque strength by rotating the test sample within an anatomical model until failure while the distal tip was not free to rotate.All samples met the acceptance criteria.
    Delivery and Retrieval ForceSubject device tested for its ability to reach and be retracted from a target site in an anatomical model with application of force below a specified value. Compared to predicate device.All samples met the acceptance criteria.
    Vacuum ResistanceDevice tested for its ability to withstand a specified vacuum pressure for a specified time without damage, lumen collapse, or kink.All samples met the acceptance criteria.
    Aspiration Flow RateAspiration flow rate of the subject device at a specified vacuum pressure measured in comparison to the predicate.All samples met the acceptance criteria.
    Elongation to FailureCatheter elongation at break obtained from the shaft tensile testing data.All samples met the acceptance criteria.
    Tip StiffnessCatheter tip tested by bending in a test fixture and measuring the maximum load that caused deflection. Compared to the predicate.All samples met the acceptance criteria.
    Atraumatic Distal TipCatheter tip inspected for smoothness.All samples met the acceptance criteria.
    Tip Shaping AbilityDistal tip shaped using the shaping mandrel supplied and assessed for damage and ability to hold tip shape.All samples met the acceptance criteria.
    Introducer Sheath CompatibilitySupplied introducer sheath tested for compatibility with the Esperance 3+ Aspiration Catheter System.All samples met the acceptance criteria.
    Device CompatibilityAppropriately sized guidewire delivered and retrieved through the catheter. Catheter delivered and retrieved through an appropriately sized sheath.All samples met the acceptance criteria.
    Surface DefectsCatheter and introducer sheath examined under magnification for extraneous matter.All samples met the acceptance criteria.
    Tensile ForcePeak tensile force of the subject catheter measured at different locations in comparison to the predicate.All samples met the acceptance criteria.
    Liquid LeakageSubject catheter tested for leakage per ISO 10555-1 and compared to the reference device.All samples met the acceptance criteria.
    Air LeakSubject catheter tested for air leakage per ISO 10555-1 and compared to the reference device.All samples met the acceptance criteria.
    Static BurstSubject catheter tested to withstand a specified static pressure.All samples met the acceptance criteria.
    Power InjectionDistal tip of the catheter blocked, and fluid injected into the lumen using a power injector until the catheter burst.All samples met the acceptance criteria.
    CorrosionSubject device visually inspected for signs of corrosion post exposure to required conditions per ISO 10555-1.All samples met the acceptance criteria.
    Particulate TestingSize and number of particulates generated during simulated use of the device in a neurovascular model were measured and calculated. Particulate generation compared to the reference device.Particulate generation was similar between the subject and reference device.
    Hub/Luer FittingCatheter hub tested as per ISO 80369-7.All samples met the acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document dedicates a section to "PERFORMANCE TESTING - BENCH" and concludes most tests with "All samples met the acceptance criteria." However, it does not specify the sample size used for each of these non-clinical tests.

    The data provenance is through non-clinical bench testing performed by the manufacturer, Wallaby Medical. There's no information about the country of origin of the data within the document, beyond the sterilization facility being in "Suzhou) Ltd., Jiangsu, China." The nature of the tests (bench testing, biocompatibility, sterilization, shelf life) indicates these are prospective tests performed specifically for this submission, rather than retrospective analysis of existing data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable as the document describes non-clinical bench performance testing, not a clinical study involving human or image-based data that would require expert ground truth labeling. The "ground truth" for these tests is based on objective measurements and established engineering standards (e.g., ISO, ASTM).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as the document describes non-clinical bench performance testing where adjudication methods typical for clinical or image-based studies are not relevant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done or reported. The device is a medical catheter and not an AI/imaging algorithm that would typically involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a physical medical catheter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests described, the "ground truth" is based on:

    • Engineering specifications and standards: Adherence to ISO standards (e.g., ISO 10993 for biocompatibility, ISO 11135 for sterilization, ISO 10555-1 and 80369 for bench tests) and relevant FDA guidance documents.
    • Objective measurements: Dimensional verification, force measurements, flow rates, visual inspections for defects, etc.
    • Comparison to predicate/reference device: Many tests involved comparing the subject device's performance (e.g., navigability, clot removal, lubricity, particulate generation) to the predicate or reference device to establish similarity and ensure similar safety and effectiveness profiles.

    8. The sample size for the training set

    This is not applicable. The document describes the testing of a physical medical device, not an AI or machine learning model that would require a training set.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1