Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240330

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-07

    (184 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HALYARD SMART-FOLD Sterilization Wrap (H450 and H650) is intended to enclose another medical device that is to be sterilized by a healthcare provider using the Advanced Sterilization Products STERRAD 100NX Sterilizer ULTRA GI Cycle.

    The H450 and H650 models of the HALYARD SMART-FOLD Sterilization Wrap have been validated for use with the STERRAD 100NX ULTRA GI Cycle as follows:
    • Hydrogen peroxide compatible flexible multi-channel duodenoscopes, with no more than 4 channels, with lumen dimensions of ≥1mm inner diameter x ≤1500mm in length, or ≥2mm inner diameter x ≤1630mm in length.
    • One flexible duodenoscope per tray, and no more than two flexible duodenoscope per cycle

    Note 1: The STERRAD 100NX Sterilizer ULTRA GI Cycle was validated using a load weight of 15.4 lbs (2 x 7.7 lbs), one endoscope per shelf.
    Note 2: Only duodenoscopes that have been cleared as compatible with vaporized hydrogen peroxide are acceptable. Check STERRAD Sterilizer Cycle Selection table for ULTRA GI Cycle compatible duodenoscopes

    The HALYARD SMART-FOLD Sterilization Wrap (H450, H650) is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s) until used.

    Device Description

    Halyard Sterilization Wrap is supplied to the customer as bulk packages of single sheets, where in accordance with standard hospital practices, two sheets are then used to wrap a medical device or a collection of medical devices for sterilization. HALYARD SMART-FOLD Sterilization Wraps are comprised of two sheets of HALYARD* Sequential Sterilization Wrap. This allows for convenient wrapping with two sheets simultaneously. The HALYARD SMART-FOLD Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using the Advanced Sterilization Products, Inc. (ASP) STERRAD 100NX Sterilizer ULTRA™ GI Cycle.

    AI/ML Overview

    The provided text describes the performance testing for the HALYARD* SMART-FOLD* Sterilization Wrap (H450 and H650).

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    PurposeTestAcceptance CriteriaResult
    Sterilant Penetration/EfficacyISO 14937AAMI TIR No 12-2010Achieving a 10$^{-6}$ sterility assurance level following processing in a worst-case half-cyclePass
    Performance Testing (Pre-Sterilization and Post-Sterilization)ANSI/AAMI/ISO 11607-1 Annex BISO 13938-2ASTM D4966-12CPSC 1610Complies with the selected physical propertiesPass
    Maintenance of Package IntegrityANSI/AAMI/ISO 11607-1ANSI/AAMI ST79AAMI TIR No 12-2010ANSI/AAMI/ISO 14937Maintain sterility for up to 90 daysPass
    BiocompatibilityISO 10993-1ISO 10993-5ISO 10993-10ISO 10993-12ISO 14937Non-cytotoxicNon-irritatingResidual level ≤ 9100 µg/cm$^2$Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for each test. The data provenance is not explicitly mentioned as a country of origin, nor is it stated whether the data is retrospective or prospective. These are non-clinical (laboratory) tests by their nature.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable as the tests performed are non-clinical, performance-based, and biocompatibility assessments, which rely on defined technical standards and measurements rather than expert human interpretation of complex data (like image analysis).

    4. Adjudication Method

    This section is not applicable for the same reasons as point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. This submission is for a medical device (sterilization wrap), not an AI/software as a medical device (SaMD). Therefore, no MRMC study or AI-related effectiveness assessment was conducted.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This section is not applicable for the same reasons as point 5.

    7. The Type of Ground Truth Used

    For the tests conducted, the "ground truth" is defined by the technical standards and specifications outlined in the referenced ISO, AAMI, ASTM, and CPSC documents. For instance:

    • Sterilant Penetration/Efficacy: The ground truth is a sterility assurance level (SAL) of 10$^{-6}$. This is an objective measurement based on microbiological testing.
    • Performance Testing: The ground truth is compliance with specified physical properties (e.g., tensile strength, tear resistance as defined by the standards).
    • Maintenance of Package Integrity: The ground truth is the ability to maintain sterility for up to 90 days, which is determined through microbial barrier testing and accelerated aging studies.
    • Biocompatibility: The ground truth is material properties meeting non-cytotoxic, non-irritating, and specific residual level criteria, evaluated through chemical extractions and biological assays.

    8. The Sample Size for the Training Set

    This section is not applicable as the device is a physical sterilization wrap and does not involve machine learning or AI models with training sets.

    9. How the Ground Truth for the Training Set was Established

    This section is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1