Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240519

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-04-17

    (54 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sphere/Asphere:
    Innova Vision Sphere and Asphere (inifilcon A) Silicone Hydrogel Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    Toric:
    Innova Vision Toric (inifilcon A) Silicone Hydrogel Soft Contact lenses are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -2.25 diopters.

    Multifocal:
    Innova Vision Multifocal (inifilcon A) Silicone Hydrogel Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with add powers from +0.25 to +4.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    Multifocal Toric:
    Innova Vision Multifocal Toric (inifilcon A) Silicone Hydrogel Soft Contact lenses are indicated for the optic correction of distance and near vision in presbyopic phakic persons with nondiseased eyes in powers of -20.00 to +20.00 diopters with add powers from +0.25 to +4.00 diopters and astigmatism corrections from -0.25 to -2.25 diopters.

    Eye Care Practitioners may prescribe the Innova Vision (inifilcon A) Silicone Hydrogel (Hydrophilic) Daily Wear Soft Contact lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement or for single-use disposable wear.

    When prescribed for frequent/planned replacement, the Innova Vision (inifilcon A) Silicone Hydrogel (Hydrophilic) Daily Wear Soft Contact Lens is to be cleaned, rinsed and disinfected each time the lens is removed. The contact lens is to be discarded after the recommended wearing period as prescribed by the Eye Care Professional. When prescribed for frequent/planned replacement wear, the lenses may be disinfected using a chemical disinfection only.

    When prescribed for single-use disposable wear, Innova Vision (inifilcon A) Silicone Hydrogel (Hydrophilic) Daily Wear Soft Contact Lens is to be discarded after each removal.

    Device Description

    The Innova Vision (inifilcon A) Silicone Hydrogel (Hydrophilic) Daily Wear Soft Contact Lens is a hemispherical shell with a molded spherical base curve and a molded front surface. The hydrophilic characteristics allow aqueous solutions to enter the lenses are fabricated from inifilcon A, which is a random co-polymer of silicone containing monomers and hydrophilic monomers. The lens consists of 60.0% inifilcon A and 40.0% water by weight when immersed in saline solution. The inifilcon A name has been adopted by the United States Adopted Names Council (USAN).

    The Innova Vision (inifilcon A) Silicone Hydrogel (Hydrophilic) Daily Wear Soft Contact Lens contains C.I. Reactive Blue 247 (21 CFR Part 73.3100 ) for visibility and handling. The Innova Vision (inifilcon A) Silicone Hydrogel (Hvdrophilic) Daily Wear Soft Contact Lens incorporates a benzotriazole UV blocking monomer to help protect against transmission of harmful UV radiation. The lens blocks >95% in the UVB range (280nm - 315nm), and >50% in the UVA range (315mm - 380mm). The Innova Vision (inifilcon A) Silicone Hydrogel (Hydrophilic) Daily Wear Soft Contact Lens is manufactured in sphere/asphere, toric, multifocal and multifocal-toric design configurations.

    AI/ML Overview

    This document is an FDA 510(k) summary for a contact lens, not an AI/ML medical device. Therefore, the information typically required for describing the acceptance criteria and study proving an AI/ML device meets those criteria (such as test sets, expert ground truth, MRMC studies, standalone performance, etc.) is not present in this document.

    The document focuses on demonstrating substantial equivalence of a new contact lens (Innova Vision (inifilcon A) Silicone Hydrogel (Hydrophilic) Daily Wear Soft Contact Lens) to legally marketed predicate devices, primarily K232139, K222954, and K210436.

    Instead of AI/ML performance metrics, the acceptance criteria for this medical device are related to its physical properties, biocompatibility, and intended use, demonstrating that it is as safe and effective as existing contact lenses.

    Here's an attempt to extract relevant information from the provided text, reinterpreting "acceptance criteria" and "study" in the context of a contact lens submission:

    Acceptance Criteria and Device Performance (Reinterpreted for a Contact Lens)

    • Acceptance Criteria: To demonstrate substantial equivalence, the new device must meet specific physical, chemical, and biological performance standards comparable to predicate devices. This includes demonstrating:

      • Absence of cytotoxicity, systemic toxicity, and ocular irritation.
      • Stability, sterility, and package integrity over the labeled shelf life.
      • Compatibility with common cleaning and disinfection solutions.
      • Consistent material properties (refractive index, water content, oxygen permeability, UV transmission, tensile strength, modulus, elongation to break, specific gravity) within acceptable tolerances.
      • Similar indications for use and general function as the predicate devices.
    • Reported Device Performance: The document states that the non-clinical performance testing (toxicology, shelf life, solution compatibility, preservative uptake and release, and bench tests) supports the claim that the Innova Vision (inifilcon A) Silicone Hydrogel (Hydrophilic) Daily Wear Soft Contact Lens is substantially equivalent to the currently marketed predicate devices. Specific numerical results demonstrating compliance with each criterion are not provided in this summary, but the conclusion confirms they were met.

    Table of Acceptance Criteria and Reported Device Performance (Reinterpreted)

    Acceptance Criterion (for Contact Lens Equivalence)Reported Device Performance
    Biocompatibility/Toxicology
    In-Vitro Cytotoxicity (ISO 10993-5)Not cytotoxic
    Systemic Toxicity (ISO 10993-11)Meets systemic injection test requirements; no acute systemic toxicity
    Acute Ocular Irritation (ISO 10993-10)No ocular irritation
    Biocompatibility of packaging materials/solutionAddressed in previously submitted assessments
    Shelf Life
    Stability, Sterility, Package IntegrityData supports establishment of proposed shelf life
    Solution Compatibility
    Physical compatibility with cleaning/disinfectionConfirmed (peroxide and MPDS)
    Preservative Uptake and ReleaseReferenced in K232139 (predicate)
    Bench Performance (Material Properties)
    Refractive IndexConsistent with predicate
    Water ContentConsistent with predicate
    Oxygen Permeability (Dk)Consistent with predicate
    % Transmission (Visible & UV)Consistent with predicate
    Tensile StrengthConsistent with predicate
    ModulusConsistent with predicate
    % Elongation to breakConsistent with predicate
    Specific GravityConsistent with predicate
    Quantification of Polymerization ResidualsConsistent with predicate

    Information Not Applicable or Not Found (as per AI/ML context):

    1. Sample size used for the test set and data provenance: Not applicable in the AI/ML sense. The "test set" here refers to the actual manufactured devices undergoing physical and biological testing. The number of batches or lenses tested for each criterion is not specified in this summary. The "data provenance" would refer to internal lab testing.
    2. Number of experts used to establish ground truth & qualifications: Not applicable. Ground truth for a contact lens is established through objective physical, chemical, and biological measurements, not expert consensus on images.
    3. Adjudication method for the test set: Not applicable. Compliance is measured objectively.
    4. Multi-Reader Multi-Case (MRMC) comparative effectiveness study: Not applicable. This is for evaluating changes in human reader performance with AI assistance.
    5. Standalone (algorithm-only) performance: Not applicable. This is a physical medical device.
    6. Type of ground truth used: For this device, the "ground truth" is derived from established standard test methods (e.g., ISO 10993 for biocompatibility, methods for Dk, water content, etc.) that yield objective measurements of the device's properties.
    7. Sample size for the training set: Not applicable. There is no AI model being trained.
    8. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document outlines the non-clinical and clinical (by reference to predicate) evidence used to demonstrate the substantial equivalence of a new contact lens. It does not contain the specific AI/ML-related details requested in the prompt because the device is a physical product, not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1