    Intended Use

    Model BB811 and Model BB811-NS: The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.

    Model BB841: Oxygenator with Integrated Arterial Filter: The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery. Cardiotomy/Venous Reservoir: The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures. The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.

    Device Description

    The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (BB811), Affinity Fusion Oxygenator with Integrated Arterial Filter and Cardiotomy/Venous Reservoir with Balance Biosurface (BB841), and Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (Non-Sterile) (B811-NS) are collectively referred to as the Affinity Fusion Oxygenator in the summary. The Affinity Fusion Oxygenator is intended to be used in an extracorporeal perfusion blood circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator contains both an integrated arterial filter and integrated heat exchanger. The Affinity Fusion Oxygenator is a microporous, hollow-fiber, gas-exchange devices available with Balance Biosurface bonded to the blood contacting surface of the device. The integrated arterial filter is designed to filter from the circuit microemboli larger than the specified micron size from the circuit for periods up to six hours during cardiopulmonary bypass surgery. Some models of the Affinity Fusion Oxygenator are packaged with an Affinity Fusion Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface which is designed to be an integral part of a cardiopulmonary bypass circuit for use during cardiac surgery. The Affinity Fusion CVR is designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Fusion CVR may be used during vacuum assisted venous drainage (VAVD) procedures and collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement during open heart surgery.

    AI/ML Overview

    The provided text discusses the Affinity Fusion Oxygenator and its acceptance criteria, specifically focusing on the addition of an alternate hollow fiber supplier.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    Test ItemAcceptance Criteria (Implied: PASS)Reported Device Performance
    O2 TransferPASSPASS
    CO2 TransferPASSPASS
    Blood Side Pressure DropPASSPASS
    6 Hour O2 TransferPASSPASS
    6 Hour CO2 TransferPASSPASS
    Pressure IntegrityPASSPASS*
    BurstPASSPASS
    Gas Pathway IntegrityPASSPASS
    Plasma BreakthroughPASSPASS
    Filtration EfficiencyPASSPASS
    Hemolysis (Max Flow)PASSPASS
    White Blood Cell RetentionPASSPASS
    Platelet (PLT) retentionPASSPASS
    Functional platelet retentionPASSNA (not specified as tested at this time point)
    Platelet PLT Function (min flow)PASSNA (not specified as tested at this time point)
    Prime VolumePASSPASS
    CoveragePASSPASS
    LeachingPASSPASS

    Biocompatibility Testing:

    Biocompatibility Test MethodTesting CategoryReported Device Performance
    ISO MEM Elution Cytotoxicity, L929 Mouse Fibroblast CellsCytotoxicityPASS
    ISO Maximization Sensitization Study, 0.9%Sodium Chloride & Sesame Oil, NF extractSensitizationPASS
    ISO Intracutaneous Study – 0.9% Sodium Chloride & Sesame Oil, NF extractIntracutaneous ReactivityPASS
    ISO Systemic Toxicity Study - 0.9% Sodium Chloride & Sesame Oil, NF extractAcute Systemic ToxicityPASS
    Pyrogen Study – Material Mediated -0.9%Sodium ChlorideMaterial Mediated Pyrogen StudyPASS
    ASTM In-vitro Hemolysis, Rabbit BloodHemocompatibilityPASS
    Complement Activation, SC5b-9, Human SerumHemocompatibilityPASS
    ASTM Partial Thromboplastin Time, Direct Contact, Human PlasmaHemocompatibilityPASS
    Genotoxicity, Bacterial Reverse Mutation StudyGenotoxicityPASS
    Genotoxicity: Mouse Lymphoma AssayGenotoxicityPASS
    Platelet and Leukocyte Count- with comparison articleHemocompatibilityPASS

    *Note on "Pressure Integrity": Devices conditioned only to factors significant for leaks (per Two Proportion statistical test).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not explicitly state the sample sizes for the test sets (number of oxygenator units tested) for design verification or biocompatibility. It also does not specify the country of origin of the data or whether the studies were retrospective or prospective. The nature of the tests (engineering performance and biocompatibility) suggests these were primarily laboratory-based prospective tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not applicable to the type of device and testing described. The "ground truth" for a medical device like an oxygenator is established through objective performance measurements and standardized biocompatibility tests, not subjective expert assessment as would be the case for image interpretation or diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies or studies involving subjective human interpretation (e.g., radiology reads) to resolve discrepancies. The tests described here are objective performance and safety tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is an oxygenator, not an AI-powered diagnostic tool requiring human reader studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable. The device is an oxygenator, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    The ground truth for this device is based on:

    • Engineering specifications and standards: For performance characteristics like O2 Transfer, CO2 Transfer, Pressure Drop, Burst, Filtration Efficiency, etc. "PASS" implies meeting predefined, objective measurable criteria.
    • International standards for biocompatibility (ISO 10993-1:2018): For tests like cytotoxicity, sensitization, genotoxicity, hemocompatibility, etc. "PASS" means meeting the established safety thresholds for these biological endpoints.

    8. The sample size for the training set:

    This is not applicable. This is not an AI/machine learning device that requires a training set. The "study" described is a series of laboratory-based design verification and biocompatibility tests.

    9. How the ground truth for the training set was established:

    This is not applicable, as there is no training set for this type of device.