510(k) Data Aggregation

    K Number
    K240238

    Validate with FDA (Live)

    Date Cleared
    2024-04-12

    (74 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vantage Fortian/Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

    MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

    ·Proton density (PD) (also called hydrogen density) ·Spin-lattice relaxation time (T1) ·Spin-spin relaxation time (T2) ·Flow dynamics ·Chemical Shift

    Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

    Device Description

    The Vantage Fortian (Model MRT-1550/ WK, WM, WQ)/Vantage Orian (Model MRT-1550/ A3, A4, A7, A8) is a 1.5 Tesla Magnetic Resonance Imaging (MRI) System. These Vantage Fortian/Orian models use 1.4 m short and 4.1 tons light weight magnet. They include the Canon Pianissimo Zen technology (scan noise reduction technology). The design of the gradient coil and the whole-body coil of these Vantage Fortian/Orian models provide the maximum field of view of 55 x 55 x 50 cm and include the standard gradient system.

    The Vantage Orian (Model MRT-1550/ UC, UG, UH, UK, UL, UO, UP, AK, AL, AO, AP, Upgrade only: A3, A4, A7, A8, AC, AD, AG, AH) is a 1.5 Tesla Magnetic Resonance Imaging (MRI) System. The Vantage Orian models MRT-1550/ UC, UG, UH, UK, UL, UO, UP, Upgrade only: A3, A4, A7, A8 use 1.4 m short and 4.1 tons light weight magnet while the Vantage Orian models MRT-1550/ AK, AL, AO, AP, Upgrade only: AC, AD, AG, AH use 1.4 m short and 3.8 tons light weight magnet. All of the aforementioned models include the Canon Pianissimo™ and Pianissimo Zen technology (scan noise reduction technology). The design of the gradient coil and the whole-body coil of these Vantage Orian models provide the maximum field of view of 55 x 55 x 50 cm. The Model MRT-1550/ UC, UD, UG, UH, UK, UL, UO, UP, AK, AL, AO, AP includes the XGO gradient system. The Model MRT-1550/ A3, A4, A7, A8, AC, AD, AG, AH include the standard gradient system.

    This system is based upon the technology and materials of previously marketed Canon Medical Systems MRI systems and is intended to acquire and display cross-sectional transaxial, coronal, sagittal, and oblique images of anatomic structures of the head or body. The Vantage Fortian/Orian MRI System is comparable to the current 1.5T Vantage Fortian/Orian MRI System (K222968), cleared October 25, 2022, with the following modifications.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    This table focuses on the specific features detailed in the "Testing" section of the document, as these are the ones with explicit performance criteria and evaluation results.

    Feature / MetricAcceptance CriteriaReported Device Performance
    Exsper 3D (artifact reduction)Reduction of artifacts caused by unfolding error compared to conventional SPEEDER.Confirmed that Exsper 3D reduced artifacts caused by unfolding error compared to conventional SPEEDER.
    Slice Shim (image quality for off-center slices)Images with Slice Shim equal to or better than those with Standard Shim, especially for off-center slices.Confirmed that images with the Slice Shim were equal to or better than those with the Standard Shim especially for off-center slices.
    UTE (Ultra Short TE) CG Recon (image resolution & SNR when scan time is reduced)Better maintenance of both image resolution and image SNR as compared to conventional grid recon when scan time is reduced.Confirmed CG recon performs better at maintaining both image resolution and image SNR as compared to conventional grid recon when scan time is reduced.
    Ringing Correction (ringing artifact reduction)Reduction of ringing artifacts.Confirmed Ringing Correction reduced ringing and met predetermined acceptance criteria.
    Auto Scan Assist (time and steps for slice positioning)Less time and fewer steps for slice positioning compared to slice positioning without Auto Scan Assist.Confirmed the operation of slice positioning utilizing Auto Scan Assist applications (NeuroLine+, SUREVOI Liver, LiverLine+, and W-SpineLine+) resulted in less time and less steps as compared to slice positioning without Auto Scan Assist.
    Ceiling Camera (patient orientation & anatomy position detection)Successful patient orientation detection and cases requiring no correction for successful patient anatomy position detection met predetermined acceptance criteria AND less or comparable patient setting time compared to conventional manual patient setting.Confirmed the percentage of successful patient orientation detection and cases requiring no correction for successful patient anatomy position detection met predetermined acceptance criteria. Additionally, testing confirmed the ceiling camera resulted in less or comparable patient setting time compared to conventional manual patient setting, regardless of the operator.
    PIQE (Precise IQ Engine) - Bench Testing (in-plane matrix, ringing, sharpness, contrast, SNR)Generates higher in-plane matrix from lower matrix images, contributes to ringing artifact reduction and increase of sharpness, sharper edges, mitigates smoothing and ringing effects, maintains similar or better contrast and SNR.Confirmed PIQE generates higher in-plane matrix from lower matrix image, PIQE contributes to ringing artifact reduction and increase of sharpness. Comparisons to standard clinical techniques confirmed PIQE generates images with sharper edges while mitigating the smoothing and ringing effects and maintaining similar or better contrast and SNR.
    PIQE (Precise IQ Engine) - Clinical Image Review (Likert score for various IQ metrics)Scores of 3 or above (clinically acceptable) for ringing, sharpness, SNR, overall IQ, and feature conspicuity.All resulting reconstructions (conventional and new) were scored at, or above, clinically acceptable by three board-certified Radiologists per anatomy. Reviewers exhibited strong agreement at "good" and "very good" level for all IQ metrics. Confirmed: (a) PIQE generates higher spatial in-plane resolution images (up to 3x matrix dimensions in both in-plane directions), (b) PIQE contributes to ringing artifact reduction, denoising and increased sharpness, (c) PIQE accelerates scanning by reducing acquisition matrix while maintaining clinical matrix size and image quality, (d) PIQE benefits obtained on regular clinical protocols.
    NeuroLine+ (angular error for slice alignment)Angular error for slice alignment similar or better as compared to the conventional method.For the angular error, NeuroLine+ met the acceptance criteria being similar or better as compared to the conventional method.
    NeuroLine+ (autopositioning success rate)Successful scan alignment (offset and angle within acceptable error defined as typical inter-rater variability) greater than 80% of the time.Yielded 97.5% success, which met the acceptance criteria.
    Iterative Motion Correction (IMC) (reduction of motion artifacts)Effective in reducing motion artifacts with metrics of peak SNR and structural similarity (SSIM).Demonstrated that IMC is effective in reducing motion artifacts and met predetermined acceptance criteria.
    IMC - Clinical Image Review (Likert score for IQ metrics)Scores of 3 or greater (clinically acceptable) for SNR, tissue contrast, image sharpness, and diagnostic confidence.Testing confirmed the IMC technique performs as expected, significantly reducing motion artifacts, and improving overall image quality metrics as evaluated via SNR, tissue contrast, image sharpness, and diagnostic confidence. IMC corrected images are the same as, or better than, images without IMC applied.
    IMC - Clinical Image Review (diagnostic information)Diagnostic information in IMC images the same or better than those without IMC applied.A second clinical image review with patients having pathology and motion further confirmed the diagnostic information in IMC images was the same or better than those without IMC applied.
    Free Breathing Dynamic DLR (arterial phase detection success rate)Automatic arterial phase detection success rate greater than or equal to 80%.Yielded 90.9% success (automatically proposed phases included the gold standard phase as manually selected by experienced radiologists), which met the acceptance criteria.
    Free Breathing Dynamic DLR - Clinical Image Review (Likert score for diagnostic importance)Average visual scores for overall SNR, overall IQ, feature conspicuity, and diagnostic confidence met acceptance criteria (3 or higher on a 5-point Likert scale).The average of visual scores for overall SNR, overall IQ, feature conspicuity and diagnostic confidence met the acceptance criteria. The results support the conclusion that Free Breathing Dynamic is a clinically acceptable option for the acquisition of free-breathing contrast enhanced dynamic liver exams providing acceptable diagnostic confidence.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • PIQE (Precise IQ Engine):

      • Sample Size: 36 unique subjects (patients).
      • Data Provenance: Two sites in France.
      • Nature of Data: Clinical cases, prospectively collected for the study (scanned to provide test data sets separately from training data).
    • NeuroLine+:

      • Sample Size: 15 clinical cases (4 male, 11 female).
      • Data Provenance: France (implied from national identities, although not explicitly stated as collected in France).
      • Nature of Data: Clinical images, newly collected, and entirely separate from the training group.
    • Iterative Motion Correction (IMC):

      • Bench Testing: 12 clinical datasets (without subject motion and with mathematically simulated motion added). Data provenance not specified.
      • Clinical Image Review (Phase 1): 18 volunteers. Data provenance not specified.
      • Clinical Image Review (Phase 2): 49 image volumes from 15 typical clinical patients with pathology and motion. Data provenance not specified.
      • Nature of Data:
        • Bench testing: Clinical datasets with simulated motion.
        • Clinical review (Phase 1): Volunteers imaged with and without motion.
        • Clinical review (Phase 2): Clinical patients with pathology and motion.
        • All testing data acquired separately and independently from training data.
    • Free Breathing Dynamic DLR:

      • Arterial Phase Detection: 11 clinical cases (5 male, 5 female, 1 unknown).
      • Clinical Image Review: 29 contrast-enhanced Free Breathing Dynamic liver studies from 29 patients (14 male, 14 female, 1 unknown).
      • Data Provenance: France and USA for arterial phase detection; United States, France, and Japan for clinical image review.
      • Nature of Data:
        • Arterial phase detection: Clinical images from patients receiving clinically indicated contrast.
        • Clinical image review: Contrast-enhanced Free Breathing Dynamic liver studies.
        • All testing data acquired separately and independently from training data.
    • Exsper 3D, Slice Shim, UTE CG Recon, Ringing Correction, Auto Scan Assist, Ceiling Camera:

      • The text states "phantom images" or "clinical images" were used, but specific sample sizes and data provenance for these components (beyond PIQE, NeuroLine+, IMC, and Free Breathing Dynamic DLR, which are highlighted deeper) are not detailed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • PIQE (Precise IQ Engine):

      • Number of Experts: 6 radiologists (3 per anatomy - brain and knee).
      • Qualifications: USA board certified radiologists.
    • NeuroLine+:

      • Number of Experts: 2.
      • Qualifications: Experienced ARRT licensed MR technologists (for manual annotation of target planes).
    • Iterative Motion Correction (IMC):

      • Number of Experts: 3.
      • Qualifications: US board certified radiologists, specializing in neuro imaging.
    • Free Breathing Dynamic DLR:

      • Arterial Phase Detection: "Experienced radiologists" (number not specified for ground truth selection).
      • Clinical Image Review: 2 US board certified radiologists.

    4. Adjudication Method for the Test Set

    • PIQE (Precise IQ Engine): Randomized, blinded to the reviewers. Scored by 3 reviewers per anatomy. The text implies a consensus or averaging approach for reaching conclusions, as it states "all scored at, or above, clinically acceptable by three board-certified Radiologists per anatomy" and "reviewers exhibited a strong agreement".
    • NeuroLine+: Manual annotation by two experienced ARRT licensed MR technologists. The text implies their annotations served as the reference for accuracy, but does not detail an adjudication process if disagreements occurred.
    • Iterative Motion Correction (IMC): Randomized and blinded review by 3 US board certified radiologists.
    • Free Breathing Dynamic DLR:
      • Arterial Phase Detection: "Gold standard phase as manually selected by experienced radiologists." (Implies expert consensus or selection, but not specific adjudication).
      • Clinical Image Review: 2 US board certified radiologists read and scored the images. The conclusion is based on the "average of visual scores."

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance.

    • Yes, MRMC studies were done for PIQE, IMC, and Free Breathing Dynamic DLR.
      • These studies involved multiple readers assessing multiple cases, and comparing images processed with the new AI/DL features (PIQE, IMC, Free Breathing Dynamic DLR) against conventional methods or images without the feature applied.
      • **However, these studies were designed as a comparison of image quality from the new AI-enhanced reconstruction/correction methods versus conventional methods, not a comparative effectiveness study measuring human reader improvement with AI assistance vs. without AI assistance (i.e., human-in-the-loop performance). The radiologists were evaluating the reconstructed images themselves, not their diagnostic performance with and without AI tools integrated into their workflow.
      • Effect Size: The document does not provide a quantitative effect size in terms of how much human readers improve with AI assistance. Instead, it reports on the quality of the AI-processed images relative to conventional images, often concluding that the AI-processed images are "same or better," "meet acceptance criteria" for clinical acceptability (e.g., Likert scores), or provide "significant reduction" in artifacts.
        • PIQE: "Confirmed PIQE generates higher spatial in-plane resolution images...contributes to ringing artifact reduction, denoising and increased sharpness... is able to accelerate scanning by reducing the acquisition matrix only, while maintaining clinical matrix size and image quality."
        • IMC: "Significantly reducing motion artifacts, and improving overall image quality metrics... IMC corrected images are the same as, or better than, images without IMC applied."
        • Free Breathing Dynamic DLR: "Providing acceptable diagnostic confidence."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, standalone performance was evaluated for several features, primarily through bench testing and quantitative metrics.
      • Exsper 3D: Evaluated using phantom images, confirmed artifact reduction compared to conventional method.
      • Slice Shim: Evaluated using clinical images, confirmed image quality equal to or better than standard shim.
      • UTE CG Recon: Evaluated using phantom and clinical images, confirmed better maintenance of resolution and SNR.
      • Ringing Correction: Evaluated using phantom and clinical images, confirmed ringing reduction.
      • PIQE: Underwent performance (bench testing) using ACR phantom images and typical clinical images (brain and knee). Metrics included SNR, signal intensity profiles for ringing and sharpness, Edge Slope Width, Ringing Variable Mean, Signal-to-Noise ratio, and Contrast Change Ratio. This is a clear standalone evaluation of the algorithm's output metrics.
      • NeuroLine+: Underwent performance (bench) testing using clinical images. Autopositioning success rate was evaluated against manually annotated ground truth. This is a standalone evaluation of the algorithm's accuracy.
      • Iterative Motion Correction (IMC): Underwent performance (bench testing) using clinical datasets with simulated motion. Metrics of peak SNR and structural similarity (SSIM) were used. This is a standalone evaluation of the algorithm's effectiveness.
      • Free Breathing Dynamic DLR (Arterial Phase Detection): Underwent performance (bench) testing using clinical images, assessing the success rate of the automatic detection against manually selected gold standard phases. This is a standalone evaluation of the algorithm's accuracy.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Expert Consensus/Annotation:

      • PIQE: Clinically relevant categories (ringing, sharpness, SNR, overall IQ, feature conspicuity) scored by board-certified radiologists on a Likert scale, with >=3 considered clinically acceptable.
      • NeuroLine+: Manual annotation of angle and position of target planes by experienced ARRT licensed MR technologists.
      • IMC: SNR, tissue contrast, image sharpness, and diagnostic confidence scored by board-certified radiologists on a Likert scale.
      • Free Breathing Dynamic DLR:
        • Arterial phase detection: "Gold standard phase as manually selected by experienced radiologists."
        • Clinical image review: Overall SNR, overall IQ, feature conspicuity, and diagnostic confidence scored by board-certified radiologists on a Likert scale.
    • Quantitative/Objective Metrics:

      • Exsper 3D: Artifact reduction compared to conventional method.
      • Slice Shim: Image quality comparison to standard shim.
      • UTE CG Recon: Maintenance of image resolution and SNR.
      • Ringing Correction: Reduction of ringing.
      • PIQE (bench): SNR, signal intensity profiles for ringing and sharpness, Edge Slope Width, Ringing Variable Mean, Signal-to-Noise ratio, Contrast Change Ratio.
      • NeuroLine+ (bench): Angular error and autopositioning success rate (comparison to manual annotations).
      • IMC (bench): Peak SNR and structural similarity (SSIM).
      • Ceiling Camera: Percentage of successful patient orientation detection and cases requiring no correction.
    • Pathology/Outcomes Data: Not explicitly mentioned as a direct ground truth for the performance evaluation of these specific software features. The IMC study did use clinical patients with pathology, but the evaluation was still based on image quality and diagnostic information as assessed by radiologists, not a direct comparison to pathology reports or long-term outcomes.


    8. The sample size for the training set

    • The document consistently states for PIQE, NeuroLine+, IMC, and Free Breathing Dynamic DLR that "All testing data were acquired separately and independently from the training data after the machine learning training was completed."
    • However, the specific sample sizes for the training datasets themselves are not provided in this document.

    9. How the ground truth for the training set was established

    • Again, the document states that training data was used, particularly for features utilizing Deep Learning (e.g., PIQE, IMC, Free Breathing Dynamic DLR, NeuroLine+ via Machine Learning).
    • However, the document does not detail how the ground truth for these training sets was established. It only clarifies that the test data was independent of the training data.