    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ID Free® Personal Lubricant is for or penile, anal and/or vaginal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprenecondoms. This product is not compatible with polyurethane condoms.

    Device Description

    ID Free® Personal Lubricant is a personal lubricant, for penile, anal and/or vaginal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. The product is intended for over-the-counter use.

    The device is a non-sterile water based personal lubricant with a clear, odorless formulation. The device is packaged in nonsterile 8.5 fl. oz/250 ml plastic/PET bottles with a screw-on cap and flip top closure, or in 0.14 fl. oz/4 ml foil sachets.

    The device is compatible with natural rubber latex and polyisoprene condoms. It is not compatible with polyurethane condoms. This product is not a contraceptive and does not contain a spermicide.

    The device formulation consists of water, propanediol, hydroxyethyl cellulose, carbomer, PEG-45M, tetrahydroxypropyl ethylenediamine, EDTA, caprylhydroxamic acid, and sucralose.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets those criteria:

    Device: ID Free® Personal Lubricant
    Regulation Name: Condom (as a personal lubricant, product code NUC)


    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Specification)Reported Device Performance (as implied by successful testing)
    ColorNo color to slight yellow tintMet specification (implied, no specific value given)
    AppearanceClear viscous gelMet specification (implied, no specific value given)
    OdorOdorlessMet specification (implied, no specific value given)
    Viscosity (Spindle #63 @ 30 RPM)2,200 - 4,400 cpsMet specification (implied, no specific value given)
    pH5.0 - 5.5Met specification (implied, no specific value given)
    Osmolality250 - 500 mOsm/kgMet specification (implied, no specific value given)
    Total yeast/mold count (TYMC) (USP <61>)< 10 cfu/mLMet specification (implied, no specific value given, "Met USP <51> acceptance criteria for Category 2 products" implies passing USP <61> as well)
    Total aerobic microbial count (TAMC) (USP <61>)< 100 cfu/mLMet specification (implied, no specific value given, "Met USP <51> acceptance criteria for Category 2 products" implies passing USP <61> as well)
    Presence of Pathogenic Organisms (USP <62>) (Staphylococcus aureus, Pseudomonas aeruginosa, and Candida albicans)AbsentMet specification (implied, no specific value given, "Met USP <51> acceptance criteria for Category 2 products" implies passing USP <62> as well)
    Antimicrobial effectiveness (USP <51>)Meets USP <51> acceptance criteria for Category 2 productsMet specification (explicitly stated: "Meets USP <51> acceptance criteria for Category 2 products")
    Biocompatibility: Cytotoxicity (ISO 10993-5:2009)Non-cytotoxicDemonstrated non-cytotoxic
    Biocompatibility: Guinea Pig Maximization Sensitization (ISO 10993-10:2010)Non-sensitizingDemonstrated non-sensitizing
    Biocompatibility: Vaginal Irritation (ISO 10993-10:2010)Non-irritatingDemonstrated non-irritating
    Biocompatibility: Acute Systemic Toxicity (ISO 10993-11:2017)Not systemically toxicDemonstrated not systemically toxic
    Shelf Life (24 months at 25°C)Device met specifications in Table 1 at all time pointsMet specifications in Table 1 across 24-month shelf-life
    Condom Compatibility (Natural Rubber Latex) (ASTM D7661-18)CompatibleCompatible
    Condom Compatibility (Polyisoprene) (ASTM D7661-18)CompatibleCompatible
    Condom Compatibility (Polyurethane)Not compatible (stated directly)Not compatible

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify exact sample sizes for each test in phrases like "The results of this testing demonstrate..." or "The results of the testing demonstrate that...". It refers to the standards (e.g., USP <61>, USP <62>, USP <51>, ISO 10993 series, ASTM D7661-18) which would define the sample sizes required for each specific test.

    • Data Provenance: The studies are non-clinical performance tests conducted for regulatory submission. There is no mention of country of origin for the data or whether it was retrospective or prospective in terms of patient data, as these are in vitro and in vivo animal (for some biocompatibility) tests, not human trials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for this medical device (a personal lubricant) is established by adherence to recognized international and national standards and test methods (e.g., ISO, ASTM, USP) for specific physical, chemical, microbiological, and biological properties, rather than expert consensus on a diagnostic outcome. The "experts" would be the scientists and technicians conducting the tests according to the specified methodologies.

    4. Adjudication Method for the Test Set

    Not applicable for this type of device and testing. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies or human-reader performance evaluations, particularly for image-based diagnostics. The performance of this device is determined by quantitative and qualitative measurements against established specifications in laboratory settings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is not relevant for this device. It is a type of study used to evaluate diagnostic imaging systems where multiple human readers interpret cases, often with and without AI assistance, to assess the impact of AI on diagnostic performance. This device is a personal lubricant, not a diagnostic tool.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This device is a physical product (personal lubricant), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by:

    • Defined Specifications: Physical and chemical properties (color, appearance, odor, viscosity, pH, osmolality) have explicit numerical ranges or descriptions that must be met.
    • Standardized Test Methods and Criteria:
      • Microbiological: Adherence to USP <61> (Total microbial counts) and USP <62> (Absence of specified objectionable microorganisms), and USP <51> (Antimicrobial Effectiveness Test) acceptance criteria. These standards define the "ground truth" for microbiological safety.
      • Biocompatibility: Adherence to ISO 10993 series standards outcomes (non-cytotoxic, non-sensitizing, non-irritating, not systemically toxic).
      • Condom Compatibility: Adherence to ASTM D7661-18 for compatibility with specific condom materials.
    • Real-time Stability Data: The device maintained its specifications over a 24-month shelf-life at a specified condition (25°C).

    8. The Sample Size for the Training Set

    Not applicable. This is a physical product, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.